Appendix B: Equipment Specifications
myCordella™ Patient Reader
Manufacturer: Endotronix®
CAUTION--Investigational device. Limited by Federal (or United States) law to
investigational use.
Method of measurement: Wireless interrogation of implanted Cordella Sensor
Pulmonary artery pulse pressure maximum range: 40-100 mmHg
Pulmonary artery pressure range at sea level: 0-100 mmHg
Pressure Transducer Accuracy: +/- 7.8 mmHg over full scale range, for
operating conditions between 15°C and 30°C and relative humidity between
6% and 93%.
Patient safety/use: Typical reading time is 30 seconds.
Calibration: At implant and when deemed necessary by a medical professional
Expected service life (of Reader only): One year
Safety standards: Meets all relevant parts of IEC 60601-1 Ed. 3.1
EMC standards: Meets all relevant parts of IEC 60601-1-2 Ed. 4.0
Operating frequency: 12.88-14.12 MHz, 2.45 GHz
Essential performance is maintained for the service life of the product. For
additional information regarding essential performance contact Endotronix
customer service.
Physical
Approximate dimensions
Width: 6.44 in / 16.35 cm
Height: 2.0 in / 5.1 cm
Depth: 5.63 in / 14.3 cm
Weight: 1.1 lbs / 0.5 kg
Power
Input of 4.2V/16.8V 925mA/250mA
Environment
The Reader may not meet its performance specifications if stored or used
outside the temperature and humidity ranges listed below.