Configuration and Use Manual MMI-20019023, Rev AA March 2012 Micro Motion® Model 1500 Transmitters with Analog Outputs Configuration and Use Manual
Safety messages Safety messages are provided throughout this manual to protect personnel and equipment. Read each safety message carefully before proceeding to the next step. Micro Motion customer service Location Telephone number Email U.S.A. 800-522-MASS (800-522-6277) (toll free) flow.support@emerson.com Canada and Latin America +1 303-527-5200 (U.S.A.) Asia Japan 3 5769-6803 All other locations +65 6777-8211 (Singapore) U.K.
Contents Contents Part I Getting Started Chapter 1 Before you begin ............................................................................................................3 1.1 1.2 1.3 1.4 Chapter 2 About this manual .......................................................................................................................3 Transmitter model code ..............................................................................................................
Contents 4.6 4.7 Chapter 5 Configure device options and preferences ....................................................................59 5.1 5.2 5.3 Chapter 6 Configure response time parameters ........................................................................................ 59 5.1.1 Configure Update Rate ............................................................................................... 59 5.1.2 Configure Calculation Speed (Response Time) .......................................
Contents Chapter 7 Completing the configuration ....................................................................................101 7.1 7.2 7.3 Test or tune the system using sensor simulation ......................................................................101 7.1.1 Sensor simulation .....................................................................................................102 Back up transmitter configuration ............................................................................
Contents 10.3 10.4 10.5 10.6 10.7 10.8 10.9 10.10 10.11 10.12 10.13 10.14 10.15 10.16 10.17 10.18 10.19 10.20 10.21 10.22 10.23 10.24 10.25 10.26 10.27 10.28 10.29 10.30 Flow measurement problems ..................................................................................................156 Density measurement problems .............................................................................................158 Temperature measurement problems ....................................................
Contents B.3 Appendix C Using the Field Communicator with the transmitter ...................................................217 C.1 C.2 C.3 Appendix D Default values and ranges ........................................................................................................235 Transmitter components and installation wiring ........................................................241 E.1 E.2 E.3 Appendix F Basic information about the Field Communicator .........................................
Contents vi Micro Motion® Model 1500 Transmitters with Analog Outputs
Getting Started Part I Getting Started Chapters covered in this part: • Before you begin • Quick start Configuration and Use Manual 1
Getting Started 2 Micro Motion® Model 1500 Transmitters with Analog Outputs
Before you begin 1 Before you begin Topics covered in this chapter: • • • • 1.1 About this manual Transmitter model code Communications tools and protocols Additional documentation and resources About this manual This manual provides information to help you configure, commission, use, maintain, and troubleshoot the Micro Motion Model 1500 transmitter.
Before you begin Table 1-1: Communications tools, protocols, and related information Communications tool Supported protocols Scope In this manual For more information ProLink II • HART/Bell 202 • Modbus/RS-485 • Service port Complete configuration Basic user information. and commissioning See Appendix A. User manual • Installed with software • On Micro Motion user documentation CD • On Micro Motion web site (www.micromotion.
Before you begin Table 1-2: Additional documentation and resources (continued) Topic Document Hazardous area installation See the approval documentation shipped with the transmitter, or download the appropriate documentation from the Micro Motion web site at www.micromotion.com. All documentation resources are available on the Micro Motion web site at www.micromotion.com or on the Micro Motion user documentation CD.
Before you begin 6 Micro Motion® Model 1500 Transmitters with Analog Outputs
Quick start 2 Quick start Topics covered in this chapter: • • • • • • 2.1 Power up the transmitter Check flowmeter status Make a startup connection to the transmitter Characterize the flowmeter (if required) Verify mass flow measurement Verify the zero Power up the transmitter The transmitter must be powered up for all configuration and commissioning tasks, or for process measurement. 1. Ensure that all transmitter and sensor covers and seals are closed.
Quick start Immediately after power-up, the transmitter runs through diagnostic routines and checks for error conditions. During the power-up sequence, Alarm A009 is active. This alarm should clear automatically when the power-up sequence is complete. 2. Check the status LED on the transmitter. Table 2-1: Transmitter status reported by status LED LED state Description Recommendation Green No alarms are active. You can continue with configuration or process measurement.
Quick start • To change the protocol, baud rate, parity, or stop bits, choose ProLink > Configuration > RS-485. • To change the address, choose ProLink > Configuration > Device. To change the communications parameters using ProLink III, choose Device Tools > Configuration > Communications. To change the communications parameters using the Field Communicator, choose On-Line Menu > Configure > Manual Setup > Inputs/Outputs > Communications.
Quick start • For curved-tube sensors, set Flow Cal (Flow Calibration Factor). 3. Set the density characterization parameters. • For straight-tube sensors, set D1, D2, DT, DTG, K1, K2, FD, DFQ1, and DFQ2. • For curved-tube sensors, set D1, D2, TC, K1, K2, and FD. (TC is sometimes shown as DT.) 2.4.1 Sources and formats for characterization parameters Different sensor tags display characterization parameters differently. On older sensors, the sensor tag may not contain all the required parameters.
Quick start Figure 2-2: Tag on newer curved-tube sensors (all sensors except T-Series) Figure 2-3: Tag on older straight-tube sensor (T-Series) Figure 2-4: Tag on newer straight-tube sensor (T-Series) Configuration and Use Manual 11
Quick start Density calibration parameters (D1, D2, K1, K2, FD, DT, TC) If your sensor tag does not show a D1 or D2 value: • For D1, enter the Dens A or D1 value from the calibration certificate. This value is the line-condition density of the low-density calibration fluid. Micro Motion uses air. If you cannot find a Dens A or D1 value, enter 0.001 g/cm3. • For D2, enter the Dens B or D2 value from the calibration certificate.
Quick start 2.5 Verify mass flow measurement Check to see that the mass flow rate reported by the transmitter is accurate. You can use any available method. • Connect to the transmitter with ProLink II and read the value for Mass Flow Rate in the Process Variables window (ProLink > Process Variables). • Connect to the transmitter with ProLink III and read the value for Mass Flow Rate in the Process Variables panel.
Quick start Important Do not verify the zero or zero the flowmeter if a high-severity alarm is active. Correct the problem, then verify the zero or zero the flowmeter. You may verify the zero or zero the flowmeter if a lowseverity alarm is active. Procedure 1. Prepare the flowmeter: a. Allow the flowmeter to warm up for at least 20 minutes after applying power. b. Run the process fluid through the sensor until the sensor temperature reaches the normal process operating temperature. c.
Quick start Important Do not verify the zero or zero the flowmeter if a high-severity alarm is active. Correct the problem, then verify the zero or zero the flowmeter. You may verify the zero or zero the flowmeter if a lowseverity alarm is active. Procedure 1. Prepare the flowmeter: a. Allow the flowmeter to warm up for at least 20 minutes after applying power. b. Run the process fluid through the sensor until the sensor temperature reaches the normal process operating temperature. c.
Quick start Table 2-2: Terminology used with zero verification and zero calibration (continued) Term Definition Live Zero The real-time bidirectional mass flow rate with no flow damping or mass flow cutoff applied. An adaptive damping value is applied only when the mass flow rate changes dramatically over a very short interval. Unit = configured mass flow measurement unit. Zero Stability A laboratory-derived value used to calculate the expected accuracy for a sensor.
Configuration and commissioning Part II Configuration and commissioning Chapters covered in this part: • Introduction to configuration and commissioning • • • • Configure process measurement Configure device options and preferences Integrate the meter with the control system Completing the configuration Configuration and Use Manual 17
Configuration and commissioning 18 Micro Motion® Model 1500 Transmitters with Analog Outputs
Introduction to configuration and commissioning 3 Introduction to configuration and commissioning Topics covered in this chapter: • • • • 3.1 Configuration flowchart Default values and ranges Disable write‐protection on the transmitter configuration Restore the factory configuration Configuration flowchart Use the following flowchart as a general guide to the configuration and commissioning process. Some options may not apply to your installation.
Introduction to configuration and commissioning Figure 3-1: Configuration flowchart Configure process measurement Configure mass flow measurement Configure fault handling parameters Configure volume flow meaurement Configure sensor parameters Volume flow type Gas Liquid Configure device options and preferences Configure device parameters Test and move to production Test or tune transmitter using sensor simulation Back up transmitter configuration Enable write-protection on transmitter configura
Introduction to configuration and commissioning 3.2 Default values and ranges See Section D.1 to view the default values and ranges for the most commonly used parameters. 3.
Introduction to configuration and commissioning 22 Micro Motion® Model 1500 Transmitters with Analog Outputs
Configure process measurement 4 Configure process measurement Topics covered in this chapter: • • • • • • • 4.1 Configure mass flow measurement Configure volume flow measurement for liquid applications Configure gas standard volume (GSV) flow measurement Configure Flow Direction Configure density measurement Configure temperature measurement Configure pressure compensation Configure mass flow measurement The mass flow measurement parameters control how mass flow is measured and reported.
Configure process measurement Options for Mass Flow Measurement Unit The transmitter provides a standard set of measurement units for Mass Flow Measurement Unit, plus one user-defined special measurement unit. Different communications tools may use different labels for the units.
Configure process measurement Overview A special measurement unit is a user-defined unit of measure that allows you to report process data, totalizer data, and inventory data in a unit that is not available in the transmitter. A special measurement unit is calculated from an existing measurement unit using a conversion factor. Procedure 1. Specify Base Mass Unit. Base Mass Unit is the existing mass unit that the special unit will be based on. 2. Specify Base Time Unit.
Configure process measurement 4.1.2 Configure Flow Damping ProLink II ProLink > Configuration > Flow > Flow Damp ProLink III Device Tools > Configuration > Process Measurement > Flow Field Communicator Configure > Manual Setup > Measurements > Flow > Flow Damping Overview Damping is used to smooth out small, rapid fluctuations in process measurement. Damping Value specifies the time period (in seconds) over which the transmitter will spread changes in the reported process variable.
Configure process measurement Table 4-2: Valid values for Flow Damping Core processor type Update Rate setting Valid damping values Standard Normal 0, 0.2, 0.4, 0.8, ... 51.2 Special 0, 0.04, 0.08, 0.16, ... 10.24 Not applicable 0, 0.2, 0.4, 0.8, ... 51.2 Enhanced Effect of Flow Damping on volume measurement Flow Damping affects volume measurement for liquid volume data. Flow Damping also affects volume measurement for gas standard volume data.
Configure process measurement Effect of Mass Flow Cutoff on volume measurement Mass Flow Cutoff does not affect volume measurement. Volume data is calculated from the actual mass data rather than the reported value. Interaction between Mass Flow Cutoff and AO Cutoff Mass Flow Cutoff defines the lowest mass flow value that the transmitter will report as measured. AO Cutoff defines the lowest flow rate that will be reported via the mA output.
Configure process measurement 4.2 Configure volume flow measurement for liquid applications The volume flow measurement parameters control how liquid volume flow is measured and reported. The volume flow measurement parameters include: • Volume Flow Type • Volume Flow Measurement Unit • Volume Flow Cutoff Restriction You cannot implement both liquid volume flow and gas standard volume flow at the same time. You must choose one or the other. 4.2.
