Installation Instructions

FieldPort SWA50
12 Endress+Hauser
Environment
Ambient temperature range
–40 to +85 °C (–40 to +185 °F)
Ex version for temperature class T4: –40 to +70 °C (–40 to +158 °F)
Storage temperature range
–40 to +85 °C (–40 to +185 °F)
Climate class
As per IEC 60068-2-38 test Z/AD
Degree of protection
When housing is closed, tested according to:
IP68 / NEMA 6P (24 h at 1 m under water)
IP66 / NEMA 4X
Vibration resistance
As per IEC 60068-2-64:2008
a(RMS) = 50 m/s²
f = 5 to 2000 Hz
t = 3 planes x 2 h
Vibration resistance tested only for "remote mounting" version.
Shock resistance
As per IEC 60068-2-27:2008
300 m/s² [= 30 gn] + 18 ms
Electromagnetic
compatibility (EMC)
Electromagnetic compatibility as per EN 61326 series and NAMUR recommendation EMC (NE 21)
Mechanical construction
Design, dimensions "Remote mounting" version
1 1
2
Ø 41.9 (1.65)
35.1 (1.38)
114.6 (4.51)
79.5 (3.13)
Ø 39.9 (1.57)
SW 36
(36 AF)
SW 27
(27 AF)
A0040484
 12 Dimensions of "remote mounting" version in mm (in)
1 M20 cable gland
2 Transmission window