Instruction Manual 748297-E June 2003 Model NGA2000 HFID Heated Flame Ionization Detector Analyzer Module http://www.processanalytic.
ESSENTIAL INSTRUCTIONS READ THIS PAGE BEFORE PROCEEDING! Rosemount Analytical designs, manufactures and tests its products to meet many national and international standards. Because these instruments are sophisticated technical products, you MUST properly install, use, and maintain them to ensure they continue to operate within their normal specifications.
Instruction Manual Model NGA2000 HFID 748297-E June 2003 TABLE OF CONTENTS PREFACE...........................................................................................................................................P-1 Definitions ...........................................................................................................................................P-1 Safety Summary .................................................................................................................
Instruction Manual 748297-E June 2003 4-0 4-1 4-2 4-3 ii Model NGA2000 HFID MAINTENANCE AND TROUBLESHOOTING .....................................................................4-1 Overview................................................................................................................................4-1 Component Replacement......................................................................................................4-3 a. Oven.....................................................
Instruction Manual 748297-E June 2003 Model NGA2000 HFID LIST OF ILLUSTRATIONS Figure 1-1. Figure 1-2. Figure 2-1. Figure 2-2. Figure 2-3. Figure 2-4. Figure 2-5. Figure 2-6. Figure 3-1. Figure 3-2. Figure 3-3. Figure 4-1. Figure 4-2. Figure 4-3. Figure 4-4. Figure 4-5. Figure 4-6. Figure 4-7. Figure 4-8. Figure 4-9. Figure 4-10. Figure 4-11. Figure 4-12. Figure 4-13. Figure 4-14. Figure 4-15. Figure 4-16. Figure 4-17. Figure 4-18. Figure 4-19. Figure 4-20. Figure 4-21. Figure 4-22.
Instruction Manual 748297-E June 2003 iv Contents Model NGA2000 HFID Rosemount Analytical Inc.
Instruction Manual 748297-E June 2003 Model NGA2000 HFID PREFACE The purpose of this manual is to provide information concerning the components, functions, installation and maintenance of the NGA2000 HFID Analyzer. Some sections may describe equipment not used in your configuration. The user should become thoroughly familiar with the operation of this module before operating it. Read this instruction manual completely.
Instruction Manual 748297-E June 2003 Model NGA2000 HFID SAFETY SUMMARY If this equipment is used in a manner not specified in these instructions, protective systems may be impaired. AUTHORIZED PERSONNEL To avoid explosion, loss of life, personal injury and damage to this equipment and on-site property, all personnel authorized to install, operate and service the this equipment should be thoroughly familiar with and strictly follow the instructions in this manual. SAVE THESE INSTRUCTIONS. DANGER.
Instruction Manual 748297-E June 2003 Model NGA2000 HFID WARNING. HIGH TEMPERATURE This equipment is used in the analysis of sample gases at temperatures of up to 250°C. All components and material in contact with the sample, the oven and the burner can reach this temperature level. Operate this equipment only when covers are secured. Servicing requires access to "hot" parts which can cause serious injury. Refer servicing to qualified personnel.
Instruction Manual 748297-E June 2003 Model NGA2000 HFID WARNING POSSIBLE EXPLOSION HAZARD Ensure that all gas connections are made as labeled and are leak free. Improper gas connections could result in explosion or death.
Instruction Manual Model NGA2000 HFID 748297-E June 2003 GENERAL PRECAUTIONS FOR HANDLING AND STORING HIGH PRESSURE GAS CYLINDERS Edited from selected paragraphs of the Compressed Gas Association's "Handbook of Compressed Gases" published in 1981 Compressed Gas Association 1235 Jefferson Davis Highway Arlington, Virginia 22202 Used by Permission 1. Never drop cylinders or permit them to strike each other violently. 2.
Instruction Manual 748297-E June 2003 Model NGA2000 HFID DOCUMENTATION The following NGA2000 HFID instruction materials are available. local representative to order.
Instruction Manual 748297-E June 2003 Model NGA2000 HFID GLOSSARY Analyzer Module The module that contains all sensor/detector components for development of a Primary Variable signal; includes all signal conditioning and temperature control circuitry. Backplane The interconnect circuit board which the Controller Board, Power Supply, Analyzer Module power and network cables, I/O Modules and Expansion Modules plug into. Control Module The Operator Interface plus the Controller Board.
Instruction Manual 748297-E June 2003 Model NGA2000 HFID Softkeys The five function keys located below the front panel display; they assume the function displayed directly above each on the display, a function dictated by software. System Any collection of Analyzer Module(s), Platform(s), I/O Module(s) and Expansion Module(s). P-8 Preface Rosemount Analytical Inc.
Instruction Manual 748297-E June 2003 Model NGA2000 HFID SECTION 1 DESCRIPTION AND SPECIFICATIONS 1-1 OVERVIEW This manual describes the Heated Flame Ionization Detector (HFID) Analyzer Module of Rosemount Analytical's NGA2000 Series of gas analysis components. See Figure 1-1 below and Figure 1-2 on page 3. from internal combustion engines are examples of typical applications for the HFID Analyzer Module.
Instruction Manual 748297-E June 2003 1-4 GAS SAFETY FEATURES The HFID Analyzer Module is designed with a factory-installed continuous dilution purge system in accordance with standard ANSI/NFPA 496 - 1993, Chapter 6. Front-panel LEDs indicate that the burner flame is lit and that the purge system is enabled. In addition, fuel gas is automatically shut off when a flame-out condition occurs or the safety system is disabled.
Instruction Manual 748297-E June 2003 Model NGA2000 HFID Gas Lines (See Figure 2-2) Purge Air Outlet Fuel Shutoff Solenoid Valve Detector (Burner) Oven Preamp Board (In Shield) Case Pressure Switch Air Circulation Fan Computer Board Case Heater Sensor Board Sample Flow Controller Safety Board Purge Air Flow Switch DC-DC Module Power Supply Board Burner Air Solenoid Valve Purge Air Regulator Burner Air & Fuel Regulators (Stacked Vertically) Network & Power Module Figure 1-2.
Instruction Manual 748297-E June 2003 1-5 Model NGA2000 HFID SPECIFICATIONS a. General Measurement Species ....................Total hydrocarbons Ranges (H2/He Fuel) Low range ............................0 to 10 ppm, CH4, through 0 to 1%, CH4 at an oven setpoint between 113°C and 191°C High range ...........................0 to 50 ppm, CH4, through 0 to <5%, CH4 at an oven setpoint between 113°C and 191°C Analysis Temperature .....................
Instruction Manual 748297-E June 2003 Model NGA2000 HFID b. Gas Requirements Sample .......................................... Non-flammable, below 100% of LEL Flow rate ..............................1.0 to 2.5 L/min. Supply pressure...................345 to 620 hPa-gauge (5 to 9 psig) Temperature ........................110oC to 230oC (230oF to 446oF), <20oC variance/24 hours, <10oC variance/hr. Particulates ..........................filtered to <2 microns Dewpoint..............................
Instruction Manual 748297-E June 2003 d. Model NGA2000 HFID Gas Connections Sample In ...................................... 1/4" O.D. tube fitting, stainless steel Regulated Air In ...............................1/4" O.D. tube fitting, brass Burner Air In .....................................1/4" O.D. tube fitting, brass Fuel In...............................................1/4" O.D. tube fitting, stainless steel Purge Air In ......................................3/8" O.D.
Instruction Manual 748297-E June 2003 Model NGA2000 HFID SECTION 2 INSTALLATION 2-1 2-2 UNPACKING Carefully examine the shipping carton and contents for signs of damage. Immediately notify the shipping carrier if the carton or its contents are damaged. Retain the carton and packing material until the instrument is operational. ASSEMBLY If the Analyzer Module requires assembly with other components, do so at this time.
