User Manual – 1929LM 
SW200158 Rev A      - Page 8 of 31 
Guidance and manufacturer’s declaration-electromagnetic immunity –   
for all ME EQUIPMENT AND ME SYSTEMS 
Guidance and manufacturer’s declaration-electromagnetic immunityError! Bookmark not 
defined. 
The 1929LM Touchmonitor for Healthcare Applications is intended for use in the electromagnetic 
environment specified below. The customer, or the user of the 1929LM Touchmonitor for Healthcare 
Applications, should assure that it is used in such an environment. 
Immunity Test   
IEC 60601 
test level 
Compliance level 
Electromagnetic environment – guidance   
Electrostatic 
discharge (ESD) 
IEC 61000-4-2 
±    6 kV contact 
±    8 kV air 
± 6 kV contact 
± 8 kV air 
Floors should be wood, concrete or ceramic 
tile. If floors are covered with synthetic 
material, the relative humidity should be at 
least 30%. 
Electrical fast 
transient/burst 
IEC 61000-4-4 
± 2 kV for power 
supply lines 
± 1 kV for 
input/output lines 
± 2 kV for power 
supply lines 
± 1 kV for 
input/output Lines 
Mains power quality should be that of a 
typical commercial or hospital environment. 
Surge 
IEC 61000-4-5 
± 1 kV line(s) to 
line(s) 
± 2 kV line(s) to 
earth 
± 1 kV line(s) to 
line(s) 
± 2 kV line(s) to 
earth 
Mains power quality should be that of a 
typical commercial or hospital environment. 
Voltage dips, short 
interruption and 
voltage variations 
on power supply 
input lines 
IEC 61000-4-11 
<5% U
T
(>95% dip in U
T
) 
for 0.5 cycle 
40% U
T
(60% dip in U
T
) 
for 5 cycles 
70% U
T
(30% dip in U
T
) 
for 25 cycles 
<5% U
T
(>95% dip in U
T
) 
for 250 cycles 
<5% U
T
(>95% dip in U
T
) 
for 0.5 cycle 
40% U
T
(60 % dip in U
T
) 
for 5 cycles 
70% U
T
(30% dip in U
T
) 
for 25 cycles 
<5% U
T
(>95% dip in U
T
) 
for 250 cycles 
Mains power quality should be that of a 
typical commercial or hospital environment. 
If the user of the 1929LM Touchmonitor for 
Healthcare Applications requires continued 
operation during power mains interruptions, 
It is recommended that the 1929LM 
Touchmonitor for Healthcare Applications be 
powered from an uninterruptible power 
supply or a battery. 
Power Frequency 
(50/60 Hz) 
Magnetic Field 
IEC 61000-4-8 
3 A/m 
3 A/m 
Power frequency magnetic fields should be 
at levels characteristic of a typical location in 
a typical commercial or hospital 
environment. 
NOTE      U
T
    is the a.c. mains voltage prior to application of the test level. 