Instruction Manual HAS9xE-IM-ATEX 05/2007 Instruction Manual CAT 100 & CAT 200 Continuous Analyzer Transmitter Addendum for Hazardous Area Applications 3rd Edition 05/2007 www.EmersonProcess.
CAT 100 & CAT 200 Addendum for Hazardous Area Applications Instruction Manual HAS9XE-IM-ATEX 05/2007 ESSENTIAL INSTRUCTIONS READ THIS PAGE BEFORE PROCEEDING! Emerson Process Management (Rosemount Analytical) designs, manufactures and tests its products to meet many national and international standards. Because these instruments are sophisticated technical products, you MUST properly install, use, and maintain them to ensure they continue to operate within their normal specifications.
Instruction Manual CAT HAS9xE-IM-ATEX 05/2007 Preface This instruction manual provides information about installing, operating and maintaining/servicing CAT series gas analyzers in hazardous (classified) areas and shall be read in conjunction with the standard analyzer instruction manual only! This instruction manual covers all analyzer variations and and therefore may describe configurations and/or options not part of your specific instrument.
Instruction Manual CAT HAS9xE-IM-ATEX 05/2007 Safety Instructions Safety Instructions Intended use statement CAT series gas analyzers are intended to be used as analyzers for industrial purposes. They must not be used in medical, diagnostic or life support applications, and no independent agency certifications or approvals are to be implied as covering such applications! safety summary If this equipment is used in a manner not specified in these instructions, protective systems may be impaired.
Instruction Manual CAT HAS9xE-IM-ATEX 05/2007 Safety Instructions Possible hazardS Make sure to understand and observe all additional warnings given in the assoziated analyzer instruction manual! Possible explosion hazard When installing these instruments all legislative requirements must be taken into consideration! Take care of the special conditions given within the instruction drawing in the manual on hand! explosion hazard by modification Any addition, substitution, or replacement of components i
Instruction Manual CAT HAS9xE-IM-ATEX 05/2007 Safety Instructions POSSIBLE EXPLOSION HAZARD Do not open instrument when energized. Ensure that external circuitry is disconnected or de-energized before opening the instrument. Ensure that all gas connections are made as labeled and are leak free. Improper gas connections could result in explosion and death. Possible explosion hazard The CAT analyzer may utilize not only sample gas but one or more pressurized carrier gases and/or calibration gases.
Instruction Manual CAT HAS9xE-IM-ATEX 05/2007 Safety Instructions Emerson Process Management GmbH & Co.
Instruction Manual CAT HAS9xE-IM-ATEX 05/2007 Compliances Compliances This product is certified by several agencies for the use in hazardous (classified) areas.
Instruction Manual CAT HAS9xE-IM-ATEX 05/2007 Compliances Special CSA approved versions with extended temperature range are available, too. These versions show different markings: Emerson Process Management GmbH & Co.
CAT S-8 Instruction Manual HAS9xE-IM-ATEX 05/2007 Emerson Process Management GmbH & Co.
Instruction Manual HAS9xE-IM-ATEX 05/2007 CAT Table of Contents Preface S-1 Definitions S-1 Safety Instructions S-2 Chapter 1 Installation 1-1 Chapter 2 Startup 2-1 Chapter 3 Service and Maintenance 3-1 APPENDIX A-1 A-1 A-2 EC Declaration of Conformity................................................................................A-2 EC Type Examination Certificates.........................................................................A-3 Emerson Process Management GmbH & Co.
Instruction Manual HAS9xE-IM-ATEX 05/2007 CAT Table of Contenst TOC-2 Emerson Process Management GmbH & Co.
Instruction Manual CAT HAS9xE-IM-ATEX 05/2007 Chapter 1 Installation Possible explosion hazard Installing and wiring this instrument must comply with all relevant national legislative requirements and regulations.