Instruction Manual 748297-E June 2003 Model NGA2000 HFID tal. Otherwise, water may accumulate in the line, causing back pressure and noisy readings, or may back up in the line and flood the burner. Depending on the percent of water vapor in the sample, the sample bypass out connection may have condensation. Proper drainage may be required. All new external gas tubing (except for PURGE IN/OUT and SAMPLE BYPASS) is strongly recommended, preferably precleaned, stainless steel, gas chromatograph-grade tubing.
Instruction Manual 748297-E June 2003 Model NGA2000 HFID c. b. Pneumatic Connections WARNING HIGH TEMPERATURE The Sample In, Sample Bypass Out, and Burner Exhaust Out gases and fittings can reach temperatures of up to 250°C. Make connections to these fittings when the oven heater is disabled or the module is powered down.
Instruction Manual 748297-E June 2003 Model NGA2000 HFID regulator, 1723 hPa-gauge (25 psig) nominal for internal pressure. introduced through the sample inlet at the rear of the module. Burner Air Pressure should be : 1725 to 3450 hPa-gauge (25 to 50 psig) for cylinder regulator, 1035 hPa-gauge (15 psig) nominal for internal pressure. Zero Gas - Analysis is affected by the background gas of the sample.
Instruction Manual 748297-E June 2003 Model NGA2000 HFID BULKHEAD 3/8T BRASS PURGE AIR IN REGULATED AIR IN HEATED SAMPLE SAMPLE BYPASS IN OUT BURNER EXHAUST BULKHEAD 1/4T - 1/8T SS BULKHEAD 1/4T - 1/4T BRASS MANIFOLD, SAMPLE CHECK VALVE 1/3 PSIG PLUG 1/8NPT PLUG 1/4NPT DIFFUSER PURGE AIR OUT FILTER BULKHEAD FUEL 3/8T - 1/4MPT IN BRASS BULKHEAD CONNECTOR 1/4T - 1/8NPT SS FUEL RESTRICTOR MALE CONNECTOR 1/8T - 1/8NPT SS PORT CONNECTOR 1/4 TUBING 1/4 OD VITON SENSOR, SAMPLE TEMPERATURE TUBING
Instruction Manual 748297-E June 2003 Model NGA2000 HFID NETWORK1 2-5 ELECTRICAL CONNECTIONS Two electrical connections are required on the Analyzer Module: POWER and NETWORK (See Figure 2-3 below). On the Analyzer Module, two NETWORK connectors are available, either of which is appropriate for: 1) interconnection with the control module or 2) "daisy-chaining" with other NGA2000 components.
Instruction Manual 748297-E June 2003 Model NGA2000 HFID FRONT VIEW 8.2 [208] 6.8 [174] .7 [18] 3.6 [91] 1.9 [48] .6 [15] 1.1 [28] 4.3 [110] 2.8 [71] 2.0 [51] SIDE VIEW 6.0 [152] 1.1 [28] 22.5 [571] .3 [7] .7 [17] 1.4 [34] .25 [6] DIA .2 [6] 7.2 [183] 5.4 [137] REAR VIEW 4.2 [107] 4.0 [103] 2.0 [52] 1.1 [27] 1.4 [35] 3.3 [83] 1.0 [25] .9 [23] 2.4 [60] 3.2 [81] 3.1 3.0 [78] [75] Dimensions: INCHES [MM] Figure 2-5. Rosemount Analytical Inc.
2-8 Installation J1 J2 J3 P1 P2 P3 COMPUTER ANALYSIS BOARD P1 EI Figure 2-6. E2 J1 PREAMP BOARD E3 CHASSIS GND Rosemount Analytical Inc.
Instruction Manual 748297-E June 2003 Model NGA2000 HFID 2-6 • The bypass line connection must be slanted down a minimum of 6 degrees from horizontal for drainage of water condensation. • If required, thermal insulation around the bypass fitting to prevent condensation in the bypass restrictor. • If required, thermal insulation for the sample inlet connection to minimize the cold spot. • The heated line is at the correct temperature.
