Revision D 3-9000-745 October 2010 MON 20/20 Software for Gas Chromatographs Applies to all Emerson XA Series Gas Chromatographs
MON 20/20 Software for Gas Chromatographs User Manual NOTICE DANIEL MEASUREMENT AND CONTROL, INC. AND ROSEMOUNT ANALYTICAL (COLLECTIVELY, “SELLER”) SHALL NOT BE LIABLE FOR TECHNICAL OR EDITORIAL ERRORS IN THIS MANUAL OR OMISSIONS FROM THIS MANUAL.
WARRANTY 1. LIMITED WARRANTY: Subject to the limitations contained in Section 2 herein and except as otherwise expressly provided herein, Daniel Measurement and Control, Inc. and Rosemount Analytical, (collectively“Seller”) warrants that the firmware will execute the programming instructions provided by Seller, and that the Goods manufactured or Services provided by Seller will be free from defects in materials or workmanship under normal use and care until the expiration of the applicable warranty period.
IMPORTANT INSTRUCTIONS • Read all instructions prior to installing, operating, and servicing this product. • Follow all warnings, cautions, and instructions marked on and supplied with this product. • Inspect the equipment packing case and if damage exists, notify your local carrier for liability. • Open the packing list and carefully remove equipment and spare or replacement parts from the case. Inspect all equipment for damage and missing parts.
This page is intentionally left blank.
Table of Contents Section 1: Getting started What’s new in MON 20/20 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3 Getting started with MON 20/20 . . . . . . . . . . . . . . . . . . . . . . . . .1-6 System requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-6 Installing MON 20/20 . . . . . . . . . . . . . . . . . . . . . . . . . . .1-7 Launching MON 20/20 . . . . . . . . . . . . . . . . . . . . . . . . . .1-7 Registering MON 20/20 . . . . . . . . . . . . . . . . . . . . .
User Manual 3-9000-745 MON20/20 Software for Gas Chromatographs OCTOBER 2010 Working with a chromatogram . . . . . . . . . . . . . . . . . . . . . . . . . Editing a chromatogram trace . . . . . . . . . . . . . . . . . . . Viewing chromatogram results . . . . . . . . . . . . . . . . . . Saving a chromatogram trace . . . . . . . . . . . . . . . . . . . Removing a chromatogram trace from view . . . . . . . . Forcing a calibration . . . . . . . . . . . . . . . . . . . . . . . . . .
MON20/20 Software for Gas Chromatographs OCTOBER 2010 User Manual 3-9000-745 Managing your gas chromatograph’s discrete inputs . . . . . . . .3-24 Renaming a discrete input . . . . . . . . . . . . . . . . . . . . . .3-24 Setting a discrete input’s operational mode . . . . . . . . .3-25 Monitoring the operational status of a discrete input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-27 Inverting the polarity of a discrete input . . . . . . . . . . .
User Manual 3-9000-745 MON20/20 Software for Gas Chromatographs OCTOBER 2010 Managing Validation Data Tables . . . . . . . . . . . . . . . . . . . . . . Managing calculations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Setting standard calculations by stream . . . . . . . . . . Editing average calculations . . . . . . . . . . . . . . . . . . . . Viewing an archive of averages for a given variable . Copying stream settings . . . . . . . . . . . . . . . . . . . . . . .
MON20/20 Software for Gas Chromatographs OCTOBER 2010 User Manual 3-9000-745 Adding files to the GC . . . . . . . . . . . . . . . . . . . . . . . . . .5-17 Deleting files from the GC . . . . . . . . . . . . . . . . . . . . . . .5-18 Viewing the event log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-19 Displaying reports . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-22 Understanding report types . . . . . . . . . . . . . . . . . . . . .
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 OCTOBER 2010 Managing users . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Creating users . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Exporting a list of user profiles . . . . . . . . . . . . . . . . . . Importing a list of user profiles . . . . . . . . . . . . . . . . . . Editing users . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Removing a user . . . . . . . . . . . . . . . .
Section 1: Getting started Welcome to MON 20/20—a menu-driven, Windows-based software program designed to remotely operate and monitor the Daniel® Danalyzer™ XA series and the Rosemount® Analytical XA series of gas chromatographs. MON 20/20 operates on an IBM-compatible personal computer (PC) running the Windows XP operating system or later.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 OCTOBER 2010 MON 20/20 can generate the following reports: • 24-Hour Averages • Analysis (GPA) • Analysis (ISO) • Calibration • Final Calibration • Validation • Final Validation • Hourly Averages • Monthly Averages • GC Configuration • Raw Data • Variable Averages • Weekly Averages • Dew Temperature Calculation (optional) MON 20/20 can access and display the following GC-generated logs: • Alarm Log • Event Log • Parameter List • Maintenance
MON20/20 Software for Gas Chromatographs OCTOBER 2010 1.1 User Manual 3-9000-745 What’s new in MON 20/20 Users familiar with MON2000 or MON2000 Plus will find a few changes when using MON 20/20: • Login security is at the gas chromatograph level instead of at the software level. This means that you no longer have to log in after starting MON 20/20—but you do have to log in to the gas chromatograph to which you are trying to connect.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 OCTOBER 2010 • Enhanced Chromatogram Viewer. The following enhancements have been made to the Chromatogram Viwer: - - - - - Users can view an unlimited number of chromatograms, in any configuration. For example, a user can view an archived chromatogram and a live chromatogram. For more information, see “Viewing chromatograms” on page 2-1. The “Keep Last CGM” option.
MON20/20 Software for Gas Chromatographs OCTOBER 2010 User Manual 3-9000-745 • Baseline offsetting. In some situations that involve TCD detectors the baseline may be displayed either too high on the graph, in which case the tops of the peaks are cut off, or too low on the graph, so that the bases of the peaks are cut off. If this occurs it is possible to offset the baseline either up or down so that the entire peak can be displayed on the graph.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 1.2 OCTOBER 2010 Getting started with MON 20/20 This section covers such issues as installing, registering and setting up the software, as well as configuring MON 20/20 to meet your specific needs. 1.2.1 System requirements To achieve maximum performance when running MON 20/20, ensure your PC meets the following specifications: • Software - Windows® XP (Service Pack 2 or later), Windows® Vista, or Windows® 7. - Internet Explorer® 6.
MON20/20 Software for Gas Chromatographs User Manual OCTOBER 2010 3-9000-745 1.2.2 Installing MON 20/20 You must install MON 20/20 from the Emerson Process Management MON 20/20 Software for Gas Chromatographs CD-ROM onto your hard drive; you cannot run the program from the CD-ROM. Double-click the Setup file and follow the on-screen installation instructions. Upon successful installation, MON 20/20 creates a shortcut icon on the computer’s desktop.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 OCTOBER 2010 Figure 1-1. The Register MON 20/20 window, page 1 1. Enter your name, your company’s name, and the serial number for your copy of MON 20/20 into the appropriate fields on the Register MON 20/20 window. 2. Click Next to continue. 3. Choose the desired registration method by clicking the corresponding checkbox. Figure 1-2.
MON20/20 Software for Gas Chromatographs OCTOBER 2010 User Manual 3-9000-745 Note To delay registration, check Register later (remind me). MON 20/20 will display the Register MON 20/20 window the next time you start the program. To prevent the Register MON 20/20 window from displaying with each program startup—and without registering—check Register later (don’t remind me). Note You can register at any time by selecting Register MON 20/20... from the Help menu. 4. Click Finish. 1.2.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 OCTOBER 2010 Note Another method for changing the folder location is to type the folder’s location into the Data Folder field and press ENTER. When the “Create the folder?” message appears, click Yes. 5. The Data Folder field updates to display the new location. Figure 1-4. The Program Settings window 1.2.
MON20/20 Software for Gas Chromatographs OCTOBER 2010 User Manual 3-9000-745 If you get the “GC directory file not found” message, click OK. The GC Directory window appears and displays a table containing an inventory of the GCs to which MON 20/20 can connect. 2. If you are configuring the first GC connection for MON 20/20, there will be on one generic GC record listed in the window. To add another record, select Add from the GC Directory window’s File menu.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 OCTOBER 2010 5. Select Ethernet. The Ethernet Connection Properties for New GC window appears. 6. In the IP address field, enter the IP address of the GC to which you want to connect. Figure 1-7. The Ethernet Connection Properties for New GC window Note If you type in an invalid IP address, you will get an error message when MON 20/20 attempts to connect to the GC. 7. Click OK. When the Save changes? message appears, click Yes. 8.
MON20/20 Software for Gas Chromatographs OCTOBER 2010 User Manual 3-9000-745 15. To save the changes and keep the window open click Save from the GC Directory window. To save the changes and close the window, click OK. When the Save changes? message appears, click Yes. For more details about configuring MON 20/20 connections, see “Configuring the gas chromatograph’s Ethernet port” on page 4-95. 1.2.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 OCTOBER 2010 2. Select the checkbox for each gas chromatograph who information you want to save. If you want to save the entire list, click Select All. 3. Click OK. The Export GC Directory File save as dialog displays. 4. Choose a save location. The default location is GCXP Data. 5. The file is automatically given the name of GC_DIRECTORY_EXPORT.DAT. If you prefer a different name, type it into the File name field. 6. Click Save.
MON20/20 Software for Gas Chromatographs OCTOBER 2010 User Manual 3-9000-745 1.2.8 Launching MON 20/20 from the SNAP-ON for DeltaV This section assumes that DeltaV is installed on the PC along with MON 20/20. Note To successfully use MON 20/20 SNAP-ON for DeltaV, you must be familiar with using the DeltaV digital automation system. To start MON 20/20, do the following: 1.
User Manual 3-9000-745 MON20/20 Software for Gas Chromatographs OCTOBER 2010 Figure 1-10. The Device Connection View 3. Right-click on a connected gas chromatograph icon to display the context menu.
MON20/20 Software for Gas Chromatographs OCTOBER 2010 User Manual 3-9000-745 Figure 1-11. Right-click to view context menu 4. Select SNAP-ON/Linked Apps → Launch MON 20/20. MON 20/20 starts and connects automatically to the GC. 1.2.9 Launching MON 20/20 from the AMS Device Manager This section assumes that DeltaV and AMS are installed on the PC along with MON 20/20. To start MON 20/20, do the following: 1.
User Manual 3-9000-745 MON20/20 Software for Gas Chromatographs OCTOBER 2010 Figure 1-12. Device Explorer 2. In the Device Connection View, open device icons by clicking once on each icon. Follow the path of connections until you locate the desired gas chromatograph icon. 3. Right-click on a connected gas chromatograph icon to display the context menu.
MON20/20 Software for Gas Chromatographs OCTOBER 2010 User Manual 3-9000-745 Figure 1-13. Right-click to view the context menu 4. Select SNAP-ON/Linked Apps → Launch MON 20/20. MON 20/20 starts and connects automatically to the GC.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 OCTOBER 2010 1.2.10 The MON 20/20 user interface MON 20/20 has two areas of interaction: the Control Area, at the top of the program’s main window, and the GC Status Bar, located at the bottom of the program’s main window. Figure 1-14.
MON20/20 Software for Gas Chromatographs OCTOBER 2010 User Manual 3-9000-745 The main user interface The main user interface of the main window contains the menus and icons that allow you to control MON 20/20 and the GC to which MON 20/20 is connected. Figure 1-15. The Control Area Titlebar Menu bar Toolbar Dialog Control Tabs Titlebar - The Titlebar displays the name of the program, and well as the program’s connection status.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 OCTOBER 2010 • Toolbar - The Toolbar contains shortcut icons for the most important and/or most often used MON 20/20 commands. From the Toolbar you can do such things as connect to and disconnect from a GC, view chromatographs, and view help files. Table 1-1. Function of the shortcut icons on the Toolbar Connect to a gas chromatograph. Disconnect from a gas chromatograph. Open a configuration file. Print a GC configuration report.
MON20/20 Software for Gas Chromatographs OCTOBER 2010 User Manual 3-9000-745 • Dialog Control Tabs bar - The Dialog Control Tabs bar contains four buttons that allow you to manage the behavior of all windows that are open in the main window. The four buttons are Minimize All, Maximize All, Restore All, and Close All. Figure 1-16. The main window showing the function of the Dialog Control Tabs bar The bar also displays a button for each open window that allows you to select or deselect that window.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 OCTOBER 2010 The GC Status Bar The GC Status Bar of the main window displays useful information about the status and functioning of the gas chromatograph to which MON 20/20 is connected. Figure 1-17. The GC Status Bar The GC Status Bar contains the following sections: • GC - The first row displays the name of the GC to which MON 20/20 is connected. If MON 20/20 is not connected to a GC, “Not Connected” displays in this row.
MON20/20 Software for Gas Chromatographs User Manual OCTOBER 2010 3-9000-745 • GC System - Displays the date and time according to the GC to which MON 20/20 is connected. The date and time displayed may be different from the user’s date and time, depending on the physical location of the GC. • FID Flame Status - Displays the status of the FID flame. Options are OFF with red background, ON with green background, and OVER TEMP with red background.
User Manual 3-9000-745 MON20/20 Software for Gas Chromatographs OCTOBER 2010 Figure 1-18. The Connect to GC window Note If you want to edit the connection parameters for one or all GCs listed in the Connect to GC window, click Edit Directory. The GC Directory window will appear. See “Configuring MON 20/20 to connect to a gas chromatograph” on page 1-10 for more information. Click the Ethernet button beside the GC to which you want to connect.
MON20/20 Software for Gas Chromatographs OCTOBER 2010 User Manual 3-9000-745 3. The Login dialog appears. Figure 1-19. The Login window Note All GCs are shipped with two default user names: daniel and emerson. A user pin is not required when using either of these user names and both user names allow administrator-level access to the GC. To add a user pin to either of these user names or for information about creating and edit user names in general, see “Managing users” on page 7-17.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 OCTOBER 2010 1.2.12 Disconnecting from a gas chromatograph Disconnecting from a GC will automatically log you off of the GC. To disconnect from a gas chromatograph, do one of the following: • On the Toolbar, click . • Select Disconnect from the Chromatograph menu.
MON20/20 Software for Gas Chromatographs User Manual OCTOBER 2010 3-9000-745 Table 1-2. Frequently Used Keystrokes (Continued) Keystroke Action SPACE Toggles settings (via radio buttons or check boxes). TAB Moves to the next control element (e.g., button) in the window; to use TAB key to move to next data field, select Program Settings... from the File menu and clear the Tab from spreadsheet to next control check box. You can use the following function keys from the main window: Table 1-3.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 1.4 OCTOBER 2010 Procedures guide Use the following table to look up the related manual section, menu path and, if appropriate, the keystroke for a given procedure. Table 1-4. MON 20/20 Task List Task or Data Item Section(s) Menu Path [Keystroke] 24-hour average, component(s) measured 4.5.2 Application → Calculations → Averages... Add a gas chromatograph 1.2.6 File → GC Directory Alarms, related components 4.2 4.8 3.
MON20/20 Software for Gas Chromatographs User Manual OCTOBER 2010 3-9000-745 Table 1-4. MON 20/20 Task List Task or Data Item Section(s) Menu Path [Keystroke] Communications 4.12 Application → Communication... Application → Ethernet Ports... Component code and name 4.2 Application → Component Data... [F6] Component full scale (for output) 4.1 3.6 Application → System... Hardware → Analog Outputs... Component(s) programmed for input 3.5 3.3 Application → Analog Inputs...
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 OCTOBER 2010 Table 1-4. MON 20/20 Task List Task or Data Item Section(s) Menu Path [Keystroke] Height or area measurement method 4.2 Application → Component Data... [F6] High alarm 4.8 Application → Limit Alarms → User... (Analyzer) I.D. 4.1 Application → System... Inhibit on-off times 4.3.4 Application → Timed Events... [F5] Input(s) being used 3.5 3.3 Hardware → Analog Inputs... Hardware → Discrete Inputs...
MON20/20 Software for Gas Chromatographs User Manual OCTOBER 2010 3-9000-745 Table 1-4. MON 20/20 Task List Task or Data Item Section(s) Menu Path [Keystroke] Stream number(s) (for output) 4.8 3.6 3.4 Application → Limit Alarms → User... Hardware → Analog Outputs... Hardware → Discrete Outputs... Stream sequences skipped, number 4.1 4.10 Application → System... Application → Streams... Streams analyzed, number 4.1 4.10 Application → System... Application → Streams...
User Manual 3-9000-745 MON20/20 Software for Gas Chromatographs OCTOBER 2010 Figure 1-21. MON 20/20 in offline edit mode 4. Use the Application and Hardware menu commands to edit the GC’s settings. For more information on these commands, see Section 3 and Section 4. 5. When finished configuring the GC, click to disconnect from the GC and to save the changes to the configuration file and to leave offline edit mode.
MON20/20 Software for Gas Chromatographs OCTOBER 2010 User Manual 3-9000-745 1.5.2 Saving a gas chromatograph’s current configuration Configuration files are saved with the .xcfg extension. To save a GC’s current configuration to a PC, do the following: 1. Select Save Configuration (to PC)... from the File menu. The Save as dialog displays. 2. Give the file a descriptive name or use the pre-generated file name and navigate to the folder to which you want to save the file. 3. Click Save. 1.5.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 1.6 OCTOBER 2010 Restoring the GC to its factory settings The GC’s default timed event, component data and validation data tables are created at the factory and are not accessable by users. To restore these tables to their default values, do the following: Note The GC should be in Idle mode while performing this task. 1. With the GC idle, select Restore to Factory Settings... from the File menu.
MON20/20 Software for Gas Chromatographs OCTOBER 2010 1.7 User Manual 3-9000-745 Configuring your printer Select Print Setup... from the File menu to configure the settings for the printer connected to your PC. These settings will apply to any print job queued from MON 20/20, such as the reports that are configured by the Printer Control. See “Printing reports automatically” on page 5-44 for information. Figure 1-24. The Print Setup dialog The settings available depend on the printer model.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 1.8 OCTOBER 2010 Using online help Currently, the online help feature contains all user information and instructions for each MON 20/20 function as well as the MON 20/20 system. To access the online help, do one of the following: • Press F1 to view help topics related to the currently active dialog or function. • Select Help Topics from the Help menu to view the help contents dialog. 1.
MON20/20 Software for Gas Chromatographs OCTOBER 2010 User Manual 3-9000-745 To view hidden columns, do the following: 1. Select Program Settings... from the File menu. The Program Settings window displays. Figure 1-25. The Program Settings window 2. Select the Show Physical Names checkbox.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 OCTOBER 2010 3. Click OK. The Physical Name column now will be visible on all windows that have the column, such as the Heater window shown in the example below. Figure 1-26. The Heater window showing Physical Name column 1.11 Selecting the GC’s networking protocol MON 20/20 can connect to the GC using one of two networking protocols: PPP or SLIP. If the version level of the GC’s firmware is 1.
