Instruction Manual HASXEDE-IM-EX 03/2012 Gas Analyzer X-STREAM Enhanced Series Flameproof Variation For Use in Zone 1 and Division 2 Hazardous Areas Instruction Manual Addendum www.EmersonProcess.
ESSENTIAL INSTRUCTIONS READ THIS PAGE BEFORE PROCEEDING! Emerson Process Management (Rosemount Analytical) designs, manufactures and tests its products to meet many national and international standards. Because these instruments are sophisticated technical products, you MUST properly install, use, and maintain them to ensure they continue to operate within their normal specifications.
Instruction Manual Introduction S-1 Definitions S-1 Safety Instructions S-5 Chapter 1 Technical Description 1-1 Chapter 2 Installation 2-1 Chapter 3 Start up 3-1 Emerson Process Management GmbH & Co. OHG TOC-1 Terms used in this instruction manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . S-2 Symbols used on and inside the unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . S-3 Symbols used in this manual . .
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 Table of Contents 3.5 Front Panel Elements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5 3.5.1 Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5 3.5.2 Status Line and Text message line . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5 3.5.3 Keys 3-6 3.6 Software . . . . . . .
Instruction Manual Table of Figures TOC A.3 CSA Certificate of Compliance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-12 A.4 Block Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-19 A.5 Assignment of Terminals and Sockets . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-25 Index of Figures Fig. 1-1: Fig. 1-2: Fig. 1-3: Fig. 1-4: Fig. 1-5: Fig. 1-6: Fig.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 Index of Figures / Index of Tables Fig. 4-1: Fig. 4-2: Fig. 4-3: Fig. 4-4: Fig. 4-5: Fig. 4-6: Fig. 4-7: Fig. 4-8: Fig. 4-9: Fig. 4-10: Fig. 4-11: Fig. 4-12: Leak Testing with U-tube Manometer (Flame arrestor)...........................................4-5 Pressure Drop Test..................................................................................................4-5 Vapor recovery application setup.............................................
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 This instruction manual provides information about installing, operating and maintaining/ servicing X-STREAM series flame proof gas analyzers in hazardous (classified) areas and shall be read in conjunction with the basic analyzer instruction manual only! S This instruction manual covers several X-STREAM series flame proof gas analyzer variations and therefore may describe configurations and/or options not part of your specific analyzer.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 Terms Used In This Instruction Manual ATEX Directive 94/9/EC, commonly called the ATEX („Atmosphères Explosibles“) directive, dealing with equipment intended to be used in potentially explosive atmospheres. This directive is valid for equipment to be sold into and/or installed and operated in the European Union (EU).
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 Symbols Used On And Inside The Unit Safety Instructions Wherever one or more of the following symbols appear on or inside the instrument, be careful and read the instructions given in the accompanying manuals! Strictly observe the given warnings, instructions and information to minimize hazards! ... means dangerous voltages may be accessible.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 Symbols Used In This Manual Where one or more of the following symbols appear within this manual, carefully read the related information and instructions! Strictly observe the given warnings, instructions and information to minimize hazards! This symbol used in the manual ... ...
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 SAFETY INSTRUCTIONS Safety Instructions Intended Use Statement General Safety Notice / Residual Risk S X-STREAM XE series gas analyzers are intended to be used as analyzers for industrial purposes. They must not be used in medical, diagnostic or life support applications. Using X-STREAM XE analyzers as safety devices is prohibited where redundancy and/or SIL classification or equivalent is needed.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 Safety Instructions Authorized Personnel In-depth specialist knowledge is an absolutely necessary condition for working with and on the analyzer. Authorized personnel for installing, operating, servicing and maintaining the analyzer are instructed and trained qualified personnel of the operating company and the manufacturer.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 Safety Instructions Safety Instructions EXPLOSION HAZARD BY CONNECTIONS Do not open instrument when energized. Ensure that external circuitry is disconnected or de-energized before opening the instrument. S Consider the waiting time statement on the front door label before opening, if the analyzer is configured with selected measurement principles! Ensure that all gas connections are made as labeled and are leak free.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 Safety Instructions ELECTRICAL SHOCK HAZARD Installation, and connecting mains and signal cables are subject to qualified personnel only, taking into account all applicable standards and legislative requirements! Failure to follow may cause warranty invalidation, property damage and/or personal injury or death! Connecting mains and signal cables to internal srew terminals requires working at open housing near life parts! Installation of this instrum
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 Safety Instructions Safety Instructions EXPLOSION HAZARD BY DAMAGED FLAME PATHS Take care to not damage the housing nor threads, and to not produce scratches on the flange, as threads and flange function as flame paths! S Violation may result in explosion and personal injury! ELECTRICAL SHOCK HAZARD WHEN CONNECTING TO MAINS Before connecting the analyzer to mains power, please read the chapter on safety warnings and the following instructions ca
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 Safety Instructions EXPLOSION HAZARD BY HIGH PRESSURE Risk of internal overpressure under leakage conditions! For the purge or high sample gas option, take care to limit the total of purge gas flow and highest flow of sample gas lines into the instrument to max.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 Safety Instructions Safety Instructions EXPLOSION HAZARD: VAPOR RECOVERY APPLICATION Consider the waiting time statement on the front door label before opening! Temperatures of components for this application installed into the analyzer exceed the analyzer's temperature classification for hazardous areas! S Disregarding the waiting time statement may result in explosion! HEAVY INSTRUMENT HAZARD X-STREAM XEFD analyzers, to which this manual relat
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 Safety Instructions S-12 Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 Safety Instructions Safety Instructions EXPLOSION HAZARD BY INTERNAL BATTERY Do NOT OPEN the analyzer enclosure if explosive atmosphere may be present! S This analyzer contains an internal battery! Disregarding may cause explosion even if the analyzer is de-energized! CRUSHING HAZARD Take care of crushing hazard when closing the front door of analyzer field housings! Keep out of the closing area between enclosure cover and base! OPERATION AT L
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 General Operating Notes GENERAL OPERATING NOTES HAZARD TO LIFE AND EXPLOSION HAZARD BY EXHAUST GASES Exhaust gases may contain hydrocarbons and other toxic gases such as carbon monoxide. Faulty gas connections may lead to explosion and death. Ensure that all gas connections are connected as labelled and airtight. • The unit must be installed in a clean and dry area, protected from strong vibrations and frost.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 1.2 Design Features Packaged into a cast aluminum enclosure, the X-STREAM XEFD gas analyzer provides all the measurement options available for general purpose instruments, but for installation at locations, where explosive gas atmosphere might be present frequently (Zone 1) or occasionally (Zone 2, Division 2).
