User guide

Troubleshooting 135
Is priming complete?
If you manually prime your pump with more than 30u insulin, this message
appears. Press ESC, ACT to clear the message. If manual prime is complete,
press ESC. If the manual prime is not complete, press and hold ACT until
manual prime is complete.
Max delivery
This alarm alerts you when you have taken more insulin than expected
based on maximum bolus and maximum basal rates.
Motor error
Insulin delivery has stopped. This alarm occurs if your pump detects a
motor error.
No delivery
Insulin delivery has stopped. This alarm occurs if your pump detects a
blockage.
No reservoir
The reservoir is not inserted correctly or no reservoir has been inserted.
IS PRIMING
COMPLETE?
Disconnect from body
for instruction
See user guide
ESC, ACT to clear
MAX
DELIVERY
Exceeded 1 hour
max delivery
Check BG
3:25P
ESC, ACT to clear
MOTOR
ERROR
Delivery stopped
Disconnect set
ESC, ACT to clear
3:25P
Delivery stopped
Change entire set
Check BG
See user guide
to troubleshoot
ESC, ACT to clear
NO
DELIVERY
11:17A
NO
RESERVOIR
Delivery Stopped
Change reservoir set
ESC, ACT to clear
4:36P