MON20/20 Software for Gas Chromatographs OCTOBER 2010 User Manual 3-9000-745 Figure 1-27. The Program Settings window 2. To use the PPP protocol, make sure the Use PPP protocol for serial connection (use SLIP if unchecked) checkbox is selected; to use the SLIP protocol, make sure the Use PPP protocol for serial connection (use SLIP if unchecked) checkbox is not selected. 3. Click OK.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 OCTOBER 2010 1.12 Using the context-sensitive variable selector The MON 20/20 method for selecting variables for calculations and other purposes is based on a simple, self-contained system. Figure 1-28. Example of a context-sensitive variable selector The context-sensitive variable selector consists of a first-level element, called the context, that is followed by a series of tiered, drop-down lists.
MON20/20 Software for Gas Chromatographs OCTOBER 2010 User Manual 3-9000-745 3. Click the third-level drop-down list. The full list of available user alarm variables displays. Figure 1-30. Third-level drop-down list 4. Select the variable you want to use for the alarm. If there are components associated with the variable, the fourth-level drop-down list will display. 5. If displayed, click the fourth-level drop-down list. The full list of available components displays. Figure 1-31.
User Manual 3-9000-745 Figure 1-32.
Section 2: Using the chromatograph functions For viewing and managing chromatograms, MON 20/20 is flexible and straighforward. This chapter shows you how to connect to and disconnect from a gas chromatograph. This chapter also shows you how to access the Chromatogram Viewer, as well as to use it to view, print and manipulate various types of chromatograms. Finally, this chapter explains how to set a gas chromatograph’s date and time. 2.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 OCTOBER 2010 2.1.1 Data displayed in the chromatogram window Figure 2-1. The chromatogram window trace #1 retention time peak detection marker trace #1 trace #2 timed event marker The following elements are displayed in the chromatogram window: • The chromatogram.
MON20/20 Software for Gas Chromatographs User Manual OCTOBER 2010 3-9000-745 • Timed event markers. These markers, which correspond to events from the Timed Events table, display on the chromatogram as black marks descending from the trace-line. There are three types of timed event markers: - Valve events display as long descending marks. Integration events display as medium descending marks. Spectrum gain events display as short descending marks. • Peak detection markers.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 OCTOBER 2010 Figure 2-2. View Current CGM The chromatogram displays in the chromatogram window. If the chromatogram contains one trace, the Det1 checkbox is automatically checked; if the chromatogram contains two traces, the Det1 and Det2 checkboxes are automatically checked. To remove a trace, uncheck its detector checkbox. Each trace that displays is color-coded; use the Chromatogram pulldown menu to select a specific trace. Figure 2-3.
MON20/20 Software for Gas Chromatographs OCTOBER 2010 User Manual 3-9000-745 Figure 2-4. The Chromatogram Viewer timed events chromatogram window component data Note By default, the timed events and component data tables are configured to scroll to and highlight the next occurring event in the analysis cycle. To disable this feature, rightclick on one of the tables and uncheck the Auto Scroll option on the pop-up menu. 2.1.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 OCTOBER 2010 • Protected chromatograms - Protected chromatograms are never deleted from the GC. To protect a chromatogram, see “Protecting or unprotecting an archived chromatogram” on page 2-9. Note Protected chromatogram files have a “lock” icon ( ) displayed beside them.
MON20/20 Software for Gas Chromatographs User Manual OCTOBER 2010 3-9000-745 Final Validation chromatograms - These chromatograms are treated in the same manner as final calibration chromatogram files.To view one or more archived chromatograms, do the following: 1. Click GC Archive. The Select archive file(s) window appears. Figure 2-5. The Select archive file(s) window The files can be sorted by date, file name, analysis type, time, or stream number by clicking the appropriate column header.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 OCTOBER 2010 2. Select one or more archive files by clicking them. Use the SHIFT or CTRL key to make multiple selections. Note To save the selected files to the PC, select the Download and save selected chromatograms check box and click Download & Save. 3. Click Download & Show. The Select window displays for each chromatogram that contains data from more than one detector. Figure 2-6. The Select window 4.
MON20/20 Software for Gas Chromatographs OCTOBER 2010 User Manual 3-9000-745 Figure 2-7. The Chromatogram Viewer displaying an archived chromatogram 2.1.4 Protecting or unprotecting an archived chromatogram By default, archived chromatograms are not saved indefinitely. Once the GC’s storage capacity for archived chromatograms has been reached, the oldest archived chromatograms are deleted to make room for the newest archived chromatograms.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 OCTOBER 2010 Note Protected chromatograms have a “lock” icon ( ) displayed beside them. Note To protect an archived chromatogram you must be logged in as a supervisor or admin. To protect a chromatogram, do the following: 1. Click GC Archive. The Select Archive File(s) window appears. Figure 2-8.
MON20/20 Software for Gas Chromatographs OCTOBER 2010 User Manual 3-9000-745 By default, they are sorted by date, with the newest chromatogram listed first. Note By default, only recent chromatograms—that is, the last five runs for each stream—are displayed. To view all archived chromatograms, click All. To return to viewing only recent chromatograms, click Recent. 2. Make sure the Chromatogram tab is selected and then select the appropriate archived chromatogram by clicking it.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 OCTOBER 2010 To unprotect a protected file, do the following: 1. Click GC Archive. The Select archive file(s) window appears. Figure 2-10. The Select archive file(s) window 2. Locate and select the protected chromatogram that you want to unprotect. Use the SHIFT or CTRL key to make multiple selections. ) from 3. Click Unprotect. MON 20/20 will remove the “lock” icon ( beside the selected chromatogram.
MON20/20 Software for Gas Chromatographs OCTOBER 2010 User Manual 3-9000-745 2.1.5 Viewing a saved chromatogram To view a chromatogram that was saved to disk, do the following: 1. Click PC File. The Open dialog appears. 2. Navigate to the desired .xcgm file or .xcmp comparison file and select it. To make multiple selections, use the SHIFT or CTRL key. 3. Click OK. The Select window displays for each chromatogram that contains data for more than one detector. Figure 2-11. The Select window 4.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 Figure 2-12.
MON20/20 Software for Gas Chromatographs OCTOBER 2010 2.2 User Manual 3-9000-745 Working with the graph Right-clicking with the mouse on the graph brings up the following commands and keyboard shortcuts: Command Name Shortcut Description Zoom In “+” (NUMPAD) Zooms in on the entire graph. NOTE: Another way to zoom in is by clicking and dragging your mouse to select the region of the graph that you want to zoom in on. Zoom Out “-” (NUMPAD) Zooms out from the entire graph.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 OCTOBER 2010 Command Name Shortcut Description Toggle Mouse Position Tip CTRL + F4 The graph’s cursor follows the movement of the mouse while a hovering tooltip displays the exact coordinates of the current point. Toggle Nearest Position Tip CTRL + F9 The graph’s cursor follows the movement of the mouse cursor. Print CTRL + P Prints the chromatogram.
MON20/20 Software for Gas Chromatographs User Manual OCTOBER 2010 3-9000-745 Figure 2-13. The Edit Scales window The following table lists the parameters that can be edited: Command Description Default Value X Min Sets the minimum value, in seconds, for the X axis. 0 X Max Sets the maximum value, in seconds, for the X axis. The is value is determined by the Timed Events table. 100 Y Min Sets the minimum value for the Y axis. -10 Y Max Sets the maximum value for the Y axis.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 OCTOBER 2010 Figure 2-14. A chromatogram Y axis Y Max Y interval color-coded traces Y Min X axis X max X Min X interval To see how your changes affect the graph, click Apply. To accept your changes, click OK. • Click Cursor to toggle the cursor size from coarse movement (less accurate) to fine movement (more accurate). • Click Print to print the chromatogram window.
MON20/20 Software for Gas Chromatographs OCTOBER 2010 User Manual 3-9000-745 2.3.2 Additional plot commands In addition to the Graph bar, there are a few other commands available that allow you to manipulate the look and feel of the graph. To access the additional plot commands menu, right-click on the Chromatogram Viewer anywhere except on the graph or the timed event and component data tables.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 OCTOBER 2010 Command Description Show Mini Plot Toggles the display of a smaller version of the chromatogram in a separate, resizable window. This allows you to keep an overview of the entire graph at all times, especially when zoomed in. This window automatically displays whenever you zoom in on the original chromatogram.
MON20/20 Software for Gas Chromatographs OCTOBER 2010 2.4 User Manual 3-9000-745 Working with a chromatogram Figure 2-15. The Chromatogram bar The Chromatogram bar contains a row of buttons that allows you to manipulate a single chromatogram. Below the row of buttons is the chromatogram pull-down menu, which contains a list of all of the currently displayed chromatograms/traces. Before you can work with a chromatogram you must first select it from the pull-down menu. 2.4.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 OCTOBER 2010 2. Click Edit. The Edit Chromatogram dialog appears. Figure 2-17. The Edit Chromatogram dialog Command Description X Offset Enter a positive number to move the trace to the right, or a negative number to move the trace to the left. Y Offset Enter a positive number to move the trace up, or a negative number to move the trace down. # points Number of data points in the trace. This field is read-only.
MON20/20 Software for Gas Chromatographs OCTOBER 2010 User Manual 3-9000-745 2.4.2 Viewing chromatogram results To display a table of calculation results for a trace, do the following: 1. From the Chromatogram pull-down menu, select the appropriate trace. Figure 2-18. Chromatogram pull-down menu 2. Click Results. A window appears displaying the calculation results for the selected trace.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 OCTOBER 2010 Figure 2-19. The results window • Click Save to save these results in one of the following formats: tab-delimited (.txt), comma-delimited (.csv), Microsoft Excel (.xls), HTM (.htm), or XML (.xml). • Click Clipboard to copy the data to the Windows clipboard, where it can be pasted into another document. • Click Print to print a tab-delimited version of the results.
MON20/20 Software for Gas Chromatographs OCTOBER 2010 User Manual 3-9000-745 2.4.3 Saving a chromatogram trace To save a trace to disk, do the following: 1. From the Chromatogram pull-down menu, select the trace that you want to save. Figure 2-20. Chromatogram pull-down menu 2. Click Save. The Save As window displays. Figure 2-21.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 OCTOBER 2010 3. For convenience the file is given an auto-generated file name that includes the trace’s creation date and time; however, you can give the file any name that you choose. Click Save and the specified trace will be saved. 2.4.
MON20/20 Software for Gas Chromatographs OCTOBER 2010 User Manual 3-9000-745 2.4.5 Forcing a calibration The Forced Cal command uses an archived chromatogram’s raw data to calibrate the GC. The calculation results are stored in the component data table for the corresponding stream number. A major benefit of a forced calibration is increased efficiency.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 OCTOBER 2010 2.4.6 Controlling the display of data in the Timed Events and Components tables MON 20/20 can display two levels of information in the Timed Events and component data tables: • All timed events and all components for all open chromatograms. • Timed events and components for the currently selected chromatogram. By default, the two tables show only the timed events and components for the currently selected chromatogram.
MON20/20 Software for Gas Chromatographs OCTOBER 2010 User Manual 3-9000-745 Figure 2-25. Chromatogram pull-down menu To view all timed events and all components for all open chromatograms, click Cur/All. Figure 2-26. Timed events and component data tables showing data for all currently open traces Note The brackets ([ ]) on the Cur/All button indicate which mode is being displayed in the tables.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 OCTOBER 2010 2.4.7 Saving a comparison file A comparison file allows you to save your current view, including all open chromatograms, for later review and reuse. To save a comparison file, do the following: 1. Click Save Cmp. The Save As dialog appears. 2. Navigate to the folder in which you want to save the file. 3.
MON20/20 Software for Gas Chromatographs OCTOBER 2010 2.5 User Manual 3-9000-745 Miscellaneous commands The series of check boxes to the right of the graph have the following functions: Figure 2-27. Miscellaneous options • Keep last CGM - When viewing a live chromatogram, upon starting a new run, MON 20/20 keeps the most recently completed chromatogram on the graph for comparative purposes.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 OCTOBER 2010 2.5.1 Working with the Timed Events table The Chromatogram Viewer displays a compact version of the Timed Events table, located on the upper right side of the window. The events displayed in the table are sorted by time. See “Managing timed events” on page 4-17 for more information. The Timed Event table displays the following data for each event: Name Description Event Type The type of timed event.
MON20/20 Software for Gas Chromatographs OCTOBER 2010 User Manual 3-9000-745 2.5.2 Editing Timed Events from the Time Events window To launch the Timed Events dialog directly, right-click on the Chromatogram Viewer’s Timed Events table and select Edit Timed Events Table. The Timed Events dialog displays. See “Managing timed events” on page 4-17 for more information. 2.5.3 Editing Timed Events from the Chromatogram Viewer To edit timed events from the Chromatogram Viewer, do the following: 1.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 OCTOBER 2010 4. To insert an event above the currently select event, right-click on the table and select Insert before. To insert an event below the currently select event, right-click on the table and select Insert after. The new row will be added. The options available for configuring the new event depends upon which edit mode you are in—Valve, Integration, or Gain. Note These options are only avialable while in edit mode. 5.
MON20/20 Software for Gas Chromatographs OCTOBER 2010 User Manual 3-9000-745 2.5.4 Using the Chromatogram Viewer’s cursor to update a Timed Event Figure 2-28. Chromatograph cursor cursor coordinates cursor The Chromatogram Viewer has its own cursor that can be displayed by double-clicking within the boundaries of the graph. Once the cursor is displayed, it can be dragged to any point on the graph.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 OCTOBER 2010 To update a timed event based on the location of the Chromatogram Viewer’s cursor, do the following: 1. Select the live or archived trace that you want to use as the source for changing the timed event. 2. Double-click on the graph to display the cursor. The cursor’s coordinates display in the upper left corner of the graph. The xcoordinate represents the analysis time in seconds.
MON20/20 Software for Gas Chromatographs OCTOBER 2010 User Manual 3-9000-745 2.5.5 Working with the Component Data Table The Chromatogram Viewer displays a compact version of the Component Data table beneath the Timed Events table. See “Managing Component Data Tables” on page 4-5 for more information. The Component Data table displays the following data for each component: Name Description Componet The name of the component. Det Identifies the detector associated with the component.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 OCTOBER 2010 2.5.6 Editing retention times from the Chromatogram Viewer To edit the retention time for a component, do the following: 1. Right-click on the Component Data table and select Edit Retention Times. The Ret column turns white, indicating that its cells are editable. 2. Click on the Ret cell for the component that you want edit, and enter a new retention time value, in seconds. The value must be less than the Analysis time. 3.
MON20/20 Software for Gas Chromatographs OCTOBER 2010 User Manual 3-9000-745 2. Click Raw Data. The Raw Data window displays and shows the raw data for the selected chromatogram. Figure 2-30. The Raw Data window The following data displays for each peak from the trace: Name Description No. Numerical identifier for the peak, listed by the order of discovery. Ret Time Time, in seconds, that the component eluted. Peak Area The area under the peak. Peak Height The maximum height of the peak.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 OCTOBER 2010 Name Description Integ. Stop Time, in seconds, when integration stopped. Peak Width Half Height The width of the peak taken at half of the peak’s height. Is Partial Peak If Y, then the Partial Peak value is used in the summation calculation; if N, then the Partial Peak value is not used in the summation calculation. 2.
MON20/20 Software for Gas Chromatographs OCTOBER 2010 User Manual 3-9000-745 Figure 2-31. The View/Set Date Time window 2. Use the drop-down menus to set the date and time. To enable or adjust daylight savings, see “Adjusting daylight savings” on page 2-42. 3. Click OK.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 OCTOBER 2010 2.6.1 Adjusting daylight savings Daylight savings time is the practice of temporarily advancing clocks so that afternoons have more daylight and mornings have less. Typically clocks are adjusted forward one hour near the start of spring and are adjusted backward in autumn. Since the use of daylight savings time is not universal, you have the option of enabling or disabling it in MON 20/ 20.
MON20/20 Software for Gas Chromatographs OCTOBER 2010 User Manual 3-9000-745 Note Make sure the GC is set to the current date and time before enabling the daylight savings feature. 2. Click the Enable Daylight Savings checkbox. The Daylight Savings section will be enabled, giving you the following two options for setting the start and end times for daylight savings: • Week format. You can specify on which week day, of what week, and of what month DST to start and end. • Month/Day format.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 OCTOBER 2010 5. Set the end date for daylight savings time. 6. Set the end time and the setback time. 7. To implement your changes without closing the View/Set Date Time window, click Save. To implement your changes and close the View/ Set Date Time window, click OK.
Section 3: Using the hardware functions Many of a gas chromatograph’s hardware components—such as its heaters, valves, and discrete outputs—can be easily managed through MON 20/20. This chapter shows you how to view and administer each of a gas chromatograph’s major hardware components. This chapter also shows you how to view an inventory of all of a gas chromatograph’s installed hardware components. 3.1 Controlling the temperature of the gas chromatograph’s heaters By selecting Heaters...
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 3.1.1 Renaming a heater To assign an identifying label to a heater, do the following: 1. Select Heaters... from the Hardware menu. The Heaters window displays. Figure 3-1. The Heaters window 2. Double-click on the appropriate row under the Label column for the heater that you want to name. Note The heaters are labelled Heater 1 - Heater N by default, where N equals the total number of heaters available to the GC. 3.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 3.1.2 Setting the heater’s type To set a heater’s type, do the following: 1. Select Heaters... from the Hardware menu. Figure 3-2. The Heaters window 2. Click on the appropriate Heater Type cell and select AC or DC from the drop-down list. 3. To save the changes without closing the window, click Save. To save the changes and close the window, click OK.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 3.1.3 Monitoring the temperature of a heater To check a heater’s temperature, select Heaters... from the Hardware menu. Figure 3-3. The Heaters window The current temperature of each heater displays under the Temperature column, and updates in real time. The percentage of the GC’s power output that is being used by each heater displays under the Current PWM column.
MON20/20 Software for Gas Chromatographs User Manual SEPTEMBER 2010 3-9000-745 3.1.4 Monitoring the operational status of a heater To check a heater’s status, select Heaters... from the Hardware menu. Figure 3-4. The Heaters window The status of each heater displays under the Status column. There are four possible status states, and their meanings are as follows: OK The heater’s control card is installed and is working correctly. Not Installed The heater’s control card is not installed.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 3.1.5 Setting the desired temperature To set the desired temperature for a heater, do the following: 1. Select Heaters... from the Hardware menu. The Heaters window displays. Figure 3-5. The Heaters window 2. For each heater that you want to set, select Auto from the appropriate row under the Switch column. 3.