Instruction Manual 748297-E June 2003 2-10 Installation Model NGA2000 HFID Rosemount Analytical Inc.
Instruction Manual 748297-E June 2003 Model NGA2000 HFID SECTION 3 STARTUP AND OPERATION 3-1 OVERVIEW Prior to initial startup, the user should leak test the module as outlined in Section 2. For the remainder of this section, Analyzer Module interconnection with a control module or some interfacing component will be assumed operational. 3. Connect the LON cable(s) and the +24VDC power cable. 4. Turn power ON. 5. Check the 4 LEDs. The power green LED should be illuminated.
Instruction Manual 748297-E June 2003 Model NGA2000 HFID • EPROM test - Checks the EPROM on the Analysis Computer PCB. • RAM test - Checks the RAM on the Analysis Computer PCB. • Power supply test - Verifies that all internal DC voltages are within the required tolerances. • Network test - Checks the internal network interface. • 20 bit ADC test - Checks the 20-bit ADC on the Analysis Computer PCB by sending a DC signal through the Preamp PCB and reading the signal back with the 20-bit ADC.
Instruction Manual 748297-E June 2003 Model NGA2000 HFID Noncompliance could cause damage to the module. At the very least, the module's safety system, which requires a certain volume of purge air flowing through the case before allowing burner ignition, will not allow the instrument to operate. The lowest purge air flow/pressure setting possible during burner operation is preferable. Thus, the user should set the external purge air pressure initially at 689 hPa-gauge (10 psig).
Instruction Manual 748297-E June 2003 Model NGA2000 HFID 15. Check and re-adjust the internal pressures if required. 16. The unit is ready for operation. 3-3 3-4 BINDING To achieve full coordination between Analyzer Modules and associated I/O Modules, the user must bind those components together in the System Set Up portion of the Technical Configuration Menu in software. CALIBRATION Calibration gas setup is as follows: 3.
Instruction Manual Model NGA2000 HFID If the user is unable to calibrate the Analyzer Module (i.e., when ZERO or SPAN is initiated, nothing happens), several possible solutions present themselves. One solution relates to the use of an incorrect gas for zeroing or spanning (e.g., using a high concentration gas to zero or a zero gas to span the Analyzer Module).
Instruction Manual 748297-E June 2003 3-6 SAFETY SYSTEM The HFID Analyzer Module safety system will not allow ignition or continuous burner function unless the following conditions are present: • • 3-6 Model NGA2000 HFID The internal purge gas pressure is at least 380 hPa - gauge (5.5 psig). (Monitor display message, Purge Gas Pressure in Physical Measurements menu, for proper setting.) Flow rate for purge air in is at least 16 L/min. and case pressure is greater than 0.5 inches of water.
Instruction Manual 748297-E June 2003 Model NGA2000 HFID 1.0 0.8 0.6 AIR: 30 psig (2064 hPa) AIR: 20 psig (1376 hPa) AIR: 10 psig (688 hPa) RESPONSE (100 ppm CH4 fullscale) 0.4 SAMPLE: 100 ppm CH4 in N2 at 3.5 psig (241 hPa) 0.2 0 0 10 688 5 344 15 1032 20 1376 25 1726 30 2064 psig FUEL PRESSURE hPa Figure 3-1. Typical Curves of Module Response vs. Pressure Setting on Fuel Pressure Regulator 1.0 FUEL: 30 psig (2064 hPa) H2 0.8 FUEL: 25 psig (1726 hPa) H2 FUEL: 20 psig (1376 hPa) H2 0.
Instruction Manual 748297-E June 2003 Model NGA2000 HFID DISPLAY MESSAGE DESCRIPTION AIR FET AIR PRESS BAIR FLOW BAROMETER BFUEL FLOW BLOCK FET CASE TEMP CRUDE NOISE CURRENTRNGHI CURRENTRNGLO CURRENTSFAC FLAME TEMP FUEL PRES LIN ERROR N15 VOLTS P10 VOLTS P15 VOLTS POL VOLTS SAMP PRES CALRESULT PURGE AIR SW ERROR Table 3-1.
