Owner manual

ISA-44-2-OD ENMET Corporation
21
5.4 Trouble Shooting
Symptom Problem/Correction
When first plugged in, the unit alarms for
up to 10 - 20 minutes.
If the unit has not been used recently, this is normal. The MOS
sensor is cold and/or contaminated. Keep unit ON or PURGE (on
units so equipped) overnight with the sensor in fresh air.
Red alarm light stays on continuously, even
after allowed to operate (or PURGED)
overnight.
Either hazardous gas conditions exist, or there is a contaminated
sensor, or a bad circuit. Contact ENMET for voltage checks.
Oxygen channel is in constant alarm (amber
light and horn are on).
Probably a dead fuel cell; replace ( see Section 5.3 ). It can also be a
grossly contaminated fuel cell; replace also.
Either green or amber or red lights do not
work.
There might be a burned out lamp.
Unit fails to alarm when calibration gas is
applied.
Recalibration may be necessary. See section 6.0 If recalibration
fails, check sensor voltage, cylinder contents and pressure.
Unit gradually creeps into alarm or
sensitive.
Either hazardous gas conditions exist, or too recalibration is
necessary. On units requiring purging, rotate to PURGE once a
week for one hour to clean sensor surface.
Unit does not operate. Check the fuse next to TB-6. If the fuse element in the glass tube is
cracked or broken, replace it (p/n 64004-001). If this does not
correct the problem, contact ENMET