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 1.3 Protective Measures in Detail 1.3 Protective Measures in Detail The cast aluminum enclosure consists of two parts: base and cover, connected by hinges. The area where the two parts are in contact is designed to work as a flange, quenching flames entering the small path between them. When operated, the analyzer enclosure has to be closed and secured by 20 screws evenly arranged all over the flange.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 X-STREAM flameproof analyzers have been subjected to additional tests to support both situations: higher sample and calibration gas pressures as well as Higher gas pressure is specified to be within the range of 1100 hPa to 1500 hPa. The maximum permitted flow is 1,5 l/min, depending on the installed measurement system lower limits may apply.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 1.5 Compliances 1.5 Explosion Protection Compliances These products are available in two different variations, separately certified by agencies for the use in hazardous (classified) areas: The one variation, to be equipped with cable glands, is certified by Fyzikálně technický zkušební ústav, s.p (FTZÚ), an European Notified Body under the Directive 94/9/EC („ATEX“) and conforms to the provisions of EN 60079-0 and EN 60079-1.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 1.5 Compliances • Only screws M16x45 ISO 4762 A2-70 as specified in the maintenance section of this manual shall be used (spare part # 42716945). • Vapor Recovery application: Pressure of gases not to exceed 1100 hPa. Concentration of gases must be below 25 % LEL. • The flame joints correspond to the drawing No. 4.271-7112/1 and do not comply with the dimensions mentioned into the Tab. 1 and Tab. 2 of EN 60079-1 ed.2.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 1.6 Nameplate Label 1.6 Nameplate Label ATEX version Area Description The analyzer´s electrical data, manufacturing data and serial number CSA-C/US version Area Description Manufacturer address IECEx / EU (ATEX) II other than mines Area classification 2 Category 2 Equipm.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 1.7 Technical Data 1.7 Technical Data 1 Techn. Description Transport lugs to be removed after installation Flame arrestors with gas fittings (enclosure threads: M18 x 1.5) Cable inlets (enclosure threads; M20 x 1.5) Eyebolt detail All dimensions in mm [inches in brackets] Fig.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 1.7 Technical Data Site of installation Humidity (non-consending) Pollution degree Installation category Altitude Sourrounding atmosphere < 90 % r.h. at +20 °C (68 °F) < 70 % r.h.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 1.7 Technical Data Tab. 1-2: Standard Interfaces Data Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 1.7 Technical Data Optional interfaces Digital I/O board electrical specification 7 or 14 digital inputs function electrical specification 9 or 18 additional relay outputs function max. 30 V, internally limited to 2.3 mA HIGH: min. 4 V; LOW: max. 3 V common GND Each input can be configured to any of the functions listed in chapter 6 of the X-STREAM XE instruction manual, e. g.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 1.7 Technical Data electrical specification 2 Analog inputs function 1 Interface 1 Service interface electrical specification function electrical specification function optional (requires to fit wire bridges, Chapter 2 'Installation'): 4 (0)–20 mA ; Rin = 50 Ω optically isolated from analyzer GND protected against overload up to ± 15 V or ± 20 mA Input analog signals from external devices, such as e.g.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 1.7 Technical Data Increased sample and calibration gas pressure Above 1100 hPa to max. 1500 hPa (take Increased gas pressure care of the measurement principles limits!) Maximum flow 1.5 l/min. The gas paths need additionally to be protected by suitable inline flame arrestors, designed and approved for the applied higher gas pressure.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 1.7 Technical Data Analog and digital I/O Other Cable glands screw terminals max. 1.5 mm² (14 AWG); end sleeves not required Ethernet: RJ45 sockets; USB-connections*) Types Permissible outer cable diameter Cable entry via, IP 68, or conduits with metric-to-NPT adaptor 3…13 mm (0.11 to 0.5 inch), see cable gland / conduit specification Detailed terminals configuration 1 1 2 3 Fig. 1-5: Ethernet & USB Analog & digital I/O terminal strips Max.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 1.7 Technical Data Power Connection Power connections Power supply Power supply fuses Fuse ratings Cable glands Types Permissible outer cable diameter 1 2 3 4 Connection via internal screw terminals near cable entries, max 4mm² (10 AWG), end sleeves not required. AC 230 V / T 4 A / 5x20 mm Cable gland, classified IP 68 or suitable conduit with metricto-NPT adaptor 3…13 mm (0.11 to 0.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 1.8 Measurements Specifications Special Specs Standard Specs or Conditions (see Tab.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 1.