MON20/20 Software for Gas Chromatographs User Manual SEPTEMBER 2010 3-9000-745 5. The appropriate rows under the PID Gain, PID Integral, and PID Derivative columns can also be edited by double-clicking and entering a new value. The value ranges for each column is as follows: PID Gain 0 - 500 PID Integral 0 - 500 PID Derivative 0 - 50000 6. To save the changes and leave the window open so that you can monitor the heaters’ status, click Save.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 Figure 3-6. The Heaters window 2. For each heater that you want to set, select Fixed On from the appropriate row under the Switch column. 3. For each heater that you want to set, double-click on the appropriate row under the Fixed PWM Output column, and enter the desired percentage of output. You can enter a decimal value between 0 and 100.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 3.1.7 Removing a heater from service To remove a heater from service, do the following: 1. Select Heaters... from the Hardware menu. The Heaters window displays. Figure 3-7. The Heaters window 2. For each heater that you want to set, select Not Used from the appropriate row under the Switch column. The row turns turqoise, indicating that it is no longer in service. 3. To save the changes without closing the window, click Save.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 3.2 SEPTEMBER 2010 Configuring the valves MON 20/20 allows you to do the following from the Valves window: • Assign identifying labels to each valve. • Monitor valve operation. • Control the operation modes for each valve. Note This window contains a hidden column labelled Physical Name. For more information about this column and how to display it, see “Viewing the Physical Name column” on page 1-38. 3.2.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 Figure 3-8. The Valves window with Physical Name column 2. Double-click on the appropriate row under the Label column for the valve that you want to name. Note The valves are labelled Valve 1 - Valve N by default, where N equals the total number of valves available to the GC. 3. Type in a new descriptive name for the valve. 4. Click OK.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 3.2.2 Setting a valve’s operational mode A valve has three operational modes: Auto, On, and Off. • Setting the valve to Off means that the valve will turn off and remain off until the operational mode is changed. • Setting the valve to Auto means that the valve will turn on and off according to the Timed Events table.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 2. Select the desired mode from the drop-down menu under the Switch cloumn for the valve. 3. To save the changes and leave the window open so that you can monitor the valve’s progress, click Save. The current state of the valve displays in the State column, and is updated in real time. 4. To save the changes and close the window, click OK. 3.2.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 3.2.4 Inverting the polarity of a valve The Invert Polariy option reverses the effect of switching a valve on or off. By default, the Invert Polarity option is set to FALSE, which means that switching a valve to ON activates it, and switching the valve to OFF deactivates it. Setting Invert Polarity to TRUE means that switching a valve to ON deactivates it, and switching the valve to OFF activates it.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 3.2.5 Setting the usage mode for a valve A valve’s usage mode determines its general function, or role, during an analysis run. A valve can be assigned one of the following usage modes: • DO • FID H2 Valve • Common Alarm • Stream • Analyzer01 ... • Analyzer016 To set the usage mode for a valve, do the following: 1. Select Valves... from the Hardware menu. The Valves window displays. Figure 3-12.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 2. Select the desired mode from the drop-down menu under the Usage cloumn for the valve. 3. To save the changes and leave the window open so that you can monitor the valve’s progress, click Save. The current state of the valve displays in the State column, and is updated in real time. 4. To save the changes and close the window, click OK.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 3.3 User Manual 3-9000-745 Controlling the detectors Use the Detectors window to monitor the activity and status of the GC’s detectors. To view the Detectors window, select Detectors... from the Hardware menu. Figure 3-13. The Detectors window showing a TCD and an FID Note Before making any modifications to this window, halt the analysis. See “Halting an analysis” on page 6-1 for more information.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 Note Blue cells display read-only data; white cells display editable data. The following data displays for each detector: Name Description Det # Numerical identifier for the detector to which the following data applies. Detector Options, which depend on your GC’s configuration, are TCD, FPD, or FID. FID Temp RTD Applies to FIDs only. Select the appropriate RTD from the drop-down list.
MON20/20 Software for Gas Chromatographs User Manual SEPTEMBER 2010 3-9000-745 Name Description Pre Amplifier Voltage Output at second stage of FID preamp. Read-only. Polarizing Voltage Igniter voltage. Read-only. FID Gain Status Options are: Low and High. Status Options are: Ok, Not Installed and Internal Error. Read-only. 3.3.
User Manual 3-9000-745 MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 Figure 3-14. The Detectors window 2. Select the appropriate detector. It may be necessary to return to the Chromatogram Viewer to learn which detector is the source of the trace that needs to be offset. 3. To lower the baseline, click Left(N). Each time this button is clicked, N is incremented by -1.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 4. To raise the baseline, click Right(N). Each time this button is clicked, N is incremented by 1. For example, is this is the first time the button has been clicked, Right(0) will be increment to Right(1) and the baseline will be raised one step. If Left(N) was clicked previously, then that button will be incremented by 1 first, until it reaches Left(0); at the point, Right(N) will be incremented by 1.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 3.3.3 Resetting the preamp value To reset the Preamp Val field on the Detectors window to 0, click AutoZero. 3.3.4 Balancing the preamp In some situations that involve TCD detectors the baseline may be displayed either too high on the graph, in which case the tops of the peaks are cut off, or too low on the graph, so that the bases of the peaks are cut off.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 Figure 3-15. The Detectors window 2. Select the appropriate detector. It may be necessary to return to the Chromatogram Viewer to learn which detector is the source of the trace that needs to be offset. 3. To lower the baseline, click Left(N). Each time this button is clicked, N is incremented by -1.
User Manual 3-9000-745 MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 4. To raise the baseline, click Right(N). Each time this button is clicked, N is incremented by 1. For example, is this is the first time the button has been clicked, Right(0) will be increment to Right(1) and the baseline will be raised one step. If Left(N) was clicked previously, then that button will be incremented by 1 first, until it reaches Left(0); at the point, Right(N) will be incremented by 1.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 Figure 3-16. The Discrete Inputs window 2. Double-click on the appropriate row under the Label column for the discrete input that you want to rename. Note The discrete inputs are labelled Discrete Input 1 - Discrete Input N by default, where N equals the total number of discrete inputs available to the GC. 3. Type in a new descriptive name for the discrete input. 4. Click OK. 3.4.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 Note The GC’s switch panel settings override MON 20/20’s settings. To set a discrete input’s operational mode, do the following: 1. Select Discrete Input... from the Hardware menu. The Discrete Input window displays. Figure 3-17. The Discrete Inputs window 2. Select the desired mode from the drop-down menu under the Switch cloumn for the discrete input. 3.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 3.4.3 Monitoring the operational status of a discrete input To check a valve’s status, select Discrete Input... from the Hardware menu. Figure 3-18. The Discrete Inputs window The status of each discrete input displays under the Status column. There are three possible status readings, and their meanings are as follows: OK The discrete input is installed and is working correctly.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 3.4.4 Inverting the polarity of a discrete input The Invert Polariy option reverses the way a voltage signal is interpreted by the discrete input. By default, the Invert Polarity option is set to Normally Open, which means that a low voltage signal is interpreted by the discrete input as ON, and a high voltage signal is interpreted by the discrete input as OFF.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 3.5 User Manual 3-9000-745 Managing your gas chromatograph’s discrete outputs You can use MON 20/20 to assign labels to the GC’s discrete outputs and to control the discrete outputs’ operational modes. The number of discrete outputs available depends on the GC. Note This window contains a hidden column labelled Physical Name. For more information about this column and how to display it, see “Viewing the Physical Name column” on page 1-38. 3.5.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 Note The discrete outputs are labeled Discrete Output 1 - Discrete Output N by default, where N equals the total number of discrete outputs available to the GC. 3. Type in a new descriptive name for the discrete output. 4. Click OK. 3.5.2 Setting a discrete output’s operational mode A discrete output has three operational modes: Auto, On, and Off.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 Figure 3-21. The Discrete Outputs window 2. Select the desired mode from the drop-down menu under the Switch cloumn for the discrete output. 3. To save the changes and leave the window open so that you can monitor the discrete output’s progress, click Save. To save the changes and close the window, click OK. The current state of the discrete output displays in the State column, and is updated in real time.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 3.5.3 Monitoring the operational status of a discrete output To check a valve’s status, select Discrete Output... from the Hardware menu. Figure 3-22. The Discrete Outputs window The status of each discrete output displays under the Status column. There are three possible status readings, and their meanings are as follows: 3-32 OK The discrete output is installed and is working correctly.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 3.5.4 Setting the usage mode for a discrete output A discrete output’s usage mode determines which signals are routed to it via the Limited Alarm and Discrete Alarm functions. A discrete output can be assigned one of the following usage modes: • DO • FID H2 Valve • Common Alarm • Stream • Analyzer01 ... • Analyzer016 To set the usage mode for a discrete output, do the following: 1. Select Discrete Output...
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 2. Select the desired mode from the drop-down menu under the Usage cloumn for the discrete output. 3. If you select DO for Usage, and Auto for Switch, then you must also set the Start Time and Duration. Double-click on the appropriate row under the Start Time column and enter the time that the digital output should be turned on.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 3.6.1 Renaming an analog input Give each analog input a descriptive label to avoid confusing one unit for another. To assign an identifying label, do the following: 1. Select Analog Inputs... from the Hardware menu. The Analog Inputs window displays. Figure 3-24. The Analog Inputs window 2. Double-click on the appropriate row under the Label column for the analog input that you want to rename.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 3.6.2 Setting a analog input’s operational mode An analog input has two operational modes: Variable and Fixed. • Setting the switch to Variable means that the analog input will be set automatically, based on the signal it receives. • Setting the switch to Fixed means that the analog input will be set to the value that you enter in the appropriate row under the Fixed Value column.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 3.6.3 Setting the scale values for an analog input device To set the zero scale and full scale, which are used when converting the analog input value, do the following: 1. Select Analog Input... from the Hardware menu. The Analog Input window displays. Figure 3-26. The Analog Inputs window 2. Double-click on appropriate row under the Zero Scale column and enter a zero scale value. 3.
User Manual 3-9000-745 MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 Figure 3-27. The Analog Inputs window 2. Select the signal type from the appropriate row under the mA/Volt column. 3. To save the changes and leave the window open so that you can monitor the analog input’s progress, click Save. To save the changes and close the window, click OK. The type of signal being generated displays in the mA/Volts column, and is updated in real time.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 3.6.5 Monitoring the status of an analog input To check an analog input’s status, select Analog Input... from the Hardware menu. Figure 3-28. The Analog Inputs window The operational status of each analog input displays under the Status column. There are three possible status readings, and their meanings are as follows: OK The analog input is installed and is working correctly.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 3.6.6 Calibrating an analog input To calibrate an analog input, do the following: 1. Select Analog Input... from the Hardware menu. The Analog Input window displays. Figure 3-29. The Analog Inputs window 2. Click on the analog input that you want to calibrate. 3. Set the analog input’s Zero Scale by entering its minimum anticipated value. 4. Set the analog input’s Full Scale by entering your maximum anticipated value. 5.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 Figure 3-30. The Analog Input Calibration Wizard 6. Click Next. Step 2 of the Analog Input Calibration Wizard displays. Figure 3-31. Step 2 of the Analog Input Calibration WIZARD 7. Click Next. Step 3 of the Analog Input Calibration Wizard displays.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 Figure 3-32. Step 3 of the Analog Input Calibration Wizard 8. Click Next. Step 4 of the Analog Input Calibration Wizard displays. Figure 3-33. Step 4 of the Analog Input Calibration Wizard 9. Click Finish. The calibration is complete.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 3.7 User Manual 3-9000-745 Managing your gas chromatograph’s analog outputs With MON 20/20 you can control them in the following ways: • Assign identifying labels. • Assign scale ranges. • Calibrate analog outputs for zero and full scale values. Note This window contains a hidden column labelled Physical Name. For more information about this column and how to display it, see “Viewing the Physical Name column” on page 1-38. 3.7.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 Note The analog output devices are labelled Analog Output 1 - Analog Output N by default, where N equals the total number of analog outputs available to the GC. 12. Type in a new descriptive name for the analog output. 13. Click OK. 3.7.2 Setting a analog output’s operational mode An analog output has two operational modes: Variable and Fixed.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 2. Select the desired mode from the drop-down menu under the Switch cloumn for the analog output. 3. To save the changes and leave the window open so that you can monitor the analog output, click Save. To save the changes and close the window, click OK. The current value of the analog output displays in the Cur Val column, and is updated in real time. 3.7.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 3.7.4 Mapping a system variable to an analog output To select the system variable on which to base the signal level of the analog output, do the following: 1. Select Analog Output... from the Hardware menu. The Analog Output window displays. Figure 3-37. The Analog Outputs window 2. Select a new variable by clicking on the appropriate drop-down list under the Variable column.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 Figure 3-38. The Analog Outputs window with Variable drop-down menu 3. To save the changes and leave the window open so that you can monitor the analog output’s progress, click Save. To save the changes and close the window, click OK. 3.7.5 Monitoring the status of an analog output To check an analog output device’s status, select Analog Output... from the Hardware menu. Figure 3-39.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 The operational status of each analog output displays under the Status column. There are three possible status readings, and their meanings are as follows: OK The analog output device is installed and is working correctly. Not Installed The analog output device is not installed. Error The Heater/Solenoid board is installed but the GC cannot communicate with it.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 3. Click AutoCal...(F4) or press F4. The Analog Output Calibration Wizard runs. Figure 3-41. The Analog Output Calibration Wizard 4. Select the check box for the unit of measure you want to use for the calibration and then click Next. Step 2 of the Analog Output Calibration Wizard displays. Figure 3-42.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 5. Enter the Zero Scale Adjustment value and then click Next. If the value entered is within tolerance, it is accepted and Step 3 of the Analog Output Calibration Wizard displays. If the value is not within tolerance, an error icon ( ) appears beside the field. Tolerance is set to ±1mA of the analog output’s default zero adjustment setting, which is 4mA. Enter a different value and try again. Figure 3-43.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 Figure 3-44. Step 4 of the Analog Output Calibration Wizard 7. Click Finish. The calibration is complete.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 3.8 SEPTEMBER 2010 Reviewing the Hardware Inventory List MON 20/20 can compile an inventory table of all hardware that is installed on the GC. To view this table, select Installed Hardware... from the Hardware menu. Figure 3-45.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 The type of hardware installed is listed under the Device Description column. The other types of information available on this screen are the following: • IO Function - Describes the function of the device. • Slot Number - Describes the location of the device on the GC.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 This page is intentionally left blank.
Section 4: Using the Application functions Many of the variables that a gas chromatograph uses during an analysis run—such as timed events, stream sequence, and calculation types—can be easily managed through MON 20/20. This chapter explains how to do the following: • View and edit general information about the GC to which MON 20/20 is connected, such as name, model, and default stream sequence. • View and edit component data, validation data, and timed event tables.
User Manual 3-9000-745 4.1 MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 Managing the system Use this function to select the default GC stream sequence and to set or edit system-wide variables such as the GC’s name, serial number, and system description. See Table 5-1 for a list of the items that are available on the System window, along with their related functions. To view the System window, select System... from the Application menu. Figure 4-1.
MON20/20 Software for Gas Chromatographs User Manual SEPTEMBER 2010 3-9000-745 Table 4-1. List of fields from System window Field Name Description Analyzer Name Defines the GC name that appears in the Status Bar on the main window when MON 20/20 is connected to the GC. Can contain up to 12 characters. GC Model The model number of the GC to which MON 20/20 is connected. System Description A field to record miscellaneous reference information to further identify the currently connected system.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 Table 4-1. List of fields from System window Field Name Description Time Format Defines how the time will be displayed. The options are: • HH:MM:SS • HH:MM Time Notation Defines the cycle of time to use when displaying the time. The options are: • 12 Hr • 24 Hr CGM FCAL Archive Sets the storage behavior for final calibration chromatograms.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 4.2 User Manual 3-9000-745 Managing Component Data Tables MON 20/20 allows you to view and edit the component data tables. The number of available component data tables depends on the GC unit configuration. To assign a component data table to a stream, see “Assigning a valve to a stream and setting the relationship between the stream’s open state to the valve’s On/Off state” on page 4-76. 1. To view a component data table, select Component Data...
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 Figure 4-3. The selected component data table Note To see a different table, select it from the Choose table drown-down list. Note To sort the list of components by detector, and then by retention time, click Sort RT. 4.2.1 Editing a Component Data Table Note Table cells with a white background are editable; table cells with a turqoise background are not editable. To edit a cell, do the following: 1. Click on the cell.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 The following table lists all of the editable parameters that are available on the component data table. The standard values for these parameters were taken from the second editions of the Orifice Metering of Natural Gas and Other Related Hydrocarbon Fluids and the Compressibility Factors of Natural Gas and Other Related Hydrocarbon Gases.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 Parameter Description Anly Meth Used to determine the component’s raw data value. Options are: • Area - Raw data value is proportional to the area under the peak. • Height - Raw data value is proportional to the height of the peak. • Fixed - Raw data value is proportional to a value that is set by the user. • Analog Input - Data signal comes from an external analyzer.
MON20/20 Software for Gas Chromatographs User Manual SEPTEMBER 2010 3-9000-745 Parameter Description AGA 8 Component The name of the component according to the American Gas Association, which is used in the AGA 8 compressibility calculation. Ref Comp The component not found in the calibration gas but in the sample gas for indirect calibration. If 'none', normal (direct) calibration is used. Not editable unless the calibration type is set to Relative.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 4.2.2 Adding a component to a Component Data Table To add a component to a component data table, do the following: 1. Select Component Data... from the Application menu. The Component Data Tables window appears, displaying a list of available component data tables. Figure 4-4. The Component Data Tables window Note Other ways of accessing the component data tables are by pressing F6 or by clicking from the Toolbar. 2.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 Figure 4-5. The selected component data table Note To sort the list of components by detector, and then by retention time, click Sort RT. 3. If you want to add the component above the currently selected component, click Insert before. If you want to add the component below the currently selected component, select Insert after from the Insert arrow. 4. To save the changes without closing the window, click Save.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 4.2.3 Removing a component from a Component Data Table To remove a component from a component data table, do the following: 1. Select Component Data... from the Application menu. The Component Data Tables window appears, displaying a list of available component data tables. Figure 4-6. The Component Data Tables window Note Other ways of accessing the component data tables are by pressing F6 or by clicking from the Toolbar.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 Figure 4-7. The selected component data table Note To sort the list of components by detector, and then by retention time, click Sort RT. 3. Select the component that you want to remove. 4. Click Delete. 5. To save the changes without closing the window, click Save. To save the changes and close the window, click OK. 4.2.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 Figure 4-8. The Component Data Tables window Note Other ways of accessing the component data tables are by pressing F6 or by clicking from the Toolbar. 2. Select the table that you want to view. The selected component data table displays. Figure 4-9.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 Note To sort the list of components by detector, and then by retention time, click Sort RT. 3. Click Std Values (F3). The Standard Component Values window displays. Figure 4-10. The Standard Component Values window 4. Click Close. 4.2.5 Viewing raw data To view the raw data for the displayed component data table, do the following: 1. Click Raw Data (F4).
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 Figure 4-11. The Select dialog 2. Double-click the desired stream. The Raw Data window appears, listing the peak raw data from the last run of the stream represented by the component data table. Figure 4-12.
