MLT Analyzer Module Manual

4.1.1 Analyzer Channel Status - Status Details
Acknowledge and Clear Failures
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
NGA 2000
4 - 19
Display Status... Main... Channel Calib...
TAG
F1 F2 F3 F4 F5
Failures: Yes
Maintenance-Requests: No
Temperature: 20.0 C
Operation: Ready
0.0 100.0
Range: 1
37.50 ppm CH4
0.00 50.00
Measure Channel Back...
TAG
F1 F2 F3 F4 F5
37.50 ppm
-- Analyzer Channel Status --
Status details...
Hours of operations: 164
Operation status: Ready
Events: No
Alarms: No
Maintenance requests: No
Function control/Service: No
Operational settings...
Failures: No
RawMeas
Measure Back...
TAG
F1 F2 F3 F4 F5
37.50 ppm
-- Status Details --
Failures...
Function controls...
Measurements/Alarms...
Events...
Maintenance requests...
Acknowledge and clear failures !
Acknowledge and clear maintenance requests !
Acknowledge and clear function controls !
Change to the
menu "Analyzer
Channel Status"
Press
F2
Notes:
If you have solved the
reasons for the failures
reported, you should start
this function.
The menu "List of Possible
Failures" will be ready for
new reports !
Change to the
submenu "Status
Details"
Press
or
Change to the line
"Acknowledge and
clear failures !"
Press or
as often as necessary
to get the menu line
"Acknowledge and clear
failures !" white on black.