User Manual

10
© 2019 Endotronix® Inc. 100469-00 Rev. 2
Getting Started
The myCordella™ Patient Reader will automatically pair with the
myCordella Hub on start-up. No additional user input is required.
Upon return home with the myCordella Reader and Docking Station,
connect the Docking Station Power Cord to the power port on the
back of the Docking Station and the other end into the wall outlet.
Place the Docking Station on a flat surface near the myCordella Hub
with a minimum two-inch/five-centimeter clearance from
surrounding objects or walls.
Place the Reader in the Docking Station. The LED light on the front of
the Docking Station will blink while charging and change to solid white
when the Reader is charged. Call Endotronix customer service if this
does not occur within 24 hours of setup.
The Reader is distributed in Travel Mode. Start the Reader for the first
time by holding the travel button on the bottom for several seconds.
Reader Placement
Reliably and repeatedly finding proper Reader location is essential to
collecting valid PA pressure measurements.
Position the Reader on the prescribed
location (from training), most likely
the upper right chest. A clinician may
indicate that the Reader should be
placed in a different locationfor
example on the side of the body. The
Reader will wait for good signal
strength before gathering information
from the Sensor, so adjust the position
until the Reader plays the Sensor
Located tone and displays a quickly
flashing blue light (see the Audio &
Visual Cues section below).