Instruction Manual 748297-E June 2003 Model NGA2000 HFID SECTION 4 MAINTENANCE AND TROUBLESHOOTING 1. Remove power to the unit; shut off gases and disconnect lines. Allow module to cool. WARNING QUALIFIED PERSONNEL This equipment should not be adjusted or repaired by anyone except properly qualified service personnel. 4-1 OVERVIEW This section contains instructions and procedures for troubleshooting and maintaining the HFID analyzer module.
Instruction Manual 748297-E June 2003 Model NGA2000 HFID Rear Panel Assembly 656954 Electronics Assembly 656943 DC Power Supply Module Assembly 657413 Oven Assembly 659551 Fan Assembly 657414 Fixed Flow Controller Assembly 657434 Front Panel Assembly 656949 Figure 4-2. 4-2 Maintenance and Troubleshooting Locations of Major Assemblies of the HFID Rosemount Analytical Inc.
Instruction Manual 748297-E June 2003 Model NGA2000 HFID 4-2 cal cables, noting location of mating connectors for re-installation. COMPONENT REPLACEMENT a. Note DO NOT remove the fittings from the gas lines on the detector. Oven Though the oven can be replaced as a complete unit, all internal components are field replaceable. Remove the two hex nuts securing the oven to the chassis and the two screws securing oven to the rear panel. Lift oven assembly from analyzer.
Instruction Manual 748297-E June 2003 Model NGA2000 HFID Oven Disassembly Temperature Sensor 1. Refer to Figure 4-4A on page 4-5. Remove the four retaining screws on the oven cover, remove cover. 1. Refer to Figure 4-4A on page 4-5. Remove the four screws on the oven cover, remove cover. 2. Remove the two screws and one nut securing the outer oven front panel to the outer oven, remove front panel. 2. Refer to Figure 4-5 below. Remove the burner cap retainer. 3.
Instruction Manual 748297-E June 2003 Model NGA2000 HFID A B Oven Cover 657105 (with insulation 657346) Exhaust Nuts and Ferrules (2 ea) Inner Oven Retaining Screw (2) and Washer (2) Inner Oven Inner Oven Front Panel 657352 Preamp connector nut, washer (typ 2) Outer Oven Outer Oven Front Panel 657356 C Burner 657359 Spring, Heated Bridge 904294 Heated Bridge 659555 Heater/Cable Assembly 659643 Jam Nuts 659550 Insulating Washers 073737 For clarity, outer oven not shown in Figures B and C.
Instruction Manual 748297-E June 2003 Model NGA2000 HFID Ignitor Flameout Sensor 1. Refer to Figure 4-4A on page 4-5. Remove the four screws on the oven cover, remove cover. 1. Refer to Figure 4-4A on page 4-5. Remove the four screws on the oven cover, remove cover. 2. Refer to Figure 4-5 below. Disconnect the ignitor wiring connector, note location. 2. Refer to Figure 4-5 below. Disconnect the flameout detector wiring connector, note location. 3.
Instruction Manual 748297-E June 2003 Model NGA2000 HFID c. 2. Remove burner/thermal block from oven per Section 4-2b on page 4-4. Burner Internal Components WARNING BURNER CONTAMINATION Do not handle internal parts of the burner with bare hands. All tools used for maintenance must be free of contaminates. Disassembly of Burner/Thermal Block 3. Refer to Figure 4-6 below. Disconnect sample capillary nut at base of burner. 4. Remove screw securing thermal block to burner.
Instruction Manual 748297-E June 2003 Model NGA2000 HFID 1. Remove oven from analyzer module per Section 4-2a on page 4-3. Replacing Burner Jets Disassemble the burner only if contaminants are evident. Combustion products or other contaminates which accumulate inside the burner may form electrical leakage paths between the collector and the burner contact, resulting in noisy readings. 2. Remove burner/thermal block from oven per 4-2b on page 4-4. 3.
