- Emerson Electric Co. Wok User Manual

MON20/20 Software for Gas Chromatographs User Manual
OCTOBER 2010 3-9000-745
2-27
2.4.5 Forcing a calibration
The Forced Cal command uses an archived chromatogram’s raw data to
calibrate the GC. The calculation results are stored in the component
data table for the corresponding stream number.
A major benefit of a forced calibration is increased efficiency. Using a
previously validated calibration gas chromatogram removes the necessity
for the GC to perform a calibration and validation run before performing
an analysis.
To perform a forced calibration, do the following:
1. From the Chromatogram pull-down menu, select the trace that you
want to use to calibrate the GC.
Figure 2-23. Chromatogram pull-down menu
2. Click Forced Cal.