- Emerson Electric Co. Wok User Manual

MON20/20 Software for Gas Chromatographs User Manual
SEPTEMBER 2010 3-9000-745
6-9
6.6 Stopping an Analysis Run
Note
This function forces the system into Idle mode. If Stop Now is performed while an
analysis is in progress, the components may continue to elute from the columns during.
No analysis data will be generated.
Do not perform a Stop Now unless absolutely necessary. Whenever possible, use the
Halt function.
To immediately stop an analysis run, do the following:
1. Select Stop Now... from the Control menu. A confirmation message
displays.
Figure 6-12. Confirmation message
2. Click Yes and the current analysis stops.