Instructions
4
Mechanical and Environmental Conditions
Operating Conditions Value Range
Dimensions 96x96
Maximum Depth
(Inside the Panel)
44.5 mm
Installation Panel type
IP protection 20
Display LED display
Button 4 x Universal interface
Storage Temperature -30 / +80°C
Operating Temperature -20 / +70°C
Maximum Humidity 95% (noncondensing)
Standards
EN 61326,61000-6-4,61000-6-2 emc
EN 61010-1 Safety
EN 60529 Mechanic
EN 60068-2-1,60068-2-2 ,60068-2-30 Environmental
Technical Specications
Operating Voltage (Un) 100-270 VAC (-15%+10%)
Operating Frequency (f) 50-60 Hz
Input Power Consumption <5 VA
Measuring Inputs Power Consumption <1 VA
Measuring Voltage Input (Vin) 10-300 VAC(VLN) 10-480 VAC(VLL)
Measuring Current Input (Iin) 0.05-5.5 A
Class
Voltage 0,5%
Current 0,5%
Frequency ±0.02 Hz
cosφ ± 0.02
Demand Time Adjustable
60/120/300/600/1200/1800/3600 sec
Communication (Insulated) 4kV
Baud Rate Adjustable 2400/ 4800/
9600/19200/38400/57600/115200
Address Adjustable 1-256
Parity Adjustable Single/Double
Relay Ouputs 2x (5A 250 VAC/ 1250VA)
Connection 3P3W, 3P4W, Aron, 3P3W(B), 3P4W(B)
Ambient Temperature -20+70°C
Storage Temperature -30+80°C
Humidity 95%
Indicator LED Display
Dimensions 96x96x44.5
Device Protection Class IP 20