Configure process measurement Prerequisites Before you configure Volume Flow Measurement Unit, be sure that Volume Flow Type is set to Liquid. Procedure Set Volume Flow Measurement Unit to the unit you want to use. The default setting for Volume Flow Measurement Unit is l/sec (liters per second). Tip If the measurement unit you want to use is not available, you can define a special measurement unit.
Configure process measurement Table 4-3: Options for Volume Flow Measurement Unit for liquid applications (continued) Unit description Label ProLink II ProLink III Field Communicator Imperial gallons per minute Imp gal/min Imp gal/min Impgal/min Imperial gallons per hour Imp gal/hr Imp gal/hr Impgal/h Imperial gallons per day Imp gal/day Imp gal/day Impgal/d Barrels per second(1) barrels/sec barrels/sec bbl/s Barrels per minute(1) barrels/min barrels/min bbl/min Barrels per hour(1)
Configure process measurement a. x base units = y special units b. Volume Flow Conversion Factor = x/y 4. Enter Volume Flow Conversion Factor. 5. Set Volume Flow Label to the name you want to use for the volume flow unit. 6. Set Volume Total Label to the name you want to use for the volume total and volume inventory unit. The special measurement unit is stored in the transmitter. You can configure the transmitter to use the special measurement unit at any time.
Configure process measurement Interaction between Volume Flow Cutoff and AO Cutoff Volume Flow Cutoff defines the lowest liquid volume flow value that the transmitter will report as measured. AO Cutoff defines the lowest flow rate that will be reported via the mA output. If mA Output Process Variable is set to Volume Flow Rate, the volume flow rate reported via the mA output is controlled by the higher of the two cutoff values.
Configure process measurement The GSV flow measurement parameters include: • Volume Flow Type • Standard Gas Density • Gas Standard Volume Flow Measurement Unit • Gas Standard Volume Flow Cutoff Restriction You cannot implement both liquid volume flow and gas standard volume flow at the same time. You must choose one or the other. 4.3.
Configure process measurement Note ProLink II and ProLink III provide a guided method that you can use to calculate the standard density of your gas, if you do not know it. 4.3.
Configure process measurement Table 4-4: Options for Gas Standard Volume Measurement Unit (continued) Unit description Label ProLink II ProLink III Field Communicator Normal liter per second NLPS NLPS NLPS Normal liter per minute NLPM NLPM NLPM Normal liter per hour NLPH NLPH NLPH Normal liter per day NLPD NLPD NLPD Standard cubic feet per second SCFS SCFS SCFS Standard cubic feet per minute SCFM SCFM SCFM Standard cubic feet per hour SCFH SCFH SCFH Standard cubic feet per d
Configure process measurement Base Gas Standard Volume Unit is the existing gas standard volume unit that the special unit will be based on. 2. Specify Base Time Unit. Base Time Unit is the existing time unit that the special unit will be based on. 3. Calculate Gas Standard Volume Flow Conversion Factor as follows: a. x base units = y special units b. Gas Standard Volume Flow Conversion Factor = x/y 4. Enter the Gas Standard Volume Flow Conversion Factor. 5.
Configure process measurement Procedure Set Gas Standard Volume Flow Cutoff to the value you want to use. The default value for Gas Standard Volume Flow Cutoff is 0.0. The lower limit is 0.0. There is no upper limit. Interaction between Gas Standard Volume Flow Cutoff and AO Cutoff Gas Standard Volume Flow Cutoff defines the lowest Gas Standard Volume flow value that the transmitter will report as measured. AO Cutoff defines the lowest flow rate that will be reported via the mA output.
Configure process measurement • 4.4 The frequency output will report the actual flow rate, and the actual flow rate will be used in all internal processing. If the gas standard volume flow rate drops below 10 SLPM, both outputs will report zero flow, and 0 will be used in all internal processing.
Configure process measurement Table 4-5: Options for Flow Direction (continued) Flow Direction setting Relationship to Flow Direction arrow on sensor ProLink II ProLink III Field Communicator Bidirectional Bidirectional Bi directional Appropriate when both forward and reverse flow are expected, and forward flow will dominate, but the amount of reverse flow will be significant.
Configure process measurement Figure 4-1: Effect of Flow Direction on the mA output: Lower Range Value = 0 Flow Direction = Reverse, Negate Forward 20 12 12 12 4 -x 0 mA output 20 4 x Reverse flow • • Flow Direction = Absolute Value, Bidirectional, Negate Bidirectional 20 mA output mA output Flow Direction = Forward -x Forward flow 0 4 x Reverse flow -x Forward flow 0 Reverse flow x Forward flow Lower Range Value = 0 Upper Range Value = x Figure 4-2: Effect of Flow Direction
Configure process measurement • Under conditions of reverse flow or zero flow, the mA output is 4 mA. • Under conditions of forward flow, up to a flow rate of 100 g/sec, the mA output varies between 4 mA and 20 mA in proportion to the flow rate. • Under conditions of forward flow, if the flow rate equals or exceeds 100 g/sec, the mA output will be proportional to the flow rate up to 20.5 mA, and will be level at 20.5 mA at higher flow rates.
Configure process measurement Effect of Flow Direction on frequency outputs Flow Direction affects how the transmitter reports flow values via the frequency outputs. The frequency outputs are affected by Flow Direction only if Frequency Output Process Variable is set to a flow variable.
Configure process measurement Table 4-8: Effect of the Flow Direction parameter and actual flow direction on flow values reported via digital communications Flow Direction setting Actual flow direction Forward Zero flow Reverse Forward Positive 0 Negative Reverse Positive 0 Negative Bidirectional Positive 0 Negative Absolute Value Positive(3) 0 Positive Negate Forward Negative 0 Positive Negate Bidirectional Negative 0 Positive Effect of Flow Direction on flow totals Flow Direct
Configure process measurement 4.5.1 Configure Density Measurement Unit ProLink II ProLink > Configuration > Density > Density Units ProLink III Device Tools > Configuration > Process Measurement > Density Field Communicator Configure > Manual Setup > Measurements > Density > Density Unit Overview Density Measurement Unit specifies the units of measure that will be displayed for density measurement. Procedure Set Density Measurement Unit to the option you want to use.
Configure process measurement 4.5.
Configure process measurement The default value for Slug High Limit is 5.0 g/cm3. The range is 0.0 to 10.0 g/cm3. 3. Set Slug Duration to the number of seconds that the transmitter will wait for a slug flow condition to clear before performing the configured slug flow action. The default value for Slug Duration is 0.0 seconds. The range is 0.0 to 60.0 seconds.
Configure process measurement Procedure Set Density Damping to the value you want to use. The default value is 1.6 seconds. The range depends on the core processor type and the setting of Update Rate, as shown in the following table: Core processor type Update Rate setting Density Damping range Standard Normal 0 to 51.2 seconds Special 0 to 10.24 seconds Not applicable 0 to 40.
Configure process measurement Density Damping controls the rate of change in the density process variable. Added Damping controls the rate of change reported via the mA output. If mA Output Process Variable is set to Density, and both Density Damping and Added Damping are set to non-zero values, density damping is applied first, and the added damping calculation is applied to the result of the first calculation. 4.5.
Configure process measurement 4.6.1 Configure Temperature Measurement Unit ProLink II ProLink > Configuration > Temperature > Temp Units ProLink III Device Tools > Configuration > Process Measurement > Temperature Field Communicator Configure > Manual Setup > Measurements > Temperature > Temperature Unit Overview Temperature Measurement Unit specifies the unit that will be used for temperature measurement. Procedure Set Temperature Measurement Unit to the option you want to use.
Configure process measurement Overview Damping is used to smooth out small, rapid fluctuations in process measurement. Damping Value specifies the time period (in seconds) over which the transmitter will spread changes in the reported process variable. At the end of the interval, the reported process variable will reflect 63% of the change in the actual measured value. Procedure Enter the value you want to use for Temperature Damping. The default value is 4.8 seconds. The range is 0.0 to 76.8 seconds.
Configure process measurement Prerequisites You will need the flow factor, density factor, and calibration pressure values for your sensor. • For the flow factor and density factor, see the product data sheet for your sensor. • For the calibration pressure, see the calibration sheet for your sensor. If the data is unavailable, use 20 PSI. Procedure 1. Choose View > Preferences and ensure that Enable External Pressure Compensation is checked. 2. Choose ProLink > Configuration > Pressure. 3.
Configure process measurement Option Setup Polling for pressure a. Ensure that the primary mA output has been wired to support HART polling. b. Choose ProLink > Configuration > Polled Variables. c. Choose an unused polling slot. d. Set Polling Control to Poll As Primary or Poll as Secondary, and click Apply. e. Set External Tag to the HART tag of the external pressure device. f. Set Variable Type to Pressure. Tip • Poll as Primary: No other HART masters will be on the network.
Configure process measurement The flow factor is the percent change in the flow rate per PSI. When entering the value, reverse the sign. Example: If the flow factor is 0.000004 % per PSI, enter −0.000004 % per PSI. 5. Enter Density Factor for your sensor. The density factor is the change in fluid density, in g/cm3/PSI. When entering the value, reverse the sign. Example: If the density factor is 0.000006 g/cm3/PSI, enter −0.000006 g/cm3/PSI. 6. 7.
Configure process measurement 9. If you want to use digital communications, click Apply, then perform the necessary host programming and communications setup to write temperature data to the transmitter at appropriate intervals. Postrequisites If you are using an external pressure value, verify the setup by checking the External Pressure value displayed in the Inputs area of the main window. 4.7.
Configure process measurement Option Setup A user-configured static pressure value a. Set Pressure Unit to the desired unit. b. Set Compensation Pressure to the desired value. Polling for pressure a. Ensure that the primary mA output has been wired to support HART polling. b. Choose Online > Configure > Manual Setup > Measurements > External Pressure/Temperature > External Polling. c. Set Poll Control to Poll As Primary Host or Poll as Secondary Host. d. Choose an unused polling slot. e.
Configure process measurement Table 4-13: Options for Pressure Measurement Unit (continued) Unit description Label ProLink II ProLink III Field Communicator Millimeters water @ 68 °F mm Water @ 68°F mm Water @ 68°F mmH2O Millimeters mercury @ 0 °C mm Mercury @ 0°C mm Mercury @ 0°C mmHg Inches mercury @ 0 °C In Mercury @ 0°C In Mercury @ 0°C inHG Pounds per square inch PSI PSI psi Bar bar bar bar Millibar millibar millibar mbar Grams per square centimeter g/cm2 g/cm2 g/Sqcm K
Configure process measurement 58 Micro Motion® Model 1500 Transmitters with Analog Outputs
Configure device options and preferences 5 Configure device options and preferences Topics covered in this chapter: • • • 5.1 Configure response time parameters Configure alarm handling Configure informational parameters Configure response time parameters You can configure the rate at which process data is polled and process variables are calculated. Response time parameters include: 5.1.
Configure device options and preferences Procedure 1. Set Update Rate as desired. Option Description Normal All process data is polled at the rate of 20 times per second (20 Hz). All process variables are calculated at 20 Hz. This option is appropriate for most applications. Special A single, user-specified process variable is polled at the rate of 100 times per second (100 Hz). Other process data is polled at 6.25 Hz). Some process, diagnostic, and calibration data is not polled.
Configure device options and preferences Table 5-1: Special mode and process variable updates Always polled and updated • • • • • • • • • • • • Mass flow Volume flow Gas standard volume flow Density Temperature Drive gain LPO amplitude Status [contains Event 1 and Event 2 (basic events)] Raw tube frequency Mass total Volume total Gas standard volume total 5.1.
