program mode –wait screen
PROGRAM TAB
WAIT
Sets the amount of time the samples will be heated during a step.
Range: 1-999 minutes.
The method will wait for the set graphite temperature to be reached,
±5
0
C degrees, rack are then lowered. The method will then wait for the
specified number of minutes.
Command Sequence:
1. Select WAIT command from the main menu
2. Select the wait time by up or down buttons or by keyboard. (minutes)
3. Added User Notes. With the keyboard activated, press comment area with stylus and type.
Proceed to the continue button
COOL
Set the amount of time the samples will be cooled. Range: 1-999 minutes
The racks will be raised out of the graphite block and cool for the specified
number of minutes. The fan will automatically increase to its maximum for
efficient sample cooling.
Command Sequence:
1. Select COOL command from the main menu
2. Select range by up or down buttons or by keyboard. (minutes)
3. Added User Notes. With the keyboard activated, press comment area with stylus and type.
Proceed to the continue button
program mode –cool screen
program mode –pause screen
PAUSE
An operator will receive a pop up message. The method will not execute
the next command until the analysts responds.
Command Sequence:
1. Select PAUSE command from the main menu
2. Add message with the keyboard activated, press comment area
with stylus and type.
Example message: Insure volume level is correct
Proceed to the continue button
program mode –sample program
DELETE
Directly deletes last command from the method screen.
Command Sequence:
1. Select PAUSE command from the main menu
Proceed to the continue button