Table of Contents 1.0 INTRODUCTION................................................................................................................................................................ 1 1.1 UNPACK .............................................................................................................................................................................. 1 1.2 CHECK ORDER ...............................................................................................................
RECON/B SERIES ENMET 1.0 Introduction The RECON/B SERIES is a portable-single gas detector for detection of combustible or toxic gases. It is intended for detection of combustible or toxic gas in commercial or industrial facilities to keep workers safe, and/or detect a potentially explosive gas environment. The RECON/B SERIES, adopting excellent-quality sensor, makes detection in the way of natural diffusion. It has good sensitivity and reproducibility.
RECON/B SERIES ENMET 2.0 Features 2.1 RECON/B SERIES Features See Figure 1 for location of features: 1 FEATURE DESCRIPTION Visual Alarms Red indicators clearly visible on 3 sides and located at the top of the RECON/B SERIES LED’s activate when alarm thresholds are exceeded 2 Pushbuttons POWER, UP, SET and DOWN Buttons See section 3.
RECON/B SERIES ENMET LCD, DISPLAY NUMBER 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 FUNCTION Battery Lock sound vibration Alarming conditions Warning or the maximum level numeric value Error Not Used Clock Zero calibration Calibration point or calibration status Not Used Set-up of the parameters Indication of the gas concentration units Figure 2: RECON/B SERIES Display 3.
RECON/B SERIES ENMET 4.0 Operation 4.1 Activation of the RECON/B SERIES To activate the detector, press the POWER button and hold for 3 seconds. The detector begins a self test sequence, example listed below 1 Display Test The LCD displays all of the screen elements.
RECON/B SERIES ENMET 4.3 Checking RECON/B SERIES Status To activate the detector, press the POWER button and hold for 3 seconds. and SET When the RECON/B SERIES has finished its start up sequence. Press UP will indicate Temp., time, STEL & TWA levels(for toxic) and maximum level buttons at the same time, the display 4.
RECON/B SERIES ENMET 4.5 Alarms When alarming continuously, you can press SET button to turn off the audible alarm and vibration and are flashing. when the The following table describes RECON/B SERIES alarms and shows examples of display for each alarm.
RECON/B SERIES ENMET Time error In this condition, the device will try to renovate it by itself. If succeeded, after start-up, it enters into the menu of time set up. Please set up the time according to the local time. If failed, the device will turn off. Please contact the supplier for repair. Memorizer error When detecting, the device will try to renovate it by itself. If failed, it will turn off. Please contact the supplier for repair.
RECON/B SERIES ENMET 5.0 Maintenance of the RECON/B SERIES To maintain proper battery function and life the Recon should be operated once a week for 6 to 8 hours and then placed on charge see Section 5.3 The RECON/B SERIES maintenacne section allows access to Zero, Calibration, Calibration target gas and alarm point adjustment menus. To enter the maintenance menus section you must start with the instrument OFF.
RECON/B SERIES ENMET 5.1.3 Calibration Point Adjustment In this mode, display is the last concentration of gas used to calibrate the instrument. Press UP or Down button to modify the display. NOTE: Not used with Recon Oxygen units NOTE: 20.9% Oxygen is the only calibration level to be used with Recon Oxygen units Open the valve on the regulator Within 30 seconds, if the RECON/B SERIES detects the calibration gas, it will start up the calibration date analysis procedure by itself. Continue gas flow.
RECON/B SERIES ENMET 5.2 Alarm Levels Adjustment of the RECON/B SERIES 5.2.1 Set up of L-alarm (Low Alarm) With the RECON/B SERIES Off press and hold the POWER and SET buttons until unit activates. button to increment the digit and the SET When 0000 is displayed, enter the password, factory set to 1270, by using the UP button to accept the indicated digit and move the cursor. When the correct password is entered, the display will advance. Press the POWER button to advance to alarm set menu.
RECON/B SERIES ENMET 5.4 Sensor Replacement Sensor life in the RECON/B SERIES is typically: 2 years for Carbon Monoxide & Hydrogen Sulfide and 1year for Oxygen and Combustible. The sensor needs to be replaced when the RECON/B SERIES can no longer be calibrated or if the sensor becomes erratic. Contact ENMET for sensor replacement information.
RECON/B SERIES ENMET 7.0 Accessories and Replacement Parts 7.1 Replacement Sensor part numbers Sensor, % LEL Combustible Gasses Sensor, CO Sensor, H2S Sensor, Oxygen Sensor, H2 Sensor, NH3 Sensor, SO2 02542-005 02542-1200 02542-0200 02542-1100 02542-1500 02542-2400 02542-0500 7.
RECON/B SERIES ENMET 8.0 WARRANTY ENMET warrants new instruments to be free from defects in workmanship and material under normal use for a period of one year from date of shipment from ENMET. The warranty covers both parts and labor excluding instrument calibration and expendable parts such as calibration gas, filters, batteries, etc... Equipment believed to be defective should be returned to ENMET within the warranty period (transportation prepaid) for inspection.
PO Box 979 680 Fairfield Court Ann Arbor, Michigan 48106-0979 734.761.1270 Fax 734.761.3220 Returning an Instrument for Repair ENMET instruments may be returned to the factory or any one of our Field Service Centers for regular repair service or calibration. The ENMET Repair Department and Field Service Centers also perform warranty service work.