4 Installation
Other printers
In addition to the DPU 414, it is also possible to connect other serial printers to the
serial interface of the BioPhotometer. With the aid of an adapter cable, parallel printers
can also be connected.
BioPhotometer
Select the function "Printer serial" in the functions list, and confirm.
Printer
Requirements for the serial printer:
Parallel printers can be connected using an adapter cable which fulfills the above
requirements.
4.3 Cuvettes
Commercially available rectangular cuvettes may be used in the cuvette shaft. When
the height of the measuring window is 8.5 mm above the cuvette base and the overall
height of the cuvette is at least 36 mm (see the graphics in "Short instructions").
The light bundle in the cuvette is 1.0 mm wide and 1.5 mm high.
For measurements, cuvettes made of glass or plastic may be used on condition that
they are transparent at the respective measuring wavelength. The UVette
®
from
Eppendorf is a plastic cuvette which is transparent at wavelengths as low as 220 nm,
which means that it is also suitable for nucleic acid measurement.
Busy Control : XON/XOFF
Baud Rate(ON) : 9600 bps
Data Bit Length : 8 bits
Parity Permission : Without
Parity Conditions : Odd