User's Manual

Table Of Contents
TrackSense
®
LyoPro Version BETA Page 26 of 46
User Manual User Manual TrackSense LyoPro MAN 7
Special: at +70 °C for approximately 5 hours
8.2. In-House Calibration Solutions
8.2.1. Ellab Temperature Standard (ETS)
The ETS can be operated as an integrated part of the ValSuite software. It is a
temperature standard that can be started, stopped and used in report
analysis and calibrations.
8.2.1.1. ETS Versions and Measuring Ranges
*Accuracy between -50 and +150 °C
8.2.1.2. Proper Use
Although the ETS is quite robust, it should be handled with
care. Dropping or causing mechanical shock will loosen or
damage the electronic connections. The platinum resistor is
extremely sensitive, drift can therefore occur if not handled
properly
Only use cables (USB) provided by Ellab
The ETS system must be properly grounded
Never use the ETS outside the measuring range
Any ETS with recurring problems should be returned to Ellab
for service
In order to maintain the accuracy of the ETS, it is
recommended that they are recalibrated by Ellab at least
once every 12 months
WARNING: When returning an ETS to Ellab for service or recalibration,
please use the suitcase delivered with the product in order to avoid
damage. If the product does not arrive at Ellab in a case, all warranty is
lost and Ellab reserves the right to include a suitcase at the user’s
expense upon returning the product.
Ordering code Type Temp. Range Accuracy
20000190 ETS10 -80 to +250 °C
± 0.025 °C
± 0.015 °C*
20000191 ETS20 -50 to +150 °C ± 0.020 °C
20000192 ETS25
-196 to +420
°C
± 0.025 °C