Reference Manual Ed 21 December 2014 MAS2600 Tank Contents Transmitter
The content of this publication are presented for information purposes only, and while effort has been made to ensure their accuracy, they are not to be construed as warranties or guarantees, expressed or implied, regarding the products or services described herein or their use or applicability. All sales are governed by our terms and conditions, which are available upon request. We reserve the right to modify or improve the designs or specifications of our products at any time without notice.
MAS2600: Table of Contents December 2014 Reference Manual Ed 21 Table of Contents Description and Technical Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Description and Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Technical Specification. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 Ordering Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Reference Manual Ed 21 2 MAS2600: Table of Contents December 2014
MAS2600: Description and Technical Data December 2014 Reference Manual Ed 21 Description and Technical Data 3
Reference Manual Ed 21 4 MAS2600: Description and Technical Data December 2014
MAS2600: Description and Technical Data Reference Manual December 2014 Ed 21 Description and Operation MAS2600 Tank Contents Transmitter The MAS2600 is a 2-wire 4-20 mA level transmitter consisting of a transducer, an amplifier and an interconnecting cable. Transducer The transducer is a pressure sensitive silicon micro strain gauge sensor mounted in a glass to metal seal.
Reference Manual MAS2600: Description and Technical Data Ed 21 December 2014 Technical Specification Transducer Ranges 0 - 3.5 / 0 - 7 / 0 - 16 / 0 - 35 m H2O Absolute 0 - 20.394 / 0 - 35.690 m H2O Accuracy ± 0.25% F.S. at 20°C Stability Max. ± 0.1% / Year TEB (Total Error Band) ± 0.4% F.S. at ±0 to +50°C ± 2.0% F.S. at -20 to +80°C Overload Capability Minimum 4 x transducer range with no changes in calibration. 6 x transducer range burst pressure.
MAS2600: Description and Technical Data Reference Manual December 2014 Ed 21 Power Supply The Power Supply can vary from 17-33 VDC. Permissible load resistance is shown graphically below. Rloop max.
Reference Manual Ed 21 8 MAS2600: Description and Technical Data December 2014
MAS2600: Ordering Information December 2014 Reference Manual Ed 21 Ordering Information 9
Reference Manual MAS2600: Ordering Information Ed 21 December 2014 MAS2600 Ordering Information Type A G H Absolute Transducer Gauge Transducer High Temperature Gauge Transducer Transducer Ranges 1 3,5 mH2O gauge 2 7,0 mH2O gauge 3 16,0 mH2O gauge 4 35,0 mH2O gauge or 0,8 to 3,5 bar absolute 7 0,8 to 2,0 bar absolute Temperature Sensor 0 1 without Built-in Pt 100 (not available for IS installations) Cable (length in meters) XX XX Standard cable (-20° to 80° C) Maximum length: Hazardous area
MAS2600: Ordering Information Reference Manual December 2014 Ed 21 Type MAS2600 -XXX-XX-X/XX Type A Absolute Transducer The absolute transducer has no breather pipe, because it refers to absolute vacuum. An atmospheric pressure sensor is used to compensate for the atmospheric pressure. The absolute transducer is especially suitable for open deck applications, where the amplifier box may be exposed to flooding.
Reference Manual MAS2600: Ordering Information Ed 21 December 2014 Measuring Ranges MAS2600 -XXX-XX-X/XX Gauge Transmitter The MAS2600 Gauge Transmitter is supplied in 4 ranges. In the programmable amplifier the measuring range is divided into 8 sub-ranges and operated by means of the range select switch. The transmitter output signal can be adjusted to 20 mA within the sub-ranges as of the table below. Range no. Transducer Type Type 1: 0 - 3.5 mH2O Type 2: 0 - 7.0 mH2O Type 3: 0 - 16.
MAS2600: Ordering Information Reference Manual December 2014 Ed 21 Selecting the Measuring Range When selecting the measuring range it is important to consider the tank height, vent pipe height, the tank type and the specific gravity (SG) of the tank contents. The transducer measures the height of the fluid from the transducer diaphragm to the surface of the fluid, including the fluid in the vent pipe.