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 1.8 Measurements Specifications Oxygen Sensors Trace (tO2) Detection limit (4 σ) 1 4 Linearity 1 4 Zero-point drift 1 4 Span (sensitivity) drift 1 4 Repeatability 1 4 Response time (t90) 3 Permissible gas flow Influence of gas flow 1 4 Maximum gas pressure 7 8 Influence of pressure 2 – At constant temperature ≤ 0.10 % per hPa ≤ 0.10 % per hPa ≤ 0.10 % per hPa – With pressure compensation 6 ≤ 0.01 % per hPa ≤ 0.01 % per hPa ≤ 0.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 1.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 1.9 Vapor Recovery Application The configuration consists of a dual channel IR measurement, connected to the inlet and outlet of a converter. This converter is installed inside the X-STREAM analyzer and is heated to about 280 °C (536 °F). 1 This application is served by a special configuration of the X-STREAM XEFD flameproof analyzer.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 1.9 Vapor Recovery Application Converter efficiency The measurement accuracy is highly dependent on the converter efficiency: If this is too low, the converter material needs replacement. To measure the converter efficiency, one has to compare the measurement values of CH1 with and without having the gas flowing through the converter. This requires activating the valve V1.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 Chapter 2 Installation On receipt, check the packaging and its contents thoroughly for damage. Inform the carrier immediately of any damage to packaging or contents, and keep damaged parts until clarification. Store the instrument at a dry and clean place, considering the acceptable environmental conditions.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 2.2 2.2 Installation - Analyzer Installing the Analyzer EXPLOSION HAZARD Installing and wiring this instrument must comply with all relevant national legislative requirements and regulations.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 Installation - Analyzer Install the analyzer to a stand or a wall by means of 4 eyebolts, provided at the instruments rear side. It is recommended to install the analyzer in an upright (vertical) position; other orientations may affect the measuring results. EXPLOSION HAZARD When installing the analyzer take care to have an area of min. 40 mm (1.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 2.3 2.3 Installation - Gas Lines Connecting Gas Lines Gas inlets and outlets are protected by flame arrestors, supporting stainless steel pipes of either 3.18 mm (1⁄8“) or 6.35 mm (1⁄4“) outer diameter (OD). The 1⁄4“ fitting may optionally be supplied with a clamping ring for 6 mm OD pipes. The instrument provides up to 8 gas inlets and outlets, depending on the ordered configuration. Unused entries are closed by approved plugs.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 2.3 Installation - Gas Lines Gas fittings are accessible at the instrument‘s outer bottom side.The number and assignment of gas inlet and outlet fittings depends on the application and is given on a label attached to the analyzer‘s bottom side adjacent to the fittings. For simple installation we recommend to mark the gas lines according to the marking on the analyzer label.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 2.3 Installation - Gas Lines 2.3.1.2 High Sample and Calibration Gas Pressures The gas paths need additionally to be protected by suitable external inline flame arrestors, designed and approved for the installation area and for the applied higher gas pressure (above 1100 hPa to max. 1500 hPa; see gas parameter specification 1-12 ).
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 Installation - Gas Lines Pressure and gas flow must remain within the values given in the „Measurement Specifications“ section within this manual. If moisture cannot be avoided, it is necessary to ensure that the dew point of the gases is at least 10 °C (18 °F) below the ambient temperature to avoid condensate in the gas lines. Hints for selected gases • Calibration gases for CO and NO need to be moistured by supplying them via a cooler. 2 2.3.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 2.4 2.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 2.4 Installation - Electrical EXPLOSION HAZARD WHEN ENERGIZED Do not open instrument when energized. 2 All cables (power and signal) must end (be connected) in either a safe (nonhazardous) area or in a protecting enclosure (e.g.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 2.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 2.4 Installation - Electrical 2 Installation Type AD-U adapter installation instructions Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 2.4 Installation - Electrical ATEX approved analyzers provide 4 cable entries (M20 x 1.5) and are shipped with a set of 4 plugs. Owner of such analyzers are responsible to provid����������������������������������������� e suitable cable glands according all applicable standards (e. g. IEC/EN 60079-14). See maintenance section of this manual for recommended cable glands. Unused entries during installation have to be provided with plugs.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 Installation with conduits Installation with cable glands Ensure all required parts are available: Adaptors, suitable conduits, compound and plugs for not used entries (see accessory kit). Ensure all required parts are available: Cable glands sets and plugs for not used entries (see accessory kit).