Configure device options and preferences 5.2 Option Description Normal Transmitter calculates process variables at the standard speed. Special Transmitter calculates process variables at a faster speed. Configure alarm handling The alarm handling parameters control the transmitter’s response to process and device conditions. Alarm handling parameters include: 5.2.
Configure device options and preferences Tip ProLink II allows you to set Fault Timeout in two locations. However, there is only one parameter, and the same setting is applied to all outputs. 5.2.
Configure device options and preferences Option Description Informational Actions when fault is detected: • The alarm is posted to the Alert List. • The status LED (if available) changes to red or yellow (depending on alarm severity). Actions when alarm clears: • The status LED (if available) returns to green and may or may not flash.
Configure device options and preferences Table 5-2: Status alarms and Status Alarm Severity (continued) Alarm code Status message Default severity Notes Configurable? A023 Internal Totals Corrupt (Core Processor) Fault Applies only to flowmeters with the No standard core processor. A024 Program Corrupt (Core Processor) Fault Applies only to flowmeters with the No standard core processor.
Configure device options and preferences Table 5-2: Status alarms and Status Alarm Severity (continued) Alarm code Status message Default severity Notes Configurable? A109 Basic Event 2 On Informational Applies only to basic events. Yes A110 Frequency Output Saturated Informational Can be set to either Informational or Ignore, but cannot be set to Fault. Yes A111 Frequency Output Fixed Informational Can be set to either Informational or Ignore, but cannot be set to Fault.
Configure device options and preferences • 5.3.
Configure device options and preferences 5.3.3 Configure Date ProLink II ProLink > Configuration > Device > Message ProLink III Device Tools > Configuration > Informational Parameters > Transmitter Field Communicator Configure > Manual Setup > Info Parameters > Transmitter Info > Date Overview Date lets you store a static date (not updated by the transmitter) in transmitter memory. This parameter is not used in processing and is not required.
Configure device options and preferences Overview Sensor Material lets you store the type of material used for your sensor’s wetted parts in transmitter memory. This parameter is not used in processing and is not required. Procedure 1. Obtain the material used for your sensor’s wetted parts from the documents shipped with your sensor, or from a code in the sensor model number. To interpret the model number, refer to the product data sheet for your sensor. 2. 5.3.
Configure device options and preferences To interpret the model number, refer to the product data sheet for your sensor. 2. 70 Set Sensor Flange Type to the appropriate option.
Integrate the meter with the control system 6 Integrate the meter with the control system Topics covered in this chapter: • • • • • • 6.
Integrate the meter with the control system Postrequisites For each channel that you configured, perform or verify the corresponding input or output configuration. When the configuration of a channel is changed, the channel’s behavior will be controlled by the configuration that is stored for the selected input or output type, and the stored configuration may not be appropriate for your process. After verifying channel and output configuration, return the control loop to automatic control. 6.
Integrate the meter with the control system Procedure Set mA Output Process Variable as desired. The default setting is Mass Flow Rate. Options for mA Output Process Variable The transmitter provides a basic set of options for mA Output Process Variable, plus several application-specific options. Different communications tools may use different labels for the options.
Integrate the meter with the control system • URV is the value of mA Output Process Variable represented by an output of 20 mA. The default value for URV depends on the setting of mA Output Process Variable. Enter URV in the measurement units that are configured for mA Output Process Variable. Tips For best performance: • Set LRV ≥ LSL (lower sensor limit). • Set URV ≤ USL (upper sensor limit). • Set these values so that the difference between URV and LRV is ≥ Min Span (minimum span).
Integrate the meter with the control system Overview AO Cutoff (Analog Output Cutoff) specifies the lowest mass flow rate, volume flow rate, or gas standard volume flow rate that will be reported through the mA output. Any flow rates below AO Cutoff will be reported as 0. Restriction AO Cutoff is applied only if mA Output Process Variable is set to Mass Flow Rate, Volume Flow Rate, or Gas Standard Volume Flow Rate.
Integrate the meter with the control system Result: • • 6.2.4 If the mass flow rate drops below 15 g/sec but not below 10 g/sec: - The mA output will report zero flow. - The frequency output will report the actual flow rate. If the mass flow rate drops below 10 g/sec, both outputs will report zero flow.
Integrate the meter with the control system Table 6-3: Valid values for Added Damping (continued) Update rate in effect Setting of Update Rate Process variable Valid values for Added Damping Special 100 Hz variable (if assigned to the mA output) 100 Hz 0.0, 0.04, 0.12, 0.30, 0.64, 1.32, 2.6, 5.4, 11, 22, 44, 88, 176, 350 100 Hz variable (if not assigned to the mA output) 6.25 Hz 0.0, 0.32, 0.96, 2.40, 5.12, 10.56, 20.8, 43.
Integrate the meter with the control system Note For some faults only: If Last Measured Value Timeout is set to a non-zero value, the transmitter will not implement the fault action until the timeout has elapsed. Procedure Set mA Output Fault Action to the desired value. 1. The default setting is Downscale. 2. If you set mA Output Fault Action to Upscale or Downscale, set mA Output Fault Level as desired.
Integrate the meter with the control system The frequency output parameters include: • Frequency Output Polarity • Frequency Output Scaling Method • Frequency Output Maximum Pulse Width • Frequency Output Fault Action and Frequency Output Fault Value Restriction The process variable assigned to the primary mA output is automatically assigned to the frequency output. You cannot assign a different process variable.
Integrate the meter with the control system Table 6-5: Options for Frequency Output Polarity (continued) 6.3.
Integrate the meter with the control system Rate Factor The maximum flow rate that you want the frequency output to report. Above this rate, the transmitter will report A110: Frequency Output Saturated.
Integrate the meter with the control system 6.3.
Integrate the meter with the control system 6.3.
Integrate the meter with the control system CAUTION! If you set mA Output Fault Action or Frequency Output Fault Action to None, be sure to set Digital Communications Fault Action to None. If you do not, the output will not report actual process data, and this may result in measurement errors or unintended consequences for your process. Restriction If you set Digital Communications Fault Action to NAN, you cannot set mA Output Fault Action or Frequency Output Fault Action to None.
Integrate the meter with the control system The default setting for Discrete Output Source is Flow Direction.
Integrate the meter with the control system Configure Flow Switch parameters ProLink II ProLink III • ProLink > Configuration > Flow > Flow Switch Setpoint • ProLink > Configuration > Flow > Flow Switch Variable • ProLink > Configuration > Flow > Flow Switch Hysteresis Device Tools > Configuration > I/O > Outputs > Discrete Output Field Communicator • Configure > Manual Setup > Inputs/Outputs > Discrete Output > Flow Switch Source • Configure > Manual Setup > Inputs/Outputs > Discrete Output >
Integrate the meter with the control system Overview Discrete outputs have two states: ON (active) and OFF (inactive). Two different voltage levels are used to represent these states. Discrete Output Polarity controls which voltage level represents which state. Procedure Set Discrete Output Polarity as desired. The default setting is Active High.
Integrate the meter with the control system Figure 6-1: Typical discrete output circuit A. B. C. D. 6.4.3 15 V (Nom) 3.
Integrate the meter with the control system Procedure Set Discrete Output Fault Action as desired. The default setting is None.
Integrate the meter with the control system 6.5.1 Configure a basic event ProLink II ProLink > Configuration > Events ProLink III Device Tools > Configuration > Events > Basic Events Field Communicator Not available Overview A basic event is used to provide notification of process changes. A basic event occurs (is ON) if the real-time value of a user-specified process variable moves above (HI) or below (LO) a user-defined setpoint. You can define up to two basic events.
Integrate the meter with the control system defined setpoints. You can define up to five enhanced events. For each enhanced event, you can assign one or more actions that the transmitter will perform if the enhanced event occurs. Procedure 1. Select the event that you want to configure. 2. Specify Event Type. Options Description HI x>A The event occurs when the value of the assigned process variable (x) is greater than the setpoint (Setpoint A), endpoint not included.
Integrate the meter with the control system Options for Enhanced Event Action Table 6-11: Options for Enhanced Event Action Action Label ProLink II ProLink III Field Communicator None (default) None None None Start sensor zero Start Sensor Zero Start Sensor Zero Perform auto zero Start/stop all totalizers Start/Stop All Totalization Start/Stop All Totalization Start/stop totals Reset mass total Reset Mass Total Reset Mass Total Reset mass total Reset volume total Reset Volume Total Re
Integrate the meter with the control system 6.6.1 Configure HART/Bell 202 communications ProLink II ProLink > Configuration > Device > Digital Comm Settings ProLink III Device Tools > Configuration > Communications > Communications (HART) Field Communicator Configure > Manual Setup > Inputs/Outputs > Communications Overview HART/Bell 202 communications parameters support HART communication with the transmitter's primary mA terminals over a HART/Bell 202 network.
Integrate the meter with the control system Tip In typical installations, burst mode is disabled. Enable burst mode only if another device on the network requires burst mode communication. 4. (Optional) Configure HART Variables.
Integrate the meter with the control system Label ProLink II 3. ProLink III Field Communicator Description Dynamic vars & PV Process Variables/ current Current Process variables/ current The transmitter sends PV, SV, TV, and QV values in measurement units and the PV’s actual milliamp reading in each burst (e.g., 50 g/sec, 23 °C, 50 g/sec, 0.0023 g/cm3, 11.8 mA). Transmitter vars Fld dev var The transmitter sends four userspecified process variables in each burst.
Integrate the meter with the control system Table 6-12: Options for HART variables (continued) Process variable Primary Varia- Secondary ble (PV) Variable (SV) Third Variable (TV) Fourth Variable (QV ) Mass total ✓ Line (Gross) Volume total ✓ Mass inventory ✓ Line (Gross) Volume inventory ✓ ✓ Gas standard volume flow rate ✓ ✓ ✓ Gas standard volume total ✓ Gas standard volume inventory ✓ Interaction of HART variables and transmitter outputs The HART variables are automatically reported t
Integrate the meter with the control system • Protocol • Modbus Address (Slave Address) • Parity, Stop Bits, and Baud Rate • Floating-Point Byte Order • Additional Communications Response Delay Restriction To configure Floating-Point Byte Order or Additional Communications Response Delay, you must use ProLink II. Procedure 1. Set Disable Modbus ASCII as desired. Support for Modbus ASCII limits the set of addresses that are available for the transmitter's Modbus address. 2.
Integrate the meter with the control system Table 6-14: Bit structure of floating-point bytes Byte Bits Definition 1 SEEEEEEE S=Sign E=Exponent 2 EMMMMMMM E=Exponent M=Mantissa 6. 3 MMMMMMMM M=Mantissa 4 MMMMMMMM M=Mantissa (Optional) Set Additional Communications Response Delay in “delay units.” A delay unit is 2/3 of the time required to transmit one character, as calculated for the port currently in use and the character transmission parameters. Valid values range from 1 to 255.
Integrate the meter with the control system Options for Digital Communications Fault Action Table 6-15: Options for Digital Communications Fault Action Label Description ProLink II ProLink III Field Communicator Upscale Upscale Upscale • Process variable values indicate that the value is greater than the upper sensor limit. • Totalizers stop incrementing. Downscale Downscale Downscale • Process variable values indicate that the value is greater than the upper sensor limit.
