User guide
v 
EMC Guidance 
Guidance and manufacturer’s declaration – electromagnetic emissions 
The 1528L is intended for use in the electromagnetic environment specified below. The customer or the user of the 1528L should 
assure that it is used in such an environment. 
Emissions test  Compliance  Electromagnetic environment – guidance 
RF emissions 
CISPR 11 
Group 1  The 1528L uses RF energy only for its internal function. Therefore, its RF emissions 
are very low and are not likely to cause any interference in nearby electronic 
equipment. 
RF emissions  Class B   
Harmonic 
emissions 
IEC61000-3-2 
Not applicable   
Voltage 
fluctuations/ 
flicker emissions 
IEC61000-3-3 
Complies   
Guidance and manufacturer’s declaration – electromagnetic immunity 
The 1528L is intended for use in the electromagnetic environment specified below. The customer or the user of the 1528L should 
assure that it is used in such an environment. 
Immunity test 
IEC60601 
test level 
Compliance level 
Electromagnetic environment – 
guidance 
Electrostatic discharge 
(ESD) 
IEC61000-4-2 
±6 kV Contact 
±8 kV air 
±6 kV Contact 
±8 kV air 
Floors should be wood, concrete or 
ceramic tile. If floors are covered with 
synthetic material the relative humidity 
should be at least 30 % 
Electrical fast 
transient/burst 
IEC61000-4-4 
±2 kV for power supply 
lines 
±1 kV for input/output 
lines 
±2 kV for power supply 
lines 
±1 kV for input/output 
lines 
Main power quality should be that of a 
typical commercial or hospital 
environment. 
Surge 
IEC6100-4-5 
±1 kV differential mode 
±2 kV common mode 
±1 kV differential mode 
±2 kV common mode 
Main power quality should be that of a 
typical commercial or hospital 
environment. 
Voltage dips, short 
interruptions and voltage 
variations on the power 
supply input lines 
IEC61000-4-11 
<5 % U
T
(>95 % dip in U
T
) 
for 0,5 cycle 
40 % U
T 
(60 % dip in U
T
) 
for 5 cycles 
70 % U
T
(30 % dip in U
T
) 
for 25 cycles 
<5 % U
T
(>95 % dip in U
T
) 
for 5 sec 
<5 % U
T
(>95 % dip in U
T
) 
for 0,5 cycle 
40 % U
T 
(60 % dip in U
T
) 
for 5 cycles 
70 % U
T
(30 % dip in U
T
) 
for 25 cycles 
<5 % U
T
(>95 % dip in U
T
) 
for 5 sec 
Main power quality should be that of a 
typical commercial or hospital 
environment. If the user of the 1528L 
requires continued operation during power 
mains interruptions, it is recommended 
that the 1528L be powered from an 
uninterruptible power supply or a battery. 
Power frequency 
(50/60 Hz) 
Magnetic field 
IEC61000-4-8 
3 A/m  3 A/m, 30 A/m  Power frequency magnetic fields should 
be at levels characteristic of a typical 
location in a typical commercial or 
hospital environment. 
NOTE U
T
 is the ac. mains voltage prior to application of the test level. 