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 2.4 Installation - Electrical Analog Outputs Relay Outputs 1 - 4 Terminals for analog signals and relais outputs 1 - 4 are located at the leftmost terminal module (terminal block X1; fig. 2-9). Analog output specification: Specification of relay outputs 1-4: Electrical data: 4 (0)–20 mA (RB ≤ 500 Ω) optically isolated from each other and from analyzer electronics Dry relay change-over contacts, can be used as NO or NC. max.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 2.4 Installation - Electrical Modbus Interface Specification and interface control: Analyzer instruction manual, chapter 9 The 9 terminals on the left (28 - 36) of the strip next to the power connections carry the Modbus interface signals.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 2.4 Installation - Electrical Modbus RJ45 connection To install this connection, a cable must be fed through the cable entry without a connector. The connector can be wired on when the free end has been fed into the instrument: We recommend the VARIOSUB RJ45 QUICKON connector (PHOENIX CONTACT), which is supplied with the unit and requires no special tools. Wiring instructions can be found in the separate manual supplied with the connector.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 2.4 Installation - Electrical 7 or 14 digital inputs electrical data max. 30 VDC, 1 A, 30 W resistive dry change-over relay contacts, can be mechanical used as NO or NC data max. 30 VDC, 1 A, 30 W 2 9 or 18 additional digitaloutputs electrical data max. 30 V, internally limited to 2,3 mA, H Signal: min. 4 V; L Signal: max.
Installation - Electrical 2.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 2.4 Installation - Electrical Connecting the power cord The power cord is connected to screw-type terminals located inside the housing. Electrical Connections Schraubklemmen mit integrierten Sicherungshaltern max. 4 mm² Supported wire cross sections 0.2 to 4 mm2 (24 to 12 AWG) no need to use wire end sleeves Cable skinning length 8 mm (0.315 inch); Tightening torque, min . 0.5 Nm (4.4 in.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 2.4 2.4.1 Installation - Electrical Equipotential bonding conductor The X-STREAM XEFD enclosure provides an additional terminal for connecting an equipotential conductor, located at the base part flange´s rear side (see fig. 2-15), near the nameplate label. Screw thread: M5 x 10 mm The bonding conductor must be at least the same section as the power supply cable! Conductor cross section: min.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 2.4 Installation - Electrical ELECTRICAL SHOCK HAZARD Before completing the electrical connection of the instrument, verify cables are inserted and connected in correct manner! • • fix the cables according to the installation instruction given in the cable glands/ conduits manufacturer documentation. • Install the analyzer´s cover by means of ALL 20 screws! Tightening torque: 25 Nm (221 in.
X-STREAM XEFD 2-22 Instruction Manual HASXEDE-IM-EX 03/2012 Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 Chapter 3 Start up EXPLOSION HAZARD Ensure all covers, plugs and housing parts are in place and secured properly before supplying power and signal voltages! The sheet also gives instructions for the sequence of supplying gases during process and analyzer startup . 3 3.1 Final Check Ensure that the analyzer has been installed according to the descriptions in chapter 2, and that all covers and doors are closed and fastened.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 3.2 Leak Test 3.2 Performing a Leak Test To achieve best and proper measuring results you must ensure the gas path system does not have leaks. The following procedure describes how to perform a leak test with focus on the instrument. The gas path system should be leak tested at least on a bimonthly basis and after maintenance, replacement or repair of gas path parts. Note! We recommend to include external equipment (e.g. cooler, dust filters, etc.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 3.3 Switching On Once the unit has been unpacked and installed, we recommend to first check the settings, and if necessary adjust them to the user’s needs. e.g: What hardware is installed? • Is the unit configured to your needs (alarms, inputs, outputs, etc.) 3 • In order for the information in this chapter to be of any relevance, the unit must have been installed according to the instructions in chapter 2.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 3.3 Symbols Used 3.4 Symbols and Typographical Conventions In the following sections, the symbols and typographical conventions explained below are used to describe the software menus and navigation. Symbol Description Convention Within Process Descriptions Setup.. Menu title Setup.. Analog outputs.. Parent (SETUP) and current menu (ANALOG OUTPUTS) Analog outputs.. Output1...
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 3.5 Front Panel Elements 3.5 Front Panel Elements 3.5.1 Ch1 Ch2 Ch3 Ch4 Front Panel Display 10.000 5.000 7.000 20.000 1/12/10 13:18 3.5.2 3 Fig. 3-2: any analyzer is configured with English and German language sets, while a third can optionally be added. Currently available or under preparation: French, Italian, Portuguese and Spanish. The units are operated by six keys on the front panel.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 3.5 Front Panel Elements 3.5.3 Keys Six keys enable the use of the menu system. Depending on the operational mode (measuring, browsing menus, editing) they have the following functions: enter key: Mode Function Measuring Enter main menu Open submenu (..
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 3.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 3.6 Software 3.6 Software The analyzer software displays measurement results and status messages, allows parameters to be set and edited, and maintenance functions (e.g. calibration) to be carried out. The software is organised hierarchically: The topmost level is called MEASUREMENT DISPLAY, followed by a MAIN MENU; all other menus and submenus are arranged below.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 3.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 3.6 Software Editing The editing mode enables changing parameters. It is initiated by pressing enter. If the selection is a parameter list, the current entry is highlighted and may be changed by up and down. If the selection is a value, the cursor is placed over the last character. Use up and down to change it. Use left and right to select another character.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 3.6 Software Access Levels & Codes Level four has the highest priority and is used for factory settings — only qualified EMERSON service personnel have access to this level. Level three gives access to system admin parameters, e.g. for configuring data acquisition systems communication. Level two covers the expert settings, e.g. basic settings for calibrations and measurements.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 3.7 Start Up 3.6.2 Special Messages 3.7 Depending on the last action performed by the user, one of the following messages may be displayed to assist or inform the user. Information on incorrect entry: The value entered by the user is outside valid limits. The display indicates what limits apply. Press to return to the previous screen to enter an acceptable value. Wrong input Range 0.1 ...
