Operating Instructions

Table Of Contents
Liquiline Mobile CML18 Operation options
Endress+Hauser 19
Display structure
1
2
3
4
5
A0041052
 7 Schematic representation of the display structure
1 Menu path/title of measuring screen
2 Bluetooth status
3 Battery level, charging information
4 NAMUR indicator *
5 Measuring screen
1)
NAMUR indicator Status
OK The device and sensor are working reliably.
F Failure of device or sensor.
F status signal as per NAMUR NE107
M Device or sensor requires maintenance.
M status signal as per NAMUR NE107
C Device or sensor undergoing function check.
C status signal as per NAMUR NE107
S Device or sensor being operated out of specification.
S status as per NAMUR NE107
1) * Status according to NAMUR NE107 categories: