ENMET Corporation PO Box 979 Ann Arbor, MI 48106-0979 SPECTRUM SP Series Operation & Maintenance Manual Manual Part Number 80012-018 MCN-428, 12/23/09
Table of Contents 1.0 INTRODUCTION .......................................................................................................................................... 1 1.1 Unpack ..............................................................................................................................................................1 1.1.1 Check Order ...........................................................................................................................................1 1.1.
SPECTRUM SP Series ENMET Corporation 1.0 Introduction The SPECTRUM SP is a small portable battery operated single channel gas detection instrument featuring an internal motorized sample pump. Depending on the sensor supplied with the instrument, it can detect one of a number of potential target gases. An electrochemical cell detects the gas, and the gas concentration is displayed on an LCD.
SPECTRUM SP Series ENMET Corporation 1.2 Turn Instrument ON Turn the instrument ON, by pressing and holding the POWER / BACKLIGHT pushbutton for two seconds. The pump will automatically activate. In uncontaminated air, for most instruments, the display should read 0000 within ten seconds of turnon. NOTE: For instruments using a biased sensor, this time is extended to 4 minutes. Stabilization may take as long as 1 hour. See Table 1 Appendix A. 1.2.1 Verify The SPECTRUM SP is calibrated prior to shipment.
SPECTRUM SP Series ENMET Corporation OPTION Switch SELECT Switch POWER/BACKLIGHT Switch Alarm Indicator LED Power Option Sampling Port Select Charge Jack Alarm Green: Battery charger Connected or fully charged Red: Battery Charging Red LED Battery Charging Front View Green LED Battery Charger Right Side End Cap View Left Side End Cap View Sampling Port Charge Jack Belt Clip Belt Clip Figure 1: SECTRUM Features 3
SPECTRUM SP Series ENMET Corporation 2.1 Operation 2.1.1 Operational Menu The operation menu flow diagram is shown in Figure 2. This menu is accessed with the OPTION pushbutton switch, the middle switch of the three. Successive displays are achieved by repeatedly pushing the switch, as indicated by "O" in the menu flow diagram. The alarm acknowledgement function, and displays and function in the "see DATA" area, are accessed with the SELECT pushbutton, indicated by "S" in the operation menu flow diagram.
SPECTRUM SP Series ENMET Corporation 2.1.2 Turn ON and OFF Turn the instrument ON by pressing the POWER / BACKLIGHT pushbutton for two seconds. The pump will automatically actuate. For most instruments, the display should read "0000" within ten seconds when the instrument is in uncontaminated air. NOTE: Instruments using biased sensor, stabilization time is extended to 4 minutes. Complete stabilization may take as long as 1 hour. When instrument is turned on it may display XXXX + .
SPECTRUM SP Series ENMET Corporation 2.1.6 Battery Status and Low Battery Alarms The battery status display is accessed by pressing the OPTION pushbutton once. When this display is "Batt=OK", the battery energy level is sufficient for operation of the instrument. In this condition, the confidence beep occurs every thirty seconds when the instrument is not in alarm.
SPECTRUM SP Series ENMET Corporation 3.0 Maintenance 3.1 Maintenance Menu From the operational display, press the OPTION pushbutton four times; "go SETUP" is displayed. This is the entrance to the maintenance menu. The maintenance menu flow diagram is shown in Figure 3. From Operational Display To Operational Display go SETUP S Key=0000 INVALID O S Changes digit indicated by underscore cursor Locks underscored digit and moves cursor VALID run ZERO S O O AIR SET? S z30:0.
SPECTRUM SP Series ENMET Corporation 3.1.1 Key Entrance to the maintenance menu is guarded with a four digit numerical key. The factory default setting of the key is 1270*. When the valid numerical key is inserted, the user is allowed to enter the maintenance menu When in the "go SETUP" location, press the SELECT pushbutton; "Key=0000" is displayed. The underscore cursor is under the left hand digit.