Instruction Manual CAT HAS9xE-IM-ATEX 05/2007 1 Installation 3 signal cable glands 4 installation drill-holes A. Ex e junction box B. Threads for cable glands for analog and digital signal cables (M16 x 1,5) C. 11 gas fittings (number depending on application; flame arrestors inside gas inlets and outlets). D. Optional sample handling system, design depending on applicational needs.
Instruction Manual CAT HAS9xE-IM-ATEX 05/2007 Chapter 2 Startup Once the CAT analyzer is installed correctly and in accordance to the instructions given in chapter „Installation“ it is prepared for startup.
Instruction Manual CAT HAS9xE-IM-ATEX 05/2007 2 2-2 Startup Emerson Process Management GmbH & Co.
Instruction Manual HAS9xE-IM-ATEX 05/2007 CAT Chapter 3 Service and Maintenance Possible explosion hazard Inspection, maintenance and service must be carried out considering all related standards (e.g.
Instruction Manual HAS9xE-IM-ATEX 05/2007 CAT 3 Service and Maintenance flammable gases - Possible explosion hazard Leaks may cause explosion when measuring flammable gases! When measuring flammable gases it is recommended to perform a leak test on all gas paths, connections and components before startup or applying power. Leak tests should be carried out on a 2 month‘s regular basis and after repair/maintenance. See the analyzer instruction manual for instructions on how to carry out leak tests.
Instruction Manual HAS9xE-IM-ATEX 05/2007 CAT APPENDIX This chapter contains series CAT EC declarations of conformity and EC type examination certificates. Emerson Process Management GmbH & Co.
Instruction Manual HAS9xE-IM-ATEX 05/2007 CAT A-1 A-1 A-2 EC Declaration of Conformity EC Declaration of Conformity Emerson Process Management GmbH & Co.
Instruction Manual HAS9xE-IM-ATEX 05/2007 CAT A-2 A-2 EC Type Examination Certificates EC Type Examination Certificates Emerson Process Management GmbH & Co.
Instruction Manual HAS9xE-IM-ATEX 05/2007 CAT A-2 A-4 EC Type Examination Certificates Emerson Process Management GmbH & Co.
Instruction Manual HAS9xE-IM-ATEX 05/2007 CAT A-2 EC Type Examination Certificates Emerson Process Management GmbH & Co.
Instruction Manual HAS9xE-IM-ATEX 05/2007 CAT A-2 A-6 EC Type Examination Certificates Emerson Process Management GmbH & Co.
Instruction Manual HAS9xE-IM-ATEX 05/2007 CAT A-2 EC Type Examination Certificates Emerson Process Management GmbH & Co.
Instruction Manual HAS9xE-IM-ATEX 05/2007 CAT A-2 A-8 EC Type Examination Certificates Emerson Process Management GmbH & Co.
Instruction Manual HAS9xE-IM-ATEX 05/2007 CAT A-2 EC Type Examination Certificates Emerson Process Management GmbH & Co.
Instruction Manual HAS9xE-IM-ATEX 05/2007 CAT A-2 A-10 EC Type Examination Certificates Emerson Process Management GmbH & Co.
Instruction Manual HAS9xE-IM-ATEX 05/2007 CAT A-2 EC Type Examination Certificates Emerson Process Management GmbH & Co.
CAT A-12 Instruction Manual HAS9xE-IM-ATEX 05/2007 Emerson Process Management GmbH & Co.
Instruction Manual HAS9xE-IM-ATEX 05/2007 CAT
Instruction Manual CAT 100 & CAT 200 Addendum WORLD HEADQUARTERS ROSEMOUNT ANALYTICAL EUROPE Emerson Process Management GmbH & Co. OHG Industriestrasse 1 63594 Hasselroth Germany T 49 6055 884 0 F 49 6055 884209 Emerson Process Management Rosemount Analytical Inc. 6565 P Davis Industrial Parkway Solon, OH 44139 USA T 440.914.1261 Toll Free in US and Canada 800.433.6076 F 440.914.1271 e-mail: gas.csc@EmersonProcess.com www.raihome.