Owner's manual
53
Technical specifications
Eppendorf SOP
English (EN)
9.3.5 Xplorer and Xplorer plus
pipette
Increment
Pipette tip
Color code
Volume range
Length
Testing
volume
Error limits
Error
Systematic
error
Random error
µL µL % µL % µL
0.5 µL – 10 µL
Increment:
0.01 µL
Medium gray
0.1 µL – 20 µL
40 mm
1 µL ±5.0 ±0.05 ±3.0 ±0.03
5 µL ±3.0 ±0.15 ±1.5 ±0.075
10 µL ±2.0 ±0.2 ±0.8 ±0.08
5 µL – 100 µL
Increment:
0.1 µL
Yellow
2 µL –200 µL
53 mm
10 µL ±2.0 ±0.2 ±2.0 ±0.2
50 µL ±1.0 ±0.5 ±0.8 ±0.4
100 µL ±0.8 ±0.8 ±0.25 ±0.25
15 µL – 300 µL
Increment:
0.2 µL
Orange
15 µL –300 µL
55 mm
30 µL ±2.5 ±0.75 ±1.0 ±0.3
150 µL ±1.0 ±1.5 ±0.5 ±0.75
300 µL ±0.6 ±1.8 ±0.25 ±0.75
50 µL – 1200 µL
Increment:
1 µL
Green
50 µL–1250 µL
76 mm
120 µL ±6.0 ±7.2 ±0.9 ±1.08
600 µL ±2.7 ±16.2 ±0.4 ±2.4
1200 µL ±1.2 ±14.2 ±0.3 ±3.6