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 3.7 Start Up SETUP enables several additional configurations, e.g. • 2 pages measurement display • different font sizes The display´s bottom line shows plain text status information (errors, maintenance requests, function checks or off-spec performance). Active messages are stored in an internal buffer. If there is more than one message in the buffer, the display will cycle through.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 3.8 Language Settings 3.8 Language Settings If the analyzer is operational and it becomes clear that the incorrect language has been set, which is unintelligible to the operator, the following sequence of keypresses (starting at the measurement display) can be used to set the language. If the system has been set up accordingly, the code for access level 1 must be entered at this point to enable access to the following menu.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 3.9 Checking the Settings 3.9 Checking the Settings ) Setup 1of2 Display.. Calibration.. Measurement.. In-/Outputs.. Communication.. Alarms.. Installed Options.. Save-Load.. Display Contrast.. Language: English Phrase Version EN1.30 Measurements.. Measurement Display.. Menu Access.. 10 Min Auto Home: 10 Min Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 3.9 Checking the Settings 3.9.1 Installed Options All X-STREAM gas analyzers can be fitted with a variety of optional components: follow these steps to see which options are installed on your analyzer. Press left to return to SETUP, highlight "Installed options" and press enter. Installed Options 1of2 Licenses.. Valves: Pumps: DIO#1 installed: DIO#2 installed: Anal.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 3.9 Checking the Settings 3.9.2 Configuring the Display Press left to return to SETUP. Display 1 Assignments Line 1: Line 2: Line 3: Line 4: Line 5: Comp1 Comp2 Comp3 Comp4 Blank Labels.. Display 1 Labels Line 1: Line 2: Line 3: Line 4: Line 5: Emerson Process Management GmbH & Co. OHG Note! X-STREAM currently supports one pressure sensor only. Values Press1 to Press5 thus refer to the same sensor. When entering LABELS..
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 3.9 Checking the Settings 3.9.3 Calibration Setup Once the display settings have been checked, press left to return to SETUP and open CALIBRATION to check the calibration settings. Setup.. Calibration.. Multi-channel unit: Select the component to be set in the gas component selection menu. Note! For more detailed information about calibration procedures, X-STREAM XE instruction manual. Ch1 Calibration Gases.. Tolerances.. Procedure..
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 3.9 Checking the Settings Press left several times to return to SETUP and open MEASUREMENT. Setup.. Measurement.. Ranges.. Damping.. Linearization.. Cross Interference.. Average.. Delay.. Ch1 Measurement 2of2 Cut-off: None Pressure Compensation .. Ch1 Pressure compensation Manual Pressure: 1013 hPa Pressure Pressure Status 1013 hPa Good Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 3.9 Checking the Settings 3.9.4 Setting the Analog Outputs Press left to return to SETUP, and then open IN/OUTPUTS. and from there enter ANALOG OUTPUTS. Setup.. In/Outputs.. Analog outputs.. Analog Outputs Output1.. Output2.. Output3.. Output4.. Output5.. Signal: OutRange: Low Scale: Max Scale: AutoScale: FailMode: 0/4 mA: 20 mA: Hold: 3-20 Select the analog output you like to check.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 3.9 Checking the Settings *) Numbers 1 to 5 refer to components [channels] 1 to 5: In case of secondary parameters, this means, the selected value is that of the sensor assigned to the given component (Press2 is the pressure value of the sensor assigned to component 2). Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 3.9 Checking the Settings Signal: OutRange: Low Scale: Max Scale: AutoScale: FailMode: 0/4 mA: 20 mA: Hold: Comp1 0-20 mA 0.00 100.00 Yes Live 0.00 100.00 No Next, select the output range: • 0-20 mA (dead zero) generates a 20 mA signal, if a concentration is measured at the upper limit of the signal range. A 0 mA signal is generated if the sample gas concentration equals the value specified with "LowScale" .
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 3.9 Checking the Settings NE43 includes recommendations for such cases, but also for the configuration of analog outputs to detect other measurement states. X-STREAM analyzers incorporate these recommendations as follows: Setting "FailMode" to HIGH + 10% or LOW 10% defines specific analog output signals in case of a failure.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 3.9 Checking the Settings Signal: OutRange: Low Scale: Max Scale: AutoScale: FailMode: 0/4 mA: 20 mA: Hold: 3-24 Comp1 0-20 mA 0.00 100.00 Yes Live 0.00 100.00 No "0/4 mA" and „20 mA“ enable to finetune the analog output to compensate possible deviations based on electronics tolerances: • Set "Signal" to 0 mA or 20 mA • measure the output current, • adjust it to the expected value. Accepted range: -10,000 ...
