Owner's manual

55
Technical specifications
Eppendorf SOP
English (EN)
9.4.2 MultipetteM4
Combitip advanced Testing volume Error limits
Error
Systematic error Random error
± % ± µL ± % ± µL
0.1 mL
white
Increment: 1 µL
2 µL ±1.6 ±0.032 ±3.0 ±0.06
20 µL ±1.0 ±0.2 ±2.0 ±0.4
0.2 mL
light blue
Increment: 2 µL
4 µL ±1.3 ±0.052 ±2.0 ±0.08
40 µL ±0.8 ±0.32 ±1.5 ±0.6
0.5 mL
purple
Increment: 5 µL
10 µL ±0.9 ±0.09 ±1.5 ±0.15
100 µL ±0.8 ±0.8 ±0.6 ±0.6
1 mL
yellow
Increment: 10 µL
20 µL ±0.9 ±0.18 ±0.9 ±0.18
200 µL ±0.6 ±1.2 ±0.4 ±0.8
2.5 mL
green
Increment: 25 µL
50 µL ±0.8 ±0.4 ±0.8 ±0.4
500 µL ±0.5 ±2.5 ±0.3 ±1.5
5 mL
blue
Increment: 50 µL
100 µL ±0.6 ±0.6 ±0.6 ±0.6
1 000 µL ±0.5 ±5.0 ±0.25 ±2.5
10 mL
orange
Increment: 0.1 mL
200 µL
0.2 mL
±0.5 ±1.0 ±0.6 ±1.2
2 000 µL
2 mL
±0.5 ±10 ±0.25 ±5.0
25 mL
red
Increment: 0.25 mL
500 µL
0.5 mL
±0.4 ±2.0 ±0.6 ±3.0
5 000 µL
5 mL
±0.3 ±15 ±0.25 ±12.5
50 mL
light gray
Increment: 0.5 mL
1 000 µL
1 mL
±0.3 ±3.0 ±0.5 ±5.0
10 000 µL
10 mL
±0.3 ±30 ±0.3 ±30