Manual

Troubleshooting
Centrifuge 5427 R
English (EN)
43
Err 22 Electronics fault. Switch centrifuge off and back on
again after > 20 s.
NO E-FAN (ERR 23) Electronic fan is faulty.
Switch centrifuge off and back on
again after > 20 s.
Err 24 Failure at the compressor.
Allow the centrifuge to cool down and
repeat the run.
INT (ERR 25) Power failure during a run.
Check the mains connection.
Err 28 Error during speed check.
Switch centrifuge off and back on
again after > 20 s.
Err 29 Set g-force/speed too high, e.g. after a
rotor change (see Automatic rotor
detection on p. 27).
Check the g-force/speed.
Repeat the run.
LID (ERR 30) Centrifuge lid cannot be locked.
Centrifuge lid cannot be released.
Try again to close centrifuge lid.
Switch the centrifuge off and back on.
Press the open key.
If the error occurs repeatedly:
Switch off centrifuge.
Actuate emergency release(see
Emergency release on p. 44).
OFF Displayed after power off.
Symptom/message Cause Remedy