Instruction Manual 748297-E June 2003 Model NGA2000 HFID 5. Holding burner base, lift burner cap off of assembly, set aside. 8. Refer to Figure 4-8 below. Lift air baffle out of burner base. 6. Remove gasket. 9. Remove the sample jet and gasket from the bottom of the burner base. 7. Refer to Figure 4-7B on page 4-8. Holding burner base, lift combustion chamber off, set aside. 10. Remove the jet nut. Grasp jet assembly and lift out (along with upper gasket) of burner base. Remove bottom gasket.
Instruction Manual 748297-E June 2003 Model NGA2000 HFID 4. Install air baffle per Figure 4-8 on page 4-9. Installation WARNING BURNER CONTAMINATION Do not handle internal parts of the burner with bare hands. All tools used for maintenance must be free of contaminates. 5. See Figure 4-7B on page 4-8. Insert new o-ring into burner base. 1. Install new lower gasket, jet assembly and upper gasket into burner base, finger-tight jet nut. 6.
Instruction Manual 748297-E June 2003 Model NGA2000 HFID d. Thermal Block The sample RTD can be replaced with the thermal block attached to burner and mounted in oven. The cartridge heater and thermostat are also replaceable with thermal block secured to burner, but must be removed from the oven. 2. Disconnect the sample RTD wiring connector, note location. 3. Refer to Figure 4-9 below. Remove the two screws securing the sample RTD, pull sample RTD out. 4.
Instruction Manual 748297-E June 2003 Model NGA2000 HFID 2. Remove burner/thermal block from oven per Section 4-2b on page 4-4. Thermostat 1. Remove oven from analyzer module per Section 4-2a on page 4-3. 3. Remove burner from thermal block per Section 4-2c on page 4-7. 2. Remove burner/thermal block from oven per Section 4-2b on page 4-4. 3. Refer to Figure 4-9 on page 4-11. Remove the two retaining screws, pull thermostat out. 4. Install replacement thermostat, attach with the two retaining screws. 5.
Instruction Manual 748297-E June 2003 Model NGA2000 HFID e. 1. Remove the hex nut and screw as shown in Figure 4-11 below. Electronics Assembly The electronics assembly must be removed from the chassis if replacement of any of the following components is necessary: 2. Lay electronics assembly on bench, do not disconnect cables or tubing.
Instruction Manual 748297-E June 2003 Model NGA2000 HFID Preamp Assembly 656945 Case Pressure Purge Switch 903690 Computer Analysis Board 659149 Sensor Board 657060 Safety Board 657499 Power Supply Board 655764 Case Temperature Sensor 656026 Figure 4-12. Electronics Assembly – Exploded View Printed Circuit Boards When replacing a circuit board, the following procedure is recommended: 3. One at a time, remove the wiring connectors and attach to replacement board. 1.
Instruction Manual 748297-E June 2003 Model NGA2000 HFID Case Temperature Sensor 1. Per Section 4-2e on page 4-13, remove securing hardware from electronics assembly and lay on bench. 2. Disconnect case temperature sensor cable. 3. Remove screw securing cable clamp holder to signal board. 4. Remove case temperature sensor from cable clamp holder. 5. Per Figure 4-13 below insert replacement case temperature sensor into cable clamp holder. Re-assemble to signal board mounting screw.
Instruction Manual 748297-E June 2003 Model NGA2000 HFID Case Pressure Purge Switch 3. Disconnect tube at pressure switch. 1. Per Section 4-2e on page 4-13, remove securing hardware from electronics assembly and lay on bench. 4. Remove mounting screws (2) and washers (2). 2. Disconnect the two electrical terminals, note location. Reverse procedure for installation of replacement switch.