MON20/20 Software for Gas Chromatographs User Manual SEPTEMBER 2010 3-9000-745 The following data displays for each peak: Name Description Peak No Numerical identifier for the peak, listed by the order of discovery. Ret Time Time, in seconds, that the component eluted. Peak Area The area under the peak. Peak Height The maximum height of the peak. Det # The detector associated with the peak. Method Method of peak end detection.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 To assign a timed events table to a stream, see “Assigning a valve to a stream and setting the relationship between the stream’s open state to the valve’s On/Off state” on page 4-76. 1. Select Timed Events... from the Application menu. The Timed Events Tables selector window appears, displaying a list of available timed events tables. Figure 4-13.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 Figure 4-14. The Timed Events window Note To sort events by time, click the appropriate Sort button. 3. To see a different timed events table, select it from the Choose table drop-down list.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 4.3.1 Editing valve events Valve-related events are grouped on the upper left side of the Timed Events window. To edit valve-related events, do the following: 1. Select Timed Events... from the Application menu. The Timed Events Tables selector window appears, displaying a list of available timed events tables. Figure 4-15.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 Figure 4-16. The Timed Events window Note To sort events by time, click the appropriate Sort button. 3. Click on the cell that you want to edit. Depending on the cell type, you will either be required to select a value from a drop-down list, or you will be able to type in the value directly. 4. To save the changes without closing the window, click Save. To save the changes and close the window, click OK.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 The following table describes the valve-related parameters that are available on the timed events window. Parameter Description TEV Type The type of event. You have the following choices: • Valve # - A valve. • DO # - A discrete output. • Strm Sw - Switches to the next stream in the sequence. • FID Gain - Sets the FID to high or low gain. • FID Auto Zero - Zeros the FID preamp after a gain change.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 Figure 4-17. The Timed Events Tables selector window Note Other ways of accessing the timed event tables are by pressing F5 or by clicking from the Toolbar. Note If only one timed events table is available, it will display immediately, bypassing the Timed Events Tables selector window. 2. Select the table that you want to view. The selected timed events table displays.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 Figure 4-18. The Timed Events window Note To sort events by time, click the appropriate Sort button. 3. Double-click on the cell that you want to edit. Depending on the cell type, you will either be required to select a value from a drop-down list, or you will be able to type in the value directly. 4. To save the changes without closing the window, click Save. To save the changes and close the window, click OK.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 The following table describes the integration-related parameters that are available on the timed events window. Parameter Description TEV Type The type of integration event. You have the following options: • Inhibit: Set to Off to start a peak; set to On to end a peak. • Integrate: Set to On and Off to set a region in which the area under the trace is computed as a peak regardless of peak onset discovery.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 Parameter Description Value The values available depend on the integration type selected from the TEV Type column. • Slope Sensitivity and Peak Width: Enter the number of points, between 1 and 99, to be used. • Single Baseline: Select Off, End, Bgn. • SW Auto Zero: No options. • All other integration types: Select On or Off. Det # The ID number of the detector that will be affected by the event. Valid values are 1 and 2.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 Figure 4-19. The Timed Events Tables selector window Note Other ways of accessing the timed event tables are by pressing F5 or by clicking from the Toolbar. Note If only one timed events table is available, it will display immediately, bypassing the Timed Events Tables selector window. 2. Select the table that you want to view. The selected timed events table displays.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 Figure 4-20. The Timed Events window Note To sort events by time, click the appropriate Sort button. 3. Click on the cell that you want to edit. Depending on the cell type, you will either be required to select a value from a drop-down list, or you will be able to type in the value directly. 4. To save the changes without closing the window, click Save. To save the changes and close the window, click OK.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 The following table describes the spectrum gain-related parameters that are available on the timed events window. Parameter Description Det # The ID number of the detector that will be affected by the event. Select 1 or 2. Gain Enter a value between 0 and 64. This is the exponent value in the following expression: 2gain value.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 Note If only one timed events table is available, it will display immediately, bypassing the Timed Evetns Tables selector window. 2. Select the table that you want to view. The selected timed events table displays. The Analysis Time section is located on the lower right side of the Timed Events window. Figure 4-22. The Timed Events window Note To sort events by time, click the appropriate Sort button.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 3. Click on the Analysis Time cell and enter a value, in seconds, between 0 and 3600. 4. Click on the Cycle Time cell and enter a value, in seconds, between 0 and 3620. Note The Cycle Time must be atleast 10 seconds greater than the Analysis Time. 5. To save the changes without closing the window, click Save. To save the changes and close the window, click OK. 4.3.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 Note If only one timed events table is available, it will display immediately, bypassing the Timed Events Tables selector window. 2. Select the table that you want to view. The selected timed events table displays. Figure 4-24. The Timed Events window Note To sort events by time, click the appropriate Sort button. 3. Select the event that you want to delete. 4. Click the appropriate Delete button.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 4.3.6 Adding an event to the Timed Event Table To add an event to one of the Valve Events, Integrate Events, or Spectrum Gain Events tables on the Timed Events window, do the following: 1. Select Timed Events... from the Application menu. The Timed Events Tables selector window appears, displaying a list of available timed events tables. Figure 4-25.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 Figure 4-26. The Timed Events window Note To sort events by time, click the appropriate Sort button. 3. If you want to add the event above the currently selected event, click the appropriate Insert before button. If you want to add the event below the currently selected event, select Insert after from the Insert arrow and then click the button.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 The new event will be added to the table. 4. Select a Type, Valve/DO#, and State for the event, if necessary, and enter a new Time for the event also. 5. To save the changes without closing the window, click Save. To save the changes and close the window, click OK. 4.4 Managing Validation Data Tables Use the validation data table to hold information about the composition of the gas that is used in the validation run.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 2. If the appropriate table is not displayed, select it from the Choose Table drop-down list. 3. Select a new variable by clicking on the appropriate drop-down list under the Variable column. For a demonstration of how to use the context-sensitive variable selector, see “Using the context-sensitive variable selector” on page 1-42. 4.
MON20/20 Software for Gas Chromatographs User Manual SEPTEMBER 2010 4.5 3-9000-745 Managing calculations MON 20/20’s Calculations submenu allows you to activate and define how the output of standard or user-defined chromatograph analysis data is used in various calculations. You can configure the following types of calculations: • Control - Allows you to designate, by streams, the standard calculations that should be performed from the analysis data.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 Figure 4-29. The Control Calculations window 2. Select a check box for a given stream to turn the calculation ON for that stream; click to clear the check box for a given stream to turn the calculation OFF for that stream. You can use the arrow keys to move from one stream cell to another, and you can press the space bar to toggle the calculation on or off. 3. To save the changes without closing the window, click Save.
MON20/20 Software for Gas Chromatographs User Manual SEPTEMBER 2010 3-9000-745 Note To copy the information on this screen to the clipboard so that it can be pasted into another application such Microsoft Word or Excel, right-click on the table and select Copy to clipboard from the right-click menu. Note To print the information on this screen, right-click on the table and select Print Sheet from the right-click menu. 4.5.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 Figure 4-30. The Averages Calculations window 2. Select a new variable by clicking on the appropriate drop-down list under the Variable column. For a demonstration of how to use the context-sensitive variable selector, see “Using the context-sensitive variable selector” on page 1-42. 3. Select the type of average to be calculated from the Average Type dropdown list.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 • Weekly - Averages will start and stop once a week at the time displayed in the Reset Time field and on the day entered in the Weekday field, from the Averages Reset section. • Monthly - Averages will start and stop once a month at the time displayed in the Reset Time field and on the day of the month entered in the Day field, from the Averages Reset section.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 4.5.3 Viewing an archive of averages for a given variable To view an archive of averages for a given variable, do the following: 1. Select Applications → Calculations → Averages.... The Averages Calculations window appears. Figure 4-31. The Averages Calculations window 2. Click on the desired variable to view its history. 3. Click Archive. The archive data screen appears.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 Figure 4-32. The archive data window Note To copy the information in this table to the clipboard so that it can be pasted into another application such as Microsoft Word or Excel, select the cells that you want to copy and then press CTRL + C to copy the information to the clipboard.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 4.5.4 Copying stream settings To copy the stream settings from a highlighted row and apply them to the next row, do the following: 1. Select Applications → Calculations → Averages.... The Averages Calculations window appears. Figure 4-33. The Averages Calculations window 2. Select the row that you want to copy. 3. Click S + Copy.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 If there are no streams available, instead of copying the stream, MON 20/20 will display the following message: Figure 4-34. No streams available warning 4. To save the changes without closing the window, click Save. To save the changes and close the window, click OK. 4.5.5 Copying component settings To copy the component settings from a highlighted row and apply them to the next row, do the following: 1.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 Figure 4-35. The Averages Calculations window 2. Select the row that contains the component that you want to copy. 3. Click the arrow beside the S + Copy button to switch it to C + Copy. 4. Click C + Copy. The component will be copied to the next row and incremented to the next available component—for example, from Ammonia to Benzene. Note You can select and copy more than one component at a time.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 If there are no components available, instead of copying the component, MON 20/20 will display the following message: Figure 4-36. No components available warning 5. To save the changes without closing the window, click Save. To save the changes and close the window, click OK. 4.6 Creating Custom Calculations To create or edit a customized calculation using GC analysis data, do the following: 1.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 Figure 4-37. The User Defined Calculations window 2. Click Insert before to add a row to the User Defined Calculations table. Note To delete this--or any--row from the table, click Delete. 3. Double-click the Label cell and enter a name for the calculation you are about to create. If you want to enter a short description for the new calculation, double-click the Comment cell and enter it there. 4. Click Edit.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 Figure 4-38. The Edit User-defined Calculation window A B C E D F In MON 20/20, building a calculation is similar to building a simple program. You have constants and two types of variables available, as well as two calculation-building commands. You can also add comments that will be ignored by the application but that can help you explain the logic and structure of the calculation you are designing.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 The following is a description of the design elements of the Edit Userdefined Calculation window: • Element A - Called the Calculation Steps Viewer, this element displays the line-by-line construction of the calculation as it is being built. The following commands allow you to interact with this area: • Click Clear All to clear the content of the Calculation Steps Viewer.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 Figure 4-39. Expression Editor - No Action Figure 4-40. Expression Editor - Insert Comment Figure 4-41. Expression Editor - Insert Condition Figure 4-42.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 The following commands allow you to interact with the Expression Editor: • Click Clear to clear the content of the entire line. The line itself is not deleted. • Click Delete Item to delete the currently active token. Each mathematical function, numeric data, and mathematical operation is treated as a token. The token to the right of the current cursor location is treated as the currently active token.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 • Section 1 - This section contains the following keys: x^y x to the power of y SQRT Square Root abs Absolute Value sin Sine cos Cosine tan Tan log10 Logarithm to the base 10 log2 Logarithm to the base 2 ln Logarithm to the base e and Logical AND or Logical OR xor Logical XOR ( Open bracket ) Close bracket • Section 2 - This section contains the traditional calculator keys and can be used with your k
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 • Constants - Allows you to select constants from a drop-down list. • Temporary Variables - Allows you to select temporary, usercreated variables from a drop-down list. • System Variables - Allows you to select system variables. • Edit Temporary Variables - Allows you to create variables. • Edit Constants - Allows you to create system-wide constants that can be used in user-defined calculations.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 Figure 4-44. Edit Comment mode 2. Enter the comment into the Edit Comment textbox and then click Done. The comment will be added to the Calculation Steps Viewer. Figure 4-45.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 4.6.2 Inserting a Conditional Statement Figure 4-46. An example of a conditional statement The Expression Editor in Edit Condition mode allows you to build the first line of the conditional statement: Figure 4-47.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 Expressions are built using the Expression Editor in Edit Expression mode. To add a conditional statement, do the following: 1. Click on the Insert drop-down list and select Insert Condition. A new line is added to the Calculation Steps Viewer and the Expression Editor switches to Edit Condition mode. 2. Add an expression.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 Figure 4-49. the Variable/Constant drop-down list For information on creating variables, see page 4-63. For information on creating constants, see page 4-61. 5. Click Done. MON 20/20 validates the statement and if there are no errors, it adds it to the Calculation Steps Viewer. Figure 4-50.
MON20/20 Software for Gas Chromatographs User Manual SEPTEMBER 2010 3-9000-745 4.6.3 Inserting an Expression A mathematical expression has the following structure: Variable = Regular expression Figure 4-51. Edit Expression area regular expression variable To add an expression to a conditional statement or calculation, do the following: 1. Click on the Insert drop-down list and select Insert Expression.
User Manual 3-9000-745 MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 Figure 4-52. The Final Result tree view 3. Add a regular expression. You can use constants, temporary variables, system variables, and the calculator functions to build the expression. For information on inserting system variables, see page 464. For information on creating variables, see page 4-63. For information on creating constants, see page 4-61. Figure 4-53. The Edit Expression area 4. Click Done.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 Figure 4-54. The Calculation Steps Viewer 4.6.4 Creating a Constant To create a constant that you can use in building a calculation, do the following: 1. From the Edit User-defined Calculation window, click Edit Constants. The Edit Constants window displays, showing all the constants that have been created so far for the GC.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 Figure 4-55. The Edit Constants window 2. To create a new constant, click Insert before. A new row will be added to the USER_CALC_CONSTANTS table. Note To delete a constant, select it in the table and click Delete. 3. Double-click the Label cell and enter a name for the constant. Note To edit any cell, double-click it. 4. Double-click the Value cell and enter a value for the constant. 5.
MON20/20 Software for Gas Chromatographs User Manual SEPTEMBER 2010 3-9000-745 4.6.5 Creating or Editing a Temporary Variable To create a temporary variable that you can use in building a calculation, do the following: 1. From the Edit User-defined Calculation window, click Edit Temporary Variables. The Edit Temporary Variables window displays, showing all the temporary variables that have been created so far for the user-defined calculation. Figure 4-56. The Edit Temporary Variables window 2.
User Manual 3-9000-745 MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 4.6.6 Inserting a System Variable To insert a system variable into the Expression Editor, do the following: From the Edit User-defined Calculation window, click on the System Variables drop-down arrow. For a demonstration of how to use the context-sensitive variable selector, see “Using the context-sensitive variable selector” on page 1-42.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 4.6.7 Using User-defined Calculations You can use a previously-created user-defined calculation when building new calculations by clicking on the System Variables drop-down arrow on the Edit User-defined Calculation window. Figure 4-58. System Variables drop-down menu For a demonstration of how to use the context-sensitive variable selector, see “Using the context-sensitive variable selector” on page 1-42.
User Manual 3-9000-745 Figure 4-59.
MON20/20 Software for Gas Chromatographs User Manual SEPTEMBER 2010 4.7 3-9000-745 Setting the calculation method MON 20/20 can be configured to perform GPA calculations, ISO calculations, or both. To set which type of calculation method MON 20/20 should use, do the following: 1. Select Applications → Calculations → Configuration.... The Calculations Configuration window displays. Figure 4-60. The Calculations Configuration window 2. Select the method from the Calculation Method drop-down list.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 3. Select a unit of measure from the Base Pressure Units drop-down list. The options are: • PSI • Bar • kPa 4. If you set the calculation method to GPA or GPA & ISO, you can also set the following options: • GPA Calculator Units (U.S. or S.I.) • GPA Pressure Display (PSI, Bar or kPa) 5.
MON20/20 Software for Gas Chromatographs User Manual SEPTEMBER 2010 3-9000-745 • Primary CV Units - kilojoules per cubic meter (kJ/m3) - kilocalories per cubic meter (kCal/m3) - kilowatt hours per cubic meter (kWhrs/m3) - megajoule per cubic meter (MJ/m3) - megajoule per kilogram (MJ/kg) - megajoule per mole (MJ/mole) • Secondary CV Units (same options as Primary CV Units) 6. Click Save to accept the changes without closing the window, or click OK to accept the change and close the window. 4.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 Figure 4-61. The Limit Alarms window 2. Select a new variable by clicking on the appropriate drop-down list under the Variable column. For a demonstration of how to use the context-sensitive variable selector, see “Using the context-sensitive variable selector” on page 1-42. 3. To change the alarm type, click the appropriate cell under the Type column. You have the following the options: • Off - Turns off the alarm.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 4. If you want a discrete output to activate when the alarm triggers, click on the appropriate cell under the DO # to Set column and select it from the drop-down list. 5. To prevent or allow averaging when the alarm triggers, double-click on the appropriate cell under the Inhibit Avg column, and select one of the following options: • True - Inhibits averaging when the alarm is active.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 9. Click C + Copy. The component will be copied and incremented to the next available component--for example, from Ammonia to Benzene. If there are no more components available, instead of copying the component, MON 20/20 will display the following message: Figure 4-63. No components available warning 10. If you want the GC to halt after the current analysis when an alarm is triggered, do the following: (a.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 Figure 4-64. The Limit Alarms window 2. If you want a discrete output to activate when the alarm triggers, click on the appropriate cell under the DO # to Set column and select it from the drop-down list. 3. To prevent or allow averaging when the alarm triggers, double-click on the appropriate cell under the Inhibit Avg column, and select one of the following options: • True - Inhibits averaging when the alarm is active.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 5. If you want the GC to halt after the current analysis when an alarm is triggered, do the following: (a.) Select the Halt on Alarm? checkbox. (b.) Enter a value in the Delay column for the length of time, in seconds, that the alarm condition should exist before the Halt command is executed. You can enter a value between 0 and 1800. 6. To save the changes without closing the window, click Save.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 Figure 4-65. The Streams window 2. For the appropriate stream, select one of the following options from the Usage column: • • • • Unused - Not used Cal - Calibration Analy - Analysis Validate - Validation 3. If you select Cal or Validation, you can also edit the following parameters: • Auto - If checked, the calibration or validation will be automatic.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 • Auto Baseline - Enable or disable the automatic baseline run. The GC performs an additional calibration run (before the calibration runs to be averaged) without the calibration gas. This run evaluates the peaks caused by the GC valve action alone; any peak areas found are subtracted from the subsequent analyses. Note Disabling the Auto Baseline setting will delete existing CDT baseline data for the associated stream. 4.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 Details about the valves in the drop-down list can be viewed from the Valves window. 3. To save the changes without closing the window, click Save. To save the changes and close the window, click OK. 4. To have the stream open when the valve is ON, select the corresponding Stream Valve On to Select checkbox; to have the stream open when the valve is OFF, clear the corresponding Stream Valve On to Select checkbox. 4.10.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 4. To save the changes without closing the window, click Save. To save the changes and close the window, click OK. 4.10.4 Changing the base pressure for a stream To change the base pressure for a stream, do the following: 1. Select Streams... from the Application menu. The Streams window opens. Figure 4-68. The Streams window 2.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 Figure 4-69. The Stream Sequence window 2. Each stream sequence table can contain up to three sequences--a primary, or default, sequence, and two auxiliary sequences. The table for detector that is designated as “1” displays by default. To display a different table, select it from the Choose table: drop-down list. 3. To create a new stream sequence, click Insert before. Note There can only be three sequences per detector.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 Note No two sequences can be activated by the same discrete input. 7. Select the type of analysis the detector should perform when following the sequence. There are two options: • Analysis - The detector performs a real analysis of the streams. • Validation - The detector performs a test analysis to verify that it is working properly. 8. To save the changes without closing the window, click Save.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 RTS Off Dly User Manual 3-9000-745 The delay in milliseconds between RTS termination and the end of data transmission. Range: 0 to 1000 RTS ON Sly The delay in milliseconds between RTS activation and the start of data transmission Range: 0 to 1000 Port Resp Dly The delay in milliseconds the communication port will wait before sending a response back to device. Range: 0 to 100 Port Avail Allows you to enable or disable the communication port.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 To map GC data to Modbus registers, do the following: 1. Select Communication... from the Application menu. The Communication window appears. Figure 4-70. The Communication window 2. Click Registers. The Modbus Map Editor window appears. Figure 4-71.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 3. To view or edit registers that are contained in an existing MAP file, click on the Select MAP File drop-down list and select the appropriate file. The registers will load into the table. Figure 4-72.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 4. To edit a cell, double-click it. You can edit the following parameters: Name Description Register Number Displays the number for the Modbus register that will be polled by a connected data acquisition system. Data Type Describes the type of data that is stored in the register. SIM_2251 and User_Modbus options are: • BOOLEAN • INT • LONG • FLOAT • Bitmap(INT) • Bitmap(LONG) • SCALED_FP1 ...
