Operating Instructions

Table Of Contents
Micropilot FMR60B HART Diagnostics and troubleshooting
Endress+Hauser 47
Reset device parameter
 For details see the "Description of device parameters" documentation.
11.10.2 Resetting the device via keys on the electronic insert
Reset the password
Display
onoff
II
15 s
I I I I
I
A0050210
 11 Sequence - resetting the password
Deleting/resetting the password
1. Press operating key I three times.
The Reset Password function is started, the LED flashes.
2. Press operating key I once within 15 s.
The password is reset, the LED flashes briefly.
If operating key I is not pressed within 15 s, the action is canceled and the LED is no longer
lit.
Resetting the device to the factory setting
12 s
50 ms
Display
onoff
I
II
I
II
A0050009
 12 Operating keys on the electronic insert
Resetting the device to the factory setting
Press operating key I and operating key II simultaneously for at least 12 s.
Device data are reset to the factory setting, the LED flashes briefly.
11.11 Device information
All the device information is contained in the Information submenu.
Navigation: System → Information