Manual
43
Technical data
Multipette
®
/Repeater
®
(X)stream
English (EN)
Test conditions and test analysis in accordance with ISO 8655, part 6. Test with an
analytical balance with a moisture trap which has been inspected by the Office of Weights
and Measures.
• Number of determinations: 10
• Use of water in accordance with ISO 3696
• Inspection at 20 °C – 25 °C ±0.5 °C
• Dispensing onto the tube wall
• Volume tests in the mode Dis
• Speed levels set: 7
Combitip advanced Volume range Testing
volume
Error limits
Error
Systematic error Random error
± %± mL± %± mL
10 mL
orange
Increment:
0.01 mL
0.1 mL – 10 mL 1 mL 0.50 0.005 0.25 0.0025
5 mL 0.40 0.02 0.25 0.0125
10 mL 0.40 0.04 0.15 0.015
25 mL
red
Increment:
0.025 mL
0.25 mL – 25 mL 2.5 mL 0.30 0.0075 0.35 0.0088
12.5 mL 0.30 0.0375 0.25 0.0313
25 mL 0.30 0.075 0.15 0.0375
50 mL
light gray
Increment:
0.05 mL
0.5 mL – 50 mL 5 mL 0.30 0.015 0.50 0.025
25 mL 0.30 0.075 0.20 0.05
50 mL 0.30 0.15 0.15 0.075
The test volumes for the systematic and random error of the Multipette
(X)stream/Repeater (X)stream comply with the requirements of ISO 8655, part 5.