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 9. To check for conflicting register assignments, click Check. MON 20/ 20 will check the table and if it encounters a conflict it will display the following message: Figure 4-73. Conflicting registers warning Review the table to locate the conflicting registers and change one. 10. To save the MAP file, do the following: (a.) Click Export.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 (b.) Enter a new name for the file or select the file that you want to overwrite. (c.) Click Save. 4.12.2 Creating a new map file To create a new MAP file, do the following: 1. Select Communication... from the Application menu. The Communication window appears. Figure 4-75. The Communication window 2. Click Registers. The Modbus Map Editor window appears.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 Figure 4-76. The Modbus Map Editor window 3. Click New. A new row will be added to the table and the column headings will be empty.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 Figure 4-77. The Modbus Map Editor window 4. From the Register Type drop-down list, select the type of PLC emulation protocol you want to use. You have two options: User_Modbus, which is a PLC emulation Modbus protocol that can use scaling to convert floating point numbers to integers, and SIM_2251, which emulates the Daniel 2500 communication protocol and is a simulation of the 2251 GC controller.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 Figure 4-78. The Open window (b.) Select the file that you want to import and click Open. The registers from the selected file will load into the table. Figure 4-79.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 6. To edit a cell, double-click it. You can edit the following parameters: Name Description Register Number Displays the number for the Modbus register that will be polled by a connected data acquisition system. Data Type Describes the type of data that is stored in the register. SIM_2251 registers use only one data type: FLOAT.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 11. To check for conflicting register assignments, click Check. MON 20/ 20 will check the table and if it encounters a conflict it will display the following message: Figure 4-80. Conflicting registers warning Review the table to locate the conflicting registers and change one. 12. To save the MAP file, do the following: (a.) Click Export.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 (b.) Enter a new name for the file or select the file that you want to overwirte. (c.) Click Save. 4.12.3 Assigning a variable to a register To assign a variable to a register, from the Modbus Map Editor window, double-click the appropriate Variable(s) cell and select a new variable. For a demonstration of how to use the context-sensitive variable selector, see “Using the context-sensitive variable selector” on page 1-42.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 Figure 4-82.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 For example: RF = 65535 RZ = 0 SF = 100 (from SCALED_FP1) SZ = 0 (from SCALED_FP1) Dfp = 97.13 (scaled percent for methane) 65535 – 0 63654 = ⎛ ------------------------⎞ ( 97.13 – 0 ) + 0 ⎝ 100 – 0 ⎠ To edit or create your own scale, do the following: 1. Select Application → Communication... → Registers and click Edit Scales from the Modbus Map Editor window. Figure 4-83. The Edit Scales window 2.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 4.13 Configuring the gas chromatograph’s Ethernet port The 700XA has two ethernet ports that can be used to connect the GC with MON 20/20. To configure one or both ethernet ports, select Ethernet Ports... from the Application menu. The Ethernet Ports window displays. Figure 4-84.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 4.14 Working with local operator interface variables Use this window to select and configure up to 25 GC parameters that you would like to monitor using the LOI’s Display mode. Refer to the 700XA Gas Chromatograph System Reference Manual (P/N# 3-9000-744) for more information about the LOI. To set an LOI parameter, do the following: 1. Select LOI Status Variables... from the Application menu.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 Note If Include Mole Percents for current stream is checked, the maximum number of variables you can select is five; if Include Mole Percents for current stream is not checked, you can choose up to 25 variables. 3. To copy the stream settings from a highlighted row and apply them to the next row, click S + Copy.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 4.15 Mapping Foundation Fieldbus variables To map a GC variable to a Foundation Fieldbus process variable (PV), do the following: 1. Select FFB PV Mappings... from the Application menu. The FFB PV Mappings window displays. Figure 4-86. The FFB PV Mappings window 2. Select a new variable by clicking on the appropriate drop-down list under the Variable column.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 Note The PV Status column indicates the state of the data displayed in the PV Value column.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 This page is intentionally left blank.
Section 5: Logs and reports The options in the Logs/Reports menu allow you to do the following: •Keep a maintenance record. •Keep a parameter record. • View alarm, system and event logs. •View and print trend data. •View the GC Config report. •View relevants drawings and diagrams. •View archived analysis, calibration and averages reports. •Configure how and when certain reports are printed. 5.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 5.1.1 Viewing unacknowledged and active alarms To view unacknowledged and active alarms, select Logs/Reports → Alarms → Unack/Active Alarms.... The Unack/Active Alarms window displays. Figure 5-1. The Unack/Active Alarms window Note Double-clicking on the GC Status Bar from the main window also displays the Unack/ Active Alarms window.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 • To view active alarms only, check Active Alarms. This is the default display option. The Unack/Active Alarms window supplies the following data for each alarm: Name Description Status Indicates whether the alarm has been acknowledged or not. State Indicates whether the alarm is ACTIVE or INACTIVE. Date Indicates the date and time at the GC when the alarm condition began. Alarm Message Describes the alarm condition.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 Note An alarm triggered by a user-defined value will continue to display as an active alarm until that value is no longer in the alarm state. • To acknowledge all the alarms displayed on the window, click Ack All (F3). • To acknowledge all the alarms displayed on the window and then remove them from the table, click Clear/Ack All (F4).
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 Figure 5-2. The Alarm Log window The Alarm Log window supplies the following data for each alarm: Name Description Date Indicates the date and time at the GC when the alarm condition began. Alarm Message Describes the alarm condition. Status Indicates whether the alarm is SET (active) or CLR (inactive). Type Indicates whether a high limit or low limit alarm was trigger: • High means a high limit alarm was triggered.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 Note Discrete alarms do not display Type, Limit, or Value data. To view a list of alarms, do the following: 1. To view all alarms, select the All checkbox. Otherwise, select the Select Range checkbox and use the Start Date and End Date dropdown boxes to select a date range. 2. Click Read Records. The list of alarms display with the most recent alarm at the top and the oldest alarm at the bottom.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 • • • • User Manual 3-9000-745 Comma-Delimited (.csv) Microsoft Excel (.xls) HTML File (.html) XML File (.xml) 4. To close the window, click Close. 5.2 Viewing the maintenance log Use this function to manually record and track maintenance activities performed on a given GC unit. To view the maintenance log, select Maintenance Log... from the Log/ Reports menu. Figure 5-4.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 5.2.1 Adding an Entry to the Maintenance Log To add an entry to the maintenance log, do the following: 1. Select Maintenance Log... from the Log/Reports menu. The Maintenance Log window displays. Figure 5-5. The Maintenance Log window 2. Click Insert At Top. A new row appears on the maintenance log table. The Date field contains the GC’s current date and time, and is editable. 3.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 5.2.2 Deleting an entry from the maintenance log To delete an entry from the maintenance log, do the following: 1. Select Maintenance Log... from the Log/Reports menu. The Maintenance Log window displays. Figure 5-6. The Maintenance Log window 2. Select the entry that you want to delete. 3. Click Delete. The entry is removed from the maintenance log. 4. To save the changes and keep the window open, click Save.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 5.3 SEPTEMBER 2010 Working with the parameter list Use this feature to keep a record of the hardware components and associated parameters for a given GC. The Parameter List is a Microsoft Excel document that can be viewed and edited from MON 20/20. Before attempting to edit the document, be sure to review it first to get an idea of what sorts of data it contains.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 Figure 5-7. The Parameter List window 2. Make your changes to the Parameter List. 3. To save the changes and keep the window open, click Save. To save the changes and close the window, click OK.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 5.3.2 Importing and exporting the parameter list The Parameter List is a Microsoft Excel document and is therefore saved with the .xls extension. To import a Parameter List, do the following: 1. Select Parameter List... from the Logs/Reports menu. The Parameter List window displays. Figure 5-8.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 2. Click Import.... The Open dialog displays. 3. Locate and select the Parameter List that you want to import. 4. Click Open and the document will be imported and displayed in the Parameter List window. 5. To save the changes and keep the window open, click Save. To save the changes and close the window, click OK. This Parameter List will now be displayed by default whenever Parameter List...
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 To find out which documents are available on the GC, select Drawings/ Documents... from the Logs/Reports menu. The Drawings/Documents window displays. Figure 5-9. The Drawings/Documents window If the list of available documents does not display under Drawings/ Documents label, click the “+” beside the label.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 5.4.1 Viewing drawings or documents To view a drawing, do the following: 1. Select Drawings/Documents... from the Logs/Reports menu. The Drawings/Documents window displays. Figure 5-10. The Drawings/Documents window 2. Select the drawing to view from the drop-down list. Note If no list displays under the Drawings/Documents label, and there is no “+” beside the label, then this GC does not contain any documents. 3.
User Manual 3-9000-745 MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 Figure 5-11. The File Viewer 4. Click Close to exit the window and to return to the Drawings/ Documents window.
MON20/20 Software for Gas Chromatographs User Manual SEPTEMBER 2010 3-9000-745 5.4.2 Adding files to the GC To add files, such as new or updated drawings, to the GC, do the following: 1. Select Drawings/Documents... from the Logs/Reports menu. The Drawings/Documents window displays. Figure 5-12. The Drawings/Documents window 2. Click Add File(s) to GC. The Open dialog displays. 3. Locate and select the file to add to the GC. 4. Click Open.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 5.4.3 Deleting files from the GC To delete drawings from the GC, do the following: 1. Select Drawings/Documents... from the Logs/Reports menu. The Drawings/Documents window displays. Figure 5-13. The Drawings/Documents window 2. Select the file to delete from the GC. 3. Click Delete File from GC. The Confirm message displays. 4. Click Yes. The file will be deleted from the GC and the Drawings/ Documents list will be updated.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 5.5 User Manual 3-9000-745 Viewing the event log Use this function to track the changes that are made to the various tables within the GC. To view the Event Log, select Logs/Reports → Event Log.... The Event Log window displays. Figure 5-14.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 The Event Log window gives you the option of viewing the total list of change events, or a date-filtered list of events. The Event Log window supplies the following data for each event: Name Description User ID Indicates which user made the change. Date Indicates the date at the GC when the event occurred. Time Indicates the time at the GC when the event occurred. Event Message Provides a description of the event.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 Figure 5-15. The Event Log 3. To save the list, click Save. The list can be saved in the following formats: • • • • • Tab-Delimited (.txt) Comma-Delimited (.csv) Microsoft Excel (.xls) HTML File (.html) XML File (.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 5.6 SEPTEMBER 2010 Displaying reports This function allows you to immediately display, print, or store preconfigured reports of GC analysis data. Data is reported in real-time from the GC or from saved files. 5.6.1 Understanding report types MON 20/20 can generate the following types of reports: • Analysis: Displays a list of the components that were detected, based on raw data.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 • Final Validation: For the most recent validation run, shows the Nominal Value, Allowed Percent Deviation, and the Average Value of each variable in the Validation Data table. See page 5-30 for an example report. Note If the actual deviation is beyond the allowed amount, then the row will be flagged with an *.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 Each report begins with the following header information: • Date-Time: The GC’s date and time when the report was generated. • Analysis Time: The duration, in seconds, of the analysis. Can be configured at Application → Timed Events.... See “Setting the cycle and analysis time” on page 4-29 for more information. • Cycle Time: The duration, in seconds, between two consecutive analyses.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 Figure 5-16.
User Manual 3-9000-745 Figure 5-17.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 Figure 5-18.
User Manual 3-9000-745 Figure 5-19.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 Figure 5-20.
User Manual 3-9000-745 Figure 5-21.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 Figure 5-22.
User Manual 3-9000-745 MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 5.6.2 Viewing reports from live data To view a report created from the most recent data, do the following: 1. Select Report Displays... from the Log/Reports menu. The Report Display window appears. Figure 5-23. The Report Display window Note By default, the Update automatically checkbox is selected.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 2. Select the type of report to view from the following list: • • • • • • • • • • • • • Analysis (GPA) Analysis (ISO) Calibration Final Calibration Validation Final Validation Every Run Hourly 24 Hour Weekly Monthly Variable Raw data 3. Select the appropriate stream. 4. Click Start (F2). The report displays.
User Manual 3-9000-745 MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 Figure 5-24. The report window Note If the report doesn’t appear right away, check the status of the report generation process in the status bar, which is below the row of buttons on the report window. 5. To change the font size, click Font.. There are five preset font sizes available. Continue to click Font to cycle through the sizes until you are satisfied with the report’s readability.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 6. To save the file, click Save. The report can be saved in the following file formats: • • • • TXT HTM HTML MHT 5.6.3 Viewing a saved report To view a saved report, do the following: 1. Select Report Displays... from the Log/Reports menu. The Report Display window appears. Figure 5-25. The Report Display window 2. Click File Viewer (F3). The Report file viewer window displays.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 Figure 5-26. The Report file viewer window 3. Click Open. The Open dialog displays. 4. Locate and select the report that you want to view. Reports can be found in the following file formats: • • • • • TXT RPT HTM HTML MHT 5. Click Open. The report displays. 6. To change the font size, click Font.. There are five preset font sizes available.
MON20/20 Software for Gas Chromatographs User Manual SEPTEMBER 2010 5.7 3-9000-745 Viewing reports based on archived data based on archived GC runs. Use the Archive Report commands to generate analysis, calibration, and average reports 5.7.1 Viewing analysis and calibration reports based on archived data To generate and view an analysis or calibration report from archived data, do the following: 1. Select Logs/Reports → Archive Report → Analysis/Calibration/ Validation....
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 Figure 5-27. The Analysis/Calibration Archive Report window 2. Select a report type from the Report drop-down list. You can choose from the following report types: • • • • • • • Analysis Calibration Final Calibration Validation Final Validation Raw Data Dew Point Calculations (optional) 3. Select a stream from the Stream drop-down list.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 Figure 5-28. The Analysis/Calibration Archive Report window 4. To date-filter the list of records, select the Time Period checkbox and use the Start Date and End Date drop-down boxes to select a date range. 5. Select the record(s) that you want to view. To select several records, hold down CTRL and select each record.
User Manual 3-9000-745 MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 Figure 5-29. The report window 7. To change the font size, click Font.. There are five preset font sizes available. Continue to click Font to cycle through the sizes until you are satisfied with the report’s readability. 8. To print the report, click Print.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 9. To save the file, click Save. The report can be saved in the following file formats: • • • • TXT HTM HTML MHT 5.7.2 Viewing average reports based on archived data To generate and view an average report from archived data, do the following: 1. Select Logs/Reports → Archive Report → Average.... The Average Archive Report window displays.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 Figure 5-30. The Analysis/Calibration Archive Report window 2. Select a report type from the Report drop-down list.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 3. Select a stream from the Stream drop-down list. By default, the List of Averages table displays all records for the selected report type and stream. 4. To date-filter the list of records, select the Time Period checkbox and use the Start Date and End Date drop-down boxes to select a date range. 5. Select the record(s) that you want to view. To select several records, hold down CTRL and select each record.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 5.7.3 Printing reports automatically To configure MON 20/20 to print a report of your choosing automatically based on that report’s schedule of availability, do the following: 1. Select Printer Control... from the Logs/Reports menu. The Printer Control window displays. Figure 5-31. The Printer Control window Note MON 20/20 must be connected to the GC for the report to be printed.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 2. The following types of potential reports are listed in the Report Name column: • Analysis (GPA) - An analysis report will print after an analysis run is completed.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 4. To print a report at a fixed interval, check the appropriate checkbox from the Print At Fixed Interval? column. (a.) Enter a start time in the Start Time column. (b.) Enter an interval, in hours, in the Interval column. 5. Use the columns numbered 1 through 20 to select the streams that you want to use for data collection. 6. To save the changes without closing the window, click Save.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 Figure 5-32. The Trend Data window 2. Click Trend. The Select records for Trending window displays.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 Figure 5-33. The Select records for Trending window 3. Select the analysis or calibration records that you want to trend from the Select Analysis/Calibration Records selection menu. Click > to move your selection to the Selected Records queue. 4. If applicable, select the type of average record that you want to trend from the Select Average Records section. Click > to move your selection to the Selected Records queue. 5.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 7. Click Trend. MON 20/20 reads the data from the GC and then closes the Select records for Trending window and plots the trend data on the graph section of the Trend Data window. Figure 5-34. The Trend Data window with graphs Each trend record is color-coded; use the Trend pull-down menu to select a specific trend record.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 5.8.2 Viewing saved trend data Trend data files are saved with the XTRD file extension. To view a saved trend file, do the following: 1. Select Trend Data... from the Logs/Reports menu. The Trend Data window displays. Figure 5-35. The Trend Data window 2. Click PC File. The Open Trend File window displays. 3. Select the file that you want to view and click Open. The trend graph displays.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 Figure 5-36. The Trend Data window Note To view the chromatogram that is associated with a particular trend data point, locate the data point in the table and double-click it while pressing the SHIFT key.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 5.9 SEPTEMBER 2010 Working with the Trend Graph Right-clicking with the mouse on the graph brings up the following commands and keyboard shortcuts: Command Name Shortcut Description Zoom In “+” (NUMPAD) Zooms in on the entire graph. NOTE: Another way to zoom in is by clicking and dragging your mouse to select the region of the graph that you want to zoom in on. Zoom Out “-” (NUMPAD) Zooms out from the entire graph.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 Command Name Shortcut Description Toggle Coarse/Fine Cursor F4 Toggles the cursor from coarse and less accurate to fine and more accurate. Toggle Lines/Dots Displays F9 Toggles the trend graph from lines to dots, or dots to lines. Toggle Mouse Position Tip CTRL + F4 The graph’s cursor follows the movement of the mouse while a hovering tooltip displays the exact coordinates of the current point.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 5.10 Editing the display properties of the graph 5.10.1 The graph bar Use the graph bar buttons to change the display parameters of the graph. Figure 5-37. The graph Y axis Y Max Y interval color-coded trend graph Y Min X axis X Min X max X interval Click Edit to view or change the display properties of the X and Y axes. The Edit Graph window displays.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 Figure 5-38.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 The following table lists the parameters that can be edited: Command Description Point Sets the X-axis values to points. For the purposes of this graph, each sample run is considered a data point. Therefore, if 2500 sample runs were used to generate the trend graph, then there are 2500 data points. Default Value 0 NOTE: The X-axis value for the first sample, or point, in the trend graph is 0, not 1.