Instruction Manual 748297-E June 2003 Model NGA2000 HFID Preamp Assembly 1. Per Section 4-2e on page 4-13, remove securing hardware from electronics assembly and lay on bench. 2. Disconnect and note location of cables. 3. Remove the two screws and washers from the top bracket and slide the preamp assembly out. 4. Remove the lower bracket from the preamp assembly and install on replacement preamp assembly. 5. Slide replacement preamp assembly into top bracket and secure with mounting hardware. 6.
Instruction Manual 748297-E June 2003 f. Model NGA2000 HFID Fan Assembly 1. Disconnect and note location of cables. 2. Remove the two hex nuts securing the fan to the chassis, lift fan assembly out. Fan Assembly 657414 Figure 4-16. 4-18 Maintenance and Troubleshooting Fan Assembly Installation Rosemount Analytical Inc.
Instruction Manual 748297-E June 2003 Model NGA2000 HFID g. Flow Controller 1. Disconnect the all tubing and wiring connectors, note locations. 2. Remove the four hex nuts securing the flow controller assembly to the analyzer module chassis. Flow Controller Assembly 657434 Figure 4-17. Rosemount Analytical Inc.
Instruction Manual 748297-E June 2003 Model NGA2000 HFID Regulated Air In Regulator 250 psi 023382 FLOW Remove red cap after installation Regulated Air Out Sample Sensor 656418 Remove and discard bracket supplied with regulator, assembly as shown. Figure 4-18. 4-20 Maintenance and Troubleshooting Flow Controller Assembly – Exploded View Rosemount Analytical Inc.
Instruction Manual 748297-E June 2003 Model NGA2000 HFID h. DC Power Supply Module 1. Disconnect and note location of all wiring to DC power supply module. Figure 4-19. Rosemount Analytical Inc. 2. Remove the two hex nuts securing module to chassis, remove module.
Instruction Manual 748297-E June 2003 i.
Instruction Manual 748297-E June 2003 Model NGA2000 HFID 4. The front panel can now be pulled away from the chassis. Replacing front panel components 1. To access components, remove the four front panel mounting screws (two on front, one on each side). Note The wiring from front panel components is still connected. Do not disconnect unless replacing that component. 2. Remove the burner air regulator and fuel regulator mounting nuts. 3. Remove the purge air regulator mounting bracket screws.
Instruction Manual 748297-E June 2003 Model NGA2000 HFID LON/Power Module 1. Disconnect wiring connectors, note locations. 2. Refer to Figure 4-20 on page 4-22. From the outside of the front panel, remove the two mounting screws. 3. Install replacement module in reverse order. LED Indicator Assembly 1. Disconnect wiring connector, note location. Burner Air Sensor 1. Disconnect wiring connector, note location. 2.
Instruction Manual 748297-E June 2003 Model NGA2000 HFID Burner Air and Fuel Regulators Purge Air Flow Switch and Diffuser 1. Disconnect the two tubes and the sensor fitting on the rear of the regulator, note locations. 1. Unscrew flow switch from tee fitting. 2. Replace the Teflon pipe thread tape on the threads of the sensor fitting. 2. Replace Teflon pipe thread tape on tee fitting. 3. Remove diffuser from flow switch and install into replacement flow switch. 3. Remove the regulator and o-ring. 4.
Instruction Manual 748297-E June 2003 Model NGA2000 HFID Air Ignite Restrictor 1. On the burner air solenoid valve: a. Disconnect the tube at the top elbow fitting. b. Disconnect tube at tee fitting. c. 4-26 Lift solenoid valve from analyzer module. Maintenance and Troubleshooting 2. Disconnect tube going to air ignite restrictor. 3. Remove restrictor from solenoid valve. 4. Add Teflon pipe thread tape to replacement restrictor, install into solenoid. 5.
Instruction Manual 748297-E June 2003 Model NGA2000 HFID j.
