Owner's manual

48
Technical specifications
Eppendorf SOP
English (EN)
9.2.8 Xplorer and Xplorer plus
pipette
Increment
Pipette tip
Color code
Volume range
Length
Testing
volume
Error limits
Error
Systematic
error
Random
error
µL µL ±% ± µL ±% ± µL
0.5 µL – 10 µL
Increment:
0.01 µL
Medium gray
0.1 µL – 20 µL
40 mm
1 µL ±2.5 ±0.025 ±1.8 ±0.018
5 µL ±1.5 ±0.075 ±0.8 ±0.04
10 µL ±1.0 ±0.1 ±0.4 ±0.04
5 µL– 100 µL
Increment:
0.1 µL
Yellow
2 µL - 200 µL
53 mm
10 µL ±2.0 ±0.2 ±1.0 ±0.1
50 µL ±1.0 ±0.5 ±0.3 ±0.15
100 µL ±0.8 ±0.8 ±0.2 ±0.2
15 µL – 300 µL
Increment:
0.2 µL
Orange
15 µL – 300 µL
55 mm
30 µL ±2.5 ±0.75 ±0.7 ±0.21
150 µL ±1.0 ±1.5 ±0.3 ±0.45
300 µL ±0.6 ±1.8 ±0.2 ±0.6
50 µL – 1 000 µL
Increment:
1 µL
Blue
50 µL – 1000 µL
71 mm
100 µL ±3.0 ±3.0 ±0.6 ±0.6
500 µL ±1.0 ±5.0 ±0.2 ±1.0
1000 µL ±0.6 ±6.0 ±0.2 ±2.0
mL mL mL ±% ± mL ±% ± mL
0.25 mL – 5 mL
Increment:
0.005 mL
Purple
0.1 mL – 5 mL
120 mm
0.5 mL ±3.0 ±15.0 ±0.6 ±3.0
2.5 mL ±1.2 ±30.0 ±0.25 ±6.25
5 mL ±0.6 ±30.0 ±0.15 ±7.5
0.5 mL – 10 mL
Increment:
0.01 mL
Turquoise
1 mL – 10 mL
165 mm
1 mL ±3.0 ±30.0 ±0.60 ±6.0
5 mL ±0.8 ±40.0 ±0.20 ±10.0
10 mL ±0.6 ±60.0 ±0.15 ±15.0