Product Info
Table Of Contents
- Table of contents
- Important document information
- Function and system design
- Input
- Output
- Power supply
- Performance characteristics
- Mounting
- Environment
- Process
- Mechanical construction
- Operability
- Certificates and approvals
- Ordering information
- Application packages
- Accessories
- Weather protection cover 316L
- Plastic weather protection cover
- Mounting bracket, adjustable
- M12 socket
- Remote display FHX50B
- Gas-tight feedthrough
- Commubox FXA195 HART
- HART Loop Converter HMX50
- FieldPort SWA50
- Wireless HART adapter SWA70
- Fieldgate FXA42
- Field Xpert SMT70
- DeviceCare SFE100
- FieldCare SFE500
- Memograph M
- RN42
- Documentation
- Registered trademarks
Micropilot FMR60B HART
Endress+Hauser V. 1, Rev. 2, 26-07-2022 51
• русский язык (Russian)
• Türkçe
• 中文 (Chinese)
• 日本語 (Japanese)
• 한국어 (Korean)
• tiếng Việt (Vietnamese)
• čeština (Czech)
• Svenska
Local operation Operating keys and DIP switches on the HART electronic insert
Display
onoff
1
2
onoff
3
4
A0046129
51 Operating keys and DIP switches on the HART electronic insert
1 Operating key for reset password (for Bluetooth login and Maintenance user role)
1+2 Operating keys for device reset (as-delivered state)
2 Operating key II (only for factory reset)
3 DIP switch for alarm current
4 DIP switch for locking and unlocking the device
The setting of the DIP switches on the electronic insert has priority over the settings made via
other operation methods (e.g. FieldCare/DeviceCare).
Local display Device display (optional)
Functions:
• Display of measured values and fault and notice messages
• Background lighting, which switches from green to red in the event of an error
• The device display can be removed for easier operation
ESC
-
+ E
1
A0039284
52 Graphic display with optical operating keys (1)