Installation, Operation & Maintenance Manual
IP262/Z0, Rev. AB
February 2012
7.11.1.9 Xdr Tilt Time - D833
A diagnostic parameter indicating the time for which the transducer has been out of the water in min:secs.
7.11.1.10 Noise Level - D834
A diagnostic parameter indicating the noise level in the tank as a % of full screen.
7.11.2 Diagnostics parameters D835 to D852
7.11.2.1 Input Status - D835
A diagnostic parameter indicating the status of the digital input, 1 indicating active, 0 indicating not active.
7.11.2.2 LE Count – D836
This gives an indication as to reliability of signal.
7.11.2.3 Radio Channel - D842
This parameter is not used with this version of the MSL600.
7.11.2.4 Radio Address - D843
This parameter is not used with this version of the MSL600.
7.11.2.5 CU Temperature (Temperature of Control Unit) - D844
View this parameter to see the current temperature in °C (imperial – degrees F) inside the MSL600 control unit.
This temperature is used for control of the LCD back light, turning it on to generate a degree of internal heating
in cold weather and off when hot to avoid overheating.
7.11.2.6 Xducer Temp (Temperature of Transducer) - D845
View this parameter to see the current temperature in °C (imperial - degrees F) of the transducer. This
temperature is used in the compensation of the speed of sound in the supernatant.
7.11.2.7 Free Memory (Logging Memory Free) - D846
View this parameter to see the amount of MSL600 logging memory remaining as a % before the end of the
buffer is reached and over writing at the start begins. This parameter is compatible with Mobrey Measurement
Log View software.
7.11.2.8 Date of Change (Last date of Change) - D848
View this parameter to see the date of the last change to the configuration of the MSL600. This parameter can
be used to detect unauthorised access.
7.11.2.9 Echo Sharpness – D849
View this parameter to estimate the size of the echo. A typical reading would be between 20 and 30.