10 Short instructions
Bradford measuring procedure
BRADFORD
CALIBRATION RANGE
100 – 2000 µg/mL
Standard
BRADFORD BLANK
Blank
0 . 0 0 0 A
BRADFORD SAMPLE 001
20+200 µL
0.352 A
595
3 6 8 µg/mL
BRADFORD SAMPLE 002
20+200 µL
0.525 A
595
5 5 2 µg/mL
Select Bradford method
Measure blank
Measure sample
Measure next sample
Sample
Sample
Blank
1
Bradford
If "Bradford micro"
is to be selected:
BRAD.micro
CALIBRATION RANGE
1.0 – 20.0 µ g/mL
Standard
Blank
1
Bradford
BRADFORD STD 6
2000 µg/mL
CV: 2.8 %
CALIBRATION STORED
1 . 3 2 5 A
BRADFORD STD 1
100 µg/mL
NEXT: STD 2
200µg/mL
0 . 0 4 8 A
Measure last standard
Measure first standard
Enter
Enter
2
Lowry
Dilution
0
Sample No.
0
Sample No.
0
Sample No.
When the sample is diluted:
Example: 20 + 200 µL
2
Lowry
Standard
Standard
BRADFORD SAMPLE 001
20+200 µL
Blank
Standard
Sample
or
or
or
Blank
Standard
Sample
or