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 3.9 Checking the Settings 3.9.5 Setting Concentration Alarms Note! If concentration alarms are not being used, continue with 3-26. Press left until SETUP is displayed, then open ALARMS - CONCENTRATION Concentration Alarm Monitor: LoLo Level: Lo Level: Hi Level: HiHi Level: Hysteresis: States.. On 50.000 ppm 100.000 ppm 400.000 ppm 600.000 ppm 10.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 3.9 Checking the Settings Ch1 Concentration Alarm Monitor: LoLo Level: Lo Level: Hi Level: HiHi Level: Hysteresis: States.. 3.9.6 On 50.000 ppm 100.000 ppm 400.000 ppm 600.000 ppm 10.000 ppm You may turn the alarm function On or Off separately for each channel ("Alarm Monitor"). It´s also possible to use some of the thresholds only.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 3.9 Checking the Settings Local Backup 7/29/09 14:26 No! Yes! Startup 0 0% Save new local backup and overwrite old one! Are you sure? Confirm the operation (select "Yes!"). Wait until "Progress" shows 100 %, then press enter to return to LOCAL BACKUP. Copying Data Busy Progress For now, make a local backup: Enter LOCAL BACKUP and then select "Save..". 3 Save.. UsrBack Date Restore..
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 3.10 Web Browser 3.10 Web Browser 3.10.1 Connection Via Network Field housings Table top analyzers To gain access to the instrument´s web browser interface, first ensure the instrument is powered and connected to your network via Ethernet1 connector (Fig. 3-4) Ethernet2 connector Ethernet1 connector Fig. 3-4: Ethernet Connectors Info Firmware 1.0 DSP version 1.0 Serial no 123456789 Components.. Installed options.. Ethernet1 IP 123.456.78.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 3.10 Web Browser Ethernet2 MAC 123456789 IP: 123.456.78.9 Subnet: 255.255.255.1 Gateway: 123.456.78.0 Use DHCP: No IP Status Ready Apply Configuration! IP address for Ethernet2 connector • • • • • • • • • • Alternatively you may use the analyzer's fixed IP, that is not shown on any menu page and is accessible via Ethernet2 only: • The analyzer is now assigned the IP 192.168.1.88. • Configure your computer's IP to meet the same net (192.168.1.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 3.10 Web Browser IP address in web browser´s address line Analyzer information Analyzer serial number Analyzer work order number MAC address of interface Enter your password here Default password is: password Fig. 3-5: Web Browser Logon Screen We recommend to set new passwords, to limit access to critical submenus (see online help). After logon, the measurements screen appears.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 Chapter 4 Service and Maintenance Note! This chapter deals with service and maintenance procedures related to explosion protection only! More detailled instructions about servicing and maintaining general purpose components of X-STREAM XE gas analyzers are subject of the X-STREAM XE analyzer instruction manual. Inspection, maintenance and service must be carried out considering all related standards e.g.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 4 Service and Maintenance EXPLOSION HAZARD BY FLAMMMABLE GASES Leaks may cause explosion when measuring flammable gases! When measuring flammable gases it is recommended to perform a leak test on all gas paths, connections and components before startup or applying power. Leak tests should be carried out on a 2 month‘s regular basis and after repair/maintenance. See the analyzer instruction manual for instructions on how to carry out leak tests.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 4.1 Verifications and Tests Contents 4.1 Verifications and Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3 4.2 Replacement of Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-6 4.3 Vapor Recovery Application - Special Information . . . . . . . . . . . . . . . . . . . 4-8 4.4 Perform a calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 4.1 4.1.1.3 Verifications and Tests Tests on Flame Arrestors EXPLOSION HAZARD AND GASES HAZARDOUS TO HEALTH Maintaining the flame arrestor must be carried out considering all applicable safety and legislative rules.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 4.1 Verifications and Tests Analyzer flame arrestors stop valve overpressure approx. 0.725 psi/50 mbar Fig. 4-1: 4.1.1.3.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 4.2 Replacement of Parts 4.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 4.2 Replacement of Parts Specifically for CSA-C/US approved analyzers Any CSA-C/US certified flameproof type Male thread: M20 x 1,5 Female thread: 1⁄2“ or 3⁄4“ per request Metric-to-NPT adapters e.g. - AD-U series by Redapt Ltd. ⁄4“ version spare part # ETC01877 3 Cable entries stopping plugs Maintenance Any CSA-C/US certifed flameproof type Male thread: M20 x 1,5 e.g. - PD-U by Redapt Ltd.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 4.3 Vapor Recovery Application - Special Information 4.3 Vapor Recovery Application - Special Information EXPLOSION HAZARD BY HOT COMPONENTS Consider the waiting time statement on the front door label before opening! Do not open the instrument when energized. Ensure that external circuitry is disconnected or de-energized before opening the instrument.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 4.3 Vapor Recovery Application - Special Information Note! The analyzer is configured to measure CH4 at channel 1 (CH1), with the converter installed in front of its measurement cell. The total content of hydorcarbons (CnHm) within the sample gas is measured at channel 2 (CH2). • • Zero Calibration.. Span Calibration.. Advanced Calibration.. Apply Gas.. Lock Menus! Acknowledgements Pump 1: Pump 2: Ranges..
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 4.3 Vapor Recovery Application - Special Information 4.3.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 4.3 Vapor Recovery Application - Special Information 2 1. Loosen the lace cord that keeps the jacket closed. 3. Loosen the union nuts of the fittings that connect the gas lines from the converter to the solenoid valve (Fig. 4-7). 1 1: 2: Jacket Union nuts Fig. 4-6: Maintenance 2. Open the jacket to get access to the converter. 4 Heated jacket 4. Take the converter vessel with connected gas lines out of jacket and analyzer.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 4.3 Vapor Recovery Application - Special Information 5. Remove the end cap. 6. Remove the washer. Behind the washer, the vessel is filled with converter material. 7. Remove this old material and fill the vessel with new converter material, considering to • hold the vessel vertically while filling • completely fill the vessel • keep the inner pipe free of converter material. 8. Put the washer on top of the vessel / new material. 9.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 4.3 Vapor Recovery Application - Special Information 1 10. Insert the converter vessel into the jacket and place it into the assembly. 11. Connect the open gas lines to the fittings of the solenoid valve (Fig. 4-9). Carefully fix the union nuts. Fig. 4-9: Union nuts Jacket in converter assembly Maintenance 12. At this point we recommend to perform a leak test to ensure the gas connections are tight! If some are not, they are accessible now.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 4.3 Vapor Recovery Application - Special Information 4.3.3 Failure Situation Separated from the converter assembly an independent over temperature protection device is installed. This device is setup to take control of the jacket heaters if the measured temperature inside the jacket exceeds about 310 °C ( 590 °F), e. g. due to a failure with the temperature control board.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 4.4 Perform a Calibration 4.4 Perform a Calibration We recommend to perform at least a zero calibration, after startup of the instrument, to ensure proper measuring results. The following steps describe, how to perform a manual calibration. If your instrument features a valve block, X-STREAM XE instruction manual for a comprehensive description of calibration procedures.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 4.4 Manual Calibration 4.4.1 Manual Zero Calibration Starting from the MEASUREMENT SCREEN press down to open the MAIN MENU and enter CONTROL. Control.. Zero Calibration.. Span Calibration.. Advanced Calibration.. Apply Gas.. Lock Menus! Acknowledgements Pump 1: Pump 2: Ranges.. To start a zero calibration select the first line: Off Off Multi-channel unit: Select the channel to be calibrated in SELECT COMPONENT.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 4.4 Manual Calibration Manual Span Calibration Select "Span Calibration.." Off Off Multi-channel unit: Select the channel to be calibrated in SELECT COMPONENT. Before selecting any further line make sure the required calibration gas is applied and flowing! Span Calibration Cancel! Start! Span Gas 0.000 ppm Concentration 0.000 ppm Flow 0.00 l/min Status.. Results.. Restore.. Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 4.4 Manual Calibration 4.4.3 Manual Calibration for Vapor Recovery Applications Zero Calibration.. Span Calibration.. Advanced Calibration.. Apply Gas.. Lock Menus! Acknowledgements Pump 1: Pump 2: Ranges.. 1. The measurement system for this applications includes a converter, therefore calibration span gas must contain oxygen for the converter to work properly. 2. During channel 1 calibrations, the calibration gas must go through the converter.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 Chapter 5 Dismounting and Disposal 5.1 Dismounting and Diposal of the Analyzer HAZARDS FROM DISMOUNTING Dismounting instruments installed in hazardous area requires special instructions to be followed! Only trained personnel, observing all applicable technical and legal requirements, and aware of the possible risks is permitted to dismount these analyzers.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 5.1 Dismounting & Disposal GASES HAZARDOUS TO HEALTH Follow the safety precautions for all applied gases and gas cylinders. Before opening the gas lines, they must be purged with air or neutral gas (N2) to avoid danger from escaping toxic, flammable, explosive or hazardous gases.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 Appendix This chapter contains A.1, page A-2 ATEX EC Type Examination Certificate A.2, page A-3 CSA Certificate of Compliance A.3, page A-12 Block diagram A.4, page A-19 Assignment of Terminals and Socket A.5, page A-25 A Appendix EC Declaration of Conformity Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 A.1 EC Declaration of Conformity A.1 EC Declaration of Conformity A-2 Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 A.2 ATEX EC Type Examination Certificate A Appendix A.2 ATEX EC Type Examination Certificate Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XEFD A.2 A-4 HASXEDE-IM-EX 03/2012 ATEX EC Type Examination Certificate Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 ATEX EC Type Examination Certificate A Appendix A.2 Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XEFD A.2 A-6 HASXEDE-IM-EX 03/2012 ATEX EC Type Examination Certificate Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 ATEX EC Type Examination Certificate A Appendix A.2 Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XEFD A.2 A-8 HASXEDE-IM-EX 03/2012 ATEX EC Type Examination Certificate Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 ATEX EC Type Examination Certificate A Appendix A.2 Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XEFD A.2 A-10 HASXEDE-IM-EX 03/2012 ATEX EC Type Examination Certificate Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 ATEX EC Type Examination Certificate A Appendix A.2 Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 A.3 CSA Certificate of Compliance A.3 CSA Certificate of Compliance Certificate of Compliance Certificate: 1714037 (LR 105173) Master Contract: 185562 Project: 2507282 Date Issued: March 14, 2012 Issued to: Emerson Process Management GmbH & Co.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 A.3 CSA Certificate of Compliance Certificate: 1714037 (LR 105173) Master Contract: 185562 Project: 2507282 Date Issued: March 14, 2012 Gas analyzer, Model: X-STREAM, rated 100-240Vac, 50/60 Hz, 1, 3 – 0,7A, Class I, Pollution Degree II.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 A.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 A.3 CSA Certificate of Compliance Certificate: 1714037 (LR 105173) Master Contract: 185562 Project: 2507282 Date Issued: March 14, 2012 X-Stream FD (X2FD) has same electronics as the X-STREAM General Purpose Gas Analyzer (X2GP) with new Hazardous Locations Enclosure.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 A.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 A.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 A.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 2 3 4 5 6 7 8 10 9 11 12 N B T 6.3 A / 250V 4.300-8112 3 EMV Filter 6ET1 1-stufiges Filter EMI Filter 6ET1 1-stage Filter T 6.3 A / 250V PE X-STREAM Version 4.300-8102 Netzeingangsmodul mit Schraubklemmen, 2-polige Sicherung 5x20 mm Model XEFD C Gn/ge gn/ye Power Input AC 100 - 240 V wide range 50 / 60 Hz Power entry module with screw terminals, 2 fuseholder 5x20 mm Zeichnung/drawing 4.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 10 9 P1 XSTA 01 Intern CAN (QSPI) C P6 RJ45 P2, P3 P22 P5 P14 P9 XECB 01 P2 XDIO 01 / 02 Stromlaufplan / schematics: 4.350-0400 Sub-D USB-A/USB-Mini P15, P18 P13 Stromlaufplan / schematics: 4.300-4680 9 Weitergabe sowie Vervielfältigung dieser Unterlage, Verwertung und Mitteilung ihres Inhaltes sind nur mit unserer Zustimmung gestattet. Alle Rechte vorbehalten. A-20 xx.xx.12 TN1274: Aktualisiert 11.11.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 A.4 3 4 5 siehe Blatt 3, 5, 8 see sheet 3, 5, 8 P3 A 6 7 8 9 10 11 12 13 P26 P27 P28 15 16 XPSA 01 / 02 P32 Stromlaufplan / schematics: 4.300-4660 / 4.350-0920 XSP 01 P12 P13 P14 P15 P19 P20 P5 P4 P7 A siehe auch Blatt 3, 5, 8 see also sheet 3, 5, 8 Stromlaufplan / schematics: 4.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 A.4 Copying of this document, and giving it to others and the use or communication of the contents thereof, are only allowed with our agreement. All rights are reserved. 1 2 3 QSPI 4 siehe Blatt 3, 5, 8 5 +Power 6 Block Diagram 7 8 9 10 11 12 13 14 15 16 see sheet 3, 5, 8 P8 A XPSA 01 / 02 P3 Stromlaufplan / schematics: 4.300-4660 / 4.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 A.4 2 3 4 5 6 7 8 10 9 11 12 13 14 15 16 Symbolerklärung: Description of used icons: 2 bl A ST3 JP1 SPI 01 TCD 01 Stromlaufplan / schematics: 4.350-0200 Thermal conductivity sensor Stromlaufplan / schematics: 4.300-2000 3/C10 5/C10 8/C10 11/D4 12/D4 3/C10 5/C10 8/C10 11/D4 12/D4 4 2 XHA 01 Stromlaufplan / schematics: 4.350-0980 Stromlaufplan / schematics: 4.
Instruction Manual X-STREAM XEFD HASXEDE-IM-EX 03/2012 A.4 2 3 4 5 6 7 8 10 9 11 12 13 14 15 16 Symbolerklärung: Description of used icons: 2 bl 2-adrige Leitung / cable with 2 lines Leitungsfarbe / cable colour Steckverbindung / plug connection A A /2.4 Verweis auf Seite 2, Spalte 4 / reference to sh 2, row 4 4.
Signal } } Appendix Pin not used not used not used not used not used not used not used not used not used not used not used not used { BR 1 fit a wire bridge here to apply analog BR 1 signal in current mode to input 1 Input 1 high (+) Input 1 high (+) Input 1 low (-) Input 1 low (-) BR 2 fit a wire bridge here to apply analog BR 2 signal in current mode to input 2 Input 2 high (+) Input 2 high (+) Input 2 low (-) Input 2 low (-) { P3.1 P3.2 P3.3 P3.4 P3.5 P3.6 P3.7 P3.8 P3.9 P3.10 P3.11 P3.12 P2.
X-STREAM XEFD A-26 Instruction Manual HASXEDE-IM-EX 03/2012 Emerson Process Management GmbH & Co.
Instruction Manual HASXEDE-IM-EX 03/2012 X-STREAM XEFD
Instruction Manual HASXEDE-IM-EX 03/2012 X-STREAM XEFD EUROPE ASIA-PACIFIC Emerson Process Management GmbH & Co. OHG Rosemount Analytical Process Gas Analyzer Center of Excellence Industriestrasse 1 63594 Hasselroth, Germany T +49 6055 884 0 F +49 6055 884 209 www.RosemountAnalytical.com Emerson Process Management Asia Pacific Private Limited 1 Pandan Crescent Singapore 128461 Republic of Singapore T +65 6 777 8211 F +65 6 777 0947 analytical@ap.emersonprocess.com www.RosemountAnalytical.