MON20/20 Software for Gas Chromatographs User Manual SEPTEMBER 2010 3-9000-745 Figure 5-39. Primary axes primary Y axis secondary Y axis Note If three or more graphs are displayed, only the second graph will be plotted using the secondary Y axis; all other graphs will be plotted with the primary Y axis. Default Value Command Description Y axis Display Format • Percent - Sets the Y-axis values to a percentage of the Y Max value. • Value - Sets the Y-axis values to the sample run values.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 Default Value Command Description Print Speed Sets the number of inches per second for the x-axis while printing a chromatogram, similar to an XY plotter. N/A X Intervals Sets the number of intervals to be displayed on the graph for the X axis. 10 Display Option Determines whether the chromatograph is displayed as a solid line or as a dotted line. Lines Show labels Determines whether each axis is labelled.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 5.11.1 Editing a trend graph You can use the Edit window to change the X and Y offset values for a graph, change its color, and also set which Y axis should be used when plotting it. These changes may be necessary to make the trend more distinguishable from those that surround it, or to position a graph in relation to a different graph for comparison. To edit a trend trace, do the following: 1.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 • Add Trace to - Sets which Y axis should be used when plotting the graph. See “The graph bar” on page 5-54 for more information. 3. To accept your changes, click OK. 5.11.2 Entering a description for a trend graph To add or change description text for a trend graph, do the following: 1. From the Trend bar, click Desc. The Edit Description window displays. Figure 5-43. The Edit Description window 2.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 5.11.3 Saving a trend trace To save a trend trace to disk, do the following: 1. From the Trend pull-down menu, select the trace that you want to save.The Trend pull-down menu. Figure 5-44. The Trend pull-down menu 2. Click Save. The Save Trend File window displays. Note To save all currently displayed trend traces into one file, click Save All. 3.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 5.11.4 Removing a trend graph from view To remove a trend graph from the graph display, do the following: 1. From the Trend pull-down menu, select the graph that you want to remove. Figure 5-45. The Trend pull-down menu 2. Click Remove. 5.11.5 Displaying trend data The data used to plot the trend graphs displays in the table to the right of the graph display area. Figure 5-46.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 The table contains the following columns: Label Description TRD Indicates the identifcation number of the trend graph. Pt # For the purposes of trend graphs, each sample run is considered a data point. Therefore, if 2500 sample runs were used to generate the trend graph, then there are 2500 data points. NOTE: The first sample, or point, is counted as 0, not 1.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 Figure 5-47. The Trend Data window The table contains the following columns: Label Description TRD Indicates the identifcation number of the trend graph. Average Indicates the average data point value of the selected samples. Minimum Inidicates the lowest data point value of the selected samples. Maximum Inidicates the highest data point value of the selected samples.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 5.12 Generating a GC Configuration Report A GC Config Report displays all current settings for the GC. This section explains how to produce a GC Config Report and provides an example for reference. To generate a GC Config Report, do the following: 1. Select GC Config Report... from the Logs/Reports menu. The GC Config Report window displays. Figure 5-48.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 2. Select the checkbox for each option that you want to include in the report. Note To select all the options, click Select All (F2). To clear all options, click Clear All (F3). 3. Select the type of output you want for the report.
MON20/20 Software for Gas Chromatographs User Manual SEPTEMBER 2010 3-9000-745 System Report from Model 03/12/2009 12:51:58 PM Description [SAMPLE] Austin Value Stream Sequence 1,2,3 Analyzer Name Austin GC Model GC700XA System Description Firmware Version GC Serial Number Company Name GC Location Number of Valves 3 Number of Serial Ports 3 Daylight Saving Time CGM Analog O/P Cfg.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 SAMPLE cont.
MON20/20 Software for Gas Chromatographs User Manual SEPTEMBER 2010 3-9000-745 SAMPLE cont. n-HEPTANE n-OCTANE 205.420 233.290 Component Reference Component C6+ 47/35/17 PROPANE i-BUTANE n-BUTANE NEOPENTANE i-PENTANE n-PENTANE NITROGEN METHANE CARBON DIOXIDE ETHANE n-NONANE n-HEXANE n-HEPTANE n-OCTANE 190.390 216.370 48.474 48.289 MultiLevel Calib 'a' 44.927 44.788 MultiLevel Calib 'b' 0.0 0.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 SAMPLE cont. n-PENTANE NITROGEN METHANE CARBON DIOXIDE ETHANE H2S Component C6+ 47/35/17 PROPANE i-BUTANE n-BUTANE NEOPENTANE i-PENTANE n-PENTANE NITROGEN METHANE CARBON DIOXIDE ETHANE H2S Component C6+ 47/35/17 PROPANE i-BUTANE n-BUTANE NEOPENTANE i-PENTANE n-PENTANE NITROGEN METHANE CARBON DIOXIDE ETHANE H2S Component C6+ 47/35/17 PROPANE i-BUTANE n-BUTANE 5-70 area area area area area area 0 0 0 0 0 0 5.00 5.
MON20/20 Software for Gas Chromatographs User Manual SEPTEMBER 2010 3-9000-745 SAMPLE cont.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 SAMPLE cont. Slope Sens Peak Width Inhibit Inhibit Inhibit Inhibit Inhibit Inhibit Summation Inhibit Inhibit Summation Inhibit Peak Width Slope Sens Inhibit Inhibit Inhibit Peak Width Inhibit Peak Width Slope Sens Inhibit Inhibit Inhibit 48 4 Off Off On On Off Off On On Off Off On 2 16 Off On On 8 Off 8 48 Off On On 1 1 2 1 2 1 1 2 2 1 1 2 1 1 1 1 2 1 2 2 1 1 1 1 2 3.0 4.0 31.3 33.0 38.5 43.5 48.0 72.5 72.6 85.0 87.0 105.
MON20/20 Software for Gas Chromatographs User Manual SEPTEMBER 2010 3-9000-745 SAMPLE cont. Software TEV table 2 TEV Type Value Det # Time (Sec) 1 2 1 2 1 2 1 2 0.0 0.0 0.0 0.0 0.1 0.1 285.0 285.0 State Time (Sec) SSO_1 SSO_2 DualColumn S/BF_1 S/BF_2 On On On On On SSO_1 SSO_2 S/BF_1 S/BF_2 DualColumn DualColumn On On On On Off On 0.0 1.0 2.0 5.0 6.0 11.0 15.0 16.0 26.5 28.0 42.5 133.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 SAMPLE cont. Inhibit Inhibit Inhibit Inhibit Inhibit Inhibit Inhibit Inhibit Inhibit Inhibit Inhibit Inhibit Inhibit Inhibit Off Off On On Off On Off Off On On On On Off On Gain TEV table 3 Det # Gain 1 2 4 4 1 2 1 2 1 1 1 2 2 1 1 1 1 1 5.0 5.5 13.0 13.5 25.0 27.0 28.0 28.3 31.5 34.0 45.0 48.0 133.2 141.5 Time (Sec) 0.0 0.
MON20/20 Software for Gas Chromatographs User Manual SEPTEMBER 2010 3-9000-745 SAMPLE cont.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 SAMPLE cont. Label Cal Freq.
MON20/20 Software for Gas Chromatographs User Manual SEPTEMBER 2010 3-9000-745 SAMPLE cont. Alarm Alarm Alarm Alarm Alarm Alarm Alarm Alarm Alarm 037 038 039 040 041 042 043 044 045 Label Alarm Alarm Alarm Alarm Alarm Alarm Alarm Alarm Alarm Alarm Alarm Alarm Alarm Alarm Alarm Alarm Alarm Alarm Alarm Alarm Alarm Alarm Alarm Alarm Alarm Alarm Alarm Alarm Alarm Alarm Alarm Alarm Alarm Alarm Alarm Alarm Alarm 1-C9 Cal Gas Final Calib.RF Dev Alarm 2-C6 Chamber Final Calib.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 SAMPLE cont.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 SAMPLE cont.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 SAMPLE cont.
MON20/20 Software for Gas Chromatographs User Manual SEPTEMBER 2010 3-9000-745 SAMPLE cont.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 SAMPLE cont.
MON20/20 Software for Gas Chromatographs User Manual SEPTEMBER 2010 3-9000-745 SAMPLE cont. Serial Port Report from Model 03/12/2009 12:52:09 PM Austin Label Physical Name First PORT C Label Port Type First Slave ModBus ID Unit System MAP File 32 U.S. Customary UsrMap.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 SAMPLE cont. SCALED_FP11 SCALED_FP12 SCALED_FP13 SCALED_FP14 SCALED_FP15 SCALED_FP16 SCALED_FP17 SCALED_FP18 SCALED_FP19 SCALED_FP20 SCALED_FP21 SCALED_FP22 SCALED_FP23 SCALED_FP24 SCALED_FP25 SCALED_FP26 SCALED_FP27 SCALED_FP28 SCALED_FP29 SCALED_FP30 SCALED_FP31 SCALED_FP32 0.000000 0.000000 0.000000 0.000000 0.000000 0.000000 0.000000 0.000000 0.000000 0.000000 0.000000 0.000000 0.000000 0.000000 0.000000 0.000000 0.
MON20/20 Software for Gas Chromatographs User Manual SEPTEMBER 2010 3-9000-745 SAMPLE cont. 3055 3057 3059 3061 3063 3065 3067 3069 3071 3073 3075 3077 3079 3081 3083 3085 3087 3089 3091 3093 3095 3097 3099 3101 3103 3105 3107 3109 3111 3113 3115 3117 3119 3121 3123 3125 1-C9 1-C9 1-C9 1-C9 1-C9 1-C9 1-C9 1-C9 1-C9 1-C9 1-C9 1-C9 2-C6 1-C9 1-C9 1-C9 1-C9 1-C9 2-C6 2-C6 1-C9 1-C9 1-C9 1-C9 1-C9 1-C9 1-C9 1-C9 1-C9 1-C9 1-C9 1-C9 1-C9 1-C9 1-C9 1-C9 Cal Gas Component.Ret Time.NITROGEN Cal Gas Component.
User Manual 3-9000-745 MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 SAMPLE cont.
MON20/20 Software for Gas Chromatographs User Manual SEPTEMBER 2010 3-9000-745 SAMPLE cont.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 5.13 Deleting archived data from the gas chromatograph To delete archived data and reset the GC memory, do the following: 1. Select Reset Archive Data... from the Logs/Reports menu. The Reset Archive Data window displays. Figure 5-49. The Reset Archive Data window 2. Select the types of data that you want to delete. Note To select all the options, click Select All. To clear all options, click Deselect All. 3. Click Reset.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 4. Click Yes. MON 20/20 clears the GC’s memory. New archived records will begin accumulating again as analysis and calibration runs occur. 5.14 The molecular weight vs. response factor graph The Molecular Weight Vs. Response Factor window, which can be useful in checking valve function, displays a graph that consists of the following information: • Log (Molecular Weight) vs.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 Figure 5-50. Molecular Weight vs. Response Factor window This graph is only available for calibration streams, which can be selected from the Stream drop-down list. By default, the newest final calibration data is used to generate the graph, but any archived final calibration file can be used by selecting it from the Final Calibration Record drop-down list. To print the graph, click Print.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 This page is intentionally left blank.
User Manual 3-9000-745 5-92 MON20/20 Software for Gas Chromatographs SEPTEMBER 2010
Section 6: Controlling Analyses The options in the Control pull-down menu allow you to manage analysis runs as well as calibration, validation and baseline runs. Control menu commands also allow you to stop an analysis run immediately or at the end of the run. 6.1 Halting an analysis Before a new analysis run can be initiated, the current analysis must be stopped. To stop the current analysis at the end of its cycle, do the following: 1. There are three ways to halt an analysis run: • Select Halt...
User Manual 3-9000-745 MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 Figure 6-2. Confirmation message 3. Click Yes and the analysis will stop at the end of the current cycle. Use the Mode column on the GC Status Bar to monitor the status of the operation. When the analysis has halted, the Mode value will be “Idle”. Figure 6-3. The GC Status Bar 6.2 Auto sequencing Use this function to start continuous GC analysis runs that follow a predefined stream sequence.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 To initiate auto-sequencing, do the following: 1. There are three ways of initiating auto sequencing: • Select Auto Sequence... from the Control menu. • Press F2. • Click on the Toolbar. A confirmation message displays. Figure 6-4. Confirmation message 2. Check the Purge stream for 60 seconds check box to set the purging option. The checkbox is selected by default.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 Note To view the results of the Auto Sequence run, select Report Display from the Logs/ Reports menu. 6.3 Analyzing a single stream Note If an analysis run is in progress, it must be stopped before auto sequencing can be started. See “Stopping an Analysis Run” on page 6-9 for more information. To start an analysis run on a single calibration or sample stream, do the following: 1. Select Single Stream...
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 Note Purging allows sample gas to flow through the sample loop for 60 seconds prior to beginning the first analysis. 4. Check the Continuous operation check box to set or unset repetitive analysis. The checkbox is selected by default. 5. Click OK and the analysis starts. Use the Mode column on the GC Status Bar to monitor the status of the analysis run. Figure 6-7.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 Figure 6-8. The Start Calibration window Note If the GC is in Auto Sequence mode, calibration will not start until two or more analysis runs have been completed. This delay is required to complete the current analysis and the analysis of the stream currently purging through the valve. 2. Select a stream from the Stream menu. 3. Check the Purge stream for 60 seconds check box to set the purging option.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 (b.) Select Forced to perform a manual calibration in which the CDT for the selected stream(s) will be updated with calibration data even if that data is outside the acceptable deviations, as listed on the CDT. For more information, see “Managing Component Data Tables” on page 4-5. 5. Click OK and the calibration starts. Use the Mode column on the GC Status Bar to monitor the status of the operation. Figure 6-9.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 Figure 6-10. The Start Validation window Note If the GC is in Auto Sequence mode, validation will not start until two or more analysis runs have been completed. This delay is required to complete the current analysis and the analysis of the stream currently purging through the valve. 2. Check the Purge stream for 60 seconds check box to set the purging option. The checkbox is selected by default.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 6.6 User Manual 3-9000-745 Stopping an Analysis Run Note This function forces the system into Idle mode. If Stop Now is performed while an analysis is in progress, the components may continue to elute from the columns during. No analysis data will be generated. Do not perform a Stop Now unless absolutely necessary. Whenever possible, use the Halt function. To immediately stop an analysis run, do the following: 1. Select Stop Now...
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 This page is intentionally left blank.
Section 7: Using MON 20/20 Tools The options in the Tools pull-down menu allow you to do the following: •Use the Modbus Test program to confirm that data is being accurately relayed from the gas chromatograph to the PC. •Manage users. •Adjust the sensitivity of the LOI keys. •Install upgrades to the GC. 7.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 This section provides detailed instructions for using the Modbus program. Use this program only if you are familiar with Modbus communication protocol and the operation of MON 20/20. 7.1.1 Comparing Modbus protocols MON 20/20 and the Modbus program can accommodate two different Modbus protocols: SIM_2251 and User_Modbus.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 7.1.2 Setting communication parameters To determine or reset the communications parameters used by the Modbus program, do the following: 1. Select Modbus Test... from the Tools menu. The Modbus Test Program window displays. The current port settings display in the window’s title bar. Figure 7-1.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 2. Click Port Setup. The Port Setup window appears. Figure 7-2. The Port Setup window 3. Make the appropriate configuration changes.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 Parameter Typical Setting Protocol ASCII Modbus RTU Modbus Note For a direct connection to the GC, ensure that the port setting is the same as the Com ID number of the serial port used. 4. Click OK. 7.1.3 Getting Modbus Data To read or write register contents to the GC, or any other device, do the following: Note Before retrieving data, print a GC Config Report (see Section 5.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 2. Select the desired read or write option from the Function pull down menu. Function Code Description 1 (Read Coil) Reads one or more coil values. 2 (Read Input Status) Reads one or more input status values. 3 (Read Multiple Regs) Reads one or more register values. 4 (Read Input Regs) Reads one or more input register values.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 5. Type the desired repeat count, which is the number of times the Modbus program should read or set the specified registers before ceasing transmission, in the Repeat field. The Modbus program will accept a repeat value of 1 to 9999. A value of –1 produces an infinite polling loop that can be terminated by clicking Stop. 7.1.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 2. Select the desired read or write option from the Function pull down menu. Function Code Description 1 (Read Coil) Reads one or more coil values. 2 (Read Input Status) Reads one or more input status values. 3 (Read Multiple Regs) Reads one or more register values. 4 (Read Input Regs) Reads one or more input register values.
MON20/20 Software for Gas Chromatographs User Manual SEPTEMBER 2010 3-9000-745 5. Type the desired repeat count, which is the number of times the Modbus program should read or set the specified registers before ceasing transmission, in the Repeat field. The Modbus program will accept a repeat value of 1 to 9999. A value of –1 produces an infinite polling loop that can be terminated by clicking Stop. 6. Select the Use to decode registers check box. 7.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 7.1.5 Transmitting using a template Templates are best used when decoding mixed data types because the template contains data that the Modbus program can use to determine which data type should be assigned to which register. To create a new template or to use an existing template, do the following: Note Before retrieving data, print a GC Config Report (see Section 5.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 3. Type the starting register value in the Data Addr field. Note The data type is set automatically by the Modbus program, based on the specified data address. 4. In the Quantity field, type the number of registers to be retrieved. The Modbus program will accept a quantity value of 1 to 2016.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 The following table describes the relationship between templates and record numbers: Data Type Setting Other Setting(s) Result Register template • Enter Data Addr value. • Enter Record No. value. • Enter Quantity value. Read Quantity fields (i.e., the number of fields specified by the Quantity setting) from the specified Record No. of the register (Data Addr). Log template • Enter Record No. value.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 11. To edit the template, select a data type for each desired offset. 12. To change all offsets to the same data type, change the first offset to the desired data type, and then click Auto Reset. The data type for the remainder of the offsets switch to the data type of the first offset. 13. To save the displayed file to disk, click Save As.... The Select Template Configuration File dialog appears.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 Figure 7-4. The Log Data window 2. Select the Enable Logging ‘Data’ Registers and Values check box to enable data logging and to activate the Log Data Parameters section. 3. Select a Logging Mode from the pull-down menu.
MON20/20 Software for Gas Chromatographs User Manual SEPTEMBER 2010 3-9000-745 7.1.7 Saving Modbus data To save the data table to a separate file, do the following: 1. Click Save Data. The Save ‘Data’ Displayed As dialog appears. The file can be saved as a tab-delimited text file, an HTML file or a Microsoft Excel file. 2. Type in a filename and click Save. 7.1.8 Printing Modbus data To print Modbus data, click Print Data. The standard print dialog displays.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 1. Click Error.... The Error window appears. Figure 7-5. The Error window Note Double-click a Description cell to “scroll through” the displayed text. 2. To view all errors that have occurred in this session, click Update. 3. To delete all entries to date, click Clear.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 7.3 User Manual 3-9000-745 Managing users Use the User Administration commands to create or delete users, change passwords, and to monitor PC-to-GC connections. Login security is at the gas chromatograph level instead of at the software level. This means that you no longer have to log in after starting MON 20/20—but you do have to log in to the gas chromatograph to which you are trying to connect.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 Note Each GC ships with two administrator accounts: daniel and emerson. By default, these two accounts do not require a password, but a password can be added, if desired.
