User's Manual
Table Of Contents
- 1. Safety Information
- 2. New Order Contents
- 3. TrackSense® LyoPro 10-slot Reader Station
- 4. Activating ValSuite®
- 5. How to Start Logging Data with TrackSense LyoPro
- 6. TrackSense® LyoPro Loggers
- 6.1. TrackSense® LyoPro
- 6.1.1. Data Logger Specifications
- 6.1.2. Proper Use
- 6.1.3. Data Logger Serial Numbers
- 6.1.4. Disassembling / Assembling Data Loggers
- 6.1.5. Procedure for Cleaning TrackSense® LyoPro Data Loggers
- 6.1.6. Procedure for Changing the Data Logger O-Ring
- 6.1.7. Extended Warranty
- 6.1.8. Thermocouple Sensor
- 6.1.9. -30 to -80 C Application Recommendations
- 6.1. TrackSense® LyoPro
- 7. Battery
- 8. Calibration
- 9. Accessories
- 10. Examples on how to Configure LyoPro with Appropriate Accessories
- 11. Firmware Update
- 12. Ellab Service Contract
- 13. Technical Specifications
- 14. FAQ
- 15. Changelog
TrackSense
®
LyoPro Version BETA Page 27 of 46
User Manual User Manual TrackSense LyoPro – MAN 7
Note: In order to protect the ETS, the upper part of the sensor stem (see
picture below) should never be exposed to temperatures outside -50 °C
to +200 °C (transition limit).
8.2.1.3. Setting up the ETS
The ETS can be setup in 3 different ways. It can be used as a standalone
unit with a display, in connection with the ETS Suite software or it can
be setup in connection with the ValSuite software.
Setup – Stand Alone Unit w/ Display:
1. Connect the ETS using a USB cable to the ETS display
2. Connect the display to the power supply
Setup – ETS using ETS Suite:
Note: USB drivers must be installed first for firmware versions older
than 2.0.
1. Connect the ETS to the PC via USB cable
2. Verify that the ETS Suite is detecting the connected ETS
Setup – ETS using ValSuite:
1. If using the USB connection, install the Ellab USB driver first
2. Disconnect all the equipment
3. The driver can be found on the ValSuite USB Key
4. Run the file Ellab USBDrivers_xxx.exe