Micropilot FMR10
Endress+Hauser 13
Influence of ambient
temperature
The measurements are carried out in accordance with EN 61298-3.
• Digital (Bluetooth® wireless technology):
Standard version: average T
K
= ±3 mm (±0.12 in)/10 K
• Analog (current output):
– Zero point (4 mA): average T
K
= 0.02 %/10 K
– Span (20 mA): average T
K
= 0.05 %/10 K