4.6 Removing the Combitip
If you accidentally operate the ejector during dispensing, and the cylinder is released
because of this, an error message appears (see Error search on p. 36).
After dispensing you can remove the Combitip. If any liquid remains in the Combitip, it
needs to be emptied first.
4.6.1 Finishing the dispensing operation and removing the Combitip
1. Finish all dispensing steps.
The following information is displayed:
2. Hold the Multipette stream/Repeater stream over a tube.
3. Press the actuate key twice to move the piston to the basic position and to dispense the
liquid remaining in the Combitip in a safe manner.
The following information is briefly displayed:
4. Press the ejector button and remove the Combitip.
4.6.2 Emptying and removing the Combitip
If you would like to terminate dispensing prematurely, you need to empty the remaining
liquid out of the Combitip before you can remove it.
1. Hold the Multipette stream/Repeater stream over a tube.
2. Press the reset key to move the piston to the basic position and to dispense the liquid
remaining in the Combitip in a safe manner.
The following information is briefly displayed:
3. Press the ejector button and remove the Combitip.
CAUTION! Personal injury and damage to the Multipette stream/Repeater
stream and to the Combitip by operating the ejector at the wrong time.
If you eject the Combitip during dispensing, liquid may escape. If the ejector is
operated at the wrong time, only the cylinder of the Combitip is released. The
piston can only be released if it is in the basic position.
Only operate the ejector when the piston in the Combitip has reached its
lowest position.
2.Short = empty
1.Short = no step
Long = aspirate
Release
Combitip now
Release
Combitip now