User's Manual

Table Of Contents
DRAFT
Micropilot FMR51, FMR52
Endress+Hauser V. 1, Rev. 8, 4-10-2012 81
Ordering information
Ordering information
Detailed ordering information is available from the following sources:
In the Product Configurator on the Endress+Hauser website: www.endress.com ® Select country ®
Instruments ® Select device ® Product page function: Configure this product
From your Endress+Hauser Sales Center: www.endress.com/worldwide
Product Configurator - the tool for individual product configuration
Up-to-the-minute configuration data
Depending on the device: Direct input of measuring point-specific information such as measuring
range or operating language
Automatic verification of exclusion criteria
Automatic creation of the order code and its breakdown in PDF or Excel output format
Ability to order directly in the Endress+Hauser Online Shop
5-point linearity protocol (in
preparation)
The following notes must be taken into account if option F4 ("5 point linearity protocol") has been
selected in feature 550 ("Calibration").
The five points of the linearity protocol are evenly distributetd across the measuring range (0% to 100%). In
order to define the measuring range, Empty calibration (E) and Full calibration (F) have to be specified
8)
.
The following restrictions have to be taken into account when defining E and F:
R
A
F
200 (7.87)
mm (in)
E
A0017983
Minimum distance between reference point (R) and
100% level
Minimum span Maximum value for "empty
calibration"
A ³ Length of antenna + 200 mm (8 in)
Minimum value: 400 mm (16 in)
F ³ 400 mm (16 in) E £ 24 m (79 ft)
The linearity is checked under reference conditions.
The selected values of Empty calibration and Full calibration are only used to record the linearity
protocol and are reset to their antenna specific default values thereafter. If values different from the
default are required, they must be ordered as a customized parametrization (® ä 82).
8) If E and F are not specified, antenna dependent default values will be used instead.