Integrate the meter with the control system 100 Micro Motion® Model 1500 Transmitters with Analog Outputs
Completing the configuration 7 Completing the configuration Topics covered in this chapter: • • • 7.1 Test or tune the system using sensor simulation Back up transmitter configuration Enable write‐protection on the transmitter configuration Test or tune the system using sensor simulation Use sensor simulation to test the system's response to a variety of process conditions, including boundary conditions, problem conditions, or alarm conditions, or to tune the loop.
Completing the configuration Option Required values Sine Period Minimum Maximum 4. For density, set Wave Form as desired and enter the required values. Option Required values Fixed Fixed Value Sawtooth Period Minimum Maximum Sine Period Minimum Maximum 5. For temperature, set Wave Form as desired and enter the required values. Option Required values Fixed Fixed Value Sawtooth Period Minimum Maximum Sine Period Minimum Maximum 7.1.1 6.
Completing the configuration • All mass flow rate, temperature, and density values shown on the display or reported via outputs or digital communications • The mass total and mass inventory values • All volume calculations and data, including reported values, volume totals, and volume inventories • All mass, temperature, density, or volume values logged to Data Logger Sensor simulation does not affect any diagnostic values.
Completing the configuration 7.3 Enable write-protection on the transmitter configuration ProLink II ProLink > Configuration > Device > Enable Write Protection ProLink III Device Tools > Configuration > Write-Protection Field Communicator Configure > Manual Setup > Info Parameters > Transmitter Info > Write Protect Overview If the transmitter is write-protected, the configuration is locked and nobody can change it until it is unlocked.
Operations, maintenance, and troubleshooting Part III Operations, maintenance, and troubleshooting Chapters covered in this part: • • • Transmitter operation Measurement support Troubleshooting Configuration and Use Manual 105
Operations, maintenance, and troubleshooting 106 Micro Motion® Model 1500 Transmitters with Analog Outputs
Transmitter operation 8 Transmitter operation Topics covered in this chapter: • • • • • • • • 8.
Transmitter operation 8.2 View process variables ProLink II ProLink > Process Variables ProLink III View the desired variable on the main screen under Process Variables. See Section 8.2.1 for more information. Field Communicator Overview > Shortcuts > Variables > Process Variables Overview Process variables provide information about the state of the process fluid, such as flow rate, density, and temperature, as well as running totals.
Transmitter operation Table 8-1: Status LED states (continued) 8.4 LED behavior Alarm condition Description Solid red Active high-severity alarm Alarm condition that will cause measurement error (outputs in fault) View and acknowledge status alarms The transmitter posts status alarms whenever a process variable exceeds its defined limits or the transmitter detects a fault condition. You can view active alarms, and you can acknowledge alarms. 8.4.
Transmitter operation 8.4.2 View and acknowledge alerts using ProLink III You can view a list containing all alerts that are active, or inactive and have been unacknowleged. From this list, you can acknowlege individual alerts or choose to acknowledge all alerts at once. 1. View alerts on the ProLink III main screen under Alerts.
Transmitter operation • 8.4.4 To refresh the list of active or unacknowledged alarms, press Service Tools > Alerts > Refresh Alerts. Alarm data in transmitter memory The transmitter maintains three sets of data for every alarm that is posted.
Transmitter operation Tip You can use the inventories to keep a running total of mass or volume across multiple totalizer resets. 8.
Transmitter operation 8.
Transmitter operation Overview When you reset an inventory, the transmitter sets its value to 0. It does not matter whether the inventory is started or stopped. If the inventory is started, it continues to track process measurement. Tip When you reset a single inventory, the values of other inventories are not reset. Totalizer values are not reset. Prerequisites To use ProLink II or ProLink III to reset the inventories, the feature must be enabled. • To enable inventory reset in ProLink II: 1.
Measurement support 9 Measurement support Topics covered in this chapter: • • • • • • • 9.1 Options for measurement support Use Smart Meter Verification Zero the flowmeter Validate the meter Perform a (standard) D1 and D2 density calibration Perform a D3 and D4 density calibration (T‐Series sensors only) Perform temperature calibration Options for measurement support Micro Motion provides several measurement support procedures to help you evaluate and maintain your flowmeter's accuracy.
Measurement support 9.2.1 Smart Meter Verification requirements To use Smart Meter Verification, the transmitter must be paired with an enhanced core processor, and the Smart Meter Verification option must be ordered for the transmitter. See Table 9‐1 for the minimum version of the transmitter, enhanced core processor, and communication tool needed to support Smart Meter Verification. Table 9-1: Minimum version to support Smart Meter Verification Item Minimum version Transmitter 6.
Measurement support Smart Meter Verification has an output mode called Continuous Measurement that allows the transmitter to keep measuring while the test is in progress. If you choose to run the test in Last Measured Value or Fault modes instead, the transmitter outputs will be held constant for the two minute duration of the test. If control loops depend on transmitter outputs, take appropriate action. Avoid process instability during the test.
Measurement support Postrequisites View the test results and take any appropriate actions. Run a Smart Meter Verification test using ProLink III 1. Choose Device Tools > Diagnostics > Meter Verification > Run Test. You may need to wait a few seconds while ProLink II synchronizes its database with the transmitter data. 2. Enter any desired information on the Test Definition screen, and click Next. All information on this screen is optional. 3. 4. Choose the desired output behavior.
Measurement support Option Description Outputs Held at Last Value During the test, all outputs will report the last measured value of their assigned process variable. The test will run for approximately 140 seconds. Outputs Held at Fault During the test, all outputs will go to their configured fault action. The test will run for approximately 140 seconds. Test progress is displayed on the screen. Postrequisites View the test results and take any appropriate actions. 9.2.
Measurement support 2. (Optional) Click Next to view and print a test report. 3. (Optional) Click Export Data to CSV File to save the data to a file on your PC. View test result data using ProLink III 1. Choose Device Tools > Diagnostics > Meter Verification and click Previous Test Results. The chart shows test results for all tests stored in the ProLink III database. 2. (Optional) Click Next to view and print a test report. 3.
Measurement support • If the meter fails the second test, the flow tubes may be damaged. Use your process knowledge to determine the possibilities for damage and the appropriate actions for each. These actions might include removing the meter from service and physically inspecting the tubes. At minimum, you should perform a flow validation and a density calibration. Abort A problem occurred with the meter verification test (e.g., process instability) or you stopped the test manually.
Measurement support 2. To schedule a single test or the first test in recurring execution, specify a value for Hours Until Next Run. 3. To schedule recurring execution, specify a value for Hours Between Recurring Runs. 4. To disable scheduled execution: • To disable execution of a single scheduled test, set Hours Until Next Run to 0. • To disable recurring execution, set Hours Between Recurring Runs to 0. • To disable all scheduled execution, click Turn Off Schedule.
Measurement support Important In most cases, the factory zero is more accurate than the field zero. Do not zero the flowmeter unless one of the following is true: 9.3.1 • The zero is required by site procedures. • The stored zero value fails the Zero Verification procedure. Zero the flowmeter using the zero button Zeroing the flowmeter establishes a baseline for process measurement by analyzing the sensor's output when there is no flow through the sensor tubes.
Measurement support • Set Zero Time to a lower value, then retry. • If the zero continues to fail, contact Micro Motion. Tip You can restore the factory zero using a communications tool such as ProLink II. Restore the factory zero only if your flowmeter was purchased as a unit, it was zeroed at the factory, and you are using the original components. This function requires the enhanced core processor. 9.3.
Measurement support Postrequisites Restore normal flow through the sensor by opening the valves. Need help? If the zero fails: • Ensure that there is no flow through the sensor, then retry. • Remove or reduce sources of electromechanical noise, then retry. • Set Zero Time to a lower value, then retry. • If the zero continues to fail, contact Micro Motion.
Measurement support 4. Modify Zero Time, if desired. Zero Time controls the amount of time the transmitter takes to determine its zeroflow reference point. The default Zero Time is 20 seconds. For most applications, the default Zero Time is appropriate. 5. Click Calibrate Zero. The Calibration in Progress message is displayed. When the calibration is complete: • If the zero procedure was successful, a Calibration Success message and a new zero value are displayed.
Measurement support e. Observe the drive gain, temperature, and density readings. If they are stable, check the Live Zero or Field Verification Zero value. If the average value is close to 0, you should not need to zero the flowmeter. 2. Press Service Tools > Maintenance > Zero Calibration > Perform Auto Zero. 3. Modify Zero Time, if desired. Zero Time controls the amount of time the transmitter takes to determine its zeroflow reference point. The default Zero Time is 20 seconds.
Measurement support Overview Meter validation compares flowmeter measurements reported by the transmitter to an external measurement standard. If the transmitter value for mass flow, volume flow, or density measurement is significantly different from the external measurement standard, you may want to adjust the corresponding meter factor. The flowmeter’s actual measurement is multiplied by the meter factor, and the resulting value is reported and used in further processing.
Measurement support 3. Configure the meter factor in the transmitter. Example: Calculating the meter factor for mass flow The flowmeter is installed and validated for the first time. The mass flow measurement from the transmitter is 250.27 lb. The mass flow measurement from the reference device is 250 lb. The mass flow meter factor is calculated as follows: MeterFactorMassFlow = 1 x 250 = 0.9989 250.27 The first meter factor for mass flow is 0.9989. One year later, the flowmeter is validated again.
Measurement support 9.5 3. Ensure that the calculated meter factor is between 0.8 and 1.2, inclusive. If the meter factor is outside these limits, contact Micro Motion customer service. 4. Configure the meter factor for volume flow in the transmitter. Perform a (standard) D1 and D2 density calibration Density calibration establishes the relationship between the density of the calibration fluids and the signal produced at the sensor.
Measurement support Procedure See Figure 9‐1.
Measurement support • If LD Optimization is enabled on your meter, disable it. To do this, choose Device Tools > Configuration > LD Optimization. LD Optimization is used only with large sensors in hydrocarbon applications. In some installations, only Micro Motion customer service has access to this parameter. If this is the case, contact Micro Motion before continuing. • The calibrations must be performed without interruption, in the order shown.
Measurement support 9.5.3 Perform a D1 and D2 density calibration using the Field Communicator Prerequisites • During density calibration, the sensor must be completely filled with the calibration fluid, and flow through the sensor must be at the lowest rate allowed by your application. This is usually accomplished by closing the shutoff valve downstream from the sensor, then filling the sensor with the appropriate fluid.
Measurement support Figure 9-3: D1 and D2 density calibration using the Field Communicator D1 calibration Close shutoff valve downstream from sensor D2 calibration Fill sensor with D1 fluid On-Line Menu > Service Tools > Maintenance > Density Calibration Fill sensor with D2 fluid Service Tools > Maintenance > Density Calibration Dens Pt 2 Dens Pt 1 Calibration method executes Calibration method executes Enter density of D2 fluid Enter density of D1 fluid OK OK Calibration in Progress message Dens
Measurement support 9.6.1 • Perform the D3 calibration if you have one calibrated fluid. • Perform both the D3 and D4 calibrations if you have two calibrated fluids (other than air and water). The calibrations must be performed without interruption, in the order shown. Make sure that you are prepared to complete the process without interruption.
