MLT Analyzer Module Manual

NGA-2000 System Calibration
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98 NGA 2000 Supplement - 11
3.3 Set-up general parameters
One general parameter is "Calibration Procedures in Test-Mode". It is located in the menu
"Calibration/Test procedures".
SCTESTMOD
SCCALTYPE2
SCPROGSTEP
SCCALTIME1
SCCALTIME2
With this parameter we can run all the calibration procedures either in the defined mode
or in a test mode (set test mode to „Yes“). The test mode means that the valve switching
and waiting for purge times is done in the same manner like in the normal calibration
procedure.
The only difference is that the single calibrations of the modules and the appropriate times
the modules would need for the calibration are not done.
System configuration and diagnostics...
System calibration...
Calibration/Test procedures...
-- Calibration/Test Procedures --
Start Zeroing all ranges!
Start Zeroing and Spanning all ranges !
Start calibration program !
Cancel calibration!
Calibration Procedures in Test-Mode: No
Test Procedures...
Calibration type: ZERO-CAL
Program step: 1
Calibration time: 16 s
Previous calibration time: 57 s
Result...