Use Instructions

DRAFT 2/12/2019 SPW 26
Operating Environment
The sTMS mini is a battery-operated device.
The sTMS mini requires a regular home or office electrical outlet for proper battery recharging.
The sTMS mini operates best in typical household and office environments where the temperature is 15°C to 40°C (60°F
to 104°F); relative humidity range of 10% to 90%, non-condensing; an atmospheric pressure range of 700 hPa to 1013
hPA; operating altitude of less than or equal to 3000m.
The sTMS mini is not affected by exposure to dust or lint normally found in household and office environments.
The sTMS mini and its accessories may be exposed to sunlight for a few days duration, but prolonged or repeated
exposure to direct sunlight is not recommended.
It is recommended that a cold device (below 60°F or 15°C) be allowed to warm to room temperature for at least 15
minutes prior to use. Similarly, it is recommended that a hot device (above 104°F or 40°C) be allowed to cool for at
least 15 minutes prior to use.
The sTMS mini is suitable for home use, and complies with IEC 60601-1-11:2015.
Precaution: Do NOT actively heat or cool your sTMS mini in any way.
Storage Environment
The sTMS mini may be stored at -40°C to 60°C (-40°F to 140°F) with a relative humidity of up to 90% non-condensing
and an atmospheric pressure range of 700 hPa to 1060 hPa and storage altitude of less than or equal to 3000m.
Keep your sTMS mini dry during operation, storage and transport. When not in use, store the sTMS to protect it from
household dust and lint, accidental spills or exposure to the elements (rain, dust, insects, vermin, prolonged direct
sunlight, etc).
During extended storage periods, it is recommended that the battery be periodically recharged every 6 months, so that
the device will be available for use when needed.
Industry Standard Classification
(IEC 60601-1)
Classification
Type of Electric Shock Protection
(sTMS mini)
(Battery Charger, external)
Internally powered
Class II
Degree of Electric Shock Protection
Type B applied part
Degree of Protection Against
Harmful Ingress of Liquids and
Particulates
IP22
Safety in the Presence of
Flammable Atmosphere
Not suitable for use in
flammable environment
Mode of Operation
Short-time operation
individual 1ms pulses
EMC
Class B
Service Life (sTMS mini device, AC
Adapter and Rx SIM Card)
1800 treatments