SPECTRUM SP Series ENMET Corporation 3.1.4 Changing the Alarm Level A valid key entry sets the instrument at the "run ZERO" location of the maintenance menu. Press the OPTION pushbutton twice to access the "set ALM" display, then press the SELECT pushbutton; ALM=0000" is displayed. This is called the alarm update window, and the value displayed is the present alarm setpoint. The underscore cursor is under the far left digit.
SPECTRUM SP Series ENMET Corporation 3.2 Changing Components Changing the battery, the sensor, or the display requires that the back cover of the instrument to be removed; remove the four phillips head screws and then the back cover. See Figure 5. 3.2.1 Sensor Removal and Replacement A sensor must be replaced when it no longer responds adequately to the target gas.
SPECTRUM SP Series ENMET Corporation 3.2.2 Battery Removal and Replacement The battery is located under the circuit board. To replace the battery: Press on the locking tab and pull on the battery pack connector. Remove the circuit board and then remove the battery pack. Note that the battery pack is firmly attached to the enclosure with industrial strength adhesive tape. Replace the battery pack and circuit board. Reconnect battery pack connector.
SPECTRUM SP Series ENMET Corporation 4.
SPECTRUM SP Series ENMET Corporation 5.0 WARRANTY ENMET warrants new instruments to be free from defects in workmanship and material under normal use for a period of one year from date of shipment from ENMET. The warranty covers both parts and labor excluding instrument calibration and expendable parts such as calibration gas, filters, batteries, etc... Equipment believed to be defective should be returned to ENMET within the warranty period (transportation prepaid) for inspection.
SPECTRUM SP Series ENMET Corporation Appendix A: Calibration Data Tables Table 1: Gas Ranges, Alarm Points and Sensor Life Gas Ammonia Arsine Carbon Monoxide Chlorine Fluorine Hydrogen Hydrogen Chloride x Hydrogen Cyanide Hydrogen Fluoride Hydrogen Sulfide Nitric Oxide x Nitrogen Dioxide Ozone Phosgene Phosphine Silane Sulfur Dioxide Range Alarm lower limit Alarm upper limit Factory Alarm Set Point Expected Sensor Lifetime 0-100ppm 15ppm 50ppm 25ppm 1 year 0-0.99ppm 0.03ppm 0.
SPECTRUM SP Series ENMET Corporation Table 2: SPECTRUM SP Calibration Voltage and Countdown Times NOTE: The following are values used for production calibration. Not all gases are available for field calibration; contact ENMET customer service personnel. Gas Calibration Voltage Countdown Time 0.67 180 sec 0.10 120 sec 100ppm 1.70 120 sec Chlorine 5ppm 1.40 180 sec Fluorine 3ppm Cl2= 7.5ppm 1.80 240 sec Hydrazine 0.3ppm 0.91 240 sec Hydrogen 800ppm 2.
SPECTRUM SP Series ENMET Corporation APPENDIX C SPECTRUM SP for Hydrazine SPECTRUM SP For Hydrazine The detection of hydrazine, N2H4, is a specialized application. Generally, hydrazine gas blended in air or nitrogen in known verified amounts is not available to use as calibration gas in the field. Also, the sensor is not responsive to any other gas which could be used as a correlation gas. The sensor itself has a relatively short operating life, and should be replaced every six months.
PO Box 979 680 Fairfield Court Ann Arbor, Michigan 48106-0979 734.761.1270 Fax 734.761.3220 Returning an Instrument for Repair ENMET instruments may be returned to the factory or any one of our Field Service Centers for regular repair service or calibration. The ENMET Repair Department and Field Service Centers also perform warranty service work.
Repair Return Form Mailing Address: ENMET Corporation PO Box 979 Ann Arbor, Michigan 48106 Phone Number: FAX Number: Shipping Address: ENMET Corporation Attn: Repair Department 680 Fairfield Court Ann Arbor, Michigan 48108 734.761.1270 734.761.