Owner's manual

52
Technical specifications
Eppendorf SOP
English (EN)
9.3.4 Reference2
Reference2 Variable multi-channel (8-channel/12-channel)
Model Test tip
epT.I.P.S. Color code
Volume range
Length
Testing
volume
Eppendorf AG error limits
Error
Systematic
error
Random
error
± % ± µL ± % ± µL
0.5 µL – 10 µL Medium gray
0.1 µL – 20 µL
40 mm
0.5 µL ±12.0 ±0.06 ±8.0 ±0.04
1 µL ±8.0 ±0.08 ±5.0 ±0.05
5 µL ±4.0 ±0.2 ±2.0 ±0.1
10 µL ±2.0 ±0.2 ±1.0 ±0.1
10 µL – 100 µL Yellow
2 µL - 200 µL
53 mm
10 µL ±3.0 ±0.3 ±2.0 ±0.2
50 µL ±1.0 ±0.5 ±0.8 ±0.4
100 µL ±0.8 ±0.8 ±0.3 ±0.3
30 µL – 300 µL Orange
20 µL – 300 µL
55 mm
30 µL ±3.0 ±0.9 ±1.0 ±0.3
150 µL ±1.0 ±1.5 ±0.5 ±0.75
300 µL ±0.6 ±1.8 ±0.3 ±0.9