User manual
BeeHive204
Selftest of ISP connector
• Insert Diagnostic POD for ISP connectors #2 into ZIF socket of the programmer.
Diagnostic POD for ISP connectors #2 must be inserted as 48 pins device.
• Interconnect 20 pins connector of Diagnostic POD for ISP connectors #2 with an ISP
connector of the programmer with an ISP cable, included in delivery programmer
package. Be sure that pins are interconnected properly (i.e. 1-1, 2-2... 20-20).
• Run selftest of ISP connector in PG4UW (Programmer / Selftest ISP connector…).
Calibration test
• Insert 48 Pins Calibration test POD, Type I into ZIF socket of the programmer. 48 Pins
Calibration test POD, Type I must be inserted as 48 pins device.
• Run calibration test of programmer in PG4UW (Programmer / Calibration test).
Technical specification
Specification (BeeHive204 multiprogramming system)
• 4x universal programming module (4x 48-pin DIL ZIF sockets)
• operation result LEDs, LED power
• USB 2.0 high-speed compatible port
• line power input 100-240VAC/60W max.
23