Measurement support Figure 9-4: D3 or D3 and D4 density calibration using ProLink II D3 calibration Close shutoff valve downstream from sensor D4 calibration Fill sensor with D3 fluid Fill sensor with D4 fluid ProLink Menu > Calibration > Density cal – Point 3 ProLink Menu > Calibration > Density cal – Point 4 Enter density of D3 fluid Enter density of D4 fluid Do Cal Do Cal Calibration in Progress light turns red Calibration in Progress light turns red Calibration in Progress light turns gree
Measurement support • - Minimum difference of 0.1 g/cm3 between the density of the D4 fluid and the density of the D3 fluid. The density of the D4 fluid must be greater than the density of the D3 fluid. - Minimum difference of 0.1 g/cm3 between the density of the D4 fluid and the density of water. The density of the D4 fluid may be either greater or less than the density of water. Before performing the calibration, record your current calibration parameters.
Measurement support • • • For D3 density calibration, the D3 fluid must meet the following requirements: - Minimum density of 0.6 g/cm3 - Minimum difference of 0.1 g/cm3 between the density of the D3 fluid and the density of water. The density of the D3 fluid may be either greater or less than the density of water. For D4 density calibration, the D4 fluid must meet the following requirements: - Minimum density of 0.6 g/cm3 - Minimum difference of 0.
Measurement support Figure 9-6: D3 or D3 and D4 density calibration using the Field Communicator D3 calibration Close shutoff valve downstream from sensor D4 calibration Fill sensor with D3 fluid Fill sensor with D4 fluid Service Tools > Maintenance > Density Calibration On-Line Menu > Service Tools > Maintenance > Density Calibration Dens Pt 4 T-Series Dens Pt 3 T-Series Calibration method executes Calibration method executes Enter density of D4 fluid Enter density of D3 fluid OK OK Calibration i
Measurement support Important Consult Micro Motion before performing a temperature calibration. Under normal circumstances, the temperature circuit is stable and should not need an adjustment. Procedure See Figure 9‐7 and Figure 9‐8.
Measurement support Figure 9-8: Temperature calibration using ProLink III Temperature Offset calibration Temperature Slope calibration Fill sensor with lowtemperature fluid Fill sensor with hightemperature fluid Wait until sensor achieves thermal equilibrium Wait until sensor achieves thermal equilibrium Device Tools > Calibration > Temperature Calibration > Temperature Calibration - Offset Device Tools > Calibration > Temperature Calibration > Temperature Calibration - Slope Enter temperature of l
Measurement support 142 Micro Motion® Model 1500 Transmitters with Analog Outputs
Troubleshooting 10 Troubleshooting Topics covered in this chapter: • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • Status LED states Status alarms Flow measurement problems Density measurement problems Temperature measurement problems Milliamp output problems Frequency output problems Use sensor simulation for troubleshooting Check power supply wiring Check sensor‐to‐transmitter wiring Check grounding Perform loop tests Check for radio frequency interference (RFI) Check the HART communication
Troubleshooting 10.1 Status LED states The status LED on the transmitter indicates whether or not alarms are active. If alarms are active, view the alarm list to identify the alarms, then take appropriate action to correct the alarm condition. Table 10-1: Status LED states 10.
Troubleshooting Table 10-2: Status alarms and recommended actions (continued) Alarm code Description Recommended actions A003 No Sensor Response The transmitter is not receiving one or more basic electrical signals from the sensor. This could mean that the wiring between the sensor and the transmitter has been damaged, or that the sensor requires factory service. 1. Check the drive gain and pickoff voltage. (See Section 10.26 and Section 10.27.) 2.
Troubleshooting Table 10-2: Status alarms and recommended actions (continued) Alarm code Description Recommended actions A005 Mass Flow Rate Overrange The sensor is signaling a flow rate that is out of range for the sensor. 1. If other alarms are present, resolve those alarm conditions first. If the current alarm persists, continue with the recommended actions. 2. Check your process conditions against the values reported by the flowmeter. 3. Check for slug flow (two-phase flow). a.
Troubleshooting Table 10-2: Status alarms and recommended actions (continued) Alarm code Description Recommended actions A008 Density Overrange The sensor is signaling a density reading below 0 g/cm3 or above 10 g/cm3.
Troubleshooting Table 10-2: Status alarms and recommended actions (continued) Alarm code Description Recommended actions A010 Calibration Failure This alarm is typically caused by flow through the sensor during the zero, or by a zero offset result that is out of range. Power to the transmitter must be cycled to clear this alarm. 1. Cycle power to the meter. 2. Make sure there is no flow through the sensor. 3. Retry the zero calibration. 4. Power-cycle the transmitter.
Troubleshooting Table 10-2: Status alarms and recommended actions (continued) Alarm code Description Recommended actions A016 Sensor RTD Failure The sensor RTD is signaling a resistance that is out of range for the sensor. 1. Check the wiring between the sensor and the transmitter. a. Using the installation manual for your transmitter, verify that the transmitter is connected to the sensor according to the instructions. Obey all safety messages when opening wiring compartments. b.
Troubleshooting Table 10-2: Status alarms and recommended actions (continued) Alarm code Description Recommended actions A019 RAM Error (Transmitter) Power to the transmitter must be cycled to clear this alarm. 1. Check that all wiring compartment covers are installed properly. 2. Check that the wiring connected to the transmitter meets specifications and that shields are properly terminated. 3. Check that the sensor and transmitter are both grounded properly. 4.
Troubleshooting Table 10-2: Status alarms and recommended actions (continued) Alarm code Description Recommended actions A026 Sensor/Transmitter Communications Failure The transmitter has lost communication with the core processor on the sensor. This alarm can be an indication of a problem with the core or the transmitter requiring the replacement of one or both parts. 1. Check the wiring between the sensor and the transmitter. a.
Troubleshooting Table 10-2: Status alarms and recommended actions (continued) Alarm code Description Recommended actions A031 Low Power The core processor on the sensor is not receiving sufficient power. Check the wiring between the transmitter and the sensor. Power to the transmitter must be cycled to clear this alarm. 1. Using the installation manual for your transmitter, verify that the transmitter is connected to the sensor according to the instructions.
Troubleshooting Table 10-2: Status alarms and recommended actions (continued) Alarm code Description Recommended actions A100 mA Output 1 Saturated The calculated mA output value is outside of the meter's configured range. 1. Check the Upper Range Value and Lower Range Value parameters. Are they set correctly? 2. Check your process conditions against the values reported by the flowmeter. 3. Verify that the measurement units are configured correctly for your application. 4. Purge the flow tubes. 5.
Troubleshooting Table 10-2: Status alarms and recommended actions (continued) Alarm code Description Recommended actions A110 Frequency Output Saturated The calculated frequency output is outside the configured range. 1. Check the Frequency Output Scaling Method parameter. 2. Check your process conditions against the values reported by the flowmeter. 3. Verify process conditions, checking especially for air in the flow tubes, tubes not filled, foreign material in the tubes, or coating in the tubes. 4.
Troubleshooting Table 10-2: Status alarms and recommended actions (continued) Alarm code Description Recommended actions A117 Density Overrange (Petroleum) 1. Check your process conditions against the values reported by the flowmeter. 2. Verify the configuration of the petroleum measurement table type and density. A118 Discrete Output 1 Fixed The discrete output has been configured to send a constant value.
Troubleshooting 10.3 Flow measurement problems Table 10-3: Flow measurement problems and recommended actions Problem Possible causes Recommended actions Flow indication at no flow conditions or zero offset • Misaligned piping (especially in new installations) • Open or leaking valve • Incorrect sensor zero • Verify that all of the characterization parameters match the data on the sensor tag. • If the flow reading is not excessively high, review the live zero.
Troubleshooting Table 10-3: Flow measurement problems and recommended actions (continued) Problem Possible causes Recommended actions Erratic non-zero flow rate when flow is steady • • • • • • Slug flow Damping value too low Plugged or coated flow tube Output wiring problem Problem with receiving device Wiring problem • Verify that the sensor orientation is appropriate for your application (refer to the sensor installation manual). • Check the drive gain and the pickoff voltage. See Section 10.
Troubleshooting 10.4 Density measurement problems Table 10-4: Density measurement problems and recommended actions Problem Possible causes Recommended actions Inaccurate density reading • • • • • • • • • • Check the wiring between the sensor and transmitter. See Section 10.10. • Check grounding. See Section 10.11. • Check your process conditions against the values reported by the flowmeter. • Verify that all of the characterization parameters match the data on the sensor tag. • Check for slug flow.
Troubleshooting 10.5 Temperature measurement problems Table 10-5: Temperature measurement problems and recommended actions Problem Possible causes Recommended actions Temperature reading significantly different from process temperature • RTD failure • Wiring problem • Check junction box for moisture or verdigris. • Perform RTD resistance checks and check for shorts to case (see Section 10.28.1). • Confirm the temperature calibration factor matches the value on the sensor tag.
Troubleshooting 10.6 Milliamp output problems Table 10-6: Milliamp output problems and recommended actions Problem Possible causes Recommended actions No mA output • Wiring problem • Circuit failure • Channel not configured for desired output • Check the power supply and power supply wiring. See Section 10.9. • Check the mA output wiring. • Check the Fault Action settings. See Section 10.19. • Measure DC voltage across output terminals to verify that the output is active. • Contact Micro Motion.
Troubleshooting Table 10-6: Milliamp output problems and recommended actions (continued) Problem Possible causes mA output consistently out of range • Incorrect process variable or units assigned • Verify the output variable assignments. to output • Verify the measurement units configured • Fault condition if fault action is set to upfor the output. scale or downscale • Check the Fault Action settings. See • LRV and URV are not set correctly Section 10.19.
Troubleshooting Table 10-7: Frequency output problems and recommended actions Problem Possible causes Recommended actions No frequency output • Stopped totalizer • Process condition below cutoff • Fault condition if fault action is set to internal zero or downscale • Slug flow • Flow in reverse direction from configured flow direction parameter • Bad frequency receiving device • Output level not compatible with receiving device • Bad output circuit • Incorrect internal/external power configuration • In
Troubleshooting For more information on using sensor simulation using ProLink II, see Section 7.1. 10.9 Check power supply wiring If the power supply wiring is damaged or improperly connected, the transmitter may not receive enough power to operate properly. Prerequisites You will need the installation manual for your transmitter. Procedure 1. Before inspecting the power supply wiring, disconnect the power source.
Troubleshooting Procedure 1. Before opening the wiring compartments, disconnect the power source. CAUTION! If the transmitter is in a hazardous area, wait five minutes after disconnecting the power. 10.11 2. Verify that the transmitter is connected to the sensor according to the information provided in your transmitter installation manual. 3. Verify that the wires are making good contact with the terminals. 4. Check the continuity of all wires from the transmitter to the sensor.
Troubleshooting Procedure 1. Test the mA output(s). a. Choose ProLink > Test > Fix Milliamp 1 or ProLink > Test > Fix Milliamp 2. b. Enter 4 mA in Set Output To. c. Click Fix mA. d. Read the mA current at the receiving device and compare it to the transmitter output. The readings do not need to match exactly. If the values are slightly different, you can correct the discrepancy by trimming the output. e. Click UnFix mA. f. Enter 20 mA in Set Output To. g. Click Fix mA. h.
