- Emerson Electric Co. Wok User Manual

User Manual MON20/20 Software for Gas Chromatographs
3-9000-745 SEPTEMBER 2010
3-40
3.6.6 Calibrating an analog input
To calibrate an analog input, do the following:
1. Select Analog Input... from the Hardware menu. The Analog Input
window displays.
Figure 3-29. The Analog Inputs window
2. Click on the analog input that you want to calibrate.
3. Set the analog input’s Zero Scale by entering its minimum anticipated
value.
4. Set the analog input’s Full Scale by entering your maximum
anticipated value.
5. Click AutoCal...(F4) or press F4. The Analog Input Calibration
Wizard runs.