Instruction Manual 748297-E June 2003 Model NGA2000 HFID Fuel In 2-Way Solenoid Valve 1. Disconnect wiring solenoid valve wiring connector, note location. 2. Inside the analyzer module, disconnect the tube going to the connector on the “out” port of the solenoid valve. 3. On the rear of the analyzer module at the fuel in port: a. Disconnect the fuel in tube. b. Remove nuts and washers. 4. Remove solenoid valve from analyzer module 5.
Instruction Manual 748297-E June 2003 Model NGA2000 HFID b. 4-3 TROUBLESHOOTING CHECKLIST a. Safety System 1. Verify purge supply pressure at bulkhead is between 10 and 20 psig. 2. Check case for leaks. 3. Check burner for leaks. 4. Verify purge pressure sensor tube connection. 5. Verify purge out port is vented to atmospheric pressure. 6. Verify Safety PCB connector J2 is attached. Ignition 1. Verify that the fuel pressure/flow is correct. 2. Verify that the burner air pressure/flow is correct. 3.
Instruction Manual 748297-E June 2003 c. Model NGA2000 HFID Drift 1. Verify that the sample, burner air, and fuel supply pressures are constant. 2. Check that the tubing, regulators, pumps, fittings, and valves are clean of hydrocarbons. d. Noise 1. Check that the burner exhaust is free from water condensation. 2. Verify connection to the collector is correct. 3. Verify connection to the polarizing voltage is correct. 3. Verify that the oxygen level in the burner air and sample are constant. 4.
Instruction Manual 748297-E June 2003 Model NGA2000 HFID SECTION 5 REPLACEMENT PARTS WARNING PARTS INTEGRITY Tampering with or unauthorized substitution of components may adversely affect safety of this product. Use only factory-approved components for repair. 5-1 MATRIX Each analyzer is configured per the customer sales order. Below is the HFID sales matrix which lists the various configurations available. To identify the configuration of an analyzer, locate the analyzer name-rating plate.
Instruction Manual 748297-E June 2003 5-2 REPLACEMENT PARTS a. General 813344 903107 657029 656560 657413 657053 657414 656943 b.
Instruction Manual Model NGA2000 HFID c. 748297-E June 2003 Oven Components 659551 Oven Assembly 657359 Burner Assembly 657205 Ignitor Assembly 903736 O-Ring (Ignitor Assembly) 657063 RTD Detector 903125 Set Screw M3X0.
Instruction Manual 748297-E June 2003 5-4 Replacement Parts Model NGA2000 HFID Rosemount Analytical Inc.
Instruction Manual 748297-E June 2003 Model NGA2000 HFID SECTION 6 RETURN OF MATERIAL 6-1 RETURN OF MATERIAL If factory repair of defective equipment is required, proceed as follows: 1. Secure a return authorization from a Rosemount Analytical Inc. Sales Office or Representative before returning the equipment. Equipment must be returned with complete identification in accordance with Rosemount instructions or it will not be accepted.
Instruction Manual 748297-E June 2003 6-2 Return of Material Model NGA2000 HFID Rosemount Analytical Inc.
WARRANTY Goods and part(s) (excluding consumables) manufactured by Seller are warranted to be free from defects in workmanship and material under normal use and service for a period of twelve (12) months from the date of shipment by Seller. Consumables, glass electrodes, membranes, liquid junctions, electrolyte, o-rings, etc., are warranted to be free from defects in workmanship and material under normal use and service for a period of ninety (90) days from date of shipment by Seller.
Instruction Manual 748297-E June 2003 Model NGA2000 HFID EMERSON PROCESS MANAGEMENT Rosemount Analytical Inc. Process Analytic Division 1201 N. Main St. Orrville, OH 44667-0901 T (330) 682-9010 F (330) 684-4434 E gas.csc@emersonprocess.com EUROPEAN TECHNOLOGY CENTER Fisher-Rosemount GmbH & Co. Industriestrasse 1 63594 Hasselroth Germany T 49-6055-884 0 F 49-6055-884209 ASIA - PACIFIC Fisher-Rosemount Singapore Private Ltd.