Troubleshooting 10.12.2 • If the mA output reading was significantly off (±200 microamps), or if at any step the reading was faulty, verify the wiring between the transmitter and the remote device, and try again. • If the discrete output reading is reversed, check the setting of Discrete Output Polarity. Perform loop tests using ProLink III Prerequisites Before performing a loop test, configure the channels for the transmitter inputs and outputs that will be used in your application.
Troubleshooting e. Click UnFix FO. 3. Test the discrete output(s). a. Choose Device Tools > Diagnostics > Testing > Discrete Output Test. b. Set Fix To: to ON. c. Verify the signal at the receiving device. d. Set Fix To: to OFF. e. Verify the signal at the receiving device. f. Click UnFix. 10.12.3 Perform loop tests using the Field Communicator Tip Loop tests are not required. However, Micro Motion recommends performing a loop test for every input or output available on your transmitter.
Troubleshooting 2. Test the frequency output(s). Note If the Weights & Measures application is enabled on the transmitter, it is not possible to perform a loop test of the frequency output, even when the transmitter is unsecured. a. Press Service Tools > Simulate > Simulate Outputs > Frequency Output Test, and choose the frequency output level. b. Read the frequency signal at the receiving device and compare it to the transmitter output. c. Choose End. 3. Test the discrete output(s). a.
Troubleshooting 10.14 - Terminate the shielding at the output device. If this is impossible, terminate the shielding at the cable gland or conduit fitting. - Do not terminate the shielding inside the wiring compartment. - 360-degree termination of shielding is unnecessary. Check the HART communication loop If you cannot establish or maintain HART communications, the HART loop may be wired incorrectly.
Troubleshooting When HART Address is changed, some configuration tools will automatically change Loop Current Mode. Tip Always verify Loop Current Mode after setting or changing HART Address. Procedure 1. Set HART Address as appropriate for your HART network. The default address is 0. This is the recommended value unless the transmitter is in a multidrop network. 2. 10.16 Set Loop Current Mode to Enabled. Check HART burst mode HART burst mode can cause the transmitter to output unexpected values.
Troubleshooting 10.19 Check mA Output Fault Action mA Output Fault Action controls the behavior of the mA output if the transmitter encounters an internal fault condition. If the mA output is reporting a constant value below 4 mA or above 20 mA, the transmitter may be in a fault condition. 1. Check the status alarms for active fault conditons. 2. If there are active fault conditions, the transmitter is performing correctly.
Troubleshooting • For the relevant status alarms, change the setting of Alarm Severity to Ignore. 3. 10.23 If there are no active fault conditions, continue troubleshooting. Check Flow Direction If Flow Direction is set inappropriately for your process, the transmitter may report unexpected flow values or totals. The Flow Direction parameter interacts with actual flow direction to affect flow values, flow totals and inventories, and output behavior.
Troubleshooting 3. Monitor the density of your process fluid output under normal process conditions. 4. Check the settings of Slug Low Limit, Slug High Limit, and Slug Duration. Tip You can reduce the occurrence of slug flow alarms by setting Slug Low Limit to a lower value, Slug High Limit to a higher value, or Slug Duration to a higher value. 10.26 Check the drive gain Excessive or erratic drive gain may indicate any of a variety of process conditions, sensor problems, or configuration problems.
Troubleshooting Table 10-8: Possible causes and recommended actions for excessive (saturated) drive gain (continued) Possible cause Recommended actions Incorrect sensor characteriza- Verify the characterization parameters. tion Erratic drive gain Table 10-9: Possible causes and recommended actions for erratic drive gain Possible cause Recommended actions Wrong K1 characterization constant for sensor Verify the K1 characterization parameter.
Troubleshooting Table 10-10: Possible causes and recommended actions for low pickoff voltage Possible cause Recommended actions Air entrainment • Increase the inlet or back pressure at the sensor. • If a pump is located upstream from the sensor, increase the distance between the pump and sensor. • The sensor may need to be reoriented. Consult the installation manual for your sensor for recommended orientations.
Troubleshooting Table 10-11: Possible causes and recommended actions for electrical shorts (continued) 10.28.1 Possible cause Recommended action Liquid or moisture inside the sensor case Contact Micro Motion. Internally shorted feedthrough Contact Micro Motion. Faulty cable Replace the cable. Improper wire termination Verify wire terminations inside sensor junction box. The Micro Motion document titled 9‐Wire Flowmeter Cable Prepa‐ ration and Installation Guide may offer some assistance.
Troubleshooting Note The CMF400 fixed resistor applies only to certain specific CMF400 releases. Contact Micro Motion for more information. There should be no open circuits, that is, no infinite resistance readings. The left pickoff and right pickoff readings should be the same or very close (±5 Ω). If there are any unusual readings, repeat the coil resistance tests at the sensor junction box to eliminate the possibility of faulty cable. The readings for each coil pair should match at both ends. 4.
Troubleshooting 10.29 Check the core processor LED The core processor has an LED that indicates different meter conditions. 1. Maintain power to the transmitter. 2. Remove the core processor lid. The core processor is intrinsically safe and can be opened in all environments. 3. Check the state of the core processor LED. Postrequisites To return to normal operation, replace the core processor lid. Important When reassembling the meter components, be sure to grease all O-rings. 10.29.
Troubleshooting Table 10-13: Standard core processor LED states (continued) LED state Description Recommended actions Core processor internal failure The meter requires factory service. Table 10-14: Enhanced core processor LED states LED state Description Recommended action Solid green Normal operation No action required. Flashing yellow Zero in progress No action required. Solid yellow Low-severity alarm Check alarm status. Solid red High-severity alarm Check alarm status.
Troubleshooting 5. Terminal pair Function Expected resistance 3–4 RS-485/A and RS-485/B 40 kΩ to 50 kΩ 2–3 VDC– and RS-485/A 20 kΩ to 25 kΩ 2–4 VDC– and RS-485/B 20 kΩ to 25 kΩ If any resistance measurements are lower than specified, the core processor may not be able to communicate with a transmitter or a remote host. The meter may need factory service. Postrequisites To return to normal operation: 1. Reconnect the 4-wire cable between the core processor and the transmitter. 2.
Using ProLink II with the transmitter Appendix A Using ProLink II with the transmitter Topics covered in this appendix: • • • A.1 Basic information about ProLink II Connect with ProLink II Menu maps for ProLink II Basic information about ProLink II ProLink II is a software tool available from Micro Motion. It runs on a Windows platform and provides complete access to transmitter functions and data.
Using ProLink II with the transmitter ProLink II messages As you use ProLink II with a Micro Motion transmitter, you will see a number of messages and notes. This manual does not document all of these messages and notes. Important The user is responsible for responding to messages and notes and complying with all safety messages. A.
Using ProLink II with the transmitter • A.2.2 You cannot make concurrent connections if the connections use the same terminals. You can make concurrent connections if the connections use different terminals.
Using ProLink II with the transmitter 4. Start ProLink II. 5. Choose Connection > Connect to Device. 6. Set Protocol to Service Port. Tip Service port connections use standard connection parameters and a standard address. You do not need to configure them here. 7. Set the COM Port value to the PC COM port that you are using for this connection. 8. If required, power-cycle the transmitter to set the terminals to service port mode.
Using ProLink II with the transmitter Procedure 1. Attach the signal converter to the serial port or USB port on your PC. 2. To connect directly to the transmitter terminals: a. Connect the leads from the signal converter to terminals 21 and 22. Tip HART connections are not polarity-sensitive. It does not matter which lead you attach to which terminal. b. Add resistance as necessary. Important HART/Bell 202 connections require a voltage drop of 1 VDC.
Using ProLink II with the transmitter Figure A-3: Connection over local loop E A D R3 R2 C R1 B A. B. C. D. E. PC Signal converter Any combination of resistors R1, R2, and R3 as necessary to meet HART communication resistance requirements DCS or PLC Transmitter Note This figure shows a serial port connection. USB connections are also supported. 4. To connect over a HART multidrop network: a. Attach the leads from the signal converter to any point on the network. b. Add resistance as necessary.
Using ProLink II with the transmitter Figure A-4: Connection over multidrop network D B A C A. B. C. D. Signal converter 250–600 Ω resistance Devices on the network Master device 5. Start ProLink II. 6. Choose Connection > Connect to Device. 7. Set Protocol to HART Bell 202. Tip HART/Bell 202 connections use standard connection parameters. You do not need to configure them here. 8. If you are using a USB signal converter, enable Converter Toggles RTS. 9.
Using ProLink II with the transmitter 12. Option Description Primary Use this setting if no other host is on the network. The Field Communicator is not a host. Click Connect. Need help? If an error message appears: • Verify the HART address of the transmitter. • Ensure that you have specified the correct COM port. • Check the physical connection between the PC and the transmitter. • Increase or decrease resistance. • Ensure that there is no conflict with another HART master. A.2.
Using ProLink II with the transmitter Figure A-5: Connection to transmitter terminals A C B A. PC B. Signal converter C. Transmitter Note This figure shows a serial port connection. USB connections are also supported. 3. To connect over the RS-485 network: a. Attach the leads from the signal converter to any point on the network. b. Add resistance as necessary. Figure A-6: Connection over network A D E C B A. B. C. D. E.
Using ProLink II with the transmitter 6. Set the connection parameters to the values configured in the transmitter. If your transmitter has not been configured, use the default values shown here.
Using ProLink II with the transmitter A.
Using ProLink II with the transmitter Figure A-8: Main menu (continued) ProLink Tools Gas Unit Configurator Meter Verification Entrained Gas Analyzer Commissioning Wizard Proving Wizard Marine Bunker Transfer Options Plug-ins Data Logging* Enable/Disable Custody Transfer Configuration Output Levels Process Variables Status Alarm Log Diagnostic Information Calibration Test API Totalizer Control CM Totalizer Control Totalizer Control Core Processor Diagnostics Finger Print API Process Variables CM Proces
Using ProLink II with the transmitter Figure A-9: Configuration menu ProLink > Configuration Additional configuration options Flow • Flow Direction • Flow Damp • Flow Cal • Mass Flow Cutoff • Mass Flow units • Mass Factor • Dens Factor • Vol Factor • Flow Switch Variable • Flow Switch Setpoint • Flow Switch Hysteresis • Vol Flow Cutoff • Vol Flow Units • Vol Flow Type • Std Gas Vol Flow Units • Std Gas Vol Flow Cutoff • Std Gas Density • Gas Wizard Configuration and Use Manual Density • Dens Units • De
Using ProLink II with the transmitter Figure A-10: Configuration menu (continued) ProLink > Configuration Additional configuration options Analog Output • Primary Variable is • Lower Range Value • Upper Range Value • AO Cutoff • AO Added Damp • Lower Sensor Limit • Upper Sensor Limit • Min Span • AO Fault Action • AO Fault Level • Last Measured Value Timeout • Valve Control Options Frequency/Discrete output • Frequency • Tertiary Variable • Freq Factor • Rate Factor • Freq Pulse Width • Last Measured Valu
Using ProLink II with the transmitter Figure A-11: Configuration menu (continued) ProLink > Configuration Additional configuration options Temperature • Temp Units • Temp Cal Factor • Temp Damping • External Temperature • External RTD Configuration and Use Manual Pressure • Flow Factor • Dens Factor • Cal Pressure • Pressure Units • External Pressure 195
Using ProLink II with the transmitter Figure A-12: Configuration menu (continued) ProLink > Configuration Additional configuration options Device • Model • Manufacturer • Hardware Rev • Distributor • Software Rev • ETO • CP Software Rev • CP ETO • Option Board • Firmware Checksum • CP Firmware Checksum • Tag • Date • Descriptor • Message • Sensor type • Transmitter Serial • Floating PT Ordering • Add Comm Resp Delay • Restore Factory Configuration • Digital Comm Fault Setting • HART Address • Enable Loop C
Using ProLink II with the transmitter Figure A-13: Configuration menu (continued) ProLink > Configuration Additional configuration options RS-485 • Protocol • Parity • Baud Rate • Stop Bits Events Event 1/2 • Variable • Type • Setpoint Discrete Events • Event Name • Event Type • Process Variable • Low Setpoint (A) • High Setpoint (B) Alarm • Alarm • Severity Figure A-14: Configuration menu (continued) ProLink > Configuration Additional configuration options Special Units • Base Mass Unit • Base Mas
Using ProLink II with the transmitter Figure A-15: Configuration menu (continued) ProLink > Configuration Variable mapping • PV is • SV is • TV is • QV is 198 System • Weights and Measures Approval • Software Rev • Totalizer Reset Options Sensor • Sensor s/n • Sensor Model • Sensor Matl • Liner Matl • Flange Micro Motion® Model 1500 Transmitters with Analog Outputs
Using ProLink III with the transmitter Appendix B Using ProLink III with the transmitter Topics covered in this appendix: • • • B.1 Basic information about ProLink III Connect with ProLink III Menu maps for ProLink III Basic information about ProLink III ProLink III is a configuration and service tool available from Micro Motion. It runs on a Windows platform and provides complete access to transmitter functions and data.
