8 Commands
Most of the commands can only be performed when the data acquisition is stopped (See commands 
stopacq and startacq). Keep in mind that this will corrupt any running schedulers as the direct command 
interface   bypasses   the   scheduler   and   the   running   Frog   software.   When   operating  the   ENIGMA 
signalprocessor with the Frog/MURAN software it is virtually never necessary to perform any of these 
commands manually.
8.1 Noise sampling
Noise sampling is performed by command 'noisesample'.
Parameters: 
  startRange - in km
  prf - in Hz
It returns the following values:
IoffsetH=<val>  - Offset of horizontal I
IoffsetV=<val>  - Offset of vertical I
QoffsetH=<val>  - Offset of horizontal Q
QoffsetV=<val>  - Offset of vertical Q
noisePowerH=<val>  - Horizontal noise power
noisePowerV=<val>  - Vertical noise power
8.2 SDP status
In addition to the ray header, some status is obtained by the following commands:
8.2.1 getburst
get burst parameters. Returns:
sigpow  - signal power in dB
freq  - burst freq in Hz
phase  - burst phase in degree