Reference Manual MAS2600: Ordering Information Ed 21 December 2014 Forepeak Tank If transducer measuring range > tank height: Select the next measuring range. Example: Tank height 12 m. Based on the tank height a transducer with a measuring range of 16 mH2O should be selected. However, due to the dynamic conditions in the forepeak tank, a transducer with a measuring range of 35 mH2O, i.e. the next in line, should be selected. Turn amplifier range select switch to range No. 6. Draft Vent pipe min.
MAS2600: Ordering Information December 2014 Reference Manual Ed 21 Temperature Sensor MAS2600 -XXX-XX-X/XX The transducer can be supplied with a built-in Pt100 temperature sensor. The temperature sensor will only give a proper reading, if the transducer is an internally mounted type. Temperature Sensor 0 Without temperature sensor. 1 With temperature sensor.
Reference Manual MAS2600: Ordering Information Ed 21 December 2014 Cable MAS2600 -XXX-XX-X/XX Two types of transducer cables are available. The standard cable for applications with temperatures up to 80C, and high temperature cable for applications with temperatures up to 125C. The transducer cable is delivered in whole meters.
MAS2600: Ordering Information Reference Manual December 2014 Ed 21 Mounting Methods MAS2600 -XXX-XX-X/XX Brackets for Internal Mounting (dwg no. P022P010) 2600 SERIES 4-20 mA x EEX TWO WIRE TRANSMITTER ia C T5 S/N: 2 600 CERTIFICATE REF.: 90-C.95948 Tam b.: -40 to +85 °C Lint = 2 0 uH < 45 mA IK = Cint = 10 nF Um ax = 33 V MARINE AUTOMATION DENMAR K 40.0 55.0 38.0 Tank top Bulkhead 125.0 45.0 160.
Reference Manual MAS2600: Ordering Information Ed 21 December 2014 Pole Mounting (dwg no. P022P011) 2600 SERIES 4-20 mA x EEX ia TWO WIRE TRANSMITTER C T5 S/N: 2600 CERTIFICATE REF.: 90-C.95948 Tam b.: -40 to +85 °C Lint = 20 uH < 45 mA IK = Cint = 10 nF Umax = 33 V MARI NE AUTOMATION DE NMARK 24.0 40.0 Bulkhead 1" Thread Tank bottom The weldable collar and the nut are manufactured of ASTM 316L steel. The gaskets are made of nitrile rubber.
MAS2600: Ordering Information Reference Manual December 2014 Ed 21 1” Pipe End Mounting (dwg no. P022P015) 2 6 0 0 E S IE R S 420m A T W O W I RETRA NSM TTE I R i x EEX a C T5 S/N: 2 6 0 0 C R E T I F IC T A E E R .F 9 : 0 - C 9 . 5 9 4 8 T a b m: . 4 0 - to + 8 5 C ° L n ti = 0 2 H u C n ti = 1 0 n F U a mx 3 = 3 V I 4 5 < = K mA M A R IN E U T A O A IO M N T D N A E R M K Ta nk t op Bu lk hea d 80,0 Ta nk b ott om 55,0 24.0 41,0 26,0 34.
Reference Manual MAS2600: Ordering Information Ed 21 December 2014 Flange Mounting (dwg no. P022P013) b 2 6 0 0 E S IR E S 420m A x EE X ia W T O W I RE C T5 R TANSM I TT ER /N: S 2 6 0 d C R E T IF IC T A E E R .F : 9 0 C 9 . 5 9 4 8 T a mb .
MAS2600: Ordering Information Reference Manual December 2014 Ed 21 Flexible Rubber Tube Mounting (dwg no. P022P001) 2 6 0 0 E S IR E S 420m A TW O W I RETRANSM I TT ER C T5 /N: 2 6 0 S x EE X i a E R T C IF IC T A E E R .: F 9 0 C 9 . 5 9 4 8 T mb a .
Reference Manual MAS2600: Ordering Information Ed 21 December 2014 Flexible PTFE Tube Mounting (dwg no. P022P021) 6 0 S E 2 R IE S 4-20m A TWOW IRETRANSMITTE R X i x EE a CT 5 S/N: 2 6 0 C R E T IF IC A E T R E F .: 9 0 - C 9 . 5 9 4 8 T a mb .