Using ProLink III with the transmitter ProLink III messages As you use ProLink III with a Micro Motion transmitter, you will see a number of messages and notes. This manual does not document all of these messages and notes. Important The user is responsible for responding to messages and notes and complying with all safety messages. B.
Using ProLink III with the transmitter • B.2.2 You cannot make concurrent connections if the connections use the same terminals. You can make concurrent connections if the connections use different terminals.
Using ProLink III with the transmitter 4. Start ProLink III. 5. Choose Connect to Physical Device. 6. Set Protocol to Service Port. Tip Service port connections use standard connection parameters and a standard address. You do not need to configure them here. 7. Set the PC Port value to the PC COM port that you are using for this connection. 8. If required, power-cycle the transmitter to set the terminals to service port mode.
Using ProLink III with the transmitter Procedure 1. Attach the signal converter to the serial port or USB port on your PC. 2. To connect directly to the transmitter terminals: a. Connect the leads from the signal converter to terminals 21 and 22. Tip HART connections are not polarity-sensitive. It does not matter which lead you attach to which terminal. b. Add resistance as necessary. Important HART/Bell 202 connections require a voltage drop of 1 VDC.
Using ProLink III with the transmitter Figure B-3: Connection over local loop E A D R3 R2 C R1 B A. B. C. D. E. PC Signal converter Any combination of resistors R1, R2, and R3 as necessary to meet HART communication resistance requirements DCS or PLC Transmitter Note This figure shows a serial port connection. USB connections are also supported. 4. To connect over a HART multidrop network: a. Attach the leads from the signal converter to any point on the network. b. Add resistance as necessary.
Using ProLink III with the transmitter Figure B-4: Connection over multidrop network D B A C A. B. C. D. Signal converter 250–600 Ω resistance Devices on the network Master device 5. Start ProLink III. 6. Choose Connect to Physical Device. 7. Set Protocol to HART Bell 202. Tip HART/Bell 202 connections use standard connection parameters. You do not need to configure them here. 8. If you are using a USB signal converter, enable Toggle RTS. 9.
Using ProLink III with the transmitter 12. Option Description Primary Use this setting if no other host is on the network. The Field Communicator is not a host. Click Connect. Need help? If an error message appears: • Verify the HART address of the transmitter. • Ensure that you have specified the correct COM port. • Check the physical connection between the PC and the transmitter. • Increase or decrease resistance. • Ensure that there is no conflict with another HART master. B.2.
Using ProLink III with the transmitter Figure B-5: Connection to transmitter terminals A C B A. PC B. Signal converter C. Transmitter Note This figure shows a serial port connection. USB connections are also supported. 3. To connect over the RS-485 network: a. Attach the leads from the signal converter to any point on the network. b. Add resistance as necessary. Figure B-6: Connection over network A D E C B A. B. C. D. E.
Using ProLink III with the transmitter 6. Set the connection parameters to the values configured in the transmitter. If your transmitter has not been configured, use the default values shown here.
Using ProLink III with the transmitter B.
Using ProLink III with the transmitter Figure B-9: Configuration: I/O Figure B-10: Configuration: Events 210 Micro Motion® Model 1500 Transmitters with Analog Outputs
Using ProLink III with the transmitter Figure B-11: Configuration: Communications Figure B-12: Configuration: Informational Parameters Configuration and Use Manual 211
Using ProLink III with the transmitter Figure B-13: Device Tools: Calibration Figure B-14: Calibration: Density Calibration 212 Micro Motion® Model 1500 Transmitters with Analog Outputs
Using ProLink III with the transmitter Figure B-15: Calibration: Temperature Calibration Figure B-16: Device Tools: Configuration Transfer Configuration and Use Manual 213
Using ProLink III with the transmitter Figure B-17: Diagnostics: Testing Figure B-18: Diagnostics: Meter Verification 214 Micro Motion® Model 1500 Transmitters with Analog Outputs
Using ProLink III with the transmitter Figure B-19: Device Tools: Trending Configuration and Use Manual 215
Using ProLink III with the transmitter 216 Micro Motion® Model 1500 Transmitters with Analog Outputs
Using the Field Communicator with the transmitter Appendix C Using the Field Communicator with the transmitter Topics covered in this appendix: • • • C.1 Basic information about the Field Communicator Connect with the Field Communicator Menu maps for the Field Communicator Basic information about the Field Communicator The Field Communicator is a handheld configuration and management tool that can be used with a variety of devices, including Micro Motion transmitters.
Using the Field Communicator with the transmitter If Micro Motion is not listed, or you do not see the required device description, use the Field Communicator Easy Upgrade Utility to install the device description, or contact Micro Motion. Field Communicator menus and messages Many of the menus in this manual start with the On-Line menu. Ensure that you are able to navigate to the On-Line menu. As you use the Field Communicator with a Micro Motion transmitter, you will see a number of messages and notes.
Using the Field Communicator with the transmitter Figure C-1: Field Communicator connection to transmitter terminals B A C A. Field Communicator B. 250–600 Ω resistance C. Transmitter terminals 2. To connect to a point in the local HART loop, attach the leads from the Field Communicator to any point in the loop and add resistance as necessary. The Field Communicator must be connected across a resistance of 250–600 Ω. Figure C-2: Field Communicator connection to local HART loop C B A A.
Using the Field Communicator with the transmitter Figure C-3: Field Communicator connection to multidrop network D B C A. B. C. D. A Field Communicator 250–600 Ω resistance Devices on the network Master device 4. Turn on the Field Communicator and wait until the main menu is displayed. 5. If you are connecting across a multidrop network: a. Set the Field Communicator to poll. The device returns all valid addresses. b. Enter the HART address of the transmitter. The default HART address is 0.
Using the Field Communicator with the transmitter Figure C-4: On-Line menu On-Line Menu Configuration and Use Manual 1 Overview 1 Check Status 2 Primary Purpose Variables 3 Shortcuts 2 Configure 1 Manual Setup 2 Alert Setup 3 Service Tools 1 Alerts 2 Variables 3 Trends 4 Maintenance 5 Simulate 221
Using the Field Communicator with the transmitter Figure C-5: Overview menu On-Line Menu > 1 Overview 1 2 Check Status 1 Refresh Alerts 2 Dev Status: 3 Comm Status: Primary Purpose Variables Mass Flow Rate Volume Flow Rate Density * 222 Displayed only if meter verification is enabled. 3 Shortcuts 1 Device Information 2 Totalizer Control 3 Zero Calibration 4 Variables 5 Trends 6 Meter Verification * 1 Device Information 1 Identification 2 Revisions 3 Mat.
Using the Field Communicator with the transmitter Figure C-6: Configure menu On-Line Menu > 1 Configure Configuration and Use Manual 1 Manual Setup 1 Characterize 2 Measurements 3 Inputs/Outputs 4 Info Parameters 5 Communications 2 Alert Setup 1 Configure Alerts 2 Discrete Output 3 Discrete Events 223
Using the Field Communicator with the transmitter Figure C-7: Manual Setup menu On-Line Menu > 2 Configure > 1 Manual Setup 1 Characterize 1 Sensor Type 2 Sensor Tag Parameters 2 Measurements 1 Flow 2 Density 3 Temperature 4 Update Rate 5 LD Optimization 6 Special Units 7 External Pressure/Temperature 8 GSV 4 3 Inputs/Outputs 1 Set Up Channels 2 Set Up mA Output 3 Set Up Frequency Output 4 Set Up Discrete Output 5 Set Up RS-485 Port 6 Map Variables 224 5 Info Parameters 1 Transmitter Info 2 Sensor
Using the Field Communicator with the transmitter Figure C-8: Manual Setup menu: Characterize On-Line Menu > 2 Configure > 1 Manual Setup > 1 Characterize 1 Sensor Type 1 Curved Tube 2 Straight Tube Curved Tube Sensor Tag Parameters 1 FlowCal 2 D1 3 D2 4 TC 5 K1 6 K2 7 FD Configuration and Use Manual 2 Sensor Type Straight Tube Sensor Tag Parameters 1 Flow Parameters 2 Density Parameters 1 2 Flow Parameters 1 Flow FCF 2 FTG 3 FFQ Density Parameters 1 D1 2 D2 3 DT 4 DTG 5 K1 6 K2 7 FD 8 DFQ1 9
Using the Field Communicator with the transmitter Figure C-9: Manual Setup menu: Measurements On-Line Menu > 2 Configure > 1 Manual Setup > 2 Measurements Flow 1 Flow Direction 2 Flow Damping 3 Mass Flow Unit 4 Mass Flow Cutoff 5 Volume Flow Unit * 6 Volume Flow Cutoff * 7 Mass Factor 8 Volume Factor 1 Density 1 Density Unit 2 Density Damping 3 Density Cutoff * 4 Density Factor 5 Slug Duration 6 Slug Low Limit 7 Slug High Limit 2 Temperature 1 Temperature Unit 2 Temp Damping 3 4 Update Rate 1 Updat
Using the Field Communicator with the transmitter Figure C-10: Manual Setup menu: I/O On-Line Menu > 2 Configure > 1 Manual Setup > 4 Inputs/Outputs Additional options 1 Set Up Channels 1 Channel A 2 Channel C 2 2 3 Set Up Frequency Output 1 FO Settings 2 FO Fault Parameters 3 FO Scaling Channel C 1 Frequency Output 2 Discrete Output 1 FO Settings 1 Third Variable 2 Max Pulse Width 3 FO Polarity 2 FO Fault Parameters 1 Third Variable 2 FO Fault Action 3 FO Fault Level 3 FO Scaling * 1 FO Scalin
Using the Field Communicator with the transmitter Figure C-11: Manual Setup menu: I/O (continued) On-Line Menu > 2 Configure > 1 Manual Setup > 4 Inputs/Outputs 228 4 Discrete Output 1 DO Assignment 2 DO Polarity 3 DO Fault Action 4 Flow Switch Source 5 Flow Switch Setpoint 6 Hysteresis (0.1-10.
