Micropilot FMR60B HART Diagnostics and troubleshooting
Endress+Hauser 43
11.7 Diagnostic list
In the Diagnostic list submenu submenu, up to 5 currently pending diagnostic messages
can be displayed. If more than 5 messages are pending, the messages with the highest
priority are shown on the display.
Navigation path
Diagnostics → Diagnostic list
Calling up and closing the remedial measures
1. Press .
The message for the remedial measures for the selected diagnostic event opens.
2. Press  +  simultaneously.
The message about the remedial measures closes.
11.8 List of diagnostic events
Diagnostic
number
Short text Remedy instructions Status
signal
[from the
factory]
Diagnostic
behavior
[from the
factory]
Diagnostic of sensor
062 Sensor connection faulty Check sensor connection F Alarm
151 Sensor electronic failure Replace sensor electronic module F Alarm
168 Build-up detected Check process conditions M Warning
1)
Diagnostic of electronic
203 HART Device
Malfunction
Check device specific diagnosis. S Warning
204 HART Electronic Defect Check device specific diagnosis. F Alarm
242 Firmware incompatible 1. Check software
2. Flash or change main electronic
module
F Alarm
252 Module incompatible 1. Check if correct electronic
module is plugged
2. Replace electronic module
F Alarm
270 Main electronics
defective
Replace main electronics F Alarm
272 Main electronics faulty 1. Restart device
2. Contact service
F Alarm
273 Main electronics
defective
Replace main electronics F Alarm
282 Data storage inconsistent Restart device F Alarm
283 Memory content
inconsistent
1. Restart device
2. Contact service
F Alarm
287 Memory content
inconsistent
1. Restart device
2. Contact service
M Warning
388 Electronics and
HistoROM defective
1. Restart device
2. Replace electronics and
HistoROM
3. Contact service
F Alarm
Diagnostic of configuration
410 Data transfer failed 1. Retry data transfer
2. Check connection
F Alarm