MAS2600: Ordering Information Reference Manual December 2014 Ed 21 DN25 flange with 1” ball valve (dwg no. P022P019) 2 6 0 0 E S IR E S 420m A X x EE i a W T O W RE I C T5 R TANSM I TT ER S/ N: C R E T IF IC T A E E R .F : 9 0 C 9 . 5 9 4 8 in = L t 2 0 u H C n ti = 1 0 n F 6 0 2 mb a T .: - 4 0 o t + 8 5 C ° ma U x = 3 V I < K = 4 5 mA A IN M E R A U O M T A IO D T N E M A N R K Ta n k to p B u lkh ea d Ta nk b ot to m Flange, valve and nut are manufactured of ASTM 316L steel.
Reference Manual MAS2600: Ordering Information Ed 21 December 2014 Amplifier Box MAS2600 -XXX-XX-X/XX The MAS2600 amplifier box is manufactured of polyester, which has the strength and stability of aluminium, insulation and resistance properties of plastics. Color code RAL 7000. The amplifier box is protected to the level of IP 56 and must be mounted outside the tank, e.g. in the engine room or other dry areas.
MAS2600: Ordering Information Reference Manual December 2014 Ed 21 Amplifier PCB MAS2600 -XXX-XX-X/XX The PCB is available in different types for the different transducers. Amplifier PCB 0 Without amplifier PCB. P Programmable output range. The amplifier can be calibrated to 4-20 mA output in whatever range wanted within the transducer range. Calibration is made by means of dip switches and potentiometers. D Differential output range.
Reference Manual Ed 21 26 MAS2600: Ordering Information December 2014
MAS2600: Intrinsically Safe Installation December 2014 Reference Manual Ed 21 Intrinsically Safe Installation Description of the Ex safety information and Intrinsically Safe Installation MAS2600 27
Reference Manual Ed 21 28 MAS2600: Intrinsically Safe Installation December 2014
MAS2600: Intrinsically Safe Installation December 2014 Reference Manual Ed 21 Intrinsic Safety When connecting equipment in hazardous areas, certain requirements must be fulfilled to comply to the equipment safety certification and cable installation. There are requirements both for the equipment in the hazardous area on deck and for the associated apparatus in the nonhazardous area.
Reference Manual MAS2600: Intrinsically Safe Installation Ed 21 December 2014 Intrinsically Safe Installation The MAS2600 is designed for use in an approved intrinsically safe installation. The amplifier is fitted with a label describing the IS category: Item Explanation S/N Serial number - production year Ex identification according to guideline 94/9/EC (ATEX) II Equipment group II 2(1) Sensor mounted in zone 0 and amplifier in zone 1 Ex ia Denotes the type of protection intrinsic safety, i.
MAS2600: Intrinsically Safe Installation Reference Manual December 2014 Ed 21 Installation and Special Considerations for Safe Use Before installing intrinsically safe units within an approved installation, it is essential to ensure that there are no deviations from the conditions of the order or the requirements of the approved block schematic diagram and the following related instructions: 1. No circuits other than those shown on the drawing are permitted.
Reference Manual Ed 21 MAS2600: Intrinsically Safe Installation December 2014 The entity concept allows inter-connections of intrinsically safe apparatus to associated apparatus not specifically examined in combination as a system.
MAS2600: Installation December 2014 Reference Manual Ed 21 Installation Installation and Handling the MAS2600 33
Reference Manual Ed 21 MAS2600: Installation December 2014 Handling The MAS2600 transducer is a highly sensitive electronic piece of equipment and proper handling is most important. To avoid damaging the transducers: • NEVER lift by the cable if transducer is mounted with fittings. • NEVER remove the protection cap in front of the diaphragm before installation. • NEVER knock the transducer. • NEVER apply pressure to the diaphragm in any way whatsoever, whether using fingers, tools or sharp objects.
MAS2600: Installation Reference Manual December 2014 Ed 21 Installation As part of the installation please check that: • the serial numbers on the transducer, amplifier and lid of amplifier box complies. The serial numbers are found on the S/N on the amplifier box at the bottom line and on the transducer S/N on the range label. • the transducer is fitted in the correct tank. • the transducer is correctly connected to the amplifier. • all cable glands are properly secured to the amplifier box.
Reference Manual Ed 21 36 MAS2600: Installation December 2014
MAS2600: Testing and Recalibration December 2014 Reference Manual Ed 21 Testing and Recalibration Testing and Recalibration of the MAS2600 37
Reference Manual Ed 21 MAS2600: Testing and Recalibration December 2014 Testing Caution! Please observe that testing and recalibration of transmitters mounted in hazardous areas shall be carried out by Ex qualified personnel. Before dispatch, the MAS2600 transmitter is calibrated in accordance with the requirements specified in the order. The transducer and the amplifier are calibrated together and must, therefore, always be used together.
MAS2600: Testing and Recalibration Reference Manual December 2014 Ed 21 Recalibrating Gauge Transmitter using Pressure on the Transducer Caution! Please observe that testing and recalibration of transmitters mounted in hazardous areas shall be carried out by Ex qualified personnel. When recalibrating the MAS2600 Gauge Transmitter a pressure calibrator with a range of -1 to 3.5 bar should be used. It should have an accuracy of 0.05% FS or better, as well as a 4½ digit milliamp meter. 1.
Reference Manual MAS2600: Testing and Recalibration Ed 21 December 2014 Recalibrating Gauge Transmitter using Vacuum (for gauge transmitters only) Caution! Please observe that testing and recalibration of transmitters mounted in hazardous areas shall be carried out by Ex qualified personnel. An internally mounted transducer can also be recalibrated by using vacuum. 1. Check that the tank is empty. 2. Check that the range select switch on the amplifier is correctly set. 3.
MAS2600: Testing and Recalibration Reference Manual December 2014 Ed 21 Recalibrating the Absolute Transmitter using Pressure on the Transducer Caution! Please observe that testing and recalibration of transmitters mounted in hazardous areas shall be carried out by Ex qualified personnel. When recalibrating the MAS2600 Absolute Transmitter a pressure calibrator with a range of -1 to 3.5 bar should be used. It should have an accuracy of 0.05% FS or better, as well as a 4½ digit milliampmeter.
Reference Manual Ed 21 42 MAS2600: Testing and Recalibration December 2014
MAS2600: Maintenance and Troubleshooting December 2014 Reference Manual Ed 21 Maintenance and Troubleshooting 43
Reference Manual Ed 21 MAS2600: Maintenance and Troubleshooting December 2014 Maintenance Caution! Please observe that testing and recalibration of transmitters mounted in hazardous areas shall be carried out by Ex qualified personnel. Marine Tank Management recommends that the following to be checked annually: 1. Visual inspection of the cable, transducer and amplifier. 2. Compare signal with e.g. manual readings, and if necessary perform a zero and full-scale calibration.
MAS2600: Maintenance and Troubleshooting December 2014 Reference Manual Ed 21 Trouble Shooting Caution! Please observe that testing and recalibration of transmitters mounted in hazardous areas shall be carried out by Ex qualified personnel. If the MAS2600 transmitter is not functioning in accordance with Marine Tank Management instructions, check that: 1. the 24 VDC power supply has been correctly connected, and is between 17-33 VDC. 2. the 4 - 20 mA output signal is functioning (compared with e.g.
Reference Manual MAS2600: Maintenance and Troubleshooting Ed 21 December 2014 MAS2600 Test Sheet Name ATT. Emerson Process Management Marine Tank Management Damcos A/S, Service Department FAX.: +45 5578 7272 Ship name Yard MAS2600 Transducer serial no. NB no. MAS2600 Amplifier serial no. Tank name Liquid in tank HFO DO Ballast The tank is Full Empty Other Step Result Unit Limits Voltage should be in the range 17 to 22 V DC. Check power supply. Voltage must not exceed 2 V AC.
MAS2600: Index December 2014 Reference Manual Ed 21 Index Numerics Mounting Methods .............................................. 17 1” Pipe End Mounting.......................................... 19 O A Absolute Transducers .......................................... 11 Absolute Transmitter........................................... 12 Amplifier ............................................................... 5 Amplifier Box....................................................... 24 Amplifier PCB .......
Reference Manual Ed 21 48 MAS2600: Index December 2014
Emerson Process Management Damcos A/S Aaderupvej 41 DK-4700 Naestved T +45 5578 7200 F +45 5578 7272 ©2014 Emerson Process Management. All rights reserved. www.EmersonProcess.com/mtm of the Emerson Process Management family of companies. All other marks are the property of The Emerson logo is a trademark and service mark of Emerson Electric Co. , Rosemount® and the Rosemount logotype are registered trademarks of Rosemount Inc., Damcos and the Damcos logotype are trademarks of Damcos A/S.