Owner's manual

47
Technical specifications
Eppendorf SOP
English (EN)
9.2.6 Biomaster
9.2.7 Varipette
0.5 µL – 5 mL
Increment:
0.005 mL
Purple
0.1 µL – 5 mL
120 mm
0.5 mL ±2.4 ±12 ±0.6 ±3.0
2.5 mL ±1.2 ±30 ±0.25 ±6.0
5.0 mL ±0.6 ±30 ±0.15 ±8.0
1 µL – 10 mL
Increment:
0.01 mL
Turquoise
1 µL – 10 mL
165 mm
1.0 mL ±3.0 ±30 ±0.6 ±6.0
5.0 mL ±0.8 ±40 ±0.2 ±10
10.0 mL ±0.6 ±60 ±0.15 ±15
Model Pipette tip Testing
volume
Error limits
Error
Systematic error
Random error
± % ± µL ± % ± µL
Biomaster 4830 Mastertip 2 µL ±6.0 ±0.12 ±4.0 ±0.08
3 µL ±5.0 ±0.15 ±3.0 ±0.09
5 µL ±4.0 ±0.2 ±2.0 ±0.1
10 µL ±3.0 ±0.3 ±1.5 ±0.15
20 µL ±2.5 ±0.5 ±0.8 ±0.16
Model Pipette tip Testing
volume
Error limits
Error
Systematic error
Random error
± % ± mL ± % ± mL
Varipette 4720 Varitip S 2.5 mL ±1.0 ±0.025 ±0.2 ±0.005
5 mL ±0.4 ±0.02 ±0.2 ±0.01
10 mL ±0.3 ±0.03 ±0.2 ±0.02
Varipette 4720 Varitip P 1 mL ±0.6 ±0.006 ±0.2 ±0.002
5 mL ±0.5 ±0.025 ±0.1 ±0.005
10 mL ±0.3 ±0.03 ±0.1 ±0.01
Research plus adjustable single-channel
Model Test tip
epT.I.P.S. Color code
Volume range
Length
Testing
volume
Eppendorf AG error limits
Error
Systematic
error
Random
error
± % ± µL ± % ± µL