4 Installation
Centrifuge 5424 / 5424 R — Operating manual
4Installation
Operating manual Centrifuge 5424 / 5424 Rseeon p.Fig.Tab.p.
4.1 Selecting the location
Select the location for the device according to the following criteria:
Suitable power connection as per the name plate (230 V/120 V/100 V).
Stable, horizontal and resonance-free lab bench. Weight of the device: 13.4 kg (5424) or
21 kg (5424 R).
A well ventilated environment which is protected from direct sunlight to prevent the device
from heating up more.
4.2 Installing the instrument
Perform the following steps in the sequence described.
1. Place the device on a suitable lab bench.
2. Allow the device to warm up for at least 3 hours (5424) or 4 hours (5424 R) to the ambient
temperature to prevent damage to the electronic components from condensation and damage
to the compressor (only5424 R).
3. Check that the mains voltage and frequency match the requirements on the device type plate.
4. Connect the centrifuge to the mains and switch it on using the mains power switch on the
right side of the device.
The display is active.
Lid opens automatically
5. Use the details included in the scope of delivery to check that the delivery is complete.
6. Check all parts for any transport damage. Contact your dealer if any damage is found.
7. Only 5424 R: Insert the condensation water tray at the left side of the device into the holder
provided (see Fig. 2 on p. 8).
NOTICE!
If a fault occurs, objects in the immediate vicinity of the devices could get damaged.
In accordance with the recommendations of EN 61010-2-020, leave a safety distance of 30
cm clear around the device during operation.
NOTICE!
Damage from overheating.
Do not place the device close to sources of heat (e.g., radiator, drying cabinet).
Do not expose the device to direct sunlight.
Ensure free circulation of air by maintaining a distance of at least 30 cm on all sides of the
device from adjacent devices or the wall and keep the underside of the device clear.
Make sure that the vents in the device are always free of obstruction.
NOTICE!
Centrifuge 5424 R: Compressor damage after improper transport.
Only switch on the centrifuge 4 hours after installation.
Hint!
Retain the packaging material and the transport protection device for subsequent transport or
storage. See also the instructions relating to transport (see p. 33).