MON20/20 Software for Gas Chromatographs User Manual SEPTEMBER 2010 Menu Logs/Reports 3-9000-745 Commands Regular User Read-Only User Y Y read-only read-only Limit Alarms Y Y read-only read-only System Alarms Y Y read-only read-only Streams Y Y read-only read-only Stream Sequence Y Y read-only read-only Communication Y Y read-only read-only Unack/Active Alarms Y Y Y read-only read-only read-only read-only read-only Ack/Clear Alarms Y Y Y N Maintenance Log
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 7.3.1 Creating users Note You must be logged in as an administrator. To create a user, do the following: 1. Select Tools → Users → User Administration.... The User Administration window appears, displaying a list of current users and their role levels. Figure 7-6. The User Administration window 2. To add a user, click Add User. The Add User window displays.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 Figure 7-7. The Add User window 3. Enter the appropriate information into the text fields. 4. Click OK. MON 20/20 creates the new user and adds it to the User table on the User Administration window. 7.3.2 Exporting a list of user profiles To save a list of users, along with their role levels and passwords, do the following: 1. Select Tools → Users → User Administration....
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 Figure 7-8. The User Administration window 2. Click Export File. The Export User File window displays.
MON20/20 Software for Gas Chromatographs User Manual SEPTEMBER 2010 3-9000-745 Figure 7-9. The Export User File window 3. Navigate to where you want to save the file, if necessary. 4. Type in a file name or use the pre-generated name provided. 5. Click Save. 7.3.3 Importing a list of user profiles To load a list of users, along with their role levels and passwords, do the following: 1. Select Tools → Users → User Administration....
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 Figure 7-10. The User Administration window 2. Click Import File. The Import User File window displays.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 Figure 7-11. The Import User File window 3. Navigate to where the file is located, if necessary. Note User files have the .xusr extension. 4. Click on the file to be loaded. 5. Click Open. The users will be added to the User Administration window.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 7.3.4 Editing users Note You must be logged in as an administrator. To edit a user’s name, role level or password, do the following: 1. Select Tools → Users → User Administration.... The User Administration window appears, displaying a list of current users and their role levels. Figure 7-12. The User Administration window 2. Select the user whose role you want to edit and click Edit User. The Edit User window displays.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 Figure 7-13. The Edit User window 3. Change the appropriate information as required. 4. Click OK. MON 20/20 makes the requested changes and returns to the User Administration window. 7.3.5 Removing a user To remove a user, do the following: 1. Select Tools → Users → User Administration.... The User Administration window appears, displaying a list of current users and their role levels.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 Figure 7-14. The User Administration window 2. Select the user you want to delete and click Remove User. A confirmation message displays. 3. Click Yes. MON 20/20 deletes the user and returns to the User Administration window. 7.3.6 Changing a user’s password A user without administrator-level access can only change his or her password. 1. Select Select Tools → Users → Change User Password....
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 Figure 7-15. The Change User Password window 2. Enter the appropriate information in the text fields and click OK. 7.3.7 Resetting the adminstrator password To reset an administrator password, do the following: 1. Start MON 20/20 and select Users → Reset Administrator User / Password. The following warning displays: Figure 7-16.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 2. Click Yes. The Connect to GC window displays. Figure 7-17. The Connect to GC window 3. Click the Ethernet button that corresponds to the GC whose password you want to reset. MON 20/20 will connect to the GC and generate a Password Reset Request ID. The MON 20/20 - Password Reset window displays.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 Figure 7-18. The MON 20/20 - Password Reset window 4. Click Copy to Clipboard and email the Password Reset Request ID to tech.service@emerson.com. You will be sent a Password Reset Key. 5. After you receive the Password Reset Key, return to the Connect to GC window and again click the Ethernet button that corresponds to the GC whose password you want to reset. The Login window displays. Figure 7-19. Login window 6.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 7.3.8 Finding out who is connected to the gas chromatograph To ascertain which users are connect to the GC, select Tools → Users → Logged on Users.... The Logged on Users window displays with a list of the users who are currently logged on to the GC, along with each user’s IP address. Figure 7-20. The Logged on Users window 7.4 Upgrading the firmware This command allows you to download upgrades to the GC’s firmware.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 Figure 7-21. The Upgrade Firmware window 2. Click Open. The Open Download File dialog displays. 3. Locate and select the desired .zip file and click Open. The .zip file’s content information displays in the Upgrade section of the Upgrade Firmware window. The Information column will alert you to the new files that should be selected and downloaded to the GC.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 4. Select the check boxes for the files that you want to download to the GC and click Download. While the files are downloading, you can monitor their status in the Upgrade Progress section. Figure 7-22. The Upgrade Firmware window Note If you want to halt the download, click Cancel Download.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 5. When the download completes successfully, a confirmation message displays. Click OK. MON 20/20 disconnects from the GC and the GC reboots. 7.5 Cold booting Cold booting the GC clears all its stored analysis files and logs and resets all the tables to the default settings. This is a necessary step towards refurbishing the GC or CPU board. 7.
User Manual 3-9000-745 MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 To view the Diagnostics window, select Diagnostics... from the Tools menu. Figure 7-23.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 7.7 User Manual 3-9000-745 Adjusting the sensitivity of the LOI Keys To adjust the sensitivity of the LOI keys, do the following: 1. Select LOI Key Sensitivity from the Tools menu. The LOI Key Sensitivity window displays. Figure 7-24. The LOI Key Sensitivity window 2. Adjust the sensitivity for a key by sliding the bar up or down. Raising the bar increases the sensitivity or the key; lowering the bar decreases the sensitivity.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 SEPTEMBER 2010 Note Click Restore Factory Defaults to return the sliders to their original settings. 3. To save the changes without closing the window, click Save. To save the changes and close the window, click OK. 7.8 Setting the ROC card type To set the card type for a ROC card, do the following: 1. Select ROC Cards... from the Tools menu. The ROC Cards window displays. Figure 7-25.
MON20/20 Software for Gas Chromatographs SEPTEMBER 2010 User Manual 3-9000-745 2. Select the card type for the ROC card from the ROC Card Type dropdown list. The options are: • • • • None (Default) ROC Analog Output ROC Communication Module RS-232 ROC Communication Module RS-485 3. To save the changes without closing the window, click Save. To save the changes and close the window, click OK.
User Manual 3-9000-745 7-40 MON20/20 Software for Gas Chromatographs SEPTEMBER 2010
Appendix A, Component Data Table A This appendix provides a sample standard component data table as well as a table of the ISO-related components. • Table A-1, Example Standard Component Data Table • Table A-2, ISO Component Data Table All values depend on a base pressure of 14.73 PSIA and a base temperature of 60 oF (15.56 oC). BTU components reference GPA Standard 2145-09. Note An asterisk (*) denotes components that are assigned temporary I.D. codes, starting with 150, as they are used.
A-2 Component Data Table Table A-1 Example Standard Component Data Table AGA 8 Component Daniel Sim 2251 I.D. No. 1476.9 1426.5 Ethane 22 0.104759 0 0 AIRmix1 26 0 0 0 0 Argon 46 0.6173 5.15 0.0874 435.4 359.8 None * 2.6969 0.8844 7.373 0.2798 3750.5 3599.2 n-Hexane * 62.1055 2.0068 0.573515 4.78155 0.32117 3264.64 3012.45 n-Butane 33 56.11 63.05 1.9372 0.6013 5.013 0.2956 3087 2885.4 n-Butane 28 Butenes 56.11 55.448 1.9372 0.6097 5.0833 0.
AGA 8 Component Daniel Sim 2251 I.D. No. 0 0 None 42 0 1267 1267 None 41 6.531 0.3403 4492.1 4189.4 n-Hexane * 0.7504 6.256 0.2961 3772.4 3520.2 n-Pentane * 3.9439 0.6979 5.819 0.5185 6247.9 5793.9 n-Octane * 7.404 2.9753 0.6664 5.556 0.4096 4756 4403.1 n-Hexane * 100.21 3.492 3.4596 0.6782 5.654 0.4682 5494.6 5091.4 n-Heptane * 2,4-Dimethpenta 100.21 3.292 3.4596 0.6773 5.647 0.4686 5499.4 5096 n-Heptane * 3,3-Dimethpenta 100.2 2.773 3.
AGA 8 Component Daniel Sim 2251 I.D. No. 5976.6 5674 n-Nonane * 0.5199 6246.1 5792.2 n-Octane * 2.5 0.169487 1012.34 911.5 Methane 00 0.796 6.64 0.1275 868.7 767.9 None * 2.9057 0.7536 6.283 0.3538 4511.6 4209.1 n-Hexane * 1.609 3.39 0.774 6.453 0.4019 5228 4874.9 n-Heptane * 100.21 2.271 3.4596 0.683 5.694 0.4647 5507.3 5104 n-Heptane * 3-Methylhexane 100.21 2.13 3.4596 0.6917 5.767 0.4589 5511.3 5107.8 n-Heptane * m-Xylene 106.17 0.326 3.
Net Dry BTU AGA 8 Component Daniel Sim 2251 I.D. No. 5222 4969.7 n-Octane * 0.088739 0 0 Oxygen 16 5.383 0.3441 3835.4 3583.3 n-Pentane 37 0.50719 4.2285 0.275429 2521.92 2320.36 Propane 02 1.411 0 0 0 2254.2 2254.2 Propane 24 227.3 1.4529 0.5226 4.3571 0.255087 2338.4 2187.05 Propane 23 40.07 0 1.411 0 0 0 2246.2 2246.2 Propane 25 p-Xylene 106.17 0.342 3.6655 0.8657 7.218 0.3885 5220.8 4968.6 n-Octane * Sulfur Dioxide 64.06 88 2.2117 1.
A-6 Component Data Table Table A-2 ISO Component Data Table Molar Mass Sum Factor (0°C) Sum Factor (15°C) Sum Factor (20°C) CV Sup kJ/Mol (0°C) CV Sup kJ/Mol (15°C) CV Sup kJ/Mol (20°C) CV Sup kJ/Mol (25°C) CV Inf kJ/Mol (0°C) CV Inf kJ/Mol (15°C) CV Inf kJ/Mol (20°C) CV Inf kJ/Mol (25°C) Acetylene 26.038 0.0949 0.0837 0.0837 1301.86 1301.37 1301.21 1301.05 1256.79 1256.94 1256.98 1257.03 Air 28.9625 0 0 0 0 0 0 0 0 0 0 0 Argon 39.948 0.0316 0.0283 0.
Molar Mass Sum Factor (0°C) Sum Factor (15°C) Sum Factor (20°C) CV Sup kJ/Mol (0°C) CV Sup kJ/Mol (15°C) CV Sup kJ/Mol (20°C) CV Sup kJ/Mol (25°C) CV Inf kJ/Mol (0°C) CV Inf kJ/Mol (15°C) CV Inf kJ/Mol (20°C) CV Inf kJ/Mol (25°C) CS2 76.143 0.2145 0.1949 0.1871 1104.06 1104.32 1104.41 1104.49 1104.06 1104.32 1104.41 1104.49 Cyclohexane 84.161 0.3209 0.2864 0.2757 3960.67 3956.02 3954.47 3952.96 3690.23 3689.42 3689.13 3688.86 Cyclopentane 70.14 0.255 0.2302 0.
Molar Mass Sum Factor (0°C) Sum Factor (15°C) Sum Factor (20°C) CV Sup kJ/Mol (0°C) CV Sup kJ/Mol (15°C) CV Sup kJ/Mol (20°C) CV Sup kJ/Mol (25°C) CV Inf kJ/Mol (0°C) CV Inf kJ/Mol (15°C) CV Inf kJ/Mol (20°C) CV Inf kJ/Mol (25°C) Methyl Alcohol 32.042 0.4764 0.3578 0.3286 766.59 765.09 764.59 764.09 676.44 676.22 676.14 676.06 Methylcyclo C5 84.161 0.313 0.2811 0.2702 3977.04 3972.46 3970.93 3969.44 3705.34 3705.59 3705.86 3706.6 Methylcyclo C6 98.188 0.3808 0.
Component Name Molar Mass Sum Factor (0°C) Sum Factor (15°C) Sum Factor (20°C) CV Sup kJ/Mol (0°C) CV Sup kJ/Mol (15°C) CV Sup kJ/Mol (20°C) CV Sup kJ/Mol (25°C) CV Inf kJ/Mol (0°C) CV Inf kJ/Mol (15°C) CV Inf kJ/Mol (20°C) CV Inf kJ/Mol (25°C) Propylene 42.0797 0.1378 0.1265 0.1225 2061.57 2059.43 2058.72 2058.02 1926.35 1926.13 1926.05 1925.97 Propyne 40.065 0 0 0 0 0 0 0 0 0 0 0 p-Xylene 106.167 0 0 0 0 0 0 0 0 0 0 0 Sulfur Dioxide 64.065 0.1549 0.
A-10 Component Data Table This page is intentionally left blank.
Appendix B, Data computations A B.1 Data acquisition Every second, exactly 50 equally-spaced data samples are taken (i.e., one data sample every 20 milliseconds) for analysis by the controller assembly. As a part of the data acquisition process, groups of incoming data samples are averaged together before the result is stored for processing. Nonoverlapping groups of N samples are averaged and stored, and thus reduce the effective incoming data rate to 40/N samples per second.
User Manual 3-9000-745 B.2 MON20/20 Software for Gas Chromatographs OCTOBER 2010 Peak detection For normal area or peak height concentration evaluation, the determination of a peak's start point and end point is automatic. The manual determination of start and end points is used only for area calculations in the Forced Integration mode. Automatic determination of peak onset or start is initiated whenever Integrate Inhibit is turned off.
MON20/20 Software for Gas Chromatographs User Manual OCTOBER 2010 3-9000-745 The zero reference line will, in general, be non-horizontal, and thus compensates for any linear drift in the system from the time the peak sequence starts until it ends. In a single peak situation, peak area is the area of the component peak between the curve and the zero reference line. The peak height is the distance from the zero reference line to the maximum point on the component curve.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 OCTOBER 2010 B.3.1 Concentration analysis with response factor Calibration The concentration calculations discussed as follows require a unique response factor for each component in an analysis. These factors may be manually entered by an operator or automatically calculated by calibrating the system.
MON20/20 Software for Gas Chromatographs User Manual OCTOBER 2010 3-9000-745 where RFAVGn Area or height average response factor for component n. RFi Area or height response factor for component n from the calibration run. k Number of calibration runs actually used to calculate the response factors.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 OCTOBER 2010 Note that the average concentration of each component will also be calculated when data averaging is requested. Component concentrations may be input through analog inputs 1 to 4 or may be fixed. If a fixed value is used, the calibration for that component is the mole percent that will be used for all analyses:. CONC n - × 100 CONCN n = --------------------------k ∑ CONCi i=1 where B.
MON20/20 Software for Gas Chromatographs User Manual OCTOBER 2010 3-9000-745 where GPMn Gallons/1000 standard cubic feet of component n. CONCNn Normalized (if selected) concentration of component n. LCFn Liquid equivalent conversion factor for component n at 14.73 PSIA and 60 degrees F. BASE PRS Base (contact) Pressure specified; defaults to 14.73. B.4.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 OCTOBER 2010 where RDn Relative Density of component “n” TOTAL RD Relative Density of total gas sample CONCNn Normalized (if selected) concentration of component n, calculated from peak analysis. P Total number of components to be used in calculation of total BTU/ CF. 100 Removed the 100 factored into the calculation of the concentration earlier in the analysis.
MON20/20 Software for Gas Chromatographs User Manual OCTOBER 2010 3-9000-745 where DRYBTU Dry Gross BTU of Total Gas; see page 7 for details. Z Compressibility factor. BASE PRS Base (contract) pressure specified; defaults to 14.73 PSIA. • Corrected Saturated BTU ( DRYBTU ) ( 0.9826 ) CORRSATBTU = --------------------------------------------------Z where DRYBTU Dry Gross BTU of Total Gas; see page 7 for details.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 OCTOBER 2010 where WVC Water volume content provided by a “live analog input”. CORR ( GROSS )BTU WI = -----------------------------------------------------RD where W.I. Wobbe index value CORRGROSSBTU Corrected Dry BTU for Total Gas Sample; see page 8 for details. RD Real Relative Density; see page 8 for details. Note All components in the sample must be measured in order to calculate weight percent.
MON20/20 Software for Gas Chromatographs OCTOBER 2010 User Manual 3-9000-745 where AVGMW Average molecular weight. Sum of weights of all components in sample. k ∑ i=1 ( WTpercent n ) ÷ ( D n ) - × 100 LVpercent = ------------------------------------------------------------k ∑ ( WTpercenti ) ÷ ( Di ) i=1 Note All components in the sample must be measured in order to calculate liquid volume from mole percent. where LV percent Liquid volume. WT percent Weight percent. D Density.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 OCTOBER 2010 Note All components in the sample must be measured in order to calculate LRDT. k ∑ ( LVpercenti ) ( LRDi ) =1 LRD T = i------------------------------------------------------------100 where LRDT Liquid Relative Density of sample, relative to water at 60 °F. LRDi Liquid Relative Density of component i (GPA2145-94). LVpercent Liquid Volume Percent.
MON20/20 Software for Gas Chromatographs User Manual OCTOBER 2010 3-9000-745 where GD Gas Density in lb/1000 ft3. RD Relative Density (relative to air). 76.4976 Density of air at 14.73 PSIA and 60 °F, in lb/1000 ft3. B.4.3 Multi-level calibration The properties of each gas component can be viewed using the Component Data menu. Included with the component properties in the Component Data Table are four coefficients labeled Multi-Level Calib 'a', 'b', 'c', and 'd', for each component.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 OCTOBER 2010 where P Peak size measured in sample gas. Coefficients: a,b,c, and d Calculated offline and entered after multi-level calibration using several--typically seven--calibration gases. NOTE: If the coefficient values are correct, the response factor will be close to 1. B.4.
Appendix C, Modbus registers list A There are two GC Modbus registers that may be of interest to the developer: SIM_2251 and User_Modbus. Differences betweeb the two registers are summarized in Table C-1. Table C-1 Comparison of SIM_2251 and User_Modbus SIM_2251 User_Modbus Serial slave port. Serial slave port. Modified protocol that allows floating point numbers to be transmitted over Modbus via 2251 emulation slave type. The standard Gould Modbus protocol that accommodates PLC Emulation LO-HI.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 OCTOBER 2010 To obtain a complete list of register assignments, both SIM_2251 and User_Modbus, print a GC Config Report. See “Generating a GC Configuration Report” on page 5-65 for more information. Table C-2 List of User_Modbus Registers Slave Name USER_MODBUS Modbus Reg.
MON20/20 Software for Gas Chromatographs User Manual OCTOBER 2010 3-9000-745 Table C-2 List of User_Modbus Registers Slave Name USER_MODBUS Modbus Reg.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 OCTOBER 2010 Table C-2 List of User_Modbus Registers Slave Name USER_MODBUS Modbus Reg.
MON20/20 Software for Gas Chromatographs User Manual OCTOBER 2010 3-9000-745 Table C-2 List of User_Modbus Registers Slave Name USER_MODBUS Modbus Reg.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 OCTOBER 2010 Table C-2 List of User_Modbus Registers Slave Name USER_MODBUS Modbus Reg.
