Electrical Fast
Transient/Burst
Immunity
IEC 61000-4-4
±2 kV for power supply
lines
±2 kV for power
supply lines
Mains power quality should be
that of a typical professional
healthcare facility or home
healthcare environment.
Surge
IEC 61000-4-5
±1 kV line to line
±1 kV line to line
±0.5 kV line to line
±0.5 kV line to
line
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11
0% UT (100% dip in UT)
for 0.5 cycle
0% UT (100% dip
in UT) for 0.5
cycle
0% UT (100% dip in UT)
for 1 cycle
0% UT (100% dip
in UT) for 1 cycle
70% UT (30% dip in UT)
for 25 cycles(50Hz) and
30 cycles(60Hz)
70% UT (30% dip
in UT) for 25
cycles(50Hz) and
30 cycles(60Hz)
0% UT (100%
interruption in UT) for
250 cycles(50Hz) and
300 cycles(60Hz)
0% UT (100%
interruption in
UT) for 250
cycles(50Hz) and
300 cycles(60Hz)
Power Frequency
Magnetic Field
IEC 61000-4-8
30A/m
50Hz and 60Hz
30A/m
50Hz and 60Hz
Power frequency magnetic fields
should be at levels characteristic
of a typical
professional healthcare facility or
home healthcare environment.
Radiated RF
Immunity
IEC 61000-4-3
10V/m
80MHz-2.7GHz
80% AM at 1kHz
10V/m
Portable and mobile RF
communication equipment
should be no closer to any part of
the system, including cables, than
the recommended separation
distance calculated from the
equation applicable to the
frequency of the transmitter.
Recommended separation
distance
d= 1.5 meters
Recommended separation
distance for Rated Maximum
Output Power of Transmitter
above 20W:
d = 0.35√P
150 kHz to
80 MHz
d = 0.35√P
80MHz to
800 MHz
Conducted RF
Immunity
IEC 61000-4-6
3 Vrms
Outside the ISM Bands
6 Vrms
In the ISM and amateur
radio bands
150kHz to 80MHz