TOUCH SCREEN CONTROLLER
ABORT
Abort Button
If you need an emergency stop depress the
E-STOP immediately.
Activate the ABORT
By selecting the abort button, the abort operation will seize all communications to the control board.
The temperature will not read correctly and will not heat. All motion is stopped. The ABORT button will turn
green.
-A pop up message will appear: “DO YOU WANT TO ABORT”
Pressing the Cancel button will bring you back to the main screen and the abort command will not be activate.
Press the Abort button to activate the abort command, a pop up message will appear
“CONTINUE TO TURN ON FAN”
“CANCEL TO TURN OFF FAN”
Press the Continue button to turn the fan on.
If fumes are present or the block is greater than 40
0
C, it is recommended to keep the fan on
Pressing the Cancel button to leave fan off.
Deactivate the ABORT
After selecting to turn on the fan or leave off
-A pop up message will appear: “READY TO HOME MOTORS?”, Press the Continue button.
-Motors will activate to initialize in their home positioning.
-A pop up message will appear: “MOTORS ARE HOMED”. Press the Continue button.
The ABORT command will then be deactivated and the ABORT button will turn red.
Emergency Stop
Temperature Readouts
Actual temperature readings for samples & block are displayed in the upper right portion of the screen.
SX
Sample Temperature,
0
C
G Graphite Block Temperature,
0
C
Sample Temperature
Place the external thermocouple, located inside the front compartment, into a sample for accurate reading.
Temperature that has been set to the graphite block ,
0
C
Displayed in the upper left portion of the screen.
Reference the Manual Tab for setting the block temperature.
Manual Tab
Manual Tab
ENVIRONMENTAL EXPRESS
For technical support call 1-800-745-8218 or send email to INFO@ENVEXP.COM
5528TOUCHCREEN Rev 2 12 9/10/09