MON20/20 Software for Gas Chromatographs User Manual OCTOBER 2010 C.2 3-9000-745 SIM_2251 Modbus register list To use the Modbus Test program and view the contents of SIM_2251 registers, set the Register Mode to “Daniel,” as noted in Table C-1. For a complete list of register assignments, both SIM_2251 and User_Modbus, print a GC Config Report. See “Generating a GC Configuration Report” on page 5-65 for more information.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 OCTOBER 2010 Table C-3 List of SIM_2251 Registers Reg. No.
Reg. No. Description 15 14 13 12 11 10 3046 checksum failure 3047 spare 9 8 7 6 5 4 3 2 1 0 anlyzr failure D/A 3 high D/A 3 low D/A 2 high D/A 2 low D/A 1 high D/A 1 low spare spare A/D 2 high A/D 2 low A/D 1 high A/D 1 low spare spare spare spare spare spare spare spare spare spare spare spare spare adjust preamp preamp failure RF % dev power failure #7 low #6 low #5 low #4 low #3 low #2 low O.D.A.
Reg. No. C-10 Modbus registers list Table C-4 SIM_2251 MODBUS REGISTER LIST (BIT NUMBERS) Description 15 14 13 12 11 10 9 8 7 6 5 4 3 2 1 0 Stream #5 3056 #15 low #14 low #13 low #12 low #11 low #10 low #9 low #8 low #7 low #6 low #5 low #4 low #3 low #2 low O.D.A. #1 low #7 high #6 high #5 high #4 high #3 high #2 high O.D.A. #1 high Stream #5 3057 #15 high #14 high #13 high #12 high #11 high #10 high #9 high #8 high 3058 New data flag.
MON20/20 Software for Gas Chromatographs OCTOBER 2010 User Manual 3-9000-745 Table C-5 SIM_2251 Modbus register list (32-bit integers) Reg. No. Description 5001 Cycle time (in 1/30ths of a second) 5002 Calibration cycle time (in 1/30ths of a second) Note The following registers contain no values until the completion of atleast one analysis run. Table C-6 SIM_2251 Modbus register list (floating point) Reg. No.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 OCTOBER 2010 Table C-6 SIM_2251 Modbus register list (floating point) Reg. No.
MON20/20 Software for Gas Chromatographs OCTOBER 2010 User Manual 3-9000-745 Table C-6 SIM_2251 Modbus register list (floating point) Reg. No.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 OCTOBER 2010 Table C-6 SIM_2251 Modbus register list (floating point) Reg. No.
MON20/20 Software for Gas Chromatographs User Manual OCTOBER 2010 3-9000-745 Table C-7 SIM_2251 Modbus Communication Indices Reg. No.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 OCTOBER 2010 Table C-7 SIM_2251 Modbus Communication Indices Reg. No.
Appendix D, Basic and advanced system variables D D.
System Reference Manual 3-9000-744 700XA Gas Chromatograph OCTOBER 2010 Group Basic Calculations Advanced Calculations Analysis Stream Avg Molecular Weight Base Pressure Base Temperature GPA Z Factor GPA Real Rel Den Gas GPA Wobbe Index Gas Den lbm/1000 ft3 HV Gross BTU Dry HV Gross BTU Sat HV Net BTU Dry HV Net BTU Sat HV Gross BTU/lb Dry HV Sup MJ/m3 Dry HV Sup MJ/m3 Sat HV Inf MJ/m3 Dry HV Inf MJ/m3 Sat HV Sup MJ/kg Dry HV Inf MJ/kg Dry HV Sup Kcal/kg Dry HV Inf Kcal/kg Dry Reid Vapor Press Start T
700XA Gas Chromatograph System Reference Manual OCTOBER 2010 Group Analysis Stream (cont.
System Reference Manual 700XA Gas Chromatograph 3-9000-744 OCTOBER 2010 Group Basic Calculations Advanced Calculations Analysis Optional Base Pressures n/a Opt Base Pressure Gal/1000 SCF C2+ Gal/1000 SCF C3+ Gal/1000 SCF C4+ Gal/1000 SCF C5+ Gal/1000 SCF C6+ HV Gross BTU Dry HV Gross BTU Sat HV Gross BTU Act HV Net BTU Dry HV Net BTU Sat HV Net BTU Act HV Sup MJ/m3 Dry HV Sup MJ/m3 Sat HV Sup MJ/m3 Act HV Inf MJ/m3 Dry HV Inf MJ/m3 Sat HV Inf MJ/m3 Act HV Sup Kcal/m3 Dry HV Sup Kcal/m3 Sat HV Sup K
700XA Gas Chromatograph System Reference Manual OCTOBER 2010 3-9000-744 Group Basic Calculations Advanced Calculations Final Calibration Stream RF Dev Alarm Start Time GPA Z Factor GPA Real Rel Den Gas GPA Wobbe Index HV Gross BTU Dry HV Gross BTU Sat RF Dev Alarm Start Time Total Unnormalized Conc Last Analysis Component Mole % Weight % Weight % Carbon Liquid Volume % Gal/1000 SCF GPA Real Rel Den Gas HV Gross BTU Dry HV Net BTU Dry HV Gross BTU/lb Dry HV Sup MJ/m3 Dry HV Inf MJ/m3 Dry HV Sup MJ
System Reference Manual 700XA Gas Chromatograph 3-9000-744 OCTOBER 2010 Group Basic Calculations Advanced Calculations Last Analysis Stream Avg Molecular Weight Base Pressure Base Temperature GPA Z Factor GPA Real Rel Den Gas GPA Wobbe Index Gas Den lbm/1000 ft3 HV Gross BTU Dry HV Gross BTU Sat HV Net BTU Dry HV Net BTU Sat HV Gross BTU/lb Dry HV Sup MJ/m3 Dry HV Sup MJ/m3 Sat HV Inf MJ/m3 Dry HV Inf MJ/m3 Sat HV Sup MJ/kg Dry HV Inf MJ/kg Dry HV Sup Kcal/kg Dry HV Inf Kcal/kg Dry Is Cal Run Reid V
700XA Gas Chromatograph System Reference Manual OCTOBER 2010 Group 3-9000-744 Basic Calculations Last Analysis Stream (cont.
System Reference Manual 700XA Gas Chromatograph 3-9000-744 OCTOBER 2010 Group Basic Calculations Advanced Calculations Last Final Calibration RF Dev Alarm Start Time GPA Z Factor GPA Real Rel Den Gas Stream Stream No GPA Wobbe Index HV Gross BTU Dry HV Gross BTU Sat RF Dev Alarm Start Time Stream No Total Unnormalized Conc Hardware - Heaters Temperature Temperature Hardware - Valves Current Value Current Value Hardware - Discrete Inputs Current Value Current Value Hardware - Discrete O
700XA Gas Chromatograph System Reference Manual OCTOBER 2010 3-9000-744 Group Basic Calculations Advanced Calculations Application Validation Data Table n/a Nominal Value Percent Deviation Application Averages Min Max Avg Samples Min Max Avg Samples Application - User Defined Calculation Calc Result Calc Result Application - Limit Alarms Alarm Low On Alarm High On Alarm Low On Alarm High On Value Causing Alarm Violated Alarm Limit Date Application - System Alarms Alarm On Alarm On Value
System Reference Manual 700XA Gas Chromatograph 3-9000-744 OCTOBER 2010 Group Basic Calculations Advanced Calculations GC Status Current Stream Last Stream Cycle Clock Counter Cycle Time Anly Time Current Mode Next Stream Last Mode Calibration Failed Validation Failed Cycle Complete Pulse Current Day Current Month Current Year Current Hour Current Minute Current Second Active Alarm Flag UnAck Alarm Flag Current Stream Last Stream Cycle Clock Counter Cycle Time Anly Time Current Mode Next Stream Las
700XA Gas Chromatograph System Reference Manual OCTOBER 2010 3-9000-744 Group Basic Calculations Advanced Calculations Analysis Stream Avg Molecular Weight Base Pressure Start Time Total Unnormalized Conc ISO Temp RefC - Pri ISO Temp RefV - Pri ISO CV Sup Dry - Pri ISO CV Sup Sat - Pri ISO CV Inf Dry - Pri ISO CV Inf Sat - Pri ISO Gas Den kg/m3 - Pri ISO Real Rel Den Gas - Pri ISO Wobbe Index Sup - Pri ISO Wobbe Index Inf - Pri Avg Molecular Weight Base Pressure Start Time Total Unnormalized Conc I
System Reference Manual 700XA Gas Chromatograph 3-9000-744 OCTOBER 2010 Group Basic Calculations Advanced Calculations Component n/a Resp Factor Resp Factor % Dev Ret Time Ret Time % Dev Component Number Calibration Stream n/a Start Time Final Calibration Component Calib Conc Old Resp Factor New Resp Factor New RF Update Flag Resp Factor % Dev Old Ret Time New Ret Time New RT Update Flag Ret Time % Dev Component Number Calib Conc Old Resp Factor New Resp Factor New RF Update Flag Resp Factor
700XA Gas Chromatograph System Reference Manual OCTOBER 2010 3-9000-744 Group Basic Calculations Advanced Calculations Last Analysis Stream Avg Molecular Weight Base Pressure Is Cal Run Start Time Stream No Total Unnormalized Conc ISO Temp RefC - Pri ISO Temp RefV - Pri ISO CV Sup Dry - Pri ISO CV Sup Sat - Pri ISO CV Inf Dry - Pri ISO CV Inf Sat - Pri ISO Gas Den kg/m3 - Pri ISO Real Rel Den Gas - Pri ISO Wobbe Index Sup - Pri ISO Wobbe Index Inf - Pri Analysis Time Avg Molecular Weight Base Press
System Reference Manual 700XA Gas Chromatograph 3-9000-744 OCTOBER 2010 Group Basic Calculations Advanced Calculations Last Calibration Component n/a Area or Height Resp Factor Ret Time Component Number Last Calibration Stream n/a Start Time Stream No Last Final Calibration Component Calib Conc Old Resp Factor New Resp Factor New RF Update Flag Resp Factor % Dev Old Ret Time New Ret Time New RT Update Flag Ret Time % Dev Component Number Calib Conc Old Resp Factor New Resp Factor New RF Updat
700XA Gas Chromatograph System Reference Manual OCTOBER 2010 3-9000-744 Group Basic Calculations Advanced Calculations Application Component Data Table n/a Det # Ret Time Resp Fact Calib Conc RT Secs Dev Resp Fact % Sum Factor - Pri Sum Factor - Sec CV Sup KJ/mol - Pri CV Inf KJ/mol - Pri CV Sup KJ/mol - Sec CV Inf KJ/mol - Sec Rel Dens Gas Rel Dens Liquid Mole Weight Rel Resp Fact Multi-level Calib 'a' Multi-level Calib 'b' Multi-level Calib 'c' Multi-level Calib 'd' Application Validation Data T
System Reference Manual 700XA Gas Chromatograph 3-9000-744 OCTOBER 2010 Group Basic Calculations Advanced Calculations Application - Streams n/a Usage TEV Total Runs Avg Runs Start Time Interval Calibration Stream Base Pressure Optional Pressure 1 Optional Pressure 2 Optional Pressure 3 Next Cal/Val Time Status Validation Average Value Current Value Average Value Current Value GC Control Auto Sequence Halt Single Stream Calibration Validation Auto Sequence Halt Single Stream Calibration Valid
700XA Gas Chromatograph OCTOBER 2010 System Reference Manual 3-9000-744 This page is intentionally left blank.
System Reference Manual 3-9000-744 D-18 700XA Gas Chromatograph OCTOBER 2010
Appendix E, Creating custom calculations To create or edit a customized calculation using GC analysis data, do the following: 1. Select Applications → Calculations → User Defined.... The User Defined Calculations window appears, containing a list of all the userdefined calculations that are available to the GC. Figure E-1. The User Defined Calculations window 2. Click Insert before to add a row to the User Defined Calculations table. Note To delete this--or any--row from the table, click Delete. 3.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 OCTOBER 2010 4. Click Edit. The Edit User-defined Calculation window appears. Figure E-2. The Edit User-defined Calculation window A B C E D F In MON 20/20, building a calculation is similar to building a simple program. You have constants and two types of variables available, as well as two calculation-building commands.
MON20/20 Software for Gas Chromatographs OCTOBER 2010 User Manual 3-9000-745 The following is a description of the design elements of the Edit Userdefined Calculation window: • Element A - Called the Calculation Steps Viewer, this element displays the line-by-line construction of the calculation as it is being built. The following commands allow you to interact with this area: • Click Clear All to clear the content of the Calculation Steps Viewer.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 OCTOBER 2010 before being added to the Calculation Steps Viewer. There are four modes of the Expression Editor, depending upon what action is being performed: Figure E-3. Expression Editor - No Action Figure E-4. Expression Editor - Insert Comment Figure E-5. Expression Editor - Insert Condition Figure E-6.
MON20/20 Software for Gas Chromatographs User Manual OCTOBER 2010 3-9000-745 The following commands allow you to interact with the Expression Editor: • Click Clear to clear the content of the entire line. The line itself is not deleted. • Click Delete Item to delete the currently active token. Each mathematical function, numeric data, and mathematical operation is treated as a token. The token to the right of the current cursor location is treated as the currently active token.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 OCTOBER 2010 • Section 1 - This section contains the following keys: x^y x to the power of y SQRT Square Root abs Absolute Value sin Sine cos Cosine tan Tan log10 Logarithm to the base 10 log2 Logarithm to the base 2 ln Logarithm to the base e and Logical AND or Logical OR xor Logical XOR ( Open bracket ) Close bracket • Section 2 - This section contains the traditional calculator keys and can be used with your k
MON20/20 Software for Gas Chromatographs User Manual OCTOBER 2010 3-9000-745 • Constants - Allows you to select constants from a drop-down list. • Temporary Variables - Allows you to select temporary, usercreated variables from a drop-down list. • System Variables - Allows you to select system variables. • Edit Temporary Variables - Allows you to create variables. • Edit Constants - Allows you to create system-wide constants that can be used in user-defined calculations.
User Manual 3-9000-745 MON20/20 Software for Gas Chromatographs OCTOBER 2010 Figure E-8. Edit Comment mode 2. Enter the comment into the Edit Comment textbox and then click Done. The comment will be added to the Calculation Steps Viewer. Figure E-9.
MON20/20 Software for Gas Chromatographs User Manual OCTOBER 2010 E.2 3-9000-745 Inserting a conditional statement Figure E-10. An example of a conditional statement The Expression Editor in Edit Condition mode allows you to build the first line of the conditional statement: Figure E-11. The Expression Editor in Edit Condition mode Regular expression Variables/Constants Relational operator Expressions are built using the Expression Editor in Edit Expression mode.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 OCTOBER 2010 1. Click on the Insert drop-down list and select Insert Condition. A new line is added to the Calculation Steps Viewer and the Expression Editor switches to Edit Condition mode. 2. Add an expression. You can use constants, temporary variables, system variables, and the calculator functions to build the expression. For information on inserting system variables, see page E-16. For information on creating variables, see page E-15.
MON20/20 Software for Gas Chromatographs OCTOBER 2010 User Manual 3-9000-745 Figure E-13. the Variable/Constant drop-down list For information on creating variables, see page E-15. For information on creating constants, see page E-14. 5. Click Done. MON 20/20 validates the statement and if there are no errors, it adds it to the Calculation Steps Viewer. Figure E-14.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 OCTOBER 2010 Figure E-15. Edit Expression area regular expression variable To add an expression to a conditional statement or calculation, do the following: 1. Click on the Insert drop-down list and select Insert Expression. A new line is added to the Calculation Steps Viewer and the Expression Editor switches to Edit Expression mode. 2. Select a variable from the Variable drop-down tree view.
MON20/20 Software for Gas Chromatographs OCTOBER 2010 User Manual 3-9000-745 3. Add a regular expression. You can use constants, temporary variables, system variables, and the calculator functions to build the expression. For information on inserting system variables, see page E-16. For information on creating variables, see page E-15. For information on creating constants, see page E-14. Figure E-17. The Edit Expression area 4. Click Done.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 E.4 OCTOBER 2010 Creating a constant To create a constant that you can use in building a calculation, do the following: 1. From the Edit User-defined Calculation window, click Edit Constants. The Edit Constants window displays, showing all the constants that have been created so far for the GC. Figure E-19. The Edit Constants window 2. To create a new constant, click Insert before. A new row will be added to the USER_CALC_CONSTANTS table.
MON20/20 Software for Gas Chromatographs User Manual OCTOBER 2010 3-9000-745 4. Double-click the Value cell and enter a value for the constant. 5. Use the Comment cell to store information that is relevant for the constant. 6. To save the changes without closing the window, click Save. To save the changes and close the window, click OK. E.5 Creating or editing a temporary variable To create a temporary variable that you can use in building a calculation, do the following: 1.
User Manual 3-9000-745 MON20/20 Software for Gas Chromatographs OCTOBER 2010 Note To delete a variable from this window, select it in the table and click Delete. 3. Double-click the Name cell and enter a name for the variable. 4. Use the Comment cell to store information that is relevant for the variable. 5. To save the changes without closing the window, click Save. To save the changes and close the window, click OK. E.
MON20/20 Software for Gas Chromatographs OCTOBER 2010 User Manual 3-9000-745 Figure E-21. The Expression Editor E.7 Using user-defined calculations You can use a previously-created user-defined calculation when building new calculations by clicking on the System Variables drop-down arrow on the Edit User-defined Calculation window.
User Manual MON20/20 Software for Gas Chromatographs 3-9000-745 OCTOBER 2010 Figure E-22. System Variables drop-down menu For a demonstration of how to use the context-sensitive variable selector, see “Using the context-sensitive variable selector” on page 1-42. The selected system variable displays in the System Variables drop-down box and in the Expression Editor.
MON20/20 Software for Gas Chromatographs OCTOBER 2010 User Manual 3-9000-745 Figure E-23.
User Manual 3-9000-745 E-20 MON20/20 Software for Gas Chromatographs OCTOBER 2010
WARRANTY CLAIM PROCEDURES To make a warranty claim, you, the Purchaser, must: 1. Provide Daniel Measurement and Control, Inc. or Rosemount Analytical, Inc. with proof of the Date of Purchase and proof of the Date of Shipment of the product in question. 2. Return the product to Daniel Measurement Services (DMS) within 12 months of the date of original shipment of the product, or within 18 months of the date of original shipment of the product to destinations outside of the United States.
This page is intentionally left blank.
CUSTOMER REPAIR REPORT FOR SERVICE, COMPLETE THIS FORM, AND RETURN IT ALONG WITH THE AFFECTED EQUIPMENT TO CUSTOMER SERVICE AT THE ADDRESS INDICATED BELOW. COMPANY NAME: ____________________________________________________________________________ TECHNICAL CONTACT:_____________________________________ PHONE: __________________________ REPAIR P. O.
This page is intentionally left blank.
Daniel Measurement and Control, Inc., Daniel Measurement Services, Inc., and Rosemount Analytical Inc., Divisions of Emerson Process Management, reserves the right to make changes to any of its products or services at any time without prior notification in order to improve that product or service and to supply the best product or service possible. www.emersonprocess.