Using the Field Communicator with the transmitter Figure C-12: Alert Setup menu On-Line Menu > 2 Configure > 2 Alert Setup 1 2 Configure Alerts 1 Fault Timeout 2 MAO Fault Action 3 MAO Fault Level 4 FO Fault Action 5 FO Fault Level 6 Comm Fault Action 7 Set Up Alert Severity 8 View Alert Severity Discrete Output 1 DO Assignment 2 DO Polarity 3 DO Fault Action 4 Flow Switch Source 5 Flow Switch Setpoint 6 Hysteresis (0.1.10.
Using the Field Communicator with the transmitter Figure C-13: Service Tools menu On-Line Menu > 3 Service Tools 1 2 3 230 Alerts 1 Refresh Alerts Alert Name Additional Information for Above 4 Maintenance 1 Routine Maintenance 2 Zero Calibration 3 Density Calibration 4 Temperature Calibration 5 Diagnostic Variables 5 Simulate 1 Simulate Outputs 2 Simulate Sensor Variables 1 Variable Summary 2 Process Variables 3 Mapped Variables 4 External Variables 5 Totalizer Control 6 Outputs * If Volume Flo
Using the Field Communicator with the transmitter Figure C-14: Service Tools menu: Variables On-Line Menu > 3 Service Tools > 2 Variables 1 Variable Summary 2 Process Variables 1 Mass Flow Rate 2 Volume Flow Rate * 3 Density 4 Temperature 3 Mapped Variables 1 PV Mass Flow Rate 2 SV Mass Flow Rate 3 TV Mass Flow Rate 4 QV Mass Flow Rate 4 External Variables 1 External Temperature 2 External Pressure * 5 If Volume Flow Type = GSV, GSV variables are displayed.
Using the Field Communicator with the transmitter Figure C-15: Service Tools menu: Maintenance On-Line Menu > 3 Service Tools > 4 Maintenance 1 Routine Maintenance 1 Trim mA Output 2 Meter Verification * 2 4 Meter Verification ** 1 Run Meter Verification 2 View Test Results 3 Schedule Meter Verification 5 2 3 Zero Calibration 1 Mass Flow Rate 2 Volume Flow Rate 3 Zero Time 4 Zero Value 5 Standard Deviation 6 Perform Auto Zero 7 Restore Factory Zero Density Calibration 1 Density 2 Dens Pt1 (Air) 3
Using the Field Communicator with the transmitter Figure C-16: Service Tools menu: Simulate On-Line Menu > 3 Service Tools > 5 Simulate Configuration and Use Manual 1 Simulate Outputs 1 mA Output Loop Test 2 Frequency Output Test 3 Discrete Output Test 2 Simulate Sensor 1 Simulate Primary Purpose Variables 2 Mass Flow Rate 3 Density 4 Temperature 233
Using the Field Communicator with the transmitter 234 Micro Motion® Model 1500 Transmitters with Analog Outputs
Default values and ranges Appendix D Default values and ranges D.1 Default values and ranges The default values and ranges represent the typical factory transmitter configuration. Depending on how the transmitter was ordered, certain values may have been configured at the factory and are not represented in the default values and ranges. Table D-1: Transmitter default values and ranges Type Parameter Default Flow Flow direction Forward Flow damping 0.8 sec(1) Flow calibration factor 1.00005.
Default values and ranges Table D-1: Transmitter default values and ranges (continued) Type Slug flow Temperature Pressure T-Series sensor Special units 236 Parameter Default Range Density units g/cm3 Density cutoff 0.2 g/cm3 D1 0 g/cm3 D2 1 g/cm3 K1 1000 µsec 1000 – 50,000 µsec K2 50,000 µsec 1000 – 50,000 µsec FD 0 Temp Coefficient 4.44 Slug flow low limit 0.0 g/cm3 0.0 – 10.0 g/ cm3 Slug flow high limit 5.0 g/cm3 0. 0 – 10.0 g/ cm3 Slug duration 0.0 sec 0.0 – 60.
Default values and ranges Table D-1: Transmitter default values and ranges (continued) Type Variable mapping mA output 1 Parameter Default Base volume time sec Volume flow conversion factor 1 Primary variable Mass flow Secondary variable Density Tertiary variable Mass flow Quaternary variable Volume flow Primary variable Mass flow LRV –200.00000 g/ s URV 200.00000 g/s AO cutoff 0.00000 g/s AO added damping 0.00000 sec LSL –200 g/s Range Comments Read-only.
Default values and ranges Table D-1: Transmitter default values and ranges (continued) Type LRV URV Parameter Default Fault action Downscale AO fault level – downscale 2.0 mA 1.0 – 3.6 mA AO fault level – upscale 22 mA 21.0 – 24.0 mA Last measured value timeout 0.00 sec Mass flow rate −200.000 g/s Volume flow rate −0.200 l/s Density 0.000 g/cm3 Temperature −240.000 °C Drive gain 0.000% Gas standard volume flow rate −423.78SCFM External temperature −240.
Default values and ranges Table D-1: Transmitter default values and ranges (continued) Type Parameter Default Polarity Active low Backlight intensity 63 0 – 63 Refresh rate 200 milliseconds 100 – 10,000 milliseconds Variable 1 Mass flow rate Variable 2 Mass total Variable 3 Volume flow rate Variable 4 Volume total Variable 5 Density Variable 6 Temperature Variable 7 Drive gain Variable 8–15 None Display totalizer start/stop Disabled Display totalizer reset Disabled Display au
Default values and ranges 240 Micro Motion® Model 1500 Transmitters with Analog Outputs
Transmitter components and installation wiring Appendix E Transmitter components and installation wiring Topics covered in this appendix: • • • E.1 Installation types Power supply terminals Input/output (I/O) wiring terminals Installation types Model 1500 and Model 2500 transmitters can be installed two different ways, only one of which applies to your specific installation. • 4-wire remote – The transmitter is installed remotely from the sensor.
Transmitter components and installation wiring • Remote core processor with remote sensor – A remote core process with remote sensor installation separates all three components – transmitter, core processor, and sensor – all of which are installed separately. A 4-wire cable connects the transmitter to the core processor, and a 9-wire cable connects the core processor to the sensor.
Transmitter components and installation wiring E.2 Power supply terminals Figure E-3: Power supply wiring terminals A B A. B.
Transmitter components and installation wiring E.3 Input/output (I/O) wiring terminals Figure E-4: I/O wiring terminals A B C A. B. C. D.
NE 53 history Appendix F NE 53 history F.1 NE 53 history Date Version Type Change Operating instruction 08/2000 1.x Expansion Added writing of the device tag using Modbus 3600204 A Adjustment Improved communication handling with the HART Tri-Loop product Feature Indication of outputs option board type appears on display at power-up Expansion Added alarm A106 to indicate that HART burst mode is enabled 05/2001 2.
NE 53 history Date Version Type Change Operating instruction The display start/stop totalizers function can be enabled or disabled Petroleum measurement application improvements Live zero available as display variable Increased options for fault output settings New cryogenic application temperature algorithms Adjustment Improved frequency output stability and unit conversions Improved the handling of volume flow rate when slug flow is detected Improved handling of density values and calibrations duri
NE 53 history Date Version Type Change Operating instruction 09/2006 5.
NE 53 history Date Version Type Change Adjustment The following combinations are not allowed: • mA Output Fault Action = None and Digital Communications Fault Action = NAN • Frequency Output Fault Action = None and Digital Communications Fault Action = NAN Operating instruction Display variables set to a volume process variable automatically switch between liquid and GSV, according to current setting of Volume Flow Type Feature Configurable hysteresis for flow switch Field Verification Zero added t
Index Index A Added Damping 76 Additional Communications Response Delay 96 address HART address 93 Modbus address 96 air calibration, See calibration, density alarms alarm codes 144 configuring alarm handling 62 Status Alarm Severity configuring 63 options 64 transmitter response 111 troubleshooting 144 viewing and acknowledging using ProLink II 109 using ProLink III 110 using the Field Communicator 110 alerts, See alarms AO Cutoff 74 B backups 103 basic events, See events burst mode 94 C Calculation Spe
Index troubleshooting 172 volume flow 32 D damping Added Damping 76 density damping 47 flow damping 26 interaction between Added Damping and process variable damping 77 on mA outputs 76 temperature damping 50 Date 68 DD, See HART device description (DD) deadband, See hysteresis default values 235 density See also standard density density calibration, See calibration, density density factor, See pressure compensation density measurement configuring 44 cutoff configuring 49 effect on volume measurement 49 d
Index Fault Timeout configuring 62 effect on Fault Action 62 Field Communicator connecting to the transmitter 218 device description (DD) 217 menu maps 220 overview 217, 218 startup connection 8 Floating-Point Byte Order 96 flow damping configuring 26 effect on volume measurement 27 interaction with added damping 27 flow direction troubleshooting 172 Flow Direction configuring 39 effect on digital communications 43 effect on discrete outputs 43 effect on frequency outputs 43 effect on mA outputs 40 effect
Index M mA outputs Added Damping configuring 76 interaction with density damping 48 interaction with flow damping 27 AO cutoff configuring 74 interaction with volume flow cutoff 33 configuring 72 Fault Action configuring 77 options 78 loop testing using ProLink II 164 using ProLink III 166 using the Field Communicator 167 Lower Range Value and Upper Range Value configuring 73 default values 74 process variable configuring 72 options 73 scaling 73 troubleshooting 160, 170, 171 mass flow measurement configur
Index primary variable (PV) 95 process measurement effect of Calculation Speed 61 effect of Update Rate 59, 60 process variables See also density measurement See also gas standard volume flow measurement See also mass flow measurement See also temperature measurement See also volume flow measurement recording values 107 viewing values 108 ProLink II connecting HART/Bell 202 184 Modbus/RS-485 188 service port 183 startup connection 8 connection types 182 menu maps 191 overview 181, 182 requirements 181 ProL
Index status See also alarms status alarms, See alarms status LED 108, 144 T temperature calibration, See calibration, temperature temperature measurement configuring 49 damping configuring 50 effect on process measurement 51 measurement units configuring 50 options 50 troubleshooting 159 tertiary variable (TV) 95 testing loop testing using ProLink II 164, 166 using the Field Communicator 167 system testing using ProLink II 101 using the Field Communicator 101 totalizers resetting performing action 113 st
Index Z zero procedure using ProLink II 124 using ProLink III 125 using the Field Communicator 126 using the zero button 123 restore factory zero using ProLink II 124 Configuration and Use Manual using ProLink III 125 using the Field Communicator 126 restore prior zero using ProLink II 124 using ProLink III 125 using the Field Communicator 126 verification using ProLink II 13 using ProLink III 14 255
*MMI-20019023* MMI-20019023 Rev AA 2012 Micro Motion Inc. USA Worldwide Headquarters 7070 Winchester Circle Boulder, Colorado 80301 T +1 303-527-5200 T +1 800-522-6277 F +1 303-530-8459 www.micromotion.com Micro Motion Europe Emerson Process Management Neonstraat 1 6718 WX Ede The Netherlands T +31 (0) 318 495 555 F +31 (0) 318 495 556 www.micromotion.