CHANGING THE PROSTHETIST SETTINGS
The system is shipped with the User Maximum set at 20 and
the Vacuum Range set at 6. If these settings are not sucient
for your patient, you may adjust them as follows:
Make sure the pump and fob are both turned on and are
communicating with each other. Quickly press the Enter button
on the fob three times to enable the Prosthetist Settings function.
The three presses must be within approximately two seconds
of each other.
The display will alternate between two quick
flashes of this symbol and the current value for
the User Maximum.
Press the Up or Down buttons to adjust the User Maximum.
The digits will flash to indicate that the User Maximum is being
changed.
• The user will not be able to increase the Vacuum Set Point
above the User Maximum.
• The User Maximum cannot be set higher than 20.
Press the Enter button to apply the new User Maximum setting
and advance to the next setting.
The display will alternate between two quick
flashes of this symbol and the current value for
the Vacuum Range.
Press the Up or Down buttons to adjust the Vacuum Range.
The digits will flash to indicate that the Vacuum Range is being
changed.
• TheVacuum Range is how much the current vacuum level
can fall below the Vacuum Set Point before the pump
activates.
• The Vacuum Range cannot be set lower than 4.
Press the Enter button to apply the new Vacuum Range setting.
The “chasing” pattern will be displayed while the setting is be-
ing applied. The fob will then return to displaying the current
vacuum level.
NOTE: Adjusting the User Maximum and Vacuum Range Settings
can aect the Vacuum Level setting. After changing either the
User Maximum or Vacuum Range settings, please check to
ensure the Vacuum Level is set to the desired level.