EX-5185-PID
ENMET Corporation
10
Table 3: Examples of Standard Calibration Gas
Gas Range Alarm 1*
Alarm 2*
Alarm 3*
Sensor
Part Number
Span Gas
Isobutylene
Gas Cylinder
Part Number
Benzene 0 - 20 ppm 0.5 PPM 1.0 PPM 5.0 PPM 03028-010 10 PPM 03290-010
1, 3-Butadine 0 – 20 ppm 0.5 PPM 1.0 PPM 5.0 PPM 03028-010 10 PPM 03290-010
Carbon disulfide 0 – 20 ppm 1.0 PPM 5.0 PPM 10 PPM 03028-010 10 PPM 03290-010
Chlorobenzene 0 – 200 ppm 50 PPM 75 PPM 100 PPM 03028-011 100 PPM 03290-100
Cyclohexanone 0 – 200 ppm 25 PPM 50 PPM 75 PPM 03028-011 100 PPM 03290-100
Diethylamine 0 – 20 ppm 5.0 PPM 10 PPM 15 PPM 03028-010 10 PPM 03290-010
Dimethyl acetamide 0 – 20 ppm 5.0 PPM 10 PPM 15 PPM 03028-010 10 PPM 03290-010
Hydrazine 0 – 20 ppm 0.5 PPM 1.0 PPM 5.0 PPM 03028-010 10 PPM 03290-010
Isobutyl acetate 0 – 2000 ppm 100 PPM 250 PPM 500 PPM 03028-012 750 PPM 03290-750
Isopropyl acetate 0 – 2000 ppm 100 PPM 250 PPM 500 PPM 03028-012 750 PPM 03290-750
Isopropylamine 0 – 20 ppm 0.5 PPM 1.0 PPM 5.0 PPM 03028-010 10 PPM 03290-010
Methyl bromide 0 – 20 ppm 5.0 PPM 10 PPM 15 PPM 03028-010 10 PPM 03290-010
Methyl ethyl Ketone 0 – 2000 ppm 100 PPM 250 PPM 500 PPM 03028-012 750 PPM 03290-750
Methyl mercaptan 0 – 20 ppm 0.2 PPM 0.5 PPM 1.0 PPM 03028-010 10 PPM 03290-010
Methylamine 0 – 20 ppm 5.0 PPM 10 PPM 15 PPM 03028-010 10 PPM 03290-010
Naphthalene 0 – 20 ppm 5.0 PPM 10 PPM 15 PPM 03028-010 10 PPM 03290-010
Octane 0 – 2000 ppm 100 PPM 250 PPM 500 PPM 03028-012 750 PPM 03290-750
Phenol 0 – 20 ppm 5 PPM 10 PPM 15 PPM 03028-010 10 PPM 03290-010
Tetrachloroethylene 0 – 200 ppm 75 PPM 100 PPM 150 PPM 03028-011 100 PPM 03290-100
Toluene 0 – 2000 ppm 100 PPM 250 PPM 500 PPM 03028-012 750 PPM 03290-750
Trichloroethylene 0 – 20 ppm 5 PPM 10 PPM 15 PPM 03028-010 10 PPM 03290-010
Trimethylbenzene 0 – 20 ppm 5 PPM 10 PPM 15 PPM 03028-010 10 PPM 03290-010
Vinyl; acetate 0 – 20 ppm 0.5 PPM 1.0 PPM 5.0 PPM 03028-010 10 PPM 03290-010
Vinyl chloride 0 – 20 ppm 0.2 PPM 0.5 PPM 1.0 PPM 03028-010 10 PPM 03290-010
*
NOTE:
The detectable range and the availability of span/calibration gas dictates that for some gasses the span/calibration gas will not
trigger the high alarm point.
*
NOTE:
EX-5185
alarm points are independent of a controller. When a controller is supplied by
ENMET
Corporation the alarm
points for the
EX-5185
and the controller have been set accordingly. See the manual of the controller the
EX-5185
sensor
transmitter has been connected to for proper adjustment of the alarm points of the controller with the
EX-5185
sensor
transmitter.