User's Manual

EMS Wireless Operator’s Manual
EkoLink II
3
In this section, instructions are provided for organizing the system layout before
installation. The installer or operator must make sure that certain requirements
have been met before beginning installation. In addition, all cables and fibers must
be labeled, and the entire system configuration must be documented. Steps for
this pre-installation set-up are given below.
Frequent referral to the Functional Overview section of the Operator’s Manual with
its detailed photographs will facilitate installation of the system.
1. One optical fiber is required for each Antenna unit; each Hub can connect to
up to four Remote units. Angle-polished SC/APC connectorized optical fiber cable
must be installed in the building. Any optical connector used between the Hub and
the Remote units should be an angle-polished connector specified for return loss
>55 dB. Fusion splices to connect fibers are also suitable. Flat polished
connectors anywhere along the optical fiber path will degrade the system
performance.
2. The customer must provide the following equipment for installation:
(a) Optical connector cleaning kit: cotton swabs, alcohol, dust-free
compressed air
(b) Crimping tool, mating connectors, and pin crimps for the DC
electrical connectors and alarm outputs at the Hub.
(c) Philips screwdriver
(d) Personal computer or laptop with terminal program such as
HyperTerminal.
(e) Volt Meter
3. Not essential, but useful for diagnostics are the following equipment:
(a) RF power meter
(b) Optical power meter
(c) RF Spectrum analyzer
(d) Signal generator
1.4 Power Supply Installation
If purchased, install the power supply and connect power to the Hub. An on-off
switch is located in the rear of the chassis and should remain in the off position until
the installation is completed. Before connecting the Hub or Remotes to the power
supply, apply AC power